Abiomed AB5000 User manual

0015-7309, RevH
AB5000™ System Installation / Preventative Maintenance Procedure Page 1 of 8
ABIOMED, Inc. Proprietary and Confidential SOP-QA1000-T2, Rev. B
AB5000™ System Console S/N
AB5000™ AB Cart OR
iPulse Cart S/N
Hand Pump S/N
Time Meter Hrs.
Date
Tested By
All service and maintenance, which involves disassembly of the AB5000TM Console, must be performed by factory
trained and certified service personnel. Replacement parts obtained from sources other than ABIOMED/IMPELLA
must not be used. Possible Electrostatic discharge precautions must be observed. Prior to servicing, check with
ABIOMED/IMPELLA technical support (USA 800-422-8666) (EUROPE +49-241 8860301) for available FCN
releases and Service Manual updates.
REQUIRED EQUIPMENT
Item Specification (or ABIOMED P/N for Specialized
Tools)
Angled mirror Miltex Mirror P/N 67-696/3 and Mirror Handle P/N 011-
26641 or equivalent
Lint free swab with 6“ plastic handle Abiomed PN 7000-0030 (Lint free swab with 6” plastic
handle) or equivalent
Hydraulic Load Simulator ABIOMED/IMPELLA P/N 0005-0070
Digital Voltmeter Fluke 77 or Equivalent
Laptop Computer Running Windows 2000 or Later, Serial Port
Diagnostic Download Cable ABIOMED/IMPELLA P/N 0015-9612
Pump ID Connector ABIOMED/IMPELLA P/N 0055-0737
Pressure/Vacuum Gauge -200mmHg to +500mmHg min (Damping set to 1 sec.)
Safety Analyzer (EUROPE ONLY) IEC601-1 Tester
[ ] 1. DECONTAMINATION
Prior to servicing, the AB5000TM Console must be decontaminated. Unless otherwise known, the Console
should be considered contaminated and universal safety precautions are to be taken. Gloves, masks, eye
protection, and protective garments are to be worn as required in the handling and decontamination of the
Console.
[ ] 2. UNCRATING
Uncrate the Console and Cart. Place the Hand Pump into the mounting bracket as required.
[ ] 3. EXTERNAL INSPECTION
Perform a thorough exterior inspection of the Console and Cart. Inspect for:
[ ] Turret rotation, panel damage, mechanical damage and any loose hardware.
[ ] 4. INTERNAL INSPECTION
Remove Console from Cart. Remove the Console panels. Perform a thorough interior visual inspection.
(Inspect fittings, tubings, quick-connect tubing (insertion depth), cable connections, hardware).
[ ] Perform a general inspection of the compressor subassemblies, including but not restricted to,
black debris in the rear bearing, tubing wear, mesh wear and electrical connections as per procedure
0015-7007.
[ ] 5. PERFORM REQUIRED FIELD CHANGES (FCNS) – CONTACT ABIOMED FOR CURRENT FCN
STATUS
0015-7309, RevH
AB5000™ System Installation / Preventative Maintenance Procedure Page 2 of 8
ABIOMED, Inc. Proprietary and Confidential SOP-QA1000-T2, Rev. B
[ ] 6. CHECK TIME METER HOURS. REPLACE PARTS AS REQUIRED PER MAINTENANCE SCHEDULE
0015-3050 Humphrey® Valve Assembly Every 5,000 Hours
0015-3000 Proportional Valve Assembly Every 5,000 Hours
0015-3040 Compressor Assembly Every 5,000 Hours
0015-3060 Battery Module Assembly Every 2 Years
0015-3061P Battery Module Assembly A1 Every 2 Years
0015-7309 Perform PM Procedure Annually
[ ] 7. CONSOLE LUGGAGE HANDLE
[ ] Check Console’s luggage handle for smooth operation. Verify proper latching in raised
position.
[ ] Check the console’s luggage handle for chips, cracks, or signs of abuse. Replace if
necessary.
[ ] If available, check Bed Plate assembly for socket head cap crew, replace with flat
countersunk head cap screw (PN 2850-5127).
[ ] 8. HAND PUMP
[ ] Inspect the Hand Pump for any physical damage or loose hardware. Verify Piston Rod
Handle is tight.
Move the left and right shuttle switches to BVS Pump mode.
[ ] Verify shuttle pins engage securely into ball plungers.
[ ] Cycle the Hand Pump Handle multiple times. Check for airflow at the left and right barbs.
Move the left and right shuttle switches to Ventricle Pump mode.
[ ] Verify shuttle pins engage securely into ball plungers.
[ ] Cycle the Hand Pump Handle multiple times. Check for airflow at the left and right barbs.
[ ] Inspect Console’s Hand Pump mount for any physical damage. Place the Hand Pump
back into the mounting bracket.
[ ] 9. POWER-UP SELF TEST SEQUENCE
Plug the Console IN. Turn the Power Switch ON. Verify:
[ ] All Keypad LED indicators illuminate and that the audible alarm sounds.
[ ] An air pulse at the left barb, then an air pulse at the right barb.
[ ] The self test ends within twenty (20) seconds and then the LCD displays the message “LEFT
SYSTEM READY FOR USE” & “RIGHT SYSTEM READY FOR USE”.
[ ] 10. FAN CHECK
[ ] Verify the operation of the system cooling fan (access at the Diagnostic Access Panel).
[ ] 11. POWER SUPPLY
Setup: Console plugged in, batteries connected, console power on.
[ ] Measure voltage output in the Power Supply assembly. Verify that voltage is 32 ± 0.1 VDC.
0015-7309, RevH
AB5000™ System Installation / Preventative Maintenance Procedure Page 3 of 8
ABIOMED, Inc. Proprietary and Confidential SOP-QA1000-T2, Rev. B
[ ] 12. BATTERY CHARGER
Setup: Console plugged in, batteries connected, console power on.
[ ] Measure charger voltage. Verify that charger voltage is between 26.6 – 27.6 VDC (this assumes
that the batteries are fully charged).
[ ] Disconnect battery connector.
[ ] Measure charger voltage. Verify that the charger voltage is between 30-33 VDC
[ ] Verify “BATTERY” LED is illuminated RED.
[ ] Verify “BATTERY FAULT DO NOT UNPLUG” is displayed.
[ ] Reconnect battery.
[ ] 13. ALARM VOLUME TEST
Press the Alarm Volume button. Verify:
[ ] That the audible alarm sounds at three volume levels.
[ ] 14. PUMPING CONTROLS
Press the left “ON” button. Verify:
[ ] The Console starts pumping and that the left “RATE” LED blinks amber.
[ ] The left “ON” LED illuminates green.
Press the left “OFF” button once. Verify:
[ ] The Console continues pumping and that the left “OFF” LED blinks red.
[ ] The left “OFF” LED turns off in 10 - 14 seconds.
Press the left “OFF” button twice within 10 seconds. Verify:
[ ] The Console stops pumping, the left “OFF” LED illuminates red, the left “ON” LED turns off, and
the audible alarm turns off.
[ ] Repeat the above test for the right side.
[ ] While holding the left “ACTIVATE” button, press the left “STROKE” button. Verify that there is a
single air pulse at the left barb.
[ ] While holding the right “ACTIVATE” button, press the right “STROKE” button. Verify that there is a
single air pulse at the right barb.
[ ] 15. ALARM MUTE
Generate an audible alarm. Press the Alarm Mute button. Verify:
[ ] The audible alarm silences and that the Alarm Mute LED illuminates amber. At between 50 – 70
seconds the audible alarm starts to sound and the Alarm Mute LED turns off.
[ ] 16. PUMP ID
Power Console on open port. Verify:
[ ] Display indicates BVS Pump Mode left and right sides.
[ ] Connect Ventricle Pump connector to left drive port. Verify display indicates Ventricle Pump mode
left side.
[ ] Move Ventricle Pump connector to right drive port. Verify display indicates Ventricle Pump mode
right side.Power Console down.
0015-7309, RevH
AB5000™ System Installation / Preventative Maintenance Procedure Page 4 of 8
ABIOMED, Inc. Proprietary and Confidential SOP-QA1000-T2, Rev. B
[ ] 17. EMERGENCY SYSTEM OPERATION
Activate the Emergency System Mode. Verify:
[ ] "EMERGENCY SYSTEM ON" indicator illuminates within 15 seconds from power-up.
[ ] The Console is pumping at 75 BPM.
[ ] The LCD display is blanked.
[ ] An audio alarm is sounded.
[ ] The audio alarm is silenced when the “Mute” button is pressed.
[ ] 18. ALARMS, PNEUMATIC
With the left barb in the open port state, press the left “ON” button. Verify:
[ ] At the beginning of the 5th beat the left side “LOW PRESSURE” alarm occurs.
[ ] Within 9 beats the left side “LOW PRESSURE” & “LOW FLOW” alarms occur.
Press the left “OFF” button twice. Block the left barb then press the left “ON” button. Verify:
[ ] Within 9 beats the left side “HIGH PRESSURE” & “LOW FLOW” alarms occur.
Press the left “OFF” button twice.
With the right barb in the open port state, press the right “ON” button. Verify:
[ ] At the beginning of the 5th beat the right side “LOW PRESSURE” alarm occurs.
[ ] Within 9 beats the right side “LOW PRESSURE” & “LOW FLOW” alarms occur.
Press the right “OFF” button twice. Block the right barb then press the right “ON” button. Verify:
[ ] Within 9 beats the right side “HIGH PRESSURE” & “LOW FLOW” alarms occur.
Press the right “OFF” button twice.
[ ] 19. ADJUSTABLE LOW FLOW ALARM (IF EQUIPPED)
Connect Load Simulator to left output. Start pumping left side. Press “FUNC” key once to select Step
Down screen. Press and hold the left Activate button. Use the“UP” button to set alarm limit to 2.5 L/min.
Release “Activate” button. Verify:
[ ] Low Flow alarm is indicated when displayed flow drops below 2.5 L/min.
[ ] Repeat test for right side.
[ ] 20. WEANING
Connect the Hydraulic Load Simulator to the Console’s left side. Start pumping in BVS mode.
Allow flows to stabilize.
[ ] Set the left side weaning to 2.0 L/min. Verify flow updates to 2.0 L/min.
[ ] Exit weaning. Verify recovery back to full flow.
[ ] Repeat test for Console’s right side output.
Insert Ventricle Pump ID connector into Console’s left Turret. Connect the Hydraulic Load Simulator to the
Console’s left side. Start pumping in Ventricle Pump mode. Allow flows to stabilize.
[ ] Set the left side weaning to 2.0 L/min. Verify flow updates to 2.0 L/min.
[ ] Exit weaning. Verify recovery back to full flow.
[ ] Repeat test for Console’s right side output.
0015-7309, RevH
AB5000™ System Installation / Preventative Maintenance Procedure Page 5 of 8
ABIOMED, Inc. Proprietary and Confidential SOP-QA1000-T2, Rev. B
[ ] 21. COMPRESSOR SPEED CHECK
With Console powered on and in diagnostic terminal mode, verify compressor speed.
[ ] BVS Pump Mode: 1050 RPM +/-52
[ ] Ventricle Pump Mode: 1300 RPM +/-65
[ ] Emergency System Operation 1050 RPM +/-52
[ ] 22. PRESSURE / VACUUM CHECK (PLENUM)
With Console powered on and in diagnostic terminal mode, verify plenum pressures and
vacuums.
BVS Mode
Pressure
(Pump ON)
BVS Mode
Vacuum
(Pump ON)
Ventricle Mode
Pressure
(Pump OFF)
Ventricle Mode
Vacuum
(Pump OFF)
Left 320mmHg
+/- 15
-35mmHg
+/- 10
420mmHg
+/-15
-100mmHg
+/-10
Right 200mmHg
+/-10
-35mmHg
+/-10
300mmHg
+/-10
-100mmHg
+/-10
NOTE: Pressure Values to be measured as Prs Plen Max.
Vacuum Values to be measured as Vac Plen Min.
Ventricle ID Plug must be installed for Ventricle measurements.
[ ] 23. PRESSURE / VACUUM CHECK (BARB)
With Console powered on and using the Pressure Gauge, verify barb pressures and vacuums.
BVS Mode
Pressure
(Pump ON)
BVS Mode
Vacuum
(Pump ON)
Ventricle Mode
Pressure
(Pump ON)
Ventricle Mode
Vacuum
(Pump ON)
Left 320mmHg
+/- 15
-35mmHg
+/- 10
420mmHg
+/-15
-100mmHg
+/-10
Right 200mmHg
+/-10
-35mmHg
+/-10
300mmHg
+/-10
-100mmHg
+/-10
NOTE: Ventricle ID Plug must be installed for Ventricle measurements.
0015-7309, RevH
AB5000™ System Installation / Preventative Maintenance Procedure Page 6 of 8
ABIOMED, Inc. Proprietary and Confidential SOP-QA1000-T2, Rev. B
[ ] 24. TARE OFFSETS
Connect diagnostic download cable between Computer and Console. Power Console on. Enter
Diagnostic Terminal mode. Record initial tare offset values at ‘0 min (a)’.
If the tare value is outside of +/- 10.0 L/min, replace the flowmeter.
If the tare value is outside of +/- (3.0 – 10.0 L/min) , send the following commands to tare the
flowmeter (items within <> indicate the key to press):
<ENTER> <ENTER> *@=A <ENTER> A$$V <ENTER>
The tare offset should be reduced following the command. Repeat once more if it does not.
Record the post-tare offset into 0(b).
If the tare value is within +/- 3.0 L/min, record the offset into 0 (a), record n/a into 0(b).
Record the tare offset in the following time intervals.
Left Alicat Right Alicat
Time
(min) Flow LPM Flow LPM Acceptable Flow Out: Pass √Fail √
0 (a) +/- ( 3.0 – 10.0 ) L/min N/A N/A
0 (b) 0.0 +/- 3.0 L/min
2 T=2-T=0 < ±0.5 L/min
4 T=4-T=2 < ±0.5 L/min
6 T=6-T=4 < ±0.5 L/min
8 T=8-T=6 < ±0.5 L/min
10 T=10-T=8 < ±0.5 L/min
12 T=12-T=10 < ±0.5 L/min
14 T=14-T=0< ±0.5 L/min
16 T=16-T=14 < ±0.5 L/min
18 T=18-T=16 < ±0.5 L/min
20 T=20-T=18 < ±0.5 L/min
22 T=22-T=20 < ±0.5 L/min
24 T=24-T=22 < ±0.5 L/min
26 T=26-T=24 < ±0.5 L/min
28 T=28-T=26 < ±0.5 L/min
30 T=30-T=16< ±0.5 L/min
Specification: 0.0 +/- 3.0 L/min (excluding 0 (a))
0015-7309, RevH
AB5000™ System Installation / Preventative Maintenance Procedure Page 7 of 8
ABIOMED, Inc. Proprietary and Confidential SOP-QA1000-T2, Rev. B
[ ] 25. VACUUM ADJUST (IF EQUIPPED)
Press Console “FUNC” key twice to select Vacuum Adjust screen. Verify:
[ ] Left-side vacuum range is adjustable between -35 mmHg to –100 mmHg in 5 mmHg
increments
[ ] Right-side vacuum range is adjustable between -35 mmHg to –100 mmHg in 5 mmHg
increments
Set and verify the following left-side and right-side vacuums:
Left [ ] -35 mmHg Right [ ] -35 mmHg
[ ] -50 mm Hg [ ] -50 mm Hg
[ ] -75 mmHg [ ] -75 mmHg
[ ] -90 mmHg [ ] -90 mmHg
[ ] 26. REMOTE ALARM (IF EQUIPPED)
Plug a 2-conductor ¼” phone plug to the remote alarm jack. Connect a DVM between the plug
contacts. Verify:
[ ] Short circuit (Ohmic resistance of < 1 ohms) when alarm condition is active.
[ ] Open circuit (Ohmic resistance more than 1M is acceptable) when no alarm is active.
[ ] 27. STATUS INDICATORS
Setup: Console is plugged in and powered on.
[ ] When the console is plugged in and powered on, the following should be indicated:
Panel LEDs: “AC POWER” is GREEN
“CHARGING”maybeAMBERorOFF
LCD Display: “FULL CHARGE” or “BATTERY CHARGING”
[ ] Unplug the console while it is running. Plug it back in within ~10 seconds. The following
should be indicated momentarily:
Panel LEDs: “AC POWER” is GREEN
“CHARGING”isAMBER
LCD Display: “BATTERY CHARGING”
[ ] When the battery is fully charged (within 5 minutes), the following should be indicated:
Panel LEDs: “AC POWER” is GREEN
LCD Display: “FULL CHARGE”
0015-7309, RevH
AB5000™ System Installation / Preventative Maintenance Procedure Page 8 of 8
ABIOMED, Inc. Proprietary and Confidential SOP-QA1000-T2, Rev. B
[ ] 28. BATTERY RUN
Setup: Console fully charged, and plugged into an AC outlet. Ventricle Pump ID Connector
plugged into left drive port. Console pumping Bivad open port.
[ ] Unplug the console. The following should be indicated:
Panel LEDs: “BATTERY” LED is AMBER
LCD Display: “BATTERY IN USE”
[ ] Verify two beep audio alarm every 60 seconds.
[ ] Verify 60-minute battery run without a Battery “Low” or “Fault” condition.
[ ] 29. CONSOLE CART
[ ] Inspect the Cart for any physical damage.
[ ] Inspect castors for any damage. Check brake, directional locks and rotation.
[ ] Check storage compartment door for proper operation and latching.
[ ] Check Console-to-Cart latching mechanism for proper operation.
[ ] 30. FINAL CHECK
[ ] Reinstall the Console panels. Perform a visual inspection to ensure that all hardware and
connections have been completed.
[ ] Place the Console onto the Cart.
[ ] Remove the protective film from the LCD display.
[ ] Perform the Power up Self Test.
[ ] If a console installation is being performed, remove the temporary advisory label on the
front panel.
[ ] 31. ELECTRICAL SAFETY (EUROPE ONLY)
[ ] Earth Resistance: mΩ(<200 mΩ)
[ ] Insulation Resistance: MΩ(>20 MΩ)
[ ] Earth Leakage NC: mA (<0.5 mA)
[ ] Earth Leakage SFC: mA (<1mA)
[ ] Enclosure Leakage NC: mA (<0.1mA)
[ ] Enclosure Leakage SFC (Neutral Open); mA (<0.5 mA)
[ ] Enclosure Leakage SFC (Earth Open): mA (0.5 mA)

Other Abiomed Medical Equipment manuals

Abiomed Impella 2.5 User manual

Abiomed

Abiomed Impella 2.5 User manual

Abiomed Impella Connect User manual

Abiomed

Abiomed Impella Connect User manual

Abiomed Impella LD User manual

Abiomed

Abiomed Impella LD User manual

Abiomed Impella RP User manual

Abiomed

Abiomed Impella RP User manual

Abiomed Impella Connect User manual

Abiomed

Abiomed Impella Connect User manual

Abiomed Impella Sidearm Retainer User manual

Abiomed

Abiomed Impella Sidearm Retainer User manual

Abiomed Impella 5.5 User manual

Abiomed

Abiomed Impella 5.5 User manual

Popular Medical Equipment manuals by other brands

Esaote MyLab40 Quick reference guide

Esaote

Esaote MyLab40 Quick reference guide

Nasco Life/form LF00958U instruction manual

Nasco

Nasco Life/form LF00958U instruction manual

laerdal MegaCode Kelly manual

laerdal

laerdal MegaCode Kelly manual

PhysioFlow Lab1 Service manual

PhysioFlow

PhysioFlow Lab1 Service manual

Maico MB 11 Classic Operation manual

Maico

Maico MB 11 Classic Operation manual

hinor POCA06-5L manual

hinor

hinor POCA06-5L manual

Bistos BT-400 Operation manual

Bistos

Bistos BT-400 Operation manual

BTL VANQUISH ME Operator's manual

BTL

BTL VANQUISH ME Operator's manual

Dräger D-Vapor Instructions for use

Dräger

Dräger D-Vapor Instructions for use

Precision Medical Easy Go Vac PM66 Series user manual

Precision Medical

Precision Medical Easy Go Vac PM66 Series user manual

Real i Series user manual

Real

Real i Series user manual

Sirona Heliodent plus operating instructions

Sirona

Sirona Heliodent plus operating instructions

Accutome a-scan plus user guide

Accutome

Accutome a-scan plus user guide

Ergotron SV DC Power System user guide

Ergotron

Ergotron SV DC Power System user guide

Envirotainer Releye RAP user manual

Envirotainer

Envirotainer Releye RAP user manual

ECS Warmair 135 Operation manual

ECS

ECS Warmair 135 Operation manual

Biocare iE 101 user manual

Biocare

Biocare iE 101 user manual

R82 Gazelle PS user guide

R82

R82 Gazelle PS user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.