SenTix Commissioning, measuring, calibration
15
ba75843e01 08/2009
Commissioning, measuring, calibration
Commissioning The electrode is filled with reference electrolyte solution in the factory. Pre-
pare the electrode for measuring as follows:
zOpen the refilling opening for the reference electrolyte solution. Depend-
ing on the model, the stopper of the refilling opening is an elastomer stop-
per or a slider.
The refilling opening must always be open during calibration and
measurement!
zRemove the watering cap from the electrode tip. Possible salt deposits in
the area of the watering cap do not affect the measuring characteristics
and can easily be removed with deionized water.
Note
Please keep the watering cap. It is required for the electrode to be stored.
Always keep the watering cap clean.
zSenTix®950: Remove any gas bubbles behind the pH membrane by
shaking.
zConnect the electrode to the meter.
zCalibrate the electrode according to the operating manual of the meter
and observe the following rules while doing so:
Calibration
and measurement:
General rules
zMake sure the refilling opening for the reference electrolyte solution is
open.
zAvoid the carryover of any solution (sample or buffer solution) from one
measurement to the next by taking the following measures:
– Shortly rinse the calibration and sample beakers with the solution the
beakers are to be filled with next.
– Between measurements, rinse the electrode with the solution that fol-
lows. Alternatively, you can also rinse the electrode with deionized wa-
ter and then carefully dab it dry.
zImmerse the electrode in the solution in a vertical or slightly tilted position.
Make sure the immersion depth is correct. The junction must be complete-
ly submersed in the solution. The junction is in the area of the bottom end
of the shaft (see figure). At the same time, the level of the reference elec-
Refilling
opening
Refilling
opening