The indicator of choosen mode is on.
i) Select a suitable scaling tip as you need, screw it on the handpiece tightly by
the torque wrench (See Figure 6).
j) Step on the foot switch, the tip begins to vibrate.
k) The normal frequency is extremely high. Under the normal working state
of the scaling tip, a light touch and a certain to-and-fro motion will eliminate the
tartar without heating. Overexertion and long-time lingering are forbidden.
l) Vibration intensity: Adjust the vibration intensity as you need, generally
increase the power by pressing the Increasing Power Switch to a higher grade.
According to patient's dierent sensitivity and the rigidity of the gingival tartar,
adjust the vibration intensity during the clinical treatment.
m) Water volume adjust: Step on the foot switch, and the tip begins to vibrate,
then turn the water control knob to form ne spray to cool down the handpiece
and clean the teeth.
n) The handpiece can be handled in the same gesture as a pen in hand.
o) During the clinical treatment, be sure not to make the end of tip touch the
teeth vertically and not to make the tip overexert on the surface of the teeth in
case of hurting the teeth and damaging the tip.
p) After nishing operation, keep the machine working for 30 seconds on the
water supply condition in order to clean the handpiece and the scaling tip.
q) Unscrew the scaling tip and pull out handpiece, then sterilize them.
Notice:
① Don't pull out the handpiece when the foot switch is stepped on and
the machine is producing ultrasonic vibration.
② Please don't choose dry mode unless needed in the treatment, please
don't make the machine work under dry mode for a long time. The teeth will
be hurt if the handpiece and tips getting overheat during the treatment.
2.1.2 Instruction for main components of detachable handpiece (See Figure 5):
a) Nipple: The nipple can be removed. You can screw out the nipple and clean
the pole with alcohol termly.
b) Decorative ring: can be disassembled and cleaned with alcohol regularly,
can be autoclaved under the high temperature and pressure.
c) Handpiece: The main part of the whole handpiece, can be autoclaved under
the high temperature and pressure.
d) The connector of the cable: Connect the handpiece with the water source
and power supply of the main unit.
Notice: Keep dry when the detachable handpiece connects to the
connector of the cable.