ZURN ZEG6003EV User manual

AquaSense®EVTM
ZEG Series
Automatic Sensor-Operated,
Battery-Powered Flushometer
Installation, Operation, Maintenance,
and Parts Manual
Patented and Patents Pending
LIMITED WARRANTY
All goods sold hereunder are warranted to be free from defects in material and factory workmanship for a period of three years from the date of
purchase. Decorative finishes warranted for one year. We will replace at no costs goods that prove defective provided we are notified in writing of
such defect and the goods are returned to us prepaid at Sanford, NC, with evidence that they have been properly maintained and used in
accordance with instructions. We shall not be responsible for any labor charges or any loss, injury or damages whatsoever, including incidental or
consequential damages. The sole and exclusive remedy shall be limited to the replacement of the defective goods. Before installation and use, the
purchasershalldeterminethesuitabilityoftheproductforhisintendeduseandthepurchaserassumesallriskandliabilitywhateverinconnectiontherewith.
Wherepermittedbylaw,theimpliedwarrantyofmerchantabilityisexpresslyexcluded.Iftheproductssoldhereunderare“consumerproducts,”the
impliedwarrantyofmerchantabilityislimitedtoaperiodofthreeyearsandshallbelimitedsolelytothereplacementofthedefectivegoods.Allweights
statedinourcatalogsandlistsareapproximateandarenotguaranteed.
ZEG6003EV Flushometer for
Z5798 Low Flow Urinal
Sensor-Operated, Battery-Powered
Urinal System

PRIOR TO INSTALLATION
Prior to installing the ZEG EcoVantage urinal flushometer
valve, install the items listed below:
• Urinal fixture
• Drain line
• Water supply line
IMPORTANT:
• All Plumbing is to be installed in accordance with applicable
codes and regulations.
• Water supply lines must be sized to provide an adequate
volume of water for each fixture.
• Flush all water lines prior to operation (See Step 2).
• Dirt and debris can cause flush valve to run continuously.
• Sensor units should not be located across from each other
or in close proximity to highly reflective surfaces.
The ZEG is designed to operate with 20 to 80 psi (138 to 552 kPa) of water pressure. Protect the chrome or special finish of
this flushometer. Do not use toothed tools to install or service the valve.
1.) Install stop valve assembly (A) using proper size supply
escutcheon and sweat solder adapter kit if applicable.
Thread sealing compounds should be used on male NPT
threads only.
3.) Prior to inserting the flush valve tailpiece (B) into stop valve
(A), be certain that the O-ring seal (C) is located in O-ring
seal groove at the end of the tailpiece and that the locking
nut (D) and locking snap ring (E) are located as shown.
Care should be taken not to damage the O-ring when
inserting the tailpiece into the stop valve. If lubrication is
needed, wetting the O-ring with water will be sufficient.
1
3
E
C
A
D
B
Before the supply water is turned on, be sure all stop valves
are closed off tight. The stop valves can be opened and closed
by using the adjusting screw (S) located at the center of the
stop valve cap (T). Stop valve adjustments can only be made
by using the adjusting screw (S). It is not necessary to re-
move the stop valve cap (T) when making adjustments. If
for any reason it becomes necessary to remove the stop
valve cap (T), be certain the water is shut off at the main
supplyvalve.
T
2
2.) When all stops are connected to the water supply and
water pressure is available, it is recommended that the
supply piping be flushed to remove dirt, metal chips, etc.,
from system.
A. Before the valve is installed, open each stop fully for a brief
time and catch the water in a two gallon or larger bucket.
For multiple installations, start with the stop valve closest
tothe watersupply and work toward the most remote valve.
B. Due to the small passages and orifices, it is not possible
to flush the piping through the low volume valve.
C. Once the lines are flushed, the valve can be installed.
S
A

7
4.) Insert the flush valve tailpiece (B) into the stop valve (A)
and hand tighten the lock nut (D) to the stop valve. Plumb
the entire unit.
5.) Determine the length of vacuum breaker tube (F)
required to join the flush valve and fixture spud. Cut the
vacuum breaker tube, if required, to this length. Assemble
the vacuum breaker tube assembly and spud nut assembly
to the flush valve and fixture spud.
6.) Hand tighten spud nut (G) and vacuum breaker tube nut
(H) to fixture and flush valve.Adjust the valve assembly for
plumb. Tighten fixture spud nut (G), vacuum breaker tube
nut (H) and lock nut (D) with a wrench. Do not turn water on
until batteries are inserted – see 7.
4
5
F
7.)ACTIVATION
Using the 5/64"Allen wrench supplied with the valve, remove
the two 6-32 button head screws holding the top cover in
place. Be sure to place the screws somewhere safe so
they don’t get lost.
Remove the top cover carefully and disconnect the solenoid
wire connector (J). The cover and battery case can now be
takento an appropiate work area to installtheAAcell batteries
provided.
Load the four heavy duty AA alkaline batteries following the
battery orientation guide on the battery box. Note that the coil
springs always contact the flat (negative) end of the battery.
Secure the battery case lid and gasket with screw provided.
Place the battery box into the valve housing, and reconnect
the solenoid wire connector (J). Secure the top cover to the
valve housing using the original screws.
Once this is completed, remove and discard the protective
label (L) from in front of the lens.
Normal valve operation will occur when the valve senses an
object (person). The LED will blink brightly once and then
dimly three times about a second apart. There will be a
pause and then a double blink. At that point the valve is
charged. When object leaves view of the sensor, valve will
activate. This six second sensing cycle will prevent the
valve from flushing needlessly when someone walks by.
If special circumstances require adjustment of the sensing
distance, see Appendix A - Changing Activation Distance.
B
D
A
L
D
H
G
J

8.) The ZEG urinal valve comes preset for both flow volume
and sensing distance. Each valve is operated at the fac-
tory using water to insure proper function before being
packed for shipment. The valve does not require water
pressure regulation for variation in water pressure within
the specified pressure range (20 psi to 80 psi) due to
internal pressure regulation.
Open the stop valve to wide open position using the ad-
justing screw (S) on the front of the stop valve. The water
volume will never have to be adjusted with the stop valve
as on some conventional urinal installations.
Install vandal resistant stop cover to complete the installa-
tion.
APPENDIX A
TO CHANGE ACTIVATION DISTANCE FOR
THE OBJECT LOCK SENSOR
The valve does not have to be disabled in any way to change
the detection range. Included with each valve is a Zurn “Magic
Magnet” that is used to intiate the auto-calibration mode.
Calibration is accomplished as follows.
1. Place the Zurn “Magic Magnet” (N) on the left hand side of
the flat top surface of the valve and move around until the LED
comes on.
2. The LED will blink once brightly, then nine times dimly, then
a final bright flash. A solid red light will appear in the sensor
eye. This means the electronics are ready to calibrate.
3. Remove the magnet and stand in front of sensor about 20”
away for the rest of the calibration. The calibration process
takes about 60 seconds. Do not move during this process.
4. When calibration is completed the solid red light will dis-
appear and the LED will blink twice brightly.
8
9.) LOWBATTERYWARNING
The LED will double flash every fifteen seconds when the
batteries reach a predetermined depletion level. There
are approximately five hundred more flushes available
once the LED starts flashing, at which time the valve will
stop operating. The LED will continue the flash sequence
until the batteries are replaced.
10
10.) FILTER
When accessing filter, be certain the stop valve is closed
at adjusting screw (S) Figure 8.
The filter is accessible from the bottom of the valve using a
slot head screw driver. The filter will come out with the
plug and can be cleaned with water and a soft brush.
When replacing be sure o-ring (M) is in place and undam-
aged.
M
Filter
Plug
S
N
11

TROUBLE SHOOTING GUIDE
Problem
Nolights are visiblein
the sensor eye.
Valve does notflush.
Lightflashesrandomly in
sensoreye and valve
doesnot flush
Lightsfollownormal
sequence,valve does
notflush.
Valve flushes butdoes
notshut off.
Leaks
LEDflashesevery15
seconds.
PossibleCause
Electronics fault.
Valve could besensinga
close by surface or reflections
fromsomeoppositesurface.
Lens could be dirty or dam-
aged.
Battery leads disconnected or
brokenwire.
Batterycorrosion has caused
loss of contact.
Sensornot detecting user.
Electronicsfault
Water not turned on.
Pluggedfilter.
Stopvalveclosed.
Wiringfault.
Solenoidmalfunction.
Diaphragmdamage orplugged
diaphragmorifice.
Solenoidmalfunction
Incorrectlyinstalled.
DamagedO-ringoninlettube.
Lowbattery voltage
Diagnosis
Runthroughreset procedure
below.
Hold a flat object at an acute
angleto the valveandsee if it
flushes(threering binder
workswell)
Visual inspection of batteries
and battery case.
Checkpluginsertion and wire
continuity.
Visual inspection
Holdhandatdifferentranges
in front of valve to see if it can
bedetected.
Runthroughreset procedure
below
Closestop,disconnectvalve,
crack stop to check for water.
Closestop,removeand
inspectfilter.
Check stop screw.
Checksolenoidleads and and
insertionofsolenoid plug.
Visual inspection.
Check for leak location.
Check for cuts.
Solution
Reducesensing range
Eliminatereflective
object.
Replaceifobstructed or
damaged
Reinsertplugor repair
leads
Replacebatteriesand
cleanor replace battery
box
Resetsensorrange
(AppendixA)
Findsource valveand
turnon.
Cleanandreinstallfilter
Openstopvalve
Reinsertplugor repair
wiring
CleanP6900-SRK
components. See parts
list
Replacediaphragm
CleanP6900-SRK
components. See parts
list
Removeand reinstall.
Pay close attention to
alignmentofinlet tube
and outlet boss with
matingpieces.
Replace O-ring
Replacebatteries
Electronics Reset Procedure: Disconnect solenoid wires ((J) Figure 7). Remove batteries from battery case. Allow electron-
ics to set without batteries installed for a minimum of 4 minutes. Place batteries in battery case (watch for proper polarity).
Secure battery case lid and gasket with screw provided. Reconnect solenoid wire connector. Place battery case back in valve
housing. Place valve cover back on valve housing, carefully tuck wires into housing. Secure cover with two screws provided.

Solenoid and Valve Kits Product No.
1/8 GPF Valve Bobdy Replacement Kit, Items 9-16 PEG6003-VBA
1/8 GPF Solenoid Repair Kit, Items 9-11 PEG6003-SRK
1/8 GPF Solenoid Replacement Kit, Items 7-8 PEG6003-M
1/8 GPF Filter Replacement Kit, Items 13-16 PEG6003-FA
Parts Identification
1. Cover
2. Cover Screw
3. Electronics Assembly
4. Lens Assembly
5. Battery Tray
6. Housing
7. Solenoid
8. Screw
9. Spring
10. Plunger
11. Diaphragm
12. Valve Body
13. Filter
14. O-Ring
15. Filter Cap
16. O-Ring
17. Tailpiece O-Ring
18. Vacuum Breaker Insert
19. Duckbill
20. Vacuum Breaker Tube
21. Vacuum Breaker Tube Nut
22. 3/4" Spud Nut
23. 3/4" Spud Friction Washer
24. 3/4" Spud Sleeve
25. 3/4" Spud Escutcheon
26. Magic Magnet
27. Setscrew for Control Stop Cover
28. Vandal-Resistant Control Stop Cover
29. Stop Cap
30. Adjusting Screw
31. Guide Holder
32. Piston Guide
33. Guide O-Ring
34. Piston
35. Piston Seal
36. Stop Body
37. Sweat Solder Adapter
38. Supply Cover Tube
39. Setscrew for Cast Wall Escutcheon
40. Cast Wall Escutcheon
41. Hex Wrench
42. Cover O-ring
®
Control Stop Repair Kit and Parts Product No.
Control Stop Repair Kit for 1” and 3/4”, PEG6000-D-SD
Includes Items 30-35
3/4” Control Stop Replacement, Includes PEG6003-C-SD-
Items 29-36 CP
Seal Seat for 1” and 3/4”, Includes Item 35 P6000-D42
Sweat Solder Adapter, Includes Item 37 P6003-YBA
Sweat Kit and Eschutcheon, Includes Items 37-41 P6003-YBYC
Flush Connections and Spud Coupling Kits Product No.
3/4" Flush Connection and Spud Coupling, P6003-H
Items 22-25
Vacuum Breaker Repair Kit, Items 18-19 P6000-B
ZEG6003EV 1/8th Gallon Urinal Parts Breakdown
FV408 06/14/07
Lid and Housing Kits Product No.
1/8 GPF Cover Replacement Kit P, Item 1 PEG6003-L-1
1/8 GPF Cover Replacement Kit S, Item 1 PEG6003-L-2
1/8 GPF Housing Replacement Kit P, Items 6-17 PEG6003-HSA-1
1/8 GPF Housing Replacement Kit S, Items 6-17 PEG6003-HSA-2
1/8 GPF Lid O-ring, Item 42 PEG6003-CVR-
ORING
Replacement Parts and Repair Kits Product No.
1/8 GPF Cover screw, Item 2 PEG6003-CVR-S
1/8 GPF Electronics Assembly, Item 3 PEG6003-EL
1/8 GPF Lens Assembly, Item 4 PEG6003-SC
1/8 GPF Battery Tray, Item 5 PEG6003-BD
WaterWise Technologies International www.waterwisetech.net
P.O. Box 18959 · Asheville, NC 28804 828.252.8144
Distributed by:
Table of contents
Other ZURN Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Leica
Leica RANGEMASTER 1200 scan instructions

FORENSICS DETECTORS
FORENSICS DETECTORS FD-311 user manual

S+S Regeltechnik
S+S Regeltechnik THERMASGARD RGTM1 Series Operating Instructions, Mounting & Installation

L&T Electrical & Automation
L&T Electrical & Automation 1 Phase DIN Rail Meter user manual

Eaton
Eaton Power Xpert PXM 4000 User and installation manual

Center
Center 221 instruction manual