
Toc-2
IM 12E01B30-02EN
<CONTENTS>
2nd Edition : June 09 , 2021-00
4.1.3 Meas. damping time constant, Maint. damping time constant .......... 4-3
4.1.4 Turbidity warning High/Low limit ........................................................ 4-3
4.1.5 Air denoising setting........................................................................... 4-3
4.2 Cal./Maint. Settings........................................................................................... 4-6
4.2.1 Turbidity value of check tool............................................................... 4-6
4.2.2 Zero shift value, Correction factor...................................................... 4-6
4.2.3 Stability check .................................................................................... 4-6
4.2.4 Wash/Cal. box.................................................................................... 4-6
4.2.5 Ultrasonic wash.................................................................................. 4-6
4.2.6 Calibration setting others ................................................................... 4-7
4.3 Diagnosis setting .............................................................................................. 4-7
4.3.1 Light source diagnosis ....................................................................... 4-7
4.3.2 Dried condition diagnosis................................................................... 4-7
4.3.3 Operating days................................................................................... 4-7
4.4 Communication setting.................................................................................... 4-7
4.5 Auto Wash/Calibration setting ........................................................................ 4-8
4.5.1 Auto sequence for wash/cal............................................................... 4-8
4.5.2 Start date............................................................................................ 4-8
4.5.3 Auto Wash/Calibration function and wash type ................................. 4-8
4.5.4 Auto sequence setting ....................................................................... 4-8
5. FLXA402T sensor menu .......................................................................... 5-1
5.1 WASH ................................................................................................................. 5-2
5.2 Calibration ......................................................................................................... 5-2
5.3 Detail................................................................................................................... 5-2
5.4 Reset op. time.................................................................................................... 5-5
5.5 Setting ................................................................................................................ 5-5
5.6 Maintenance ...................................................................................................... 5-6
6. OPERATION .............................................................................................. 6-1
6.1 Turbidity measurement .................................................................................... 6-1
6.2 Air denoising ..................................................................................................... 6-1
6.3 When Sample Water Supply is Cut O ........................................................... 6-1
6.4 Automatic wash/Automatic calibration Operation........................................ 6-1
6.5 Stop and restarting operation ......................................................................... 6-1
7. CALIBRATION........................................................................................... 7-1
7.1 Turbidity calculation and calibration menu ................................................... 7-2
7.2 Turbidity standards........................................................................................... 7-3
7.3 Zero Calibration................................................................................................. 7-7
7.4 Slope calibration (Check tool) ......................................................................... 7-8
7.5 Slope calibration (Standard solution)............................................................. 7-8
7.6 Auto zero calibration ........................................................................................ 7-9
7.7 Zero shift correction ......................................................................................... 7-9
7.8 Sensitivity correction ....................................................................................... 7-9
7.9 2 points correction.......................................................................................... 7-10