3Disc FireCR Dental User manual

User Manual.Rx Only
Dental Computed Radiography Reader
Intended Use: The FireCR Dental Imaging System is indicated for capture, digitization and
processing of intra oral x-ray images stored on imaging plate recording media.
Doc No. : TM-801-EN
Rev 1.2.3 Jul 2018
Part No. : CR-FPM-11-001-EN
3DISC, FireCR, Quantor and the 3D Cube are trademarks of 3D Imaging & Simulations Corp,
South Korea, and its affiliates. All other trademarks are held by their respective owners and
are used in an editorial fashion with no intention of infringement. The data in this publication
are for illustration purposes only and do not necessarily represent standards or specifications,
which must be met by 3D Imaging & Simulations Corp. All information contained herein is
intended for guidance purposes only, and characteristics of the products and services described
in this publication can be changed at any time without notice. Products and services may not
be available in your local area. Please contact your local sales representative for availability
information. 3D Imaging & Simulations Corp. strives to provide as accurate information as
possible, but shall not be responsible for any typographical error.
© Copyright 2014 3D Imaging & Simulations Corp. All rights reserved.

User Manual
FireCR Dental TM-801-EN
2
Contact
3D Imaging & Simulations Corp.
Bldg.1, 48, Yuseong-daero 1184 beon-gil,
Yuseong-gu, Daejeon, 34109 Korea
Tel : 82-42-931-2100
Fax : 82-42-931-2299
Website : www.3DISC.com
E-mail : info@3DISC.com
3DISC Americas
22560 Glenn Dr, Suite 116
Sterling, VA 20164 USA
Tel : 1-703-430-6080
E-mail : info@3DISC.com
3DISC Europe
Gydevang, 39-41, 3450 Alleroed, Denmark
Tel : 45-88-276-650
E-mail : info@3DISC.com

User Manual
FireCR Dental TM-801-EN
3
The device complies with DHHS Radiation Safety Standards in effect as of the date of
manufacture.
The device complies with Part 15 of the FCC Rules. Operation is subject to the condition
that this device does not cause harmful interference.
NOTE : This equipment has been tested and found to comply with the limits for a Class B Digital
Device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates,
uses and can radiate radio frequency energy and, if not installed and used in accordance with
the instruction, may cause harmful interference to radio communication. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment causes
harmful interference to radio or television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the interference by one or more of
the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver
- Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.

User Manual
FireCR Dental TM-801-EN
4
Warnings and Used Symbols
To ensure the safety of patients, staff and other persons, any changes to software and hardware
delivered by 3D Imaging & Simulations Corp. may only be made with prior written permission
from 3D Imaging & Simulations Corp.
Please read the respective manuals of the connected software, such as acquisition and
diagnostic software, before starting to use the system.
The following symbols will be used throughout this manual:
DANGER
General mandatory action manual.
The functionality of the system can be destroyed in the case of
incorrect use.
If unauthorized changes have been made to delivered system and
accessories, the warranty by 3D Imaging & Simulations Corp.
becomes void. 3D Imaging & Simulations Corp. will not accept
any responsibility or liability for the improper functioning of the
product in such a case.
WARNING
The functionality of the system can be limited in the case of
incorrect use. Hints that require special attention.
NOTE
Notes represent information that is important to know but which do
not affect the functionality of the system.
DANGER
General prohibition indication.
The functionality of the system can be destroyed in the case of
incorrect use.
If unauthorized changes have been made to delivered system and
accessories, the warranty by 3D Imaging & Simulations Corp.
becomes void. 3D Imaging & Simulations Corp. will not accept
any responsibility or liability for the improper functioning of the
product in such a case.

User Manual
FireCR Dental TM-801-EN
5
General Safety Guidelines
All the safety and operating instructions should be read carefully before this device is operated.
This device has been designed and tested to meet strict safety requirements applicable to
medical equipment, and has been supplied in a safe condition. To ensure personnel and
patient safety, the device shall be operated and serviced in compliance with all procedures,
warnings and precautions during all phases of operation and service of this device. Failure to
comply with safety guidelines may result in injury to service personnel, operator, or patient.
3D Imaging & Simulations Corp. assumes no liability for failure to comply.
If this device is not used as specified, the protection provided by the device could be impaired.
This device must be used in normal conditions only.
Installation, service and operation of this device should only be undertaken by qualified and
trained personnel. The operator should study instructions and precautions carefully here and
throughout the manual before starting to use the device.
There are no user serviceable parts inside this device. The device should only be opened and
serviced by qualified service personnel. Failure to heed this warning may result in injury to
service personnel or damage to equipment, and void any and all warranties. If there is a
service problem, please contact 3D Imaging & Simulations Corp. or an authorized dealer.
Do not spill liquids on the device, and never operate the device in a wet environment.
Keep the device away from radiators and heat sources.
Use the device only with accessories supplied with this device.
This device is intended to be grounded. Plug power cord into properly grounded electrical
outlets. This cord is equipped with three-prong plugs to help ensure proper grounding.
This device contains static sensitive components. Proper static handling procedures and
equipment must be used when servicing this device.

User Manual
FireCR Dental TM-801-EN
6
Do not look inside of the device.
If any of the following conditions occur, unplug the device from the electrical outlet and contact
authorized service personnel.
The power cord or power adapter is damaged.
An object has fallen into the device.
The device has been exposed to water.
The device has been dropped or damaged.
The device does not operate correctly when the operating instructions are followed.
Federal law restricts this device to sale by or on the order or a physician
Intended Use
This device is a Dental Computed Radiography System and intended for use in producing digital
X-Ray images for dental radiography purposes. It comprises of reader, reusable imaging plate
and workstation software. It scans X-Ray exposed imaging plate and produces X-Ray image
in digital form. Then, digital image is transferred to workstation for further processing and
routing. This device is intended to be operated in a radiological environment by qualified staff.
WARNING
Pay particular attention to use, care, maintenance, and infection
control of Imaging Plate, Chapter 4.3.

User Manual
FireCR Dental TM-801-EN
7
Index of contents
Chapter 1. Introduction.................................................................... 9
Chapter 2. Unpacking.................................................................... 10
2.1. Inspection for Damage ................................................................... 10
2.2. Identify the Components................................................................. 10
Chapter 3. Setting Up.................................................................... 13
3.1. Positioning...................................................................................... 13
3.2. Identify Important Features............................................................. 14
3.2.1. Reader Connection Panel ........................................................ 14
3.2.2. Touch Display Panel................................................................. 15
3.3. Computer Requirements ................................................................ 21
3.3.1. Recommended Configuration................................................... 21
3.3.2. Minimum Requirement ............................................................. 21
3.4. Installation of Acquisition and Diagnostic Software ........................ 21
3.5. Connect the Cable and Power Cord............................................... 22
3.5.1. Connecting the USB Interface Cable........................................ 22
3.5.2. Connecting the Ethernet Cable................................................. 23
3.5.3. Connecting the Power Cord ..................................................... 24
3.5.4. Installation Report..................................................................... 25
Chapter 4. Operating..................................................................... 26
4.1. System Specifications .................................................................... 26
4.2. Operation Conditions...................................................................... 27
4.3. Use, Care, Maintenance and Infection Control............................... 28
4.3.1. Use Protective Cover................................................................ 28
4.3.2. Use Hygienic Bag..................................................................... 29
4.3.3. Cleaning of the Tray ................................................................. 30
4.4. Operating Instructions .................................................................... 30
4.4.1. Turn on the Reader................................................................... 30
4.4.2. Turn on the Computer............................................................... 30
4.4.3. X-ray exposure on imaging plate.............................................. 31

User Manual
FireCR Dental TM-801-EN
8
4.4.4. Imaging Plate Placement and Removal.................................... 31
4.4.5. Getting a scanned image.......................................................... 32
4.4.6. Circuit Functions....................................................................... 33
Chapter 5. Symbols....................................................................... 35
5.1. Manufacturer’s Declaration –Electromagnetic Emission ............ 36
5.2. Manufacturer’s Declaration - Electromagnetic Immunity ................ 37
5.3. Guidance and Manufacturer ’s Declaration –Electromagnetic
Immunity..................................................................................................... 40
5.4. Laser Safety Statement.................................................................. 41
Chapter 6. Technical Assistance.................................................... 42

User Manual
FireCR Dental TM-801-EN
9
Chapter 1. Introduction
Dear Customer
Thank you for choosing the 3DISC Imaging FireCR Dental Reader as your new dental solution.
The advanced CR technology of the FireCR Dental Reader enables you to produce high-
quality digital images for diagnosing the patients in your facility. The reader can be used as a
central reader, which distributes images throughout your facility, or as an exam-room based
solution. The reader is DICOM 3.0 compatible with existing systems and uses a full range of
low-cost, reusable bitewings and intraoral imaging plates. The design features a built-in erase
function and a color touch-screen LCD panel without physical push buttons for seamless device
operation.
Please read and follow the instructions given in this ‘User Manual’ carefully prior to using the
FireCR Dental Reader and keep this manual within reach for future reference.
The purpose of this manual is to direct you through the main functions and interfaces of the
FireCR Dental Reader. You will be guided through the procedures of ‘Unpacking’, ‘Setting Up’
and ‘Operating’ the FireCR Dental Reader. You can also learn about ‘Symbols’, ‘Warranty and
Repair Service’ and ‘Technical Assistance’. It is important to observe all safety information to
prevent potential personal injury or material damage.
Please complete and submit the ‘Installation Report’ (Appendix 1) when installing the device.

User Manual
FireCR Dental TM-801-EN
10
Chapter 2. Unpacking
2.1. Inspection for Damage
FireCR Dental Reader is shipped in a custom designed container to protect the reader from
external shock. Before unpacking the reader, inspect the shipping container for damage. In
case the container is damaged, notify the shipper immediately.
2.2. Identify the Components
Open the shipping container and identify each of these components.
Common items
Part No.
Item
CR-FP-11-001
FireCR Dental Reader
CR-FPA-01-002
Power Adapter
CR-FPA-02-001
USB 2.0 Interface Cable
CR-FPA-02-002
RJ45 CAT.5E FTP Cable 2M(Cross type)
CR-FPA-03-00X
Power Cord
CR-FPM-11-001
FireCR Dental User Manual
CR-PKM-11-004
IP Storage Box
Medical application items
Part No.
Item
CR-FP-12-010
Imaging Plate Starter Kit –Dental
Contains : 2 x IP size 0 and 4 x IP size2
CR-FPA-15-001
Size 0 IP Hygienic Bag
Box of 100 pcs
CR-FPA-15-003
Size 2 IP Hygienic Bag
Box of 100 pcs
Veterinary application items
Part No.
Item
CR-FP-12-030
Imaging Plate Starter Kit –Veterinary
Contains : 4 x IP size 2 and 1 x IP size 4c
CR-FPA-15-003
Size 2 IP Hygienic Bag
Box of 100 pcs
CR-FPA-15-005
Size 4c IP Hygienic Bag(option)
Box of 100 pcs

User Manual
FireCR Dental TM-801-EN
11
Optional items
Part No.
Item
CR-FP-12-012
Size 0 Imaging Plate Kit
Contains : 4 x IP size 0
CR-FP-12-013
Size 1 Imaging Plate Kit
Contains : 4 x IP size 1
CR-FP-12-014
Size 2 Imaging Plate Kit
Contains : 4 x IP size 2
CR-FP-12-015
Size 3 Imaging Plate Kit
Contains : 4 x IP size 3
CR-FP-12-016
Size 4c Imaging Plate Kit
Contains : 1 x IP size 4c
CR-FP-12-020
Hygienic Bags Size 0
Box of 100 pcs
CR-FP-12-021
Hygienic Bags Size 1
Box of 100 pcs
CR-FP-12-022
Hygienic Bags Size 2
Box of 100 pcs
CR-FP-12-023
Hygienic Bags Size 2
Box of 300 pcs
CR-FP-12-024
Hygienic Bags Size 3
Box of 100 pcs
CR-FP-12-025
Hygienic Bags Size 4c
Box of 100 pcs
CR-FP-12-035
Protective Cover Size 0
Box of 100 pcs
CR-FP-12-036
Protective Cover Size 1
Box of 100 pcs
CR-FP-12-037
Protective Cover Size 2
Box of 100 pcs
CR-FP-12-038
Protective Cover Size 2
Box of 300 pcs
CR-FP-12-039
Protective Cover Size 3
Box of 100 pcs
CR-FP-12-040
Protective Cover Size 4c
Box of 100 pcs
CR-FP-51-010
FireID Kit (RFID Reader, mini USB cable)
WARNING
If the FireCR Dental needs to be returned to manufacturer or one
of its representatives, the reader must be repacked in the original
container with all accessories.

User Manual
FireCR Dental TM-801-EN
12
WARNING
Use of Power Cord;
Type SJT or SVT, min. 18AWG, 3-Conductor, VW-1 125V, min 10A
(or 250V, 10A). Max 3.0m long; one end with Hospital Grade
Type, NEMA 5-15P for 125V or NEMA 6-15P for 250V. Other
end with appliance coupler. “CAUTION Grounding reliability can
only be achieved when the equipment is connected to an
equipment receptacle marked “Hospital Only”or “Hospital Grade”.
For connection to a supply not located in the USA, make sure the
power cord meets the requirements for your area.
WARNING
Improper disposal of this product may result in environmental
contamination. When disposing of this equipment, contact 3D
Imaging & Simulations Corp.’s representative or related
government agencies. Do not dispose of any part of this
equipment without consulting a 3D Imaging & Simulations Corp.
representative first.
3D Imaging & Simulations Corp. does not assume any
responsibility for damage resulting from disposal of this equipment
without consulting 3D imaging & Simulations Corp.
NOTE
AC/DC Adapter
Manufacturer : Bridge Power corp.
Model : BPM050X24XXX
This adapter meets the requirements of IEC60601-1.
WARNING
Use only devices meeting the requirements of IEC60950-1 or
IEC60601-1 when connecting to the FireCR Dental via the USB
port.

User Manual
FireCR Dental TM-801-EN
13
Chapter 3. Setting Up
3.1. Positioning
The reader must be placed on a rigid and flat desk or tabletop with at least 5 cm (2 inches) free
space on both of the sides, 10 cm (4 inches) on rear side and 15 cm (6 inches) on front side for
imaging plate insertion. Its space requirements are shown below.
Figure 1. Space Requirements (Top View)
DANGER
Never place the reader on the floor.
Install in a location that is level and stable. Installation in an
unsuitable location can cause accidents, or deterioration in image
quality.
Allow a minimum free space
of 10 cm (4 inches) on the
backside to allow the power
switch, power cord and
interface cable to be
reached by hand at all times.
Allow a minimum free space
of 15 cm (6 inches) on the
front side for imaging plate
insertion and removal.
Front Side
Backside
WARNING
Unsuitable Installation Sites
Locations with excessive humidity or dust
Locations subject to high temperature
Locations subject to shaking or vibration
Locations exposed to considerable electrical or magnetic
noise, or other forms of electromagnetic energy
Locations with poor heat radiation

User Manual
FireCR Dental TM-801-EN
14
3.2. Identify Important Features
Look over the reader and features shown in this section. User will need to know where these
features are when user operates the reader in later chapters.
3.2.1. Reader Connection Panel
Figure 2. Reader Connection Panel
WARNING
Sliding of the reader may result in internal damage or
misalignment of the optics.
External vibration or shock during scanning may affect image
quality. The reader must be placed on a rigid, flat and reinforced
desk or tabletop.
DANGER
Do not place anything on top of the reader.
WARNING
This equipment may be interfered with or may interfere with
electromagnetic or other interferences.
Assure a distance of minimum 1.0m between reader and
neighboring equipment.
USB 2.0
Port
Power Inlet
Power Switch
100Mbps Ethernet port

User Manual
FireCR Dental TM-801-EN
15
3.2.2. Touch Display Panel
Screen displays the status of the reader and control of the reader can be done through touch
display panel.
Figure 3. Touch Display Panel
Display
Status
Remark
Booting screen
When the reader is turned on,
booting screen is displayed during
system initialization.
Touch display panel

User Manual
FireCR Dental TM-801-EN
16
Sleeping
Auto Sleep mode –IP tray moves
back into the reader and door will
be closed.
Description of
icons
at status bar
, ~ : Size of
imaging plate.
: High resolution, standard
resolution. Temporarily toggles
when it is touched.
: USB, Ethernet, no
cable connection respectively.
Disconnection
USB cable is connected but no
application program is running on
computer.
Ready for
scan, but no IP
Waiting for IP to be placed. Scan
will not start with an empty tray.

User Manual
FireCR Dental TM-801-EN
17
Ready for
Scan
IP size is recognized and the
reader is ready.
Scanning
Scanning is in progress.
Erasing
Erasing is in progress.
System error
Unexpected system error.
Contact technical support.

User Manual
FireCR Dental TM-801-EN
18
Setting mode
When button is selected from
main screen, Setting mode is
displayed.
Scanning
resolution
setting
Choose HD or SD
HD: High Definition
SD: Standard Definition
System setting
System setting for Auto Sleep and
Auto Trim
Auto Sleep
setting
Choosing time duration until sleep
mode.

User Manual
FireCR Dental TM-801-EN
19
Auto Trim
setting
Choose Off or On
Auto Trim Off : Scan corresponding
area of IP detected by the reader
Auto trim on : Scan whole tray area
and crop the image automatically
Language
Choose language
Network
Network setting menu
DHCP setting
Choose DHCP mode for automatic
network setting.
When this setting is changed, the
system will restart automatically.

User Manual
FireCR Dental TM-801-EN
20
IPAddress
setting
Type in IP address manually.
When this setting is changed, the
system will restart automatically.
Subnet mask
setting
Type in Subnet mask manually.
When this setting is changed, the
system will restart automatically.
Default
Gateway
Setting
Type in Gateway address manually.
When this setting is changed, the
system will restart automatically.
Other manuals for FireCR Dental
1
Table of contents
Other 3Disc Dental Equipment manuals
Popular Dental Equipment manuals by other brands

Aseptico
Aseptico AEU-7000L-70V Operation and maintenance instruction manual

Amanngirrbach
Amanngirrbach Ceramill Base Important note on upgrading

Midwest
Midwest RDH Repair Procedure

KaVo
KaVo Globus 1078 S Instructions for use

Carestream DENTAL
Carestream DENTAL CS 1500 user guide

Dentsply Sirona
Dentsply Sirona CEREC Primemill operating instructions

SILMET
SILMET SECURApplier Instructions for use

KaVo
KaVo Everest scan Base Camp Instructions for use

Wassermann
Wassermann HSS-88 Replacement manual

DEXIS
DEXIS FOCUS user manual

Guilin Woodpecker Medical Instrument
Guilin Woodpecker Medical Instrument LED.Q user manual

3M ESPE
3M ESPE Pentamix Lite operating instructions