WPI IGS100 User manual

IGS100
Implantable Glucose Sensor
World Precision Instruments
INSTRUCTION MANUAL
Serial No._____________________
011310
www.wpiinc.com

IGS100
ii World Precision Instruments

IGS100
World Precision Instruments iii
Copyright © 2009 by World Precision Instruments, Inc. All rights reserved. No part of this publication
may be reproduced or translated into any language, in any form, without prior written permission of
World Precision Instruments, Inc.
CONTENTS
ABOUT THIS MANUAL ......................................................................................................1
INTRODUCTION ..................................................................................................................2
Notes and Warnings .....................................................................................................2
INSTRUMENT DESCRIPTION ..........................................................................................2
Parts List ............................................................................................................................2
Unpacking ........................................................................................................................2
OPERATING INSTRUCTIONS ...........................................................................................3
Conditioning the Sensor ............................................................................................3
Preparing Solution for Calibration ...........................................................................3
Calibrating the IGS100 .................................................................................................3
SENSOR CARE AND MAINTENANCE ...........................................................................5
Storing the Sensor .........................................................................................................5
Maintaining the Sensor ...............................................................................................5
Sterilizing the Sensor ....................................................................................................5
ACCESSORIES .......................................................................................................................6
TROUBLESHOOTING .........................................................................................................6
SPECIFICATIONS ..................................................................................................................6
WARRANTY ..........................................................................................................................7
Claims and Returns .......................................................................................................7
Repairs ...............................................................................................................................7

IGS100
iv World Precision Instruments

IGS100
World Precision Instruments 1
ABOUT THIS MANUAL
The following symbols are used in this guide:
This symbol indicates a CAUTION. Cautions warn against actions that
can cause damage to equipment. Please read these carefully.
This symbol indicates a WARNING. Warnings alert you to actions that
can cause personal injury or pose a physical threat. Please read these
carefully.
NOTES and TIPS contain helpful information.
Fig. 1–IGS100 sensor
Fig. 2–Tip of the sensor showing the coil holding the enzyme saturated cotton
plug. The diameter of the coiled tip is about 600µm.

IGS100
2 World Precision Instruments
INTRODUCTION
IGS100 implantable glucose sensor for real-time monitoring is an
amperometric microsensor that can be used in vivo. The implantable,
sensing tip is slightly larger than a half-millimeter in diameter. The tip is
comprised of a metal coil with an enzyme-soaked cotton core. A proprietary
coating covers the entire tip. As the sensor is used, the enzyme is consumed.
The tip design resists contamination from biological materials, which
significantly improves the life of the sensor. Depending on measurement
conditions, IGS100 may be used in vivo for weeks or even months. When
not in use, it has a six-month shelf life.
Notes and Warnings
NOTE: The sensor must be polarized for at least 4 hours in 5mM glucose
solution prior to use.
INSTRUMENT DESCRIPTION
Parts List
After unpacking, verify that there is no visible damage to the sensor. Verify
that all items are included:
(1) IGS100 Glucose sensor
(1) Instruction Manual
(1) Sensor Performance Evaluation
Unpacking
Upon receipt of this sensor, make a thorough inspection of the contents and
check for possible damage. Missing cartons or obvious damage to cartons
should be noted on the delivery receipt before signing. Concealed damage
should be reported at once to the carrier and an inspection requested.
Please read the section entitled “Claims and Returns” on page 7 of this
manual. Please contact WPI Customer Service if any parts are missing at
941.371.1003 or customerservice@wpiinc.com.
The IGS100 sensor is shipped in a sealed, rigid plastic, hinged box with
foam padding to avoid damage to your sensor during shipment.
KEEP THE SENSOR STORAGE BOX and all the documentation in a safe
place. The storage box allows you to store the sensor safely with the tip
suspended. The test date and serial number of each sensor is printed on the
bottom of its box. Use of the sensor should begin within 30 days of receipt.

IGS100
World Precision Instruments 3
Returns: Do not return any goods to WPI without obtaining prior approval
(RMA # required) and instructions from WPI’s Returns Department. Goods
returned (unauthorized) by collect freight may be refused. If a return
shipment is necessary, use the original container, if possible. If the original
container is not available, use a suitable substitute that is rigid and of
adequate size. For further details, please read the section entitled “Claims
and Returns” on page 7 of this manual.
OPERATING INSTRUCTIONS
Conditioning the Sensor
1. Turn on the free radical analyzer.
2. Set the poise voltage on the free radical analyzer to +600mV.
3. Plug the IGS100 glucose sensor into the microsensor cable (WPI
#91580) and plug the other end into the free radical analzyer.
4. Immerse the sensor element into a 5mM glucose solution with 0.05M
PBS buffer (pH=7.4). It takes about 3-4 hours to stabilize the first time.
Preparing Solution for Calibration
Prepare a 2M glucose standard solution by dissolving 7.2g glucose into
20mL distilled water. Let the solution stand overnight at room temperature
to allow equilibration of anomers. Store solution at 4ºC.
Calibrating the IGS100
1. Condition the sensor as described above.
2. Remove the sensor from the glucose solution. Rinse the sensor with
distilled water.
3. Immerse the glucose sensor tip into 20mL of 0.05M PBS solution.
4. Place a magnetic stirring bar into the solution. turn on the stirrer so that
the bar is stirring at a moderate rate.
NOTE: The rate of the stirrer should not be modified once it is set.
5. After waiting a few minutes for the current to stabilize, record its value.
Typically the current should be a few nA or less. ( It should not exceed
20nA.)
NOTE: The calibration should be carried out at the temperature at which
the measurement of glucose is to be made. This can be accomplished
by placing the vial and stand in a water bath at the appropriate

IGS100
4 World Precision Instruments
temperature, and allowing the temperature of the solution in the bottle
to equilibrate with water bath.
5. Once the current has achieved a stable value, record this value or use
zero function in the instrument to set the baseline to zero.
6. Add 20µl of the 2M stock glucose solution into the 20mL of 0.05M PBS
buffer solution. The concentration of glucose in the solution is now
2mM. The current output will rapidly increase upon addition of glucose
and reach a plateau in a couple minutes.
7. Add an additional 20µl aliquot of 2M stock glucose solution when the
signal reaches a plateau. The concentration of glucose in the solution is
increasing by 2mM with the addition of each aliquot. When the current
output reaches a plateau, record the value displayed on the free radical
analyzer.
8. Repeat step 7 another three times. A typical calibration plot looks
similar to the one shown in Fig. 3.
2mM glucose
2.86nA
10mM glucose
14.72nA
8mM glucose
11.91nA
6mM glucose
9.18nA
4mM glucose
6.00nA
Glucose (nA)
Time (seconds)
Fig. 3–Calibration plot showing five aliquots of 2M glucose injected into PBS solution
9. Plot the current readings against the total glucose concentration on
a graph. The slope of this line indicates the sensitivity of the glucose
sensor. Once the sensitivity of the glucose sensor is known, the sensor is
ready for experimental use.

IGS100
World Precision Instruments 5
Fig. 4–Calibration data plotted shows the linear rise in current with the increase of
the solution molarity
NOTE: Since the enzyme is consumed as the sensor is used, the sensor
should be re-calibrated before each experiment.
SENSOR CARE AND MAINTENANCE
Storing the Sensor
STANDBY: If the sensor is being used the next day, it should be stored with
the tip suspended in 5mM glucose solution at 4ºC.
LONG-TERM: For long-term storage of more than one week, rinse the
sensor tip with distilled water and let it dry. Protect the tip and store the
sensor dry at 4ºC. It is best to store the sensor in its original box with the tip
suspended.
Maintaining the Sensor
After each use, rinse the sensor with distilled water.
Sterilizing the Sensor
The sensor can be sterilized chemically using an appropriate disinfectant
(for example, Cidex, WPI#7364).
CAUTION: Do not use alcohol on the sensor.

IGS100
6 World Precision Instruments
ACCESSORIES
Table 1: Accessories
Part Number Description
91580 Microsensor cable
TROUBLESHOOTING
If you have a problem/issue with your IGS100 sensor, contact the WPI
Technical Support team at 941.371.1003 or technicalsupport@wpiinc.com.
SPECIFICATIONS
The IGS100 sensor conforms to the following specifications:
In Vitro Precision Coefficient of variation (CV) ≤5%
Glucose Range 36 - 450mg/dl (or 2-25mM)
Response Time 100-300 sec
Interference Species Acetaminophen, ascorbic acid, uric acid
Overall Length 5 cm
Sensor Size Diameter: 0.6 mm
Active sensor length: 1.5 mm
Flex body: 35 mm
Reference Electrode Ag/AgCl integral
Polarization Voltage 0.6V
Sensor Life 3-4 months in solutions at room
temperature under continuous
polarization; 15-30 days in vivo.
Shelf Life 6 months
Operating Conditions 20°–40° C (68°–104° F)
Storage Conditions 4°–25° C (50°–77° F)

IGS100
World Precision Instruments 7
* Electrodes, batteries and other consumable parts are warranted for 30 days only from the date on which the
customer receives these items.
WARRANTY
WPI (World Precision Instruments, Inc.) warrants to the original purchaser that this equipment, includ-
ing its components and parts, shall be free from defects in material and workmanship for a period of
30 days* from the date of receipt. WPI’s obligation under this warranty shall be limited to repair or re-
placement, at WPI’s option, of the equipment or defective components or parts upon receipt thereof
f.o.b. WPI, Sarasota, Florida U.S.A. Return of a repaired instrument shall be f.o.b. Sarasota.
The above warranty is contingent upon normal usage and does not cover products which have been
modified without WPI’s approval or which have been subjected to unusual physical or electrical stress
or on which the original identification marks have been removed or altered. The above warranty will
not apply if adjustment, repair or parts replacement is required because of accident, neglect, misuse,
failure of electric power, air conditioning, humidity control, or causes other than normal and ordinary
usage.
To the extent that any of its equipment is furnished by a manufacturer other than WPI, the foregoing
warranty shall be applicable only to the extent of the warranty furnished by such other manufacturer.
This warranty will not apply to appearance terms, such as knobs, handles, dials or the like.
WPI makes no warranty of any kind, express or implied or statutory, including without limitation any
warranties of merchantability and/or fitness for a particular purpose. WPI shall not be liable for any
damages, whether direct, indirect, special or consequential arising from a failure of this product to
operate in the manner desired by the user. WPI shall not be liable for any damage to data or property
that may be caused directly or indirectly by use of this product.
Claims and Returns
Inspect all shipments upon receipt. Missing cartons or obvious damage to cartons should be noted
on the delivery receipt before signing. Concealed loss or damage should be reported at once to the
carrier and an inspection requested. All claims for shortage or damage must be made within ten
(10) days after receipt of shipment. Claims for lost shipments must be made within thirty (30) days
of receipt of invoice or other notification of shipment. Please save damaged or pilfered cartons until
claim is settled. In some instances, photographic documentation may be required. Some items are
time-sensitive; WPI assumes no extended warranty or any liability for use beyond the date specified
on the container
Do not return any goods to us without obtaining prior approval and instructions from our Returns
Department. Goods returned (unauthorized) by collect freight may be refused. Goods accepted for
restocking will be exchanged or credited to your WPI account. Goods returned which were ordered
by customers in error are subject to a 25% restocking charge. Equipment which was built as a special
order cannot be returned.
Repairs
Contact our Customer Service Department for assistance in the repair of apparatus. Do not return
goods until instructions have been received. Returned items must be securely packed to prevent fur-
ther damage in transit. The Customer is responsible for paying shipping expenses, including adequate
insurance on all items returned for repairs. Identification of the item(s) by model number, name, as
well as complete description of the difficulties experienced should be written on the repair purchase
order and on a tag attached to the item.

World Precision Instruments, Inc.
USA
International Trade Center, 175 Sarasota Center Blvd., Sarasota FL 34240-9258
Tel: 941-371-1003 • Fax: 941-377-5428 • E-mail: sales@wpiinc.com
UK
Astonbury Farm Business Centre • Aston, Stevenage, Hertfordshire SG2 7EG
Tel: 01438-880025 • Fax: 01438-880026 • E-mail: [email protected]
Germany
Zossener Str. 55, 10961 Berlin
Tel: 030-6188845 • Fax: 030-6188670 • E-mail: [email protected]
Japan
Physio-Tech Co., Ltd.
1-6-3 Iwamoto-cho, Chiyoda-ku, Tokyo 100-0032
Tel: 81-3-3864-2781 • Fax: 81-3-3864-2787 • E-mail: [email protected]
Australia
Coherent Life Sciences Pty. Ltd.
116 Sir Donald Bradman Dr • Hilton, South Australia 5033
Tel: (03) 9887-6262 • Fax: (03) 9887-9585 • E-mail: [email protected]
China & Hong Kong
WPI Shanghai Trading Co., Ltd.
Rm 20a, No8 Dong Fang Rd., Lu Jia Zui Financial District, Shanghai PRC
Tel: +86 688 85517 • E-mail:[email protected].com
Internet
www.wpiinc.com
Table of contents
Other WPI Accessories manuals
Popular Accessories manuals by other brands

Young Living
Young Living MACARON BL-056 user manual

ARDES
ARDES AR4O160 Instructions for use

Grange Fencing
Grange Fencing PKITCAG Assembly instructions

Fire-Lite Alarms
Fire-Lite Alarms SD355CO Installation and maintenance instructions

HomeMatic
HomeMatic HM-WDS100-C6-O Mounting instruction and operating manual

CKD
CKD RAPIFLOW FSM2-D Series instruction manual