Place the scale, and ,using the bubble on the frontside, level It.
Press to turn on the scale. Following a countdown, marked also by the beep, you will see the software
version, the capacity of the scale, and then you enter the measurement mode.
Wait at least 15 minutes to warm up before the first use.
User calibration (only for no OIML products)
Each balance is calibrated in the laboratory by qualified technicians using certified sample masses.
Calibration procedure
•Turn off the balance. Keep pressed key, and press key to display shows “CAL”;
•When no-load, press key to calibrate zero point, display [C-ZERO] first;
•Then display shows “SPAN”: place 1 kg, 2 kg or 3 kg.
The display shows “on1”, “on2” or “on3” in reference with the mass on weighing plate;
•Then display shows “pass”and count backward to the weighing mode.
Unit Setting
Unit selection
When the balance is in weighing mode, keep pressed for 3 second to change the unit one by one: kg,
g, lb-oz, pcs. The led of the corresponding unit lights on.
Default unit selected is gram [g].
Zero
With no weight on the scales, press the button to take the 0.00 reading on the display. The ZERO
function is only effective if the value shown is less than 4% of total capacity. If the value is greater than 4%,
the load is regarded as a tare.
Tare
After placing the tare on the scales, press the button to reset the value displayed to 0.00. The load
used for Tara has to be subtracted by the overall capacity weighing.