Num Object Remark Type No./model No./ Manufacturer/Trademark
1 Detachable handpiece / HW-6L Woodpecker
2 Scaler tips / / Woodpecker
3 Power suplly See picture 5 ZMN-03 Guilin Woodpecker
4 F2 food switch See picture 6 F2 Woodpecker
5 Endochuck / TW-E1 Woodpecker
6 Torque wrench / TW-1L Woodpecker
7 Endo wrench / / Woodpecker
2. Installation and adjustment
2.1 Operation
2.1.1 Open the packing box, make sure that all the parts and accessories are complete according to the packing list, take the main unit out of the
box, and put it on the the stable plane facing to the operator.In addtion,install the device to an operator-friendly position.
2.1.2 Turn the water control knob to the max according to the direction showed in picture 1. Do not screw it over tight in case of damage. [ note 1]
2.1.3 Insert the plug of the foot switch to its socket. (see picture 2)
2.1.4 Connect one end of the water pipe to the water entrance, and the other end to the clean water source. (see picture 2)
2.1.5 Before treatment, make sure scaling tip, wrench and handpiece have been sterilized. Choose the scaling tip according to the requirement,
and x the scaling tip with the wrench. ( see picture 4)
2.1.6 Turn on the power switch. If the power indicator is lighted, the machine is ready for work.
2.1.7 Under normal working condition, the frequency of the tips is very high. Light touch and a certain to-and-fro motion will eliminate the tartar
without obvious heating. Overexertion and overstay are forbidden.
2.1.8 Vibrating intensity: Adjust the vibrating intensity according to your need, usually adjust to the middle grade, and adjust the vibrating during
the clinical treatment according to the patient's sensitivity and the rigidity of the tartar.
2.1.9 Step on the foot switch, the tip begins to vibrate, and the LED lamp on the top of the handpiece shines. Release the foot switch, the LED
lamp keep shining for 10 seconds.
2.1.10 Water volume adjustment: Step on the foot switch, and the tip begins to vibrate, then turn the water control switch to ne spray to cool
down the handpiece and clean the teeth.
2.1.11 The handpiece can be handled in the same gesture as a pen in hand.
2.1.12 Be sure not to make the end of the tip touch the teeth vertically, and not use too much pressure when the tip touch the surface of the teeth,
in order not to hurt the teeth and the tip.
2.1.13 After nishing operation, keep the machine working for 30 seconds with the water supply to clean the handpiece and the tip.
2.1.14 Unscrew the scaling tip and sterilize it.