19
clients connected to the server will receive a reminder of 10s countdown to
upgrade. After 10 seconds, the software will automatically close and then start
to upgrade. Click “No” and the software will not be upgraded temporarily.
4 Notes
4.1 The imaging plate scanner shall never be used in the presence of
ammable anesthetic gas, pure oxygen or nitrogen oxide to avoid any risk of
explosion.
4.2 The imaging plate scanner is a Class I laser product and turns to Class
3B after the cover is removed. Make sure to avoid exposure to laser rays to
prevent the risk of injury to eyes.
4.3 The imaging plate scanner and its accessories, especially the IP imaging
plate, protection card and protective bag, have been designed and developed
to ensure the highest safety and performance. The use of accessories not
provided by the original manufacturer may cause risks to patients, users
or the equipment itself. Do not use IP imaging plates, protection cards
and protective bags provided by other manufacturers to avoid any risk of
contamination or malfunction of imaging plate scanner.
4.4 In the handling of IP imaging plate, protection card and protective bag,
necessary health protection measures must be taken to avoid the risk of cross
infection.
4.5 The imaging plate scanner complies with the IEC 60601-1 standard.
Only peripheral equipment conforming to IEC 60950-1 can be connected to
the imaging plate scanner to avoid any risk of failure of the scanner.
4.6 Our company is specialized in the production of medical devices. We
are responsible for the safety of the device only when the maintenance, repair
and modication of the machine are carried out by our company or by our
authorized dealers, and the replacement parts are our Woodpecker accessories
and operated according to the operating instructions.
4.7 Other safety information can be found in each chapter of this instruction
manual. Please read the whole manual carefully.
4.8 In order to ensure safe and correct operation and the use of the imaging
plate scanner, it is very important to use the power cord provided by the
scanner. The power cord of the imaging plate scanner system must only be
replaced by the same type of cord.
4.9 Due to the electromagnetic compatibility of imaging plate scanner,
other equipment nearby may be affected during the use. There is a risk of
malfunction of nearby equipment.
4.10 Due to electromagnetic compatibility, the use of other equipment may
interfere with the imaging plate scanner.