York Survey Supply 343660 User manual

Code: 343660
Operating
Instructions
Ref:.. \operat98\instructions 18\343660.qxp 05-01-18 ©York Survey Supply Centre 2018
Datalogging
Light Meter
Prospect ouse, George Cayley Drive, Clifton Moor, York, England YO30 4XE
Tel: +44 (0) 1904 692723 Fax: +44 (0) 1904 690385
E-Mail: sales@yorksurvey.co.uk www.yorksurvey.co.uk
If the software still displays ‘OFFLINE’ there is no connection. To create
a connection, go to Com ort (C) icon (top menu line), click on the
drop down menu, select Com3 (M). The light meter image on the
software should now be displaying the same information as your light
meter. In some cases this com port will not work and Com4 (4) or
another may have to be selected, by selecting Other Com and
choosing the appropriate com port where the information is being
displayed on the light meter. This works exactly as it would with
Com3 (M) selected.
When the connection is made, on the top right side of the instrument
display you will see this symbol, and the word “Scout”. The
software will show an image of the instrument and display measured
values the same as the instrument.
Reading the Recall 99 Memory Function via the Software
On the top menu line, click on Mem(M). The software will now
download the memorised values and, when complete, auto open a new
window listing: - The Reading Number, Measured Value with the Time
and Date the measurement was memorised. This data can be saved or
printed.
General Recording Using the Software
Open the software and connect the instrument. On the top menu icons,
click on RealTime(R). On the drop down menu, click Run. A new
window will open “Sample Rate Setup”.
The sample rate can be set from 1 second up to 99 seconds. Click OK.
The software will now log results from the instrument. Note: Data is
captured based on the date and time of the connected computer
irrespective of the date and time settings on the instrument.
To stop recording, click on RealTime(R). On te drop down menu, click
Stop.
Alternatively, recordings can be made using the Stop/Go buttons
located at the top of the software.
Viewing Saved Data
When stopped, the data can be viewed as a graph and by moving the
mouse over the data line, the time and measured value will be
displayed at the bottom of the software.
The graph has a zoom facility. old the right button of te mouse near to
the area, ten drag and draw a box to cover the data line. On releasing
the mouse button, the zoom will auto focus to the selected area. This
can be actioned 3 times for in-depth analysis.
To undo the zoom, click this icon in the top menu.
Datalogging
Ensure date, time and sampling rate are set correctly. Sample rate can
be set from 1 second up to 99 seconds.
Auto power off has to be disabled.
To set the instrument to record mode press and hold REC/Setup
button (No 4) until, in the top right corner of the display, MEM01
appears. The instrument is now logging measured values as per set
sampling time. To stop datalogging, press and hold REC/Setup until
the MEM01 disappears. To take a second set of data, repeat until
MEM02 appears. Multiple or single data logging sets can be stored up
to a maximum of 16,000 measurements. If the datalogger is full the OL
will appear in the MEM sector.
To View Logged Data
Connect the instrument to the computer and open the software.
Click on Data Logger(D) icon on the menu bar at the top of the screen.
A loading window saying ‘Logging...’ will appear. This window will auto
close when the download is complete and open a new window giving a
line listing of data sets.
Double click on the data that you wish to view and a red line will appear
on the graph. The zoom facility and file save work as previously
detailed.
To delete the data files, press and hold REC/Setup, then turn the
instrument on. The instrument display will show “del” to indicate the
memory has had its data sets deleted.
©York Survey Supply Centre 2018Ref:.. \operat98\instructions 18\343660.qxp 05-01-18
Datalogging Light Meter
Introduction
This digital datalogging light meter is a precision instrument used to
measure luminance by Lux or Footcandles (FC). It meets the CIE
photopic spectral response standard. When not in use, the instrument
should be kept in the carry case and lens cap put onto the lens to
protect it from damage.
Features
> 4 Measuring Ranges Lux
> 4 Measuring Ranges Footcandles
> Silicon Photo Diode Sensor with a Spectral Response Filter Fully
Cosine Corrected
> igh accuracy and rapid response
> Data old, Peak old, Min/Max Memory, Relative Function
> 56mm ight Contrast LCD with 41 Segment Bargraph, Function
Indicators and Backlight
> Auto-Power Off after 15 minutes with User Disable
> USB output
> 99 Individual Record Memory independent of the Datalogger
> Datalogger Capacity of 16,000 Readings
> Accessories Included: Carry Case, Software and USB Connection
Cable
Specifications
> 4 Measuring Ranges for Lux: 0Lux to 400Lux, 4KLux, 40KLux,
400KLux
> 4 Measuring Ranges for Footcandles: 0FC to 40FC, 400FC, 4KFC,
40KFC
> Accuracy: ±3% rdg ±0.5% f.s. (<10,000Lux). ±4% rdg ±10d.
(>10,000Lux)
> Sampling rate of 1.5 times per second (selectable rate for data
logging)
> Operating temperature and humidity: 0°C ~ 40°C and
0%R ~ 80%R
> Storage conditions: -10°C ~ 50°C and 0%R ~ 70%R
> Power source: 1x 9V battery
> Length of Connection Lead for Light Sensor: 1m
> Light Sensor Dimensions: 115L x 80W x 20 (mm)
> Meter Dimensions: 230L x 80W x 50 (mm)
> Weight of Unit: 390g
Function Buttons
1. ower On/Off Button
3. UNITS Button
4. Backlight and LOAD Button
5. REC/SETU Button
6. EAK Button
7. HOLD Button
8. RANGE/A O Button
9. MIN/MAX Button
10. REL Button
Operating Instructions
ower On/Off, Auto ower Off and Display Backlight
To turn the meter on, press the yellow power button (No 1) on the
bottom of the unit. If auto power off is disabled, press the button again
to turn the meter off.
At first power on the instrument defaults to a 15 min “Auto Power OFF”.
To disable auto power off, press and hold REC/Setup button (No 5)
then press RANGE/A O button (No 8) until this symbol disappears
from the bottom left of the display. Repeat to re-enable the auto power
off.
To illuminate the backlight, press the LOAD button (No 4). Repeat to
turn off the backlight.
Selection of Measurements and Ranges in Lux or Footcandles
To change measurement units, press the UNITS button (No 2). This
changes the displayed values from Lux to Footcandles or vice versa.
1FC = 10.76Lux.
There are 4 measuring ranges for Lux and Footcandles. Press
RANGE/A O button (No 8) to scroll through the measuring ranges.
Should the display show “OL” then the range selected is too low for the
measured value. Select the next range up. Repeat until the instrument
displays a value. The selected range is displayed below the main
reading.
Connect the sensor to the instrument. Remove the black plastic cover
protecting the sensor and the instrument will display the measured
value.
To Set Time, Date and Datalogging Sampling Time
The time, date and sampling rate has to be set for recording and
storing measurements.
To enter this setup mode, press and hold down for 5 seconds the
REC/Setup button (No 5) and the UNITS button (No 3) simultaneously.
The word ‘Time’ and the hour digits should be flashing. You are now
in the setup mode.
The 4 keys (No’s 6, 7, 10 and 9) with arrows are used to change the
data and scroll through the setup menu. The our is the first part of the
program and is flashing. Use the up/down arrow buttons (No’s 7 and 9)
to increase or decrease the displayed hour. When the hour is correct,
press the right arrow button (No 10) which then moves the setup to
Minutes. Again using up/down arrow buttons (No’s 7 and 9) change the
displayed digits to the correct value. Repeat this process to set
Seconds, Sampling Time, Month, Date and Year. Note: The sampling
time is for the datalogger recording only. The true sampling rate of the
instrument is 1.5 times per second. To exit the setup program, press
and hold down for 5 seconds the REC/Setup button (No 5) and the
UNITS button (No 3) simultaneously.
Data Hold
This function allows the current reading to be frozen on the display.
Press HOLD button (No 7) to activate. An indicator in the bottom of the
display will show Manu Hold. Press HOLD again to cancel data hold
and the instrument will display current measured values.
Minimum/Maximum Values
This function is activated by pressing MIN/MAX button (No 9) and will
display the minimum and maximum values since the instrument was
switched on. The first value is the maximum measured value with a
Manu Max indicator shown in the bottom of the display. Press the
MIN/MAX button again. The displayed value with Manu Min indicator
shown in the bottom of the display. Press MIN/MAX again to cancel
data hold and the instrument will display current measured values.
eak Hold
This function’s purpose is to display only the highest or lowest
measured value and is activated by pressing the EAK button (No 6).
The first value is the highest measured value with a Manu max
indicator shown in the bottom of the display. Press the EAK button
again. The displayed value is the lowest measured value with Manu
min indicator shown in the bottom of the display. Press EAK again
to cancel peak measuring and the instrument will display current
measured values.
Relative Readings
This function is used to read measured values compared to a
memorised value. When a value is established, press the REL button
(No 10). The displayed values will then be deviations to the relative
stored value and shown as plus or negative value with a Manu
indicator shown in the bottom of the display. Press the REL button
again to exit this function and the instrument will display current
measured values.
Recall 99 Memory Function
The instrument has an independent memory separate to the
datalogger. It is used to memorise individual measured values for recall
without using a computer connection. To memorise a reading, press
REC/Setup button (No 5). In the top right hand corner of the display an
indicator will appear “MEM 01”. This confirms the measured value has
been memorised as record number 1. Press REC/Setup again and the
indicator will display as “MEM 02”. This confirms the measured value
has been memorised as record number 2. This can be repeated for up
to 99 readings.
To recall the readings from the memory, press and hold for 5 seconds
the LOAD button (No 4). Using the up and down buttons (No’s 7 and 9)
will allow you to scroll through each memorised value.
To clear the memory, press and hold for 5 seconds the LOAD and
REC/Setup buttons at the same time. MEM CL will appear in the top
right hand corner of the display whilst the memory is cleared. When the
memory is clear, the instrument will revert back to normal measuring
mode.
To read the “Recall 99 Memory” via the software. On the top menu line
click on Mem(M). The software will now download the memorised
values and, when complete, auto open a new window listing the
reading Number, Measured Value with the Date and Time the
measurement was memorised. This data can be saved or printed.
Low Battery Indication
When the battery power falls below the required voltage, a battery
symbol will appear on the display. To change the battery, slide the
back cover off, disconnect the expired battery and replace.
E ROM Memory Backup
The EPROM has a 2nd battery backup so no data is lost due to main
battery changes.
Recommended Illumination Values
Maintenance and Calibration
In order to ensure accurate readings, the lens should be cleaned using
a damp cloth to remove any dust or dirt. Do not store the instrument
where the temperatures are excessively low or high or the humidity is
excessively high. This should be considered especially when leaving
instruments in vehicles.
Re-calibration of this instrument will vary due to operating conditions
and regulations. It is recommended that the meter is re-calibrated at
least once every 12 months. York Survey Supply Centre offers a fully
traceable calibration service to national standards. Please call our
technical helpline for current prices: 01904 692723.
Installing the Software
Start Windows and insert the CD into the CD drive. Install Wizard
should detect and install the software.
If install wizard does not appear automatically, go to START at the
bottom of the desktop, then Run... Browse for Light Meter (D:) which
is the name of the software.
Click on Light Meter (D:). This will open up the files stored within the
CD. Open the SETU file and follow the instructions to install the
software.
Once this is complete you will see this icon,
appear on your desktop. The installation
was successful, but before you can use the
software you need to install the hardware
drivers. Do not remove the CD.
Installing the Hardware Drivers
You do not need to install the hardware drivers if you already have
installed it - the following Datalogging instruments all use the
same program:
SL-8851 USB Logging Sound Level Meter
SL-8852 Datalogging Sound Level Meter
LX-1309 USB Logging Lux Meter
With the CD for the software still in the CD drive, turn the instrument
on, then connect the meter to the computer using the USB cable
supplied.
A message will appear to show the computer has detected new
hardware in the USB port. Double click on this message. A new window
will appear asking you to install the hardware. On this window click ‘No,
not this time’, as you do not need to connect to the windows update.
Follow the instructions to install the hardware, but before you press
FINIS check that it reads ‘C 2101 USB Composite Device’ in the
window (this is the first hardware driver). Click FINISH.
The window will now ask to install the hardware; this is to install the
second driver. Install in the same way as you did for the previous driver.
In the last window it should show ‘C 2101 USB to UART Bridge
Controller’. Click FINIS and both hardware drivers should now be
installed.
To un-install the drivers (in the “add or remove programs”) you will not
find two separate drivers. You will find one driver called ‘C 210x USB
to UART Bridge Controller’ . This is because both drivers are saved
together, so by un-installing this you are, in fact, un-installing both the
drivers.
Using the Software
Connect the instrument to the computer with the USB cable. Open the
software by clicking on the icon.
If there is no connection the software will display OFFLINE in the
instrument image. Check the connections are correct and plugged in.
LOCATIONS Lux FC
OFFICE
Conference, Reception room 200 ~ 750 18 ~ 70
Clerical work 700 ~ 1,500 65 ~ 140
Typing drafting 1,000 ~ 2,000 93 ~ 186
FACTORY
Visual work at production line 300 ~ 750 28 ~ 70
Inspection work 750 ~ 1,500 70 ~ 140
Electronic parts assem ly line 1,500 ~ 3,000 140 ~ 279
Packing work, Entrance passage 150 ~ 300 14 ~ 28
HOTEL
Pu lic room, Cloakroom 100 ~ 200 9 ~ 18
Reception 200 ~ 500 18 ~ 47
Cashier 750 ~ 1,000 70 ~ 93
STORE
Indoors Stairs Corridor 150 ~ 200 14 ~ 18
Show window, Packing ta le 750 ~ 1,500 70 ~ 140
Forefront of show window 1,500 ~ 3,000 140 ~ 279
HOSPITAL
Sickroom, Warehouse 100 ~ 200 9 ~ 18
Medical Examination Room 300 ~ 750 28 ~ 70
Operating room, Emergency treatment 750 ~ 1,500 70 ~ 140
SCHOOL
Auditorium, Indoor Gymnasium 100 ~ 300 9 ~ 28
Classroom 200 ~ 750 18 ~ 70
La oratory, Li rary, Drafting room 500 ~ 1,500 47 ~ 140

Other York Survey Supply Measuring Instrument manuals

York Survey Supply 261640 User manual

York Survey Supply

York Survey Supply 261640 User manual

York Survey Supply 354910 User manual

York Survey Supply

York Survey Supply 354910 User manual

York Survey Supply 343090 User manual

York Survey Supply

York Survey Supply 343090 User manual

York Survey Supply 330550 User manual

York Survey Supply

York Survey Supply 330550 User manual

York Survey Supply 354420 User manual

York Survey Supply

York Survey Supply 354420 User manual

York Survey Supply 263090 User manual

York Survey Supply

York Survey Supply 263090 User manual

York Survey Supply 343000 User manual

York Survey Supply

York Survey Supply 343000 User manual

York Survey Supply 346010 User manual

York Survey Supply

York Survey Supply 346010 User manual

York Survey Supply 343300 User manual

York Survey Supply

York Survey Supply 343300 User manual

York Survey Supply 354820 User manual

York Survey Supply

York Survey Supply 354820 User manual

York Survey Supply 34302 User manual

York Survey Supply

York Survey Supply 34302 User manual

York Survey Supply 343010 User manual

York Survey Supply

York Survey Supply 343010 User manual

York Survey Supply 354920 User manual

York Survey Supply

York Survey Supply 354920 User manual

York Survey Supply 343680 User manual

York Survey Supply

York Survey Supply 343680 User manual

York Survey Supply Pro Sport 850 User manual

York Survey Supply

York Survey Supply Pro Sport 850 User manual

York Survey Supply 33235 User manual

York Survey Supply

York Survey Supply 33235 User manual

York Survey Supply 33245 User manual

York Survey Supply

York Survey Supply 33245 User manual

York Survey Supply 343520 User manual

York Survey Supply

York Survey Supply 343520 User manual

York Survey Supply 37135 User manual

York Survey Supply

York Survey Supply 37135 User manual

York Survey Supply 33065 User manual

York Survey Supply

York Survey Supply 33065 User manual

York Survey Supply S-Digit Mini+ 263040 User manual

York Survey Supply

York Survey Supply S-Digit Mini+ 263040 User manual

Popular Measuring Instrument manuals by other brands

Vivax Metrotech HL5000 Operation manuals

Vivax Metrotech

Vivax Metrotech HL5000 Operation manuals

Milwaukee MI405 user manual

Milwaukee

Milwaukee MI405 user manual

GASLOCK GASLEVEL installation guide

GASLOCK

GASLOCK GASLEVEL installation guide

F&F LE-02d CT quick start guide

F&F

F&F LE-02d CT quick start guide

Metrohm MIRA Cal DS Tutorial

Metrohm

Metrohm MIRA Cal DS Tutorial

Performance Tool W89719 owner's manual

Performance Tool

Performance Tool W89719 owner's manual

Anritsu MS9740A Operation manual

Anritsu

Anritsu MS9740A Operation manual

netvox DirekTronik R718KA user manual

netvox

netvox DirekTronik R718KA user manual

shelyak UVEX PF0079 user manual

shelyak

shelyak UVEX PF0079 user manual

Bosch UniversalDistance 50 Original instructions

Bosch

Bosch UniversalDistance 50 Original instructions

Spectrum Technologies Field Scout TCM 500 manual

Spectrum Technologies

Spectrum Technologies Field Scout TCM 500 manual

Lasertec CX2G instruction manual

Lasertec

Lasertec CX2G instruction manual

Proceq Zehntner ZG8000 Quick Start Guide & Product Certificates

Proceq

Proceq Zehntner ZG8000 Quick Start Guide & Product Certificates

GREISINGER GMH 3161-01 user manual

GREISINGER

GREISINGER GMH 3161-01 user manual

IKA HELISTAND manual

IKA

IKA HELISTAND manual

GDS GASMAX II Operation and maintenance manual

GDS

GDS GASMAX II Operation and maintenance manual

HiWeigh X2-BMI user manual

HiWeigh

HiWeigh X2-BMI user manual

geo-FENNEL Selection PRO FL 40-PowerCross Plus SP user manual

geo-FENNEL

geo-FENNEL Selection PRO FL 40-PowerCross Plus SP user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.