Accora FloorBed 2 User manual


2© Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com
Contents
Title Page
Introduction 2
1. Means of delivery 4
2. Safety Instructions 4
3. Intended use 4
4. Technical specification 4
5. Accessories 5
6. Electrical specification 5
7. Assembly 6
8. Positioning 8
9. Functionality check 9
10. Using the castor brakes 9
11. Mattresses selection 10
12. Siderail selection 10
13. Disassembly 10
14. Moving and Repositioning 11
15. Cleaning & Disinfection 11
16. Troubleshooting 11
17. Storage 12
18. Daily inspection 12
19. General maintenance 12
20. Guarantee 12
21. Disposal 12
Introduction
Dear Customer,
Thank you for purchasing an Accora healthcare
product. We feel sure that this product will
exceed your expectations.
This manual is important for realising the full
potential of your new FloorBed; please read it
before use. If you need further information,
please contact us at:
Tel: 01223 206100
Fax: 01223 206120
General
The FloorBed is classified as a Medical
Device Class 1 in accordance with
93/42/EEC.
The FloorBed nursing bed is designed to be
used by patients in both low and high need
care environments where medical supervision
is required and monitoring is provided if
necessary.
Before operating the bed, you must become
acquainted in detail with this manual. All
actions and handling of the bed must be
performed in accordance with the instructions
in this manual.
Any actions that are inconsistent with the
manual are performed at your own risk and
Accora shall not be liable for any injury or
damage. Please ensure that the manual is
available to operators throughout the bed’s
service life.

GENERAL WARNINGS
1. Keep this Instruction Manual available for
future reference.
2. These instructions must be adhered to
ensure the safe and effective use of this
bed and the safety of users and carers.
3. This bed must be assembled, positioned
and used in accordance with these
instructions.
4. The safety features for operating the bed
and instructions concerning the bed must
be strictly observed.
5. This bed must not be exposed to smoke,
naked flame, extreme temperature,
flammable gasses or other hazardous
substances or situations.
6. Accora shall not be held liable for any
damage, injuries or accidents arising from
unauthorised modifications, non-genuine
spare parts, negligence or use that is at
variance with this manual.
7. Electrical equipment can be hazardous if
misused or abused. Ensure the electrical
supply cable is not damaged by crushing
and does not create a trip hazard.
8. Only use side rails, and other accessories,
that are compatible with this bed as
supplied by Accora. Incompatible side
rails can create serious hazards.
9. Keep children and pets away from this
bed unless supervised by an adult as there
is a risk of injury and/or choking on small
parts.
10. Do not lower the bed while a hoist that
extends beneath the bed is being used.
Hoist access is obtained when the bed is
raised to 45cm measured from the floor to
the mattress platform base.
11. When routing cables for other electronic
equipment used with the bed (e.g. air
mattress pump), please ensure cables
cannot be squeezed, crushed or
damaged by the moving parts on the
bed.
12. The hand control should be positioned to
avoid strangulation risk. Inappropriate use
of the hand control (e.g. kinking, shearing)
could lead to dangerous electric hazards.
The bed must not be used if there is any
visible damage to the handset or cable.
13. Inappropriate use of the power supply
cord (e.g. kinking, shearing) could lead to
dangerous electric hazards. The bed must
not be used if there is any visible damage
to this cable.
14. Inappropriate routing of accessory cables,
e.g. mattress air pump cable, could lead
to dangerous electric hazards if the
squeezed or crushed between moving
parts. The bed must not be used if there is
any visible damage to any cables.
15. The bed should be left in the Floor-level
position when the patient is unattended in
order to reduce risk of injury due to falls.
16. If using the electronic functions adversely
affects the health of the patient,
disconnect the power supply and only use
the bed in the static mode.
17. Do not move the bed when it is in the
Floor-level position.
18. This bed should not be used for
transporting patients.
19. This bed is typically not suitable for
children unless a risk assessment has been
carried out.
20. Do not modify this bed without the
authorisation Accora.
21. Before operating this bed, ensure the
patient is safely positioned to remove the
risks of bed fall, entrapment and
imbalance.
22. Always check for any entrapment risks
under the bed before lowering to the
Floor-level position.
23. Electrical installations must meet local
requirements. It is recommended that the
bed is disconnected from the mains
during exceptional cases (i.e. a storm).
24. Patients, or users, should be risk assessed to
ensure they are able to understand this
manual and to operate FloorBed safely
and without risk to themselves and those
around.
25. Patients, or users, should only be allowed
to operate the bed independently if they
are able to understand the safety
instructions in this manual and have been
risk assessed as appropriate to do so.
26. If the combined weight of the mattress
and accessories exceeds 35kg, the
maximum patient weight must be
reduced accordingly.

4© Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com
1. Means of Delivery
The bed is supplied on a transport bracket. An
inspection must take place upon receipt to ensure the
delivery is complete and undamaged.
Any missing parts, faults or damage must be reported
immediately to the carrier, and Accora, in writing.
When moving the bed, or loading and unloading it,
care must be taken that the bed castors can rotate
freely and the castor brakes are unlocked. The bed
castors are designed for use in an indoor environment
and for travel on even, smooth and clean floors, (e.g.
ceramic floor tiles, linoleum, cast floors). The castors
may become damaged when moving the bed along
a rough, uneven or dirty surface.
2. Safety Instructions
1. Before using the bed, you must read the manual
and use the bed in accordance with it.
2. The bed may not be used if faults have been
detected on it that may injure the patient, staff or
a third person, the bed or the surroundings.
3. The bed may only be operated by persons who
are able to operate it in accordance with the
manual.
4. The operating staff should acquaint the patient
with the control functions that apply to the patient.
5. Before using the bed, the operator should
understand the bed and its functionality.
6. The safe working load, as specified in section 4,
must never be exceeded.
7. If there is a patient on the bed, the bed castors
must be locked as an unlocked bed castor can
cause injury to a patient who leaves the bed or
changes position.
8. The height of the mattress platform must be
adjusted to the correct height for the condition of
the patient.
9. Only one person should occupy the bed at any
time.
10. When operating the moving parts of the bed, care
must be taken to ensure that the patient, other
people and objects do not become trapped.
11. If a lifting pole or infusion stand is fixed to the bed,
increased attention must be taken during
movement, lifting or tipping, to the space around
the lifting pole and infusion stand, so that the
equipment is not damaged.
12. Before cleaning the bed, the electrical supply must
be disconnected.
13. The bed may not be used where there is a danger
of explosion or in the presence of inflammable
liquids.
14. When repairing the bed, only original materials
and components may be used, otherwise the
manufacturer cannot guarantee against any
damage that might occur.
3. Intended use
This bed is intended for use in the following application
environments only:
1. Application Environment 3 - Long-term care in a
medical area where medical supervision is
required and monitoring is provided if necessary
and ME equipment used in medical procedures
may be provided to help maintain or improve the
condition of the patient. Note, this includes use in
nursing homes and in rehabilitation and geriatric
facilities.
2. Application Environment 4 - Care provided in a
domestic area where ME equipment is used to
alleviate or compensate for an injury, disability or
disease
It is essential to consult Accora in advance if you wish
to use the bed for any purpose outside the use
detailed in this manual. Electrical installations must
meet local requirements. It is recommended that the
bed is disconnected from the mains during
exceptional cases (i.e. a storm).
4. Technical Specification
Environmental information:
Condition
Temperature
Range
Relative
Humidity
Atmospheric
pressure
Operating
+10°C to +40°C
(+50°F to +104°F)
30% to 75%
(Non-
condensing)
700 hPa to
1060 hPa
Transport/
storage
-20°C to +50°C
(-4°F to +122°F)
If the bed is stored in conditions outside the normal
operating range, it should be allowed time to stabilise,
in normal operating conditions, before use.
WARNING
Extreme caution must be taken when moving
the bed on the transport bracket to prevent
the bed tipping over or moving unexpectedly.

Description
Value
Overall dimensions
940mm W × 2290mm L
(37in W x 90in L)
Mattress platform dimensions*
900mm W × 2000mm L
(35.5in W x 78.75in L)
Bed castor
4 x 75mm with brake
(4 x 3in with brake)
Mattress platform height
71mm to 800mm
(2.8in to 31.5in)
Max self-help pole lifting load
75kg (12st) (165lb)
Safe Working Load**
185kg (29st) (407lb)
Maximum Patient Weight**
150kg (23.5st) (330lb)
Audible noise
53Db (A) typical TBC
Mass of Bed (excluding transport bracket)
93.2kg (14.5st) (205lb)
- Headboard
26.5kg (4st) (58.5lb)
- Footboard
26.5kg (4st) (58.5lb)
- Mattress platform, head end
23.0kg (3.5st) (50lb)
- Mattress platform, foot end
17.2kg (2.7st) (38lb)
Liquid Ingress Protection
IPX4 Splash resistant
Trendelenburg function
15 degrees
Expected service life
Typically 5 years
FloorBed mattress platform range including maximum
angles:
Key parts of the bed:
* The recommended patient height is 1460 –1850mm.
Taller patients may be accommodated by using a
mattress platform extension.
** The safe working load is calculated as follows (as
specified by EN 60601-2-52):
Maximum patient weight: 150kg 23.5st 330lb
Mattress 20kg 3st 44lb
Accessories 15kg 2.5st 33lb
TOTAL (Safe working load) 185kg 29st 407lb
5. Accessories
1. Mattress platform extension –extends the mattress
platform from 2000mm to 2200mm.
2. Siderails - integrated side rails made of soft
material.
3. Fixed and rotating bed levers –assists with mobility
and patient transfers.
4. Headboard and footboard bumpers - reduce
impact and minimise the risk of injury.
5. Mattress platform safety sleeve –reduces the risk of
entrapment in the mattress platform.
6. Electrical specification
Duty Cycle: Intermittent operation 2 min/18 min; this
implies that after the max. continuous action of two
minutes, there must be a break of 18 minutes.
The B symbol indicates this product has a
degree of protection against electric shock
for type B equipment.
Caution, read the instructions before use.
Degree of protection against liquid ingress.
Do not dispose of in household waste.
Degree of protection against electric shock:
Class II Double Insulated.
For indoor use only
Voltage supply
Voltage input frequency
Headboard
Footboard
Mattress platform, head section
Mattress platform, foot section
1
2
3
4

6© Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com
7. Assembly
1. Check that the delivery is complete and whether
any visible damage has occurred to the bed
during transport.
2. Identify all components:
3. Remove the components from the transport
bracket in the following order. Remove the fixing
pins and loosen the bolts (shown in red), one
component at a time, in the sequence as shown
below, a –e. Ensure the transport bracket castors
are positioned and locked as shown below.
a. Box including mattress guides, handset,
mattress platform joining bars and power
supply.
b. Mattress platform, head section
c. Mattress platform, foot section
d. Footboard
e. Headboard
WARNING
Assembly MUST be carried out by a
competent person.
All functions MUST be tested and approved by
a competent person after assembly.
Assembly MUST take place in a clear,
uncluttered area and children and pets
should be kept away.
Only power supply supplied with bed should
be used.
If bed has become soiled or contaminated
during transit refer to cleaning and
disinfection instructions.
Ensure headboard and footboard are
assembled as shown below so that the
Trendelenburg function works in a safe way.
1
2
4
3
5
6
7
8
Headboard
Footboard
Mattress platform, head section
Mattress platform, foot section
Mattress guides
Handset
Power supply
Mattress platform joining bars
1
2
3
4
5
6
7
8

4. Assemble the headboard and mattress platform
head end as shown below. Insert locking pin and
R-clip, as shown, in two positions.
5. Assemble the footboard and mattress platform
foot end as 4 above. Insert locking pin and R-clip,
as shown, in two positions.
6. Insert two mattress platform joining bars and
secure with locking pins and R-clips as shown in
two positions.
7. Join the two parts of the bed together using the
joining bars and secure with locking pins and R-
clips as shown in two positions.
8. The bed should now be assembled with 8 locking
pins and R-clips located as shown below.
9. Plug in four actuators and fit sliding protective
caps as shown below.

8© Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com
10. Inspect all wiring for damage and risk of crushing
before plugging in the power supply.
11. Using the handset, raise the mattress platform
approximately 30cm. Tighten four locking screws
under the mattress platform as shown.
12. If the bed has been supplied with tie wraps / hook
and loop tape (or similar) securing any of the
sections in place, ensure they are removed prior to
operation.
8. Positioning
The handset can be used for positioning the sections
and for adjusting the height of the mattress platform of
the bed by either the patient or the carer.
Always check for obstruction before raising and
lowering. Before using the control, the operating staff
should explain to the patient how the bed can be
positioned. If the medical staff state that the patient’s
medical condition is inappropriate for the patient to
be able to adjust the bed independently, the bed’s
position must be adjusted by the carer.
Please check for any entrapment risks under the
bed before lowering to the lowest position.
WARNING
Bed positioning MUST be carried out by a
competent person.
Check for obstructions around, above and
below the bed frame and position the bed so
that it can operate through the full height
range without any chance of obstruction or
entrapment.
The head end of the bed must be a minimum
of 300mm from the wall and always engage
the brakes when the bed is stopped or left
unattended.
The patient should only operate bed if they
are competent to do so.
2
4
6
7
9
1
3
5
8
Backrest elevating/lowering
Legrest elevating/lowering
Bed elevating/lowering
Floor-level safety button (must be
pressed with button number six)
Reverse Trendelenburg and
Trendelenburg function
1
3
5
8
2
4
6
7
9

To activate these options, the magnetic security key
must be swiped across the lower part of the handset,
as shown above by the red box:
First swipe - Handset status indicator light green. The
top 6 buttons, Floor-level safety button and Reverse
Trendelenburg will function. Trendelenburg function will
remain locked.
Second swipe - Handset status indicator light
orange. All 9 buttons now functional, including
Trendelenburg. If the handset is not used it will auto-
lock after a pre-set time. This is called the “SAFETY TIME
OUT” function. The handset status indicator light will
turn off.
Third swipe - Handset status indicator light off. The
handset is now fully locked.
The cycle for light changing is:
No light –handset locked
First swipe –green light
Second swipe –orange light
Third swipe –no light
The locked and unlocked options may be chosen at
any time depending on the needs of the user.
9. Functionality check
Using the handset test all bed functions and check all
cables for risk of crushing. Refer to section 8 above,
Positioning, for handset controls.
a. Raise the bed to full height (button 5)
b. Lower the bed until it stops in the Safety Stop
position (button 6)
c. Use the safety double button function to lower the
bed to the Floor-level position (buttons 6 & 7
pressed together)
d. Check all cables for risk of crushing.
e. Lift and lower backrest (buttons 1 & 2)
f. Lift and lower legrest (buttons 3 & 4)
g. Check the reverse Trendelenburg function (head
up, feet down) (button 8). Trendelenburg function
(head down, feet up) (button 9) should not
activate unless this function is unlocked as
described in section 8, Positioning.
Check the correct function of the castors and brake
control. Check all explained functions of the handset.
10.Using the castor brakes
All four castors can be locked by pressing down on the
lower lever of the castors; they can then be unlocked
by pressing down on the top lever of the castors. Care
must be taken to make sure the castor brakes are
always locked when the bed is in use, being
assembled or dismantled so that the bed does not
move accidentally.
Brake locked Brake unlocked
WARNING
Functionality check MUST be carried out by a
competent person.
Check for obstructions around, above and
below the bed frame and position the bed so
that it can operate through the full height
range without any chance of obstruction or
entrapment.
The head end of the bed must be a minimum
of 300mm from the wall and always engage
the brakes when the bed is stopped or left
unattended.

10 © Accora: 01223 206100 | [email protected]om | www.accora.uk.com
11.Mattresses selection
All mattresses must be fitted and used in accordance
with the mattresses manufacturer or supplier’s
instructions.
12.Siderail selection
13.Disassembly
1. Remove all accessories, e.g. mattress, side rails, bed
levers etc.
2. Using the handset, raise the mattress platform
approximately 30cm. Loosen the 4 mattress
platform joining bar locking screws. (Do not remove
the R-clips and locking pins at this stage)
3. Fix the backrest an legrest in the flat position with
the Velcro straps provided.
4. Lower the mattress platform to the Floor-level
position and disconnect the power supply.
5. Remove the control box cover and unplug
headboard and footboard lift actuators, legrest
actuator, handset and power supply. Ensure wires
are left free so that the bed can be dismantled
without risk of damage to the wires.
6. Remove the central joining bar R-clips and locking
pins and separate the bed into two halves.
7. Dismantle the headboard/mattress platform head
end and footboard/mattress platform foot end by
removing the R-clips and locking pins.
8. Ensure the disassembled parts are fitted to the
transport bracket in the following order and
positioned as shown in Section 7:
a. Footboard
b. Headboard
c. Mattress platform foot section
d. Mattress platform head section
9. Ensure all R-clips and pins are refitted so that the
bed on the transport bracket is safe to move.
10. Ensure all wires are secure to prevent damage.
WARNING
Incompatible mattresses can create hazards.
Read instructions for use.
Mattress dimensions must not exceed 2000 x
900 x 200mm. The distance from the top of the
uncompressed mattress to the top of the
siderail must always be greater than 220mm
Where a speciality mattress or mattress
overlay is used and the distance from the top
of the uncompressed mattress to the top of
the siderail is less than 220mm a risk
assessment must be performed to assure
equivalent safety.
WARNING
Only use side rails that are compatible with
this bed as supplied by Accora.
Incompatible siderails can create hazards.
WARNING
Disassembly MUST be carried out by a
competent person.
The bed must be disconnected from the
power supply before disassembly.
Disassembly must take place in a clear,
uncluttered area and children and pets
should be kept away.
If bed has become soiled or contaminated
during use refer to cleaning and disinfection
instructions.
Extreme caution must be taken when moving
the bed on the transport bracket to prevent
the bed tipping over or moving unexpectedly.
Please contact
Accora for
compatible
mattresses.
Incompatible
mattresses can
create hazards.

11
14.Moving and Repositioning
1. Ensure the bed is at the Safety Stop position (see
Section 9, point b).
2. Disconnect the power.
3. Secure the handset, power supply and all cables to
prevent damage.
4. Unlock the castors and move the bed.
5. When the bed has been moved or repositioned
lock al the castors and follow the instruction in
Section 8
6. When the bed has been positioned correctly, follow
the instructions in Section 9.
15.Cleaning & Disinfection
Cleaning Information:
In order to disinfect the bed, only use detergents
designed for use in healthcare. Do not use abrasives,
scourers or other materials that could damage the
coating. Do not use corrosives, caustics or strong acids.
Do not use detergents that could alter the structure or
behaviour of the plastics (petrol etc.).
Clean by wiping with a damp and properly wrung
cloth.
The bed is not designed for maintenance in automatic
bed washers or for cleaning with pressurised water,
spraying, showering or steam cleaning.
Accora cannot be liable for any damage or risk of
damage if inappropriate cleaning or disinfectant
agents are used.
Cleaning procedure:
1. Remove all accessories, mattress etc.
2. Adjust the mattress platform to the highest position
and adjust the position of the Backrest and Footrest
to provide access for cleaning all the platform
parts.
3. Disconnect the bed from the power supply.
4. Move the bed to where cleaning will take place
and lock the bed brakes.
5. Clean as described in the “Cleaning Information”.
16. Troubleshooting
The FloorBed does not function correctly:
1. Is the ‘Power on’ indicator light on the handset lit? If
not:
a. Is the power supply plugged in and turned on?
b. Is the power lead in-line socket plugged in
correctly?
c. If the ‘Power on’ indicator light is still not lit,
contact Accora for further advice.
2. If the ‘Power on’ indicator light is on, has the
handset been unlocked? If not, refer to Section 8.
3. If the handset has been unlocked and the FloorBed
still does not function correctly, contact Accora for
further advice.
If the FloorBed does not stop at the Safety-stop
position, check the main lift actuators are correctly
plugged in at all points between the actuator and the
control box.
If the FloorBed still does not function correctly, contact
Accora for further advice.
WARNING
Moving or repositioning MUST be carried out
by a competent person.
All functions MUST be tested and approved by
a competent person after moving or
repositioning
Only power supply supplied with bed should
be used.
Do not move the bed in the Floor-Level
position
Do not move or reposition the bed with
service user or patient on the bed.
WARNING
Cleaning and disinfection MUST be carried out
by a competent person.
The bed must be disconnected from the
power supply when being cleaned or
disinfected.
All functions MUST be tested and approved by
a competent person after cleaning or
disinfection.
WARNING
Troubleshooting MUST be carried out by a
competent person.
Do not attempt to open any electrical part
enclosures.
Do not attempt to repair any electrical parts.
All functions MUST be tested and approved by
a competent person after troubleshooting

12 © Accora: 01223 206100 | [email protected]om | www.accora.uk.com
17. Storage
For problem-free storage we recommend:
1. Disconnecting the bed from the electric supply
2. Remove the accessories
3. Wrapping the bed and accessories or covering
them so that the coating and plastic parts are not
damaged
4. Bed should be stored in a temperature between
+10°C and +40°C
5. Bed should be stored somewhere where the
relative humidity is between 30% and 75%
18. Daily Inspection
Daily visual inspection is strongly recommended and
may be carried out by Carer, user or other (not
necessarily a competent) person.
The following checks must be carried out:
1. Does the bed operate as per its intended purpose
without unexpected noise or motion?
2. Are there any signs of abuse or excessive wear?
3. Are all fixtures and fittings tight and secure?
4. Does the bed frame appear stable and secure?
5. Are all accessories fitted in line with the accessory
manufacturer or accessory supplier’s instructions?
6. Are all the castors brakes in the locked position?
7. Are all electrical cables (including accessories, e.g.
mattress air pump) secured and routed to prevent
damage?
8. Does the handset locking function work correctly?
(See Section 8)
9. Does the bed stop at the Safety Stop position? (See
Section 9)
10. Is the area around, above and below bed clear of
possible obstruction?
11. Is there any risk of entrapment or patient injury?
If any damage, performance issue or cause for
concern is noted during this inspection the bed should
be withdrawn from service and appropriate steps
should be taken.
19. General maintenance
General maintenance must be carried out once yearly
as a minimum and by a competent person or person
trained to do so. The following checks must be carried
out in addition to the checks listed in Section 20 –Daily
Inspection.
Note: It is not current practice for lifting beds to be
considered to be lifting equipment within the scope of
the Lifting Equipment Operations Regulations (LOLER),
and so there is no formal legal requirement for regular
inspection by a (LOLER) competent person.
The following checks must be carried out:
1. Do the castors operate correctly, including the
brakes?
2. Do all electrical bed positioning functions work
correctly? (Backrest motion etc.).
3. Does the power supply, or handset, show any sign
of damage or abuse?
4. Are any electrical cables pinched, crushed or
damaged in any other way?
5. Does the bed frame show any signs of abuse,
damage or breakage?
6. Is the bed frame mechanically sound without any
cracking at welds etc?
7. Is the control box plastic clip on cover holding the
actuator plugs in place fitted?
8. Are the plastic retaining covers fitted to the head
and footboard actuators securing the wires in
place?
9. Is the control box metal cover fitted correctly?
10. Are all fixtures, fittings, nuts, bolts etc. tight and
secure as appropriate?
11. Are all actuators secured using a 10mm clevis pin
and R-clip?
12. Is the mattress platform secured to the head and
footboards using 4 x 10mm clevis pin and R-clips?
13. Are the mattress platform joining bars clamped in
place (with M8 bolts from below) and secured
with 4 x 10mm clevis pin and R-clips?
If any damage, performance issue or cause for
concern is noted during this inspection the bed should
be withdrawn from service and appropriate steps
should be taken.
20. Guarantee
The warranty period for the FloorBed is 24 months,
covering parts only.
If the bed was acquired for a care home, the care
home maintenance department should contact
Accora to obtain the necessary parts.
21. Disposal of the FloorBed
In the event of the disposal of materials from the bed,
end-of-life parts must be disposed of in accordance
with current environmental regulations.
WARNING
Maintenance MUST be carried out by a
competent person.
All functions MUST be tested and approved by
a competent person after maintenance.
Refer to Sections 8 and 9.
Only power supply supplied with bed should
be used.
Do not carry out maintenance with service
user or patient on the bed.

Guidance and manufacturer’s declaration-electromagnetic emissions
The bed is intended for use in the electromagnetic environment specified below.
The customer or the user of the bed should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The bed uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The bed is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations
/flicker emissions
IEC 61000-3-3
Compliance

14 © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com
Guidance and manufacturer’s declaration-electromagnetic immunity
The bed is intended for use in the electromagnetic environment specified below.
The customer or the user of the bed should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment-
guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
+ 6 kV contact
+ 8 kV air
+ 6 kV contact
+ 8 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%
Electrical fast
transient/burst
IEC 61000-4-4
+ 2kV for power
supply
lines
+ 1kV for
input/output
lines
+ 2kV for power
supply
lines
Not applicable
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge IEC 61000-4-5
+ 1kV line(s) to line(s)
+ 2kV line(s) to earth
+ 1kV differential
mode
Not applicable
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage Dips, short
interruptions and
voltage
variations on power
supply
input lines IEC 61000-4-
11
<5% UT(>95% dip in
UT) for 0,5 cycle
40% UT(60% dip in
UT) for 5 cycles
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
<5% UT(>95% dip in
UT) for 0,5 cycle
40% UT(60% dip in
UT) for 5 cycles
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the bed
requires continued
operation during
power mains
interruptions, it is
recommended that
the bed be powered
from an uninterruptible
power supply or a
battery.
Power frequency(50,
60 Hz)
magnetic field IEC
61000-4-8
3 A/m
3 A/m
The bed power
frequency magnetic
fields should be at
levels characteristic of
a typical location in a
typical commercial or
hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.

Guidance and manufacturer’s declaration-electromagnetic immunity
The bed is intended for use in the electromagnetic environment specified below.
The customer or the user of the bed should assure that is used in such and environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment-
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 KHz to 80 MHz
3 V/m
80MHz to 2,5 GHz
3 Vrms
3 V/m
Portable and mobile RF
communications equipment should be
used no closer to any part of the bed
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey, a should be less than the
compliance level in each frequency
range Interference may occur in the
vicinity of equipment marked with the
following symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
b
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the bed is
used exceeds the applicable RF compliance level above, the bedshould be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the bed.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

16 © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com
Recommended separation distance between portable and mobile RF communications
equipment and the bed.
The bed is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled.
The customer or the user of the bed can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the bed as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d =1,2√P
80 MHz to 800 MHz
d =1,2√P
800 MHz to 2,5 GHz
d =2,3√P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can
be estimated using the equation applicable to the frequency of the transmitter, where p is
the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.

EC Declaration of Conformity
In accordance with EN ISO 17050-1:2010
We Accora Ltd
Of Charter House, Barrington Road, Orwell, Cambridge, SG8 5QP
hereby declare that:
Model reference: FloorBed 2
Serial No: Refer to product label
conforms to the Requirements of Directive 93/42/EEC relating to Medical Devices
Class 1 Product.
Signed by: 01.01.2017
Name: Laurie Drake
Position: Operations Director

20 © Accora: 01223 206100 | info@accora.uk.com | www.accora.uk.com
IFU-FB2-001 EN Rev 1 JAN 2016

• •
Table of contents
Other Accora Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Oticon Medical
Oticon Medical Digisonic SP Instructions for use

R82
R82 Rabbit user guide

AGA Medical
AGA Medical Amplatzer TorqVue LP Instructions for use

Covidien
Covidien Nellcor OxiMax N-600x Operator's manual

NeuroMetrix
NeuroMetrix NC-stat DPNCheck user manual

Otto Bock
Otto Bock Infinity Double Air Walker high 50S24 Instructions for use