11
4.8 Technical data
WARNING - Do not exceed the total permissible weight and the maximum person weight, as
this may lead to defects or an overload the V-MAX2.
iNOTE - Observe the technical data.
Technology V-MAX2
Lithium-ion (ALPHA box) Lead gel (GAMMA box)
Drive dimensions Refer to chapter "Dimensioned sketch"
Own weight
Drive unit incl. handles = 9.3 kg
Battery pack weight = 4.2 kg
Interface / power cable weight = 0.3 kg
Battery bag weight = 0.2 kg
Total weight V-MAX2 = 14.0 kg
Drive unit incl. handles = 9.3 kg
Battery pack weight = 10.2 kg
Interface / power cable weight = 0.3 kg
Battery bag weight = 0.2 kg
Total weight V-MAX2 = 20.0 kg
Total permissible weight 210 kg (wheelchair + person + V-MAX2)1, 4
Widening of wheelchair by attachment Min. 3.6 7
Limit values during construction From seat width 38 cm ( max. wheel camber 0°)
Possible wheel sizes in inches 20"; 24"
Speed forward I backward 6 km I 3 km
Range per battery charge for a total
weight of 130 kg Up to 26 km 2Up to 19 km 2
Range per battery charge for a total
weight of 131 to 210 kg Up to 13.6 km 2Up to 10.5 km 2
Braking / rated slope for max. person
weight
20 % (11,3°) with 96 kg 3, 4, 5 20 % (11,3°) with 90 kg 3, 4, 5
12 % (6,8°) with 176 kg 3, 4, 5 12 % (6,8°) with 170 kg 3, 4, 5
Batteries 28.8 V / 16 Ah 2 x 12 V /15 Ah
Direct current motor 24 V / 280 W
Noise information The A-weighted sound pressure level is typically below 56 dB(A)
Switching/actuating forces Rocker forward/ backward: 3N - 4N
Static stability downhill Observe the information in the wheelchair manufacturer's user manual.
Static stability sideways Observe the information in the wheelchair manufacturer's user manual.
Maximum surmountable obstacle height Observe the information in the wheelchair manufacturer's instructions for use and the
chapter "Overcoming obstacles".
Storing/operating temperature: Device -40 °C to +65 °C / -25 °C to +50 °C
Storing/operating temperature: Battery
pack -20 °C to +65 °C 6 / -25 °C to +50 °C 6
Charging temperature: Battery pack 0 °C to +40 °C
Vibration information The weighted RMS value of acceleration, to which the upper body is subjected, is typically
below 2.5 m/s2.
Degrees of protection, IP class IPX4
Adhesive label with serial number Back of device
(1) Observe the maximum load according to the wheelchair manufacturer's information.
(2) According to ISO 7176-4, the distance range depends, among other factors, on the weight of the transported person, the ground, the terrain prole, the
ambient temperature, etc.
(3) The rated slope and braking ability also depend on the surface, terrain prole, ambient temperature etc.
(4) The person weight on the type plate and in the technical data were calculated using an assumed wheelchair weight of 20 kg.
(5) The person weight on the type plate was calculated using device type B with lead gel battery pack [GAMMA BOX].
(6) At temperatures below 0 °C the performance of the battery pack will decrease; therefore we recommend a storage temperature of at least 0 °C.
(7) The attachment width can vary further depending on the wheelchair model.