11
Setup
Sensor Installation
To use the ProSwap Logger with a ProDSS Sensor, proceed with the
following instructions:
1. Remove the port cover shipped with the cable.
2. Apply a thin coat of lubricant to the sensor o-rings.
3. Carefully align the sensor and bulkhead connectors by
inserting the sensor into the port then gently rotating the
sensor until the connectors align. Once aligned, push the
sensor toward the bulkhead until the sensor seats in the port.
4. Finger-tighten the retaining nut clockwise. If any resistance
is felt, loosen the retaining nut completely to prevent cross-
threading.
5. Use the sensor installation tool to tighten the retaining
nut clockwise until snug. Be careful not to over-tighten the
retaining nut.
6. Install the probe guard.
NOTICE: Incorrect installation or over-tightening can cause
damage to the sensor or bulkhead that is not
covered by the warranty.
Sensor Removal
Make sure the probe guard is removed first.
1. To remove a sensor, insert the sensor installation/removal
tool into the retaining nut, then rotate the retaining nut
counterclockwise to loosen.
2. After the retaining nut has been completely unscrewed from the
bulkhead, pull the sensor straight out of the port and place it on
a clean surface.
3. Use a lint-free cloth or compressed air to ensure that the port is
clean and dry. Be careful not to damage the connector pins.
4. Install a port plug if not reinstalling a sensor in the exposed port.
a. Apply a thin coat of o-ring lubricant to the o-rings on the plug
port.
b. Remove any excess lubricant from the o-rings and port plug
with a lint-free cloth.
c. Insert the port plug into the empty port and press until firmly
seated.
d. Finger-tighten the port plug clockwise to install. If necessary,
use the sensor installation tool to make sure that the plug is fully
seated into the port. The o-rings will not be visible if a port plug
is correctly installed. Do not over-tighten the port plug.
NOTICE: Exposure to water can cause
damage or corrosion to the
bulkhead connector not covered by
the warranty.
A port plug and a tube of o-ring
lubricant are included in the
maintenance kit that ships with all
ProSwap Loggers.