AccuVein AV400 User manual

English 1
Español 17
Français 33
Deutsch 49
Dansk 57
Italiano 65
Nederlands 73
Português 81
Suomi 89
Svenska 97
Türkçe 105
Русс кий 113
121
129
137
145
A health professional’s guide for use
and operation of the AccuVein AV400
AccuVein AV400 User Manual
AV400_compilation_covers_RevB.indd • May 22, 2013 7:48 AM • right page
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • right page
-AV400_RevB_compiled.indd 1 5/22/13 7:48 AM

AccuVein® is the registered trademark of AccuVein Inc. The content of this manual
including all photographs and drawings is the proprietary information of AccuVein
Inc.
This manual last revised: April 2013
Warnings and Cautions
Read all instructions, cautions, and warnings prior to use. This product should be
operated only by qualied medical professionals. The AV400 should not be used as
the sole vein location method, and it is not a substitute for sound medical judgment
and the visual and tactile location of veins.
Before using the AV400 on a patient, qualied medical professionals must read and
understand this AV400 User’s Manual. Before rst use, users should compare how
the AV400 detects veins with visual detection and palpation techniques.
AV400 User Manual
© AccuVein Inc. 2013
AccuVein Inc.
40 Goose Hill Road
Phone: (816) 997-9400
Cold Spring Harbor, NY 11724
United States of America
www.accuvein.com
AV400 is manufactured for:
AccuVein Inc.
40 Goose Hill Road
Cold Spring Harbor, NY 11724
United States of America
Device fully complies with
European Directive 2002/364/
EC. Conformité Européenne
For United States of America
customers: IEC/EN 60825-1 Safety
of Laser Product: Complies with 21
CFR 1040.10 and 1040.11 except for
deviations pursuant to Laser Notice
No. 50, dated June 24, 2007.
EMERGO EUROPE
Molenstraat 15
2513 BH, The Hague
The Netherlands
Phone: +31.70.345.8570
Fax: +31.70.346.7299
This AV400 User Manual and additional information can also be
downloaded from www.accuvein.com
Wavelength nm/ or nm
Peak Power mW / mW
Pulse Energy nJ / nJ
Pulse Length . µs
IEC/EN - (/)
VISIBLE AND INVISIBLE
LASER RADIATION.
DO NOT STARE
INTO BEAM. CLASS LASER
PRODUCT.
AccuVein AV400 User Manual
A health professional’s guide for use and
operation of the AccuVein AV400
English
AV400
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
AV400_compilation_covers_RevB.indd • May 22, 2013 7:48 AM • left page
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • left page
-AV400_RevB_compiled.indd 2 5/22/13 7:48 AM

AccuVein® is the registered trademark of AccuVein Inc. The content of this manual
including all photographs and drawings is the proprietary information of AccuVein
Inc.
This manual last revised: April 2013
Warnings and Cautions
Read all instructions, cautions, and warnings prior to use. This product should be
operated only by qualied medical professionals. The AV400 should not be used as
the sole vein location method, and it is not a substitute for sound medical judgment
and the visual and tactile location of veins.
Before using the AV400 on a patient, qualied medical professionals must read and
understand this AV400 User’s Manual. Before rst use, users should compare how
the AV400 detects veins with visual detection and palpation techniques.
AV400 User Manual
© AccuVein Inc. 2013
AccuVein Inc.
40 Goose Hill Road
Phone: (816) 997-9400
Cold Spring Harbor, NY 11724
United States of America
www.accuvein.com
AV400 is manufactured for:
AccuVein Inc.
40 Goose Hill Road
Cold Spring Harbor, NY 11724
United States of America
Device fully complies with
European Directive 2002/364/
EC. Conformité Européenne
For United States of America
customers: IEC/EN 60825-1 Safety
of Laser Product: Complies with 21
CFR 1040.10 and 1040.11 except for
deviations pursuant to Laser Notice
No. 50, dated June 24, 2007.
EMERGO EUROPE
Molenstraat 15
2513 BH, The Hague
The Netherlands
Phone: +31.70.345.8570
Fax: +31.70.346.7299
This AV400 User Manual and additional information can also be
downloaded from www.accuvein.com
Wavelength nm/ or nm
Peak Power mW / mW
Pulse Energy nJ / nJ
Pulse Length . µs
IEC/EN - (/)
VISIBLE AND INVISIBLE
LASER RADIATION.
DO NOT STARE
INTO BEAM. CLASS LASER
PRODUCT.
AccuVein AV400 User Manual
A health professional’s guide for use and
operation of the AccuVein AV400
English
AV400
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
AV400_compilation_covers_RevB.indd • May 22, 2013 7:48 AM • left page
1
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • right page
-AV400_RevB_compiled.indd 1 5/22/13 7:48 AM

Contents
Basic Use
and Operation 6
Setup
Intended Use
Product Description
User controls and screens
Operating the AV400
Charging, Cleaning
and Troubleshooting 13
Charging the AV400
Cleaning and Disinfecting
Troubleshooting
Warranty
and Liability 16
Warranty Summary
Limitation of Liability
Product
Specication 17
Specications
Product labels
AccuVein AV400 User Manual
A health professional’s guide for use and
operation of the AccuVein AV400
English
AV400
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
Contents
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • left page
-AV400_RevB_compiled.indd 2 5/22/13 7:48 AM

Contents
Basic Use
and Operation 6
Setup
Intended Use
Product Description
User controls and screens
Operating the AV400
Charging, Cleaning
and Troubleshooting 13
Charging the AV400
Cleaning and Disinfecting
Troubleshooting
Warranty
and Liability 16
Warranty Summary
Limitation of Liability
Product
Specication 17
Specications
Product labels
AccuVein AV400 User Manual
A health professional’s guide for use and
operation of the AccuVein AV400
English
AV400
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
Contents
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
613 14 15
3
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • right page
-AV400_RevB_compiled.indd 3 5/22/13 7:48 AM

Warnings and Cautions
Patient and User Safety—Warnings
WARNING: The AV400 enables location of certain supercial veins and is not a
substitute for sound medical judgment based on the visual and tactile location and
assessment of veins. The AV400 should be used only as a supplement to the judgment
of a qualied professional.
WARNING: The AV400 should only be operated when its battery is suciently
charged (indicated when the battery icon at the upper right of the LCD screen is
green), or when the device is operated in a powered hands-free accessory to ensure
AV400 is available for use.
WARNING: AV400 vein location is dependent on a variety of patient factors
and may not display veins on patients with deep veins, skin conditions, hair, scarring
or other highly contoured skin surface, and adipose (fatty) tissue.
WARNING: The AV400 displays only superficial veins, and does so only
to limited depths dependent on a variety of patient factors. The AV400 does not
indicate vein depth.
WARNING: The AV400 emits Visible and Invisible laser radiation. Do not stare
into beam. Class 2 laser product.
WARNING: Do not shine vein display light in eyes.
WARNING: Blink reexes can be inhibited or reduced by disease, drugs, or other
medical conditions. For patients who might have inhibited or reduced blink reexes,
AccuVein recommends outtting the patient with a protective eye shield. A protective
eye shield capable of blocking 642 nm and 785 nm should be used in this case.
WARNING: Stop using the AV400 if the red light does not turn on when the
blue button is pushed.
WARNING: Do not hold the AV400 while performing venipuncture or other
medical procedures.
WARNING: For external use only.
WARNING: Keep the AV400 and its battery out of the reach of children.
WARNING: To view vein location accurately, you must position the AV400 at
the proper height, and directly over the center of the vein being assessed.
Patient and User Safety—Contraindications
WARNING: The AV400 should not be used to locate veins in or near eyes.
WARNING: The AV400 is not intended to be used as a diagnostic device or for
treatment of any kind.
Patient and User Safety—Cautions
CAUTION: United States Federal law restricts this device to sale by or on the order
of a physician or other qualied medical professional.
CAUTION: Operation or use of the AV400 in a manner dierent than specied in this
AV400 User’s Manual may result in hazardous radiation exposure.
CAUTION: Do not make any connections to the USB port on the AV400 device.
Equipment Care— Warnings
WARNING: Do not immerse the AV400 or the AV400 charging cradle in liquid
or wet the AV400 such that liquid spills o.
WARNING: Do not attempt to open, disassemble, or service the battery pack.
Do not crush, puncture, or dispose of the battery in re or water. Do not short external
contacts. Do not expose to temperature above 60°C / 140°F.
WARNING: Do not modify in any way the interior or exterior components of
the AV400.
WARNING: To reduce the risk of re or shock hazard and annoying interference,
use only the recommended accessories and do not expose this equipment to rain or
heavy moisture.
WARNING: Use only AccuVein accessories and replacement parts with the
AV400. The use of non AccuVein accessories may degrade safety.
Equipment Care—Cautions
CAUTION: Use only AccuVein approved battery charging accessories, AccuVein BA300
and BA400 batteries, and AccuVein brand additional accessories. If the AV400 will not
be used for an extended time, remove and store battery in a safe, dust-free location.
The battery should be replaced only with an approved AccuVein battery. Removing
or replacing the battery should not to be done in the patient environment.
CAUTION: Electromagnetic Interference (EMI) can aect the proper performance
of the device. Normal operation can be restored by removing the source of the
interference.
CAUTION: Do not attempt to sterilize the AV400 with heat or pressure sterilization
methods.
CAUTION: The AV400 will not display veins if operated outside its temperature range.
CAUTION: The AV400 may not display veins if operated under bright light such as
bright sunlight.
CAUTION: Dispose of used batteries according to manufacturer’s instructions.
A recyclable lithium ion battery powers the AV400. Please call AccuVein at
(816) 997-9400 for information on how to dispose or recycle this battery.
CAUTION: Do not disassemble or modify the AV400 or any of its charging accessories.
CAUTION: Do not service. The AV400 contains no customer serviceable components.
The AV400 and its accessories should be serviced only by an authorized AccuVein
repair department.
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
Caution
Consult operating instructions
Refer to instruction manual/booklet
The AV400 emits laser radiation
Type B Applied Part
Conformité Européenne (European Conformity).
This symbol means that the device fully complies
with European Directive 93/42/EEC
ETL Classied means that most system
elements conform to UL Std 60601-1
Product Serial Number
Reference to Catalog Model Number
Universal Product Number
European Representative
Manufacturer
European Union Waste Electrical and Electronic Equipment
Directive Logo. Return is allowed for proper disposal.
Temperature limitation
Humidity limitation
Do not use if package is damaged
Keep dry
Fragile, handle with care
Reusable packaging
Recyclable packaging
No water ingress protection
US federal law restricts sale of this device only to physicians or
other qualied medical professionals.For external use only.
Direct Current. Example: 3.7 V indication of battery voltage
Date of Manufacture
UPN
Rx Only
IPX0
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
4
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • left page
-AV400_RevB_compiled.indd 4 5/22/13 7:48 AM

Warnings and Cautions
Patient and User Safety—Warnings
WARNING: The AV400 enables location of certain supercial veins and is not a
substitute for sound medical judgment based on the visual and tactile location and
assessment of veins. The AV400 should be used only as a supplement to the judgment
of a qualied professional.
WARNING: The AV400 should only be operated when its battery is suciently
charged (indicated when the battery icon at the upper right of the LCD screen is
green), or when the device is operated in a powered hands-free accessory to ensure
AV400 is available for use.
WARNING: AV400 vein location is dependent on a variety of patient factors
and may not display veins on patients with deep veins, skin conditions, hair, scarring
or other highly contoured skin surface, and adipose (fatty) tissue.
WARNING: The AV400 displays only superficial veins, and does so only
to limited depths dependent on a variety of patient factors. The AV400 does not
indicate vein depth.
WARNING: The AV400 emits Visible and Invisible laser radiation. Do not stare
into beam. Class 2 laser product.
WARNING: Do not shine vein display light in eyes.
WARNING: Blink reexes can be inhibited or reduced by disease, drugs, or other
medical conditions. For patients who might have inhibited or reduced blink reexes,
AccuVein recommends outtting the patient with a protective eye shield. A protective
eye shield capable of blocking 642 nm and 785 nm should be used in this case.
WARNING: Stop using the AV400 if the red light does not turn on when the
blue button is pushed.
WARNING: Do not hold the AV400 while performing venipuncture or other
medical procedures.
WARNING: For external use only.
WARNING: Keep the AV400 and its battery out of the reach of children.
WARNING: To view vein location accurately, you must position the AV400 at
the proper height, and directly over the center of the vein being assessed.
Patient and User Safety—Contraindications
WARNING: The AV400 should not be used to locate veins in or near eyes.
WARNING: The AV400 is not intended to be used as a diagnostic device or for
treatment of any kind.
Patient and User Safety—Cautions
CAUTION: United States Federal law restricts this device to sale by or on the order
of a physician or other qualied medical professional.
CAUTION: Operation or use of the AV400 in a manner dierent than specied in this
AV400 User’s Manual may result in hazardous radiation exposure.
CAUTION: Do not make any connections to the USB port on the AV400 device.
Equipment Care— Warnings
WARNING: Do not immerse the AV400 or the AV400 charging cradle in liquid
or wet the AV400 such that liquid spills o.
WARNING: Do not attempt to open, disassemble, or service the battery pack.
Do not crush, puncture, or dispose of the battery in re or water. Do not short external
contacts. Do not expose to temperature above 60°C / 140°F.
WARNING: Do not modify in any way the interior or exterior components of
the AV400.
WARNING: To reduce the risk of re or shock hazard and annoying interference,
use only the recommended accessories and do not expose this equipment to rain or
heavy moisture.
WARNING: Use only AccuVein accessories and replacement parts with the
AV400. The use of non AccuVein accessories may degrade safety.
Equipment Care—Cautions
CAUTION: Use only AccuVein approved battery charging accessories, AccuVein BA300
and BA400 batteries, and AccuVein brand additional accessories. If the AV400 will not
be used for an extended time, remove and store battery in a safe, dust-free location.
The battery should be replaced only with an approved AccuVein battery. Removing
or replacing the battery should not to be done in the patient environment.
CAUTION: Electromagnetic Interference (EMI) can aect the proper performance
of the device. Normal operation can be restored by removing the source of the
interference.
CAUTION: Do not attempt to sterilize the AV400 with heat or pressure sterilization
methods.
CAUTION: The AV400 will not display veins if operated outside its temperature range.
CAUTION: The AV400 may not display veins if operated under bright light such as
bright sunlight.
CAUTION: Dispose of used batteries according to manufacturer’s instructions.
A recyclable lithium ion battery powers the AV400. Please call AccuVein at
(816) 997-9400 for information on how to dispose or recycle this battery.
CAUTION: Do not disassemble or modify the AV400 or any of its charging accessories.
CAUTION: Do not service. The AV400 contains no customer serviceable components.
The AV400 and its accessories should be serviced only by an authorized AccuVein
repair department.
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
Caution
Consult operating instructions
Refer to instruction manual/booklet
The AV400 emits laser radiation
Type B Applied Part
Conformité Européenne (European Conformity).
This symbol means that the device fully complies
with European Directive 93/42/EEC
ETL Classied means that most system
elements conform to UL Std 60601-1
Product Serial Number
Reference to Catalog Model Number
Universal Product Number
European Representative
Manufacturer
European Union Waste Electrical and Electronic Equipment
Directive Logo. Return is allowed for proper disposal.
Temperature limitation
Humidity limitation
Do not use if package is damaged
Keep dry
Fragile, handle with care
Reusable packaging
Recyclable packaging
No water ingress protection
US federal law restricts sale of this device only to physicians or
other qualied medical professionals.For external use only.
Direct Current. Example: 3.7 V indication of battery voltage
Date of Manufacture
UPN
Rx Only
IPX0
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
5
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • right page
-AV400_RevB_compiled.indd 5 5/22/13 7:48 AM

Basic Use and Operation Basic Use and Operation
Product Description
The AV400 operates by using infrared light to detect veins be-
neath the skin, then projecting the position of the veins on the
skin surface directly above the veins. Qualied medical per-
sonnel can observe the vasculature as displayed to assist them
in nding a vein of the right size and position for venipuncture
and other medical procedures requiring the location of super-
cial veins. No training is required to operate the AV400.
The AV400 only shows supercial vasculature. The maximum
depth that veins are displayed varies by patient. In addition,
some patients’ veins or a portion of their veins might not be
displayed well or at all. Causes for less than optimal or lack of
vein display include, but are not limited to, vein depth, skin
conditions (e.g., eczema, tattoos), hair, scarring or other high-
ly contoured skin surface, and adipose (fatty) tissue.
When held directly overhead, the AV400 accurately locates
the center of a vein. Increasing the displacement from direct-
ly overhead results in an oset in the displayed vein position.
Width of displayed vein may dier from the actual width de-
pending on patient to patient dierences and vein width. The
center line of the vein is accurate when the device is being
used correctly and should always be used as the target when
performing venipuncture or other medical procedures.
The AV400 requires no routine or preventative maintenance.
The AV400 is portable internally powered by battery and ap-
proved for continuous operation. The AV400 is considered a
Type B applied part and is not protected by ingress of water.
The technology in the AV400 is covered by the following pat-
ents: US 8,073,531; US 7,904,138; US 8,150,500; US 7,983,738;
US 8,244,333; US 8,255,040; US 8,295,904; US 8,328,368 Mex-
ico 0357588. Other US and international patents pending. Visit
www.accuvein.com for the most current information.
9
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
Basic Use and Operation
Basic Use and Operation
8
Setup
The AV400 Vein Illumination System includes an AV400 vein
illuminator with rechargeable battery, a charging cradle, pow-
er supply and power adaptors, and instructional documents.
AccuVein recommends fully charging the AV400 before rst
use. Connect the charging cradle to the power supply. The
AV400 comes with power supply country adaptors to ensure
outlet compatibility in your country. If necessary, secure a dif-
ferent power adaptor to the power supply and before plug-
ging it into an outlet.
Remove the plastic protective cover from the LCD before use
WARNING: Inspect power cords and supplies for dam-
age. If any damage is found do not use until replacement is
obtained.
CAUTION: Inspect the power adaptor on the power supply to
conrm it is properly secured.
AccuVein also offers optional hands-free stands that many
practitioners find useful (HF410, HF470). For information
about these and other optional accessories visit the AccuVein
web site www.accuvein.com.
Intended Use
The AccuVein® AV400 is a portable, hand-held instrument that
helps medical professionals to locate certain supercial veins.
The AV400 is intended to be used as a supplement to appro-
priate medical training and experience. The AV400 should not
be used as the sole method for locating veins, and should be
used only by a qualied medical professional, who should do
so either prior to palpation to help identify the location of a
vein, or afterwards to conrm or refute the perceived location
of a vein. When using the AV400, medical practitioners should
always follow the appropriate medical protocols and practic-
es as required by their medical facility, as well as exercising
sound medical judgment.
When used properly, the AV400 enables users to locate cer-
tain supercial veins in connection with medical procedures,
such as venipuncture.
The AV400 can be used whenever the determination of vein
location is appropriate.
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
6
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • left page
-AV400_RevB_compiled.indd 6 5/22/13 7:48 AM

Basic Use and Operation
Basic Use and Operation
Product Description
The AV400 operates by using infrared light to detect veins be-
neath the skin, then projecting the position of the veins on the
skin surface directly above the veins. Qualied medical per-
sonnel can observe the vasculature as displayed to assist them
in nding a vein of the right size and position for venipuncture
and other medical procedures requiring the location of super-
cial veins. No training is required to operate the AV400.
The AV400 only shows supercial vasculature. The maximum
depth that veins are displayed varies by patient. In addition,
some patients’ veins or a portion of their veins might not be
displayed well or at all. Causes for less than optimal or lack of
vein display include, but are not limited to, vein depth, skin
conditions (e.g., eczema, tattoos), hair, scarring or other high-
ly contoured skin surface, and adipose (fatty) tissue.
When held directly overhead, the AV400 accurately locates
the center of a vein. Increasing the displacement from direct-
ly overhead results in an oset in the displayed vein position.
Width of displayed vein may dier from the actual width de-
pending on patient to patient dierences and vein width. The
center line of the vein is accurate when the device is being
used correctly and should always be used as the target when
performing venipuncture or other medical procedures.
The AV400 requires no routine or preventative maintenance.
The AV400 is portable internally powered by battery and ap-
proved for continuous operation. The AV400 is considered a
Type B applied part and is not protected by ingress of water.
The technology in the AV400 is covered by the following pat-
ents: US 8,073,531; US 7,904,138; US 8,150,500; US 7,983,738;
US 8,244,333; US 8,255,040; US 8,295,904; US 8,328,368 Mex-
ico 0357588. Other US and international patents pending. Visit
www.accuvein.com for the most current information.
9
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
Basic Use and Operation Basic Use and Operation
8
Setup
The AV400 Vein Illumination System includes an AV400 vein
illuminator with rechargeable battery, a charging cradle, pow-
er supply and power adaptors, and instructional documents.
AccuVein recommends fully charging the AV400 before rst
use. Connect the charging cradle to the power supply. The
AV400 comes with power supply country adaptors to ensure
outlet compatibility in your country. If necessary, secure a dif-
ferent power adaptor to the power supply and before plug-
ging it into an outlet.
Remove the plastic protective cover from the LCD before use
WARNING: Inspect power cords and supplies for dam-
age. If any damage is found do not use until replacement is
obtained.
CAUTION: Inspect the power adaptor on the power supply to
conrm it is properly secured.
AccuVein also offers optional hands-free stands that many
practitioners find useful (HF410, HF470). For information
about these and other optional accessories visit the AccuVein
web site www.accuvein.com.
Intended Use
The AccuVein® AV400 is a portable, hand-held instrument that
helps medical professionals to locate certain supercial veins.
The AV400 is intended to be used as a supplement to appro-
priate medical training and experience. The AV400 should not
be used as the sole method for locating veins, and should be
used only by a qualied medical professional, who should do
so either prior to palpation to help identify the location of a
vein, or afterwards to conrm or refute the perceived location
of a vein. When using the AV400, medical practitioners should
always follow the appropriate medical protocols and practic-
es as required by their medical facility, as well as exercising
sound medical judgment.
When used properly, the AV400 enables users to locate cer-
tain supercial veins in connection with medical procedures,
such as venipuncture.
The AV400 can be used whenever the determination of vein
location is appropriate.
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
7
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • right page
-AV400_RevB_compiled.indd 7 5/22/13 7:48 AM

Basic Use and Operation
Display indications
Do not shine in eyes
Battery charge status
Next selection
Select item
Device in powered hands-free accessory
AV400 controls at a glance
Power AV400 ON or OFF
Turn vein display light ON or OFF
Invert the projection
About Product
Select time-out
Blue Button
Left Button
Center Button
Power Button
11
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
Basic Use and Operation
User controls and screens
LCD Screen
Lens
Lens
Vein
Illumination
Window
Battery
Compartment
Door
Keypad
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
8
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • left page
-AV400_RevB_compiled.indd 8 5/22/13 7:48 AM

Basic Use and Operation
Display indications
Do not shine in eyes
Battery charge status
Next selection
Select item
Device in powered hands-free accessory
AV400 controls at a glance
Power AV400 ON or OFF
Turn vein display light ON or OFF
Invert the projection
About Product
Select time-out
Blue Button
Left Button
Center Button
Power Button
11
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
Basic Use and Operation
User controls and screens
LCD Screen
Lens
Lens
Vein
Illumination
Window
Battery
Compartment
Door
Keypad
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
9
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • right page
-AV400_RevB_compiled.indd 9 5/22/13 7:48 AM

Basic Use and Operation Basic Use and Operation
Hold the device from 4–18" (10 to 45 cm) over the surface
of the skin. Scan the area of interest to view . Once a vein is
selected, make sure the vein display light is centered direct-
ly above the vein’s center line. Don’t move or tilt the device
to either side of the vein or this will oset the projected vein
from its true location beneath the skin. You can then often en-
hance display quality by slightly adjusting the height and an-
gle to the skin. In particular, moving the device closer or fur-
ther from the skin can help bring additional veins into view,
depending on the patient’s vasculature, room lighting, and
depth of the veins.
WARNING: To view vein location accurately, you must
position the AV400 directly over the center of the vein being
assessed.
CAUTION: The AV400 projection may not display veins if op-
erated under bright light such as bright sunlight
Time-out setting:
The time-out setting determines the amount of time until the
vein display light turns o. This setting may help increase bat-
tery operation time of the AV400. To set the time, press the
left button to go into the time-out menu. The time out can be
set to 0 (no time-out), 2 minutes, 5 minutes, or 15 minutes.
Blue button
The blue button is located on the left side of the AV400. Press
the blue button to power on the AV400 and turn on the vein
display light. Press the button again to turn the vein display
light o. This leaves the device on so you can navigate the de-
vice settings. Press the blue side button again to turn the vein
display light back on and begin locating veins again.
Operating the AV400
Positioning the AV400
Power on/o button
Press the on/o button to power on or o the AV400.
13
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
Basic Use and Operation
Basic Use and Operation
Operating screen display
Projection o, ready for use
Projection on—dark veins
over illuminated background
Projection on—illuminated
veins over dark background
Selected time-out
About product
Non-operating screen display
Unit too cold. Bring the
device above 4°C/39°F
Unit too hot. Bring the
device below 33°C/90°F
Less than 2 minutes
of battery charge
remaining. Charge
battery immediately.
Unit faulted. Call
AccuVein Support at
(816) 997-9400.
12
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
10
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • left page
-AV400_RevB_compiled.indd 10 5/22/13 7:48 AM

Basic Use and Operation
Basic Use and Operation
Hold the device from 4–18" (10 to 45 cm) over the surface
of the skin. Scan the area of interest to view . Once a vein is
selected, make sure the vein display light is centered direct-
ly above the vein’s center line. Don’t move or tilt the device
to either side of the vein or this will oset the projected vein
from its true location beneath the skin. You can then often en-
hance display quality by slightly adjusting the height and an-
gle to the skin. In particular, moving the device closer or fur-
ther from the skin can help bring additional veins into view,
depending on the patient’s vasculature, room lighting, and
depth of the veins.
WARNING: To view vein location accurately, you must
position the AV400 directly over the center of the vein being
assessed.
CAUTION: The AV400 projection may not display veins if op-
erated under bright light such as bright sunlight
Time-out setting:
The time-out setting determines the amount of time until the
vein display light turns o. This setting may help increase bat-
tery operation time of the AV400. To set the time, press the
left button to go into the time-out menu. The time out can be
set to 0 (no time-out), 2 minutes, 5 minutes, or 15 minutes.
Blue button
The blue button is located on the left side of the AV400. Press
the blue button to power on the AV400 and turn on the vein
display light. Press the button again to turn the vein display
light o. This leaves the device on so you can navigate the de-
vice settings. Press the blue side button again to turn the vein
display light back on and begin locating veins again.
Operating the AV400
Positioning the AV400
Power on/o button
Press the on/o button to power on or o the AV400.
13
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
Basic Use and Operation Basic Use and Operation
Operating screen display
Projection o, ready for use
Projection on—dark veins
over illuminated background
Projection on—illuminated
veins over dark background
Selected time-out
About product
Non-operating screen display
Unit too cold. Bring the
device above 4°C/39°F
Unit too hot. Bring the
device below 33°C/90°F
Less than 2 minutes
of battery charge
remaining. Charge
battery immediately.
Unit faulted. Call
AccuVein Support at
(816) 997-9400.
12
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
11
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • right page
-AV400_RevB_compiled.indd 11 5/22/13 7:48 AM

Basic Use and Operation Charging, Cleaning and Troubleshooting
Verify the charging cradle is properly connected (see the Set-
up section). Place the AV400 in the charging cradle with the
LCD facing outward to begin charging. While the AV400 is
in its charging cradle, the vein display light does not operate
and the device cannot be used. An animated battery charge
icon will display when the AV400 is charging. The light on the
charging cradle will only turn on when a device is detected. If
the light turns on when there is no device placed in the charg-
ing cradle please call AccuVein support for assistance.
When the AV400 battery gets low the battery indicator will
ash yellow or red and beep repeatedly. Before the battery is
completely depleted, an alert requesting to charge the battery
will be displayed on the screen. The AV400 will resume nor-
mal operation once it has been charged.
In normal operation, the AV400 battery needs to be replaced
about every two years. Longer charge cycles or shorter operat-
ing periods are indications that the battery should be replaced.
To replace the battery, open the battery compartment on the
narrow end of the device’s underside. Use a screwdriver or
edge of a small coin to unscrew the cover. Be careful not to
touch or scratch the lenses on the back of the device. Do NOT
use a tool to remove the battery.
To ensure the battery remains charged and ready for use,
store the AV400 in its charging cradle or a powered Accu-
Vein accessory.
CAUTION: Use only the AccuVein® PS310 power sup-
ply with the CC300 cradle and other AccuVein approved
accessories.
Charging the AV400
15
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
Basic Use and Operation
Inverse setting:
The inverse setting inverts the dark (shadow) and light (red)
projection areas of the AV400 such that the veins can be de-
picted by either light or dark. Changing the inverse setting
may improve vein visibility and eye comfort level. AccuVein
recommends reversing the vein display to show light veins in
cases where veins are very small. When operating the device
outdoors or in bright surroundings, use the dark veins settings
to enhance visibility.
WARNING: Do not shine vein display light in eyes.
WARNING: Blink reexes can be inhibited or reduced by
disease, drugs, or other medical conditions. For patients who
might have inhibited or reduced bright light blink reexes, Ac-
cuVein recommends outtting the patient with a protective
eye shield. A protective eye shield capable of blocking 642 nm
and 785 nm should be used.
Conrm the Vein
Assess the Vasculature
While the vein display light is on, shine the AV400 over the pa-
tient’s skin. You can do this before palpation, scanning quickly
over the skin to help narrow down possible locations, or after
palpation has conrmed vein location and suitability.
After assessing the patient’s vasculature, conrm the site for
your procedure by verifying the location and suitability of
the vein using normal medical techniques and good medi-
cal judgment, such as vein visualization, palpation, and oth-
er medical techniques.
If performing venipuncture
If you are performing venipuncture, you may use an AccuVein
hands-free accessory so both your hands remain free to per-
form the procedure. You can also hand the AV400 to an assis-
tant during the procedure. The AV400 should be used only by
qualied medical professionals and only in conjunction with
appropriate visual and tactile vein assessment techniques.
Operating the AV400 (continued)
You can then often see veins better by rotating the AV400
slightly on its axis, and moving the device closer to or further
away from the skin.
14
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
12
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • left page
-AV400_RevB_compiled.indd 12 5/22/13 7:48 AM

Basic Use and Operation
Charging, Cleaning and Troubleshooting
Verify the charging cradle is properly connected (see the Set-
up section). Place the AV400 in the charging cradle with the
LCD facing outward to begin charging. While the AV400 is
in its charging cradle, the vein display light does not operate
and the device cannot be used. An animated battery charge
icon will display when the AV400 is charging. The light on the
charging cradle will only turn on when a device is detected. If
the light turns on when there is no device placed in the charg-
ing cradle please call AccuVein support for assistance.
When the AV400 battery gets low the battery indicator will
ash yellow or red and beep repeatedly. Before the battery is
completely depleted, an alert requesting to charge the battery
will be displayed on the screen. The AV400 will resume nor-
mal operation once it has been charged.
In normal operation, the AV400 battery needs to be replaced
about every two years. Longer charge cycles or shorter operat-
ing periods are indications that the battery should be replaced.
To replace the battery, open the battery compartment on the
narrow end of the device’s underside. Use a screwdriver or
edge of a small coin to unscrew the cover. Be careful not to
touch or scratch the lenses on the back of the device. Do NOT
use a tool to remove the battery.
To ensure the battery remains charged and ready for use,
store the AV400 in its charging cradle or a powered Accu-
Vein accessory.
CAUTION: Use only the AccuVein® PS310 power sup-
ply with the CC300 cradle and other AccuVein approved
accessories.
Charging the AV400
15
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
Basic Use and Operation
Inverse setting:
The inverse setting inverts the dark (shadow) and light (red)
projection areas of the AV400 such that the veins can be de-
picted by either light or dark. Changing the inverse setting
may improve vein visibility and eye comfort level. AccuVein
recommends reversing the vein display to show light veins in
cases where veins are very small. When operating the device
outdoors or in bright surroundings, use the dark veins settings
to enhance visibility.
WARNING: Do not shine vein display light in eyes.
WARNING: Blink reexes can be inhibited or reduced by
disease, drugs, or other medical conditions. For patients who
might have inhibited or reduced bright light blink reexes, Ac-
cuVein recommends outtting the patient with a protective
eye shield. A protective eye shield capable of blocking 642 nm
and 785 nm should be used.
Conrm the Vein
Assess the Vasculature
While the vein display light is on, shine the AV400 over the pa-
tient’s skin. You can do this before palpation, scanning quickly
over the skin to help narrow down possible locations, or after
palpation has conrmed vein location and suitability.
After assessing the patient’s vasculature, conrm the site for
your procedure by verifying the location and suitability of
the vein using normal medical techniques and good medi-
cal judgment, such as vein visualization, palpation, and oth-
er medical techniques.
If performing venipuncture
If you are performing venipuncture, you may use an AccuVein
hands-free accessory so both your hands remain free to per-
form the procedure. You can also hand the AV400 to an assis-
tant during the procedure. The AV400 should be used only by
qualied medical professionals and only in conjunction with
appropriate visual and tactile vein assessment techniques.
Operating the AV400 (continued)
You can then often see veins better by rotating the AV400
slightly on its axis, and moving the device closer to or further
away from the skin.
14
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
13
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • right page
-AV400_RevB_compiled.indd 13 5/22/13 7:48 AM

Product Specication
Guidance and Manufacturer’s Declara on
The AV400, CC300, and HF470 are intended for use in the electromagnetic environment
specifi ed below. The customer or the user of the AV400, CC300, or HF470 should assure that
it is used in such an environment.
IMMUNITY Test IEC 60601
Test L evel
Compliance Level Electromagne c environment -
guidance
Electrosta c
Discharge (ESD) IEC
61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete,
or ceramic le. If fl oors are covered
with synthe c material, the rela ve
humidity should be at least 30%.
Electrical fast
transient/burst
EC 61000-4-4
± 2 kV for power
supply lines
± 1 kV input/output
lines
± 2 kV for power
supply lines
Not applicable
Mains power quality shouldbe that
of a typical commercial or hospital
environment
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV diff eren al
mode
± 2 kV common
mode
Mains power quality shouldbe that
of a typical commercial or hospital
environment
Voltage dips, short
interrup ons and
voltage varia ons on
power supply input
lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT) for
0.5 cycles
40 % UT
(60 % dip in UT) for
5 cycles
70 % UT
(30 % dip in UT) for
25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
<5 % UT
(>95 % dip in UT)
for 0.5 cycles
40 % UT
(60 % dip in UT) for
5 cycles
70 % UT
(30 % dip in UT) for
25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
Mains power quality should be
that of a typical commerial or
hospital environment. If the user
of the AV400 requires con nued
opera on during power mains
interrup ons, it is recomended
that the AV400 be powered from
an uninterrup ble power suply or
a ba ery
Power frequency
(50/60 Hz) magne c
fi e l d
IEC 61000-4-8
3 A/m 3 A/m Power frequency magne c
fi elds should be at levels
characteris c of a typical loca on
in a typical commercial or hospital
environment.
NOTE: UT is the AC mains voltage prior to applica on of the test level.
Guidance and Manufacturer’s Declara on – Electromagne c Immunity
The AV400, CC300, and HF470 are intended for use in the electromagnetic environment
specifi ed below. The customer or the user of the AV400, CC300, or HF470 should assure that it
is used in such an environment.
IMMUNITY TEST IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
Electromagne c Environment –
Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 Vrms
Portable and mobile RF communica ons
equipment should be used no closer to
any part of the AV400, including cables,
than the recommended separa on
distance calculated from the equa on
applicable to the frequency of the
transmi er.
Recommended separa on distance
d =1,2√—
P
d =1.2√—
P 80 MHz to 800 MHz
d =2.3√—
P 800 MHz to 2.5 GHz
where P is the maximum output
power ra ng of the transmi er in
wa s (W) according to the transmi er
manufacturer and d is the recommended
separa on distance in meters (m). Field
strengths from fi xed RF transmi ers,
as determined by an electromagne c
site survey, should be less than the
compliance level in each frequency
range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situa ons. Electromagne c propaga on is aff ected by
absorp on and refl ec on from structures, objects, and people.
Field strengths from fi xed transmi ers, such as base sta ons for radio (cellular/cordless) telephones
and land mobile radios amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theore cally with accuracy. To assess the electromagne c environment due to fi xed RF
transmi ers, an electromagne c site survey should be considered. If the measured fi eld strength in
the loca on in which the AV400 is used exceeds the applicable RF compliance level above the AV400
should be observed to verify normal opera on. If abnormal performance is observed, addi onal
measures may be necessary, such as re-orien ng or reloca ng the AV400. Over the frequency range
150 kHz to 80 MHz, fi eld strengths should be less than 3 V/m.
Guidance and Manufacturer’s Declara on - Electromagne c Emissions
The AV400, CC300, and HF470 are intended for use in the electromagnetic environment
specifi ed below. The customer or the user of the AV400, CC300, or HF470 should assure
that it is used in such an environment.
Emissions test Compliance Electromagne c Environment Guidance
RF Emissions
CISPR 11
Group 1 The AV400 uses RF energy only for its internal
func on. Therefore, its RF emissions are very
low and not likely to cause any interference in
nearby electronics
RF Emissions
CISPR 11
Class A The AV400 is suitable for use in all
establishments other than domes c and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domes c purposes
Harmonic Emissions
IEC 6100-3-2
Complies
Voltage Fluctua ons
IEC 6100-3-3
Complies
Recommended separa on distances between
portable and mobile RF communica ons equipment and the AV400
The AV400, CC300, and HF470 are intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the AV400, CC300, or HF470
can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the AV400 as recommended
below, according to the maximum output power of the communications equipment.
Rated Maximum
Output
power of transmi er
W
Separa on distance according to frequency of transmi er
m
150 kHz to 80 MHz
d =1.2√—
P
80 MHz to 800 MHz
d =1.2√—
P
800 MHz to 2.5 GHz
d =2.3√—
P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
11.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmi ers rated at a maximum output power not listed above, the recommended separa on
distance d in meters (m) can be es mated using the equa on applicable to the frequency of the
transmi er, where P is the maximum output power ra ng of the transmi er in wa s (W) according to
the transmi er manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separa on distance for the higher
frequency range applies. NOTE 2 These guidelines may not apply in all situa ons. Electromagne c
propaga on is aff ected by absorp on and refl ec on from structures, objects, and people.
17
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
Charging, Cleaning and Troubleshooting
Cleaning and Disinfecting
WARNING: Users should inspect the AV400 and clean and disinfect the
AV400 as required by their institution’s policies to ensure that it is sucient-
ly clean before each use.
WARNING: Do not use AV400 if the vein illumination window is
scratched or dirty. The AV400 should be returned to AccuVein for servicing if
the vein light window is scratched.
For peak performance, the optical surfaces and lenses on the back of the AV400
should be kept clean. Use alcohol wipes or any soft lens wipe with several drops
of 70% isopropyl alcohol. Dirt or scratches on the vein display window show up
as dark shadows in the vein projection.
To clean the body of the AV400 and its charging cradle, you may use a cloth
moistened with 70% isopropyl alcohol, a 10% dilution of chlorine bleach in
distilled water, or soap and water. You can also use common household dis-
infectants and wipes.
Do not immerse the AV400, the AV400 charging cradle, or the AV400 hands-
free stand in liquid or get the AV400 or its components wet such that liquid
spills o. Do not attempt to sterilize the AV400 with heat or pressure steril-
ization methods.
Do not clean the AV400 while in a charging cradle or hands-free accessory.
Do not clean the AV400 when the battery cover is removed.
Unplug the charging cradle before cleaning the cradle.
Troubleshooting
If the AV400 detects that it cannot operate properly, it turns o the vein dis-
play light and displays an alert or fault screen (See Non-Operating Display un-
der User Screens). When the AV400 is not operating properly turn the device o
using the round Power on/o button.
A yellow header on the screen indicates an alert. Alerts appear if you oper-
ate the device in an environment that is too cold (less than 4°C/39°F) or too
hot (more than 33°C/90°F). Should a temperature alert occur, simply bring the
AV400 back within its operating temperature range.
Other alerts indicate when the battery needs to be charged or replaced.
The fault screen has a red header and is picture above. If the fault screen ap-
pears, you should stop using the AV400 immediately. Turn o the device and
contact AccuVein support via telephone at (816) 997-9400 option 3.
Warranty Summary
AccuVein warrants the AV400, when purchased new from an authorized sell-
er, is of sound materials and workmanship and conforms to the manufacturer’s
specications for a period of one year from the date of purchase from AccuVein
or its authorized distributors or agents. The warranty period for accessories is
as follows: Charging Cradle – 1 year; Power Supply – 1 year; Battery – 90 days.
AccuVein will repair or replace any product that does not satisfy these warran-
ties within this time limit at no cost to the customer. You must obtain a Re-
turn of Material Authorization (RMA) number from AccuVein support. The re-
paired unit has a 90-day warranty from the date it is received by the customer.
Contact AccuVein support at (816) 997-9400 option 3 for service or addition-
al information.
These warranties do not cover any misuse or abuse. These warranties are also
void if the instrument is repaired by other than AccuVein or an authorized
agent. There are no other warranties, either express or implied, other than
those set forth in AccuVein’s written warranty.
This is only a summary of the AccuVein warranty coverage. Contact AccuVein
or see www.accuvein.com/legal/warranty for full details of warranty coverage.
Limitation of Liability
In no event shall either AccuVein or the customer be liable to the other for any
incidental, indirect, special or consequential damages (including without lim-
itation lost prots or lost business opportunity) that the other party may incur
by reason of its having entered into or relied on a purchase or lease agreement,
or arising out of the performance or breach of such an agreement, even if the
party against whom a claim is made was advised or knew of the possibility of
such damages. The foregoing limitation shall apply regardless of the form of
the claim in which such liability may be asserted, including breach of contract,
tort (including negligence) or otherwise.
Warranty and Liability
16
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
14
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • left page
-AV400_RevB_compiled.indd 14 5/22/13 7:48 AM

Product Specication
Guidance and Manufacturer’s Declara on
The AV400, CC300, and HF470 are intended for use in the electromagnetic environment
specifi ed below. The customer or the user of the AV400, CC300, or HF470 should assure that
it is used in such an environment.
IMMUNITY Test IEC 60601
Test L evel
Compliance Level Electromagne c environment -
guidance
Electrosta c
Discharge (ESD) IEC
61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete,
or ceramic le. If fl oors are covered
with synthe c material, the rela ve
humidity should be at least 30%.
Electrical fast
transient/burst
EC 61000-4-4
± 2 kV for power
supply lines
± 1 kV input/output
lines
± 2 kV for power
supply lines
Not applicable
Mains power quality shouldbe that
of a typical commercial or hospital
environment
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV diff eren al
mode
± 2 kV common
mode
Mains power quality shouldbe that
of a typical commercial or hospital
environment
Voltage dips, short
interrup ons and
voltage varia ons on
power supply input
lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT) for
0.5 cycles
40 % UT
(60 % dip in UT) for
5 cycles
70 % UT
(30 % dip in UT) for
25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
<5 % UT
(>95 % dip in UT)
for 0.5 cycles
40 % UT
(60 % dip in UT) for
5 cycles
70 % UT
(30 % dip in UT) for
25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
Mains power quality should be
that of a typical commerial or
hospital environment. If the user
of the AV400 requires con nued
opera on during power mains
interrup ons, it is recomended
that the AV400 be powered from
an uninterrup ble power suply or
a ba ery
Power frequency
(50/60 Hz) magne c
fi e l d
IEC 61000-4-8
3 A/m 3 A/m Power frequency magne c
fi elds should be at levels
characteris c of a typical loca on
in a typical commercial or hospital
environment.
NOTE: UT is the AC mains voltage prior to applica on of the test level.
Guidance and Manufacturer’s Declara on – Electromagne c Immunity
The AV400, CC300, and HF470 are intended for use in the electromagnetic environment
specifi ed below. The customer or the user of the AV400, CC300, or HF470 should assure that it
is used in such an environment.
IMMUNITY TEST IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
Electromagne c Environment –
Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 Vrms
Portable and mobile RF communica ons
equipment should be used no closer to
any part of the AV400, including cables,
than the recommended separa on
distance calculated from the equa on
applicable to the frequency of the
transmi er.
Recommended separa on distance
d =1,2√—
P
d =1.2√—
P 80 MHz to 800 MHz
d =2.3√—
P 800 MHz to 2.5 GHz
where P is the maximum output
power ra ng of the transmi er in
wa s (W) according to the transmi er
manufacturer and d is the recommended
separa on distance in meters (m). Field
strengths from fi xed RF transmi ers,
as determined by an electromagne c
site survey, should be less than the
compliance level in each frequency
range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situa ons. Electromagne c propaga on is aff ected by
absorp on and refl ec on from structures, objects, and people.
Field strengths from fi xed transmi ers, such as base sta ons for radio (cellular/cordless) telephones
and land mobile radios amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theore cally with accuracy. To assess the electromagne c environment due to fi xed RF
transmi ers, an electromagne c site survey should be considered. If the measured fi eld strength in
the loca on in which the AV400 is used exceeds the applicable RF compliance level above the AV400
should be observed to verify normal opera on. If abnormal performance is observed, addi onal
measures may be necessary, such as re-orien ng or reloca ng the AV400. Over the frequency range
150 kHz to 80 MHz, fi eld strengths should be less than 3 V/m.
Guidance and Manufacturer’s Declara on - Electromagne c Emissions
The AV400, CC300, and HF470 are intended for use in the electromagnetic environment
specifi ed below. The customer or the user of the AV400, CC300, or HF470 should assure
that it is used in such an environment.
Emissions test Compliance Electromagne c Environment Guidance
RF Emissions
CISPR 11
Group 1 The AV400 uses RF energy only for its internal
func on. Therefore, its RF emissions are very
low and not likely to cause any interference in
nearby electronics
RF Emissions
CISPR 11
Class A The AV400 is suitable for use in all
establishments other than domes c and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domes c purposes
Harmonic Emissions
IEC 6100-3-2
Complies
Voltage Fluctua ons
IEC 6100-3-3
Complies
Recommended separa on distances between
portable and mobile RF communica ons equipment and the AV400
The AV400, CC300, and HF470 are intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the AV400, CC300, or HF470
can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the AV400 as recommended
below, according to the maximum output power of the communications equipment.
Rated Maximum
Output
power of transmi er
W
Separa on distance according to frequency of transmi er
m
150 kHz to 80 MHz
d =1.2√—
P
80 MHz to 800 MHz
d =1.2√—
P
800 MHz to 2.5 GHz
d =2.3√—
P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
11.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmi ers rated at a maximum output power not listed above, the recommended separa on
distance d in meters (m) can be es mated using the equa on applicable to the frequency of the
transmi er, where P is the maximum output power ra ng of the transmi er in wa s (W) according to
the transmi er manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separa on distance for the higher
frequency range applies. NOTE 2 These guidelines may not apply in all situa ons. Electromagne c
propaga on is aff ected by absorp on and refl ec on from structures, objects, and people.
17
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • right page
Charging, Cleaning and Troubleshooting
Cleaning and Disinfecting
WARNING: Users should inspect the AV400 and clean and disinfect the
AV400 as required by their institution’s policies to ensure that it is sucient-
ly clean before each use.
WARNING: Do not use AV400 if the vein illumination window is
scratched or dirty. The AV400 should be returned to AccuVein for servicing if
the vein light window is scratched.
For peak performance, the optical surfaces and lenses on the back of the AV400
should be kept clean. Use alcohol wipes or any soft lens wipe with several drops
of 70% isopropyl alcohol. Dirt or scratches on the vein display window show up
as dark shadows in the vein projection.
To clean the body of the AV400 and its charging cradle, you may use a cloth
moistened with 70% isopropyl alcohol, a 10% dilution of chlorine bleach in
distilled water, or soap and water. You can also use common household dis-
infectants and wipes.
Do not immerse the AV400, the AV400 charging cradle, or the AV400 hands-
free stand in liquid or get the AV400 or its components wet such that liquid
spills o. Do not attempt to sterilize the AV400 with heat or pressure steril-
ization methods.
Do not clean the AV400 while in a charging cradle or hands-free accessory.
Do not clean the AV400 when the battery cover is removed.
Unplug the charging cradle before cleaning the cradle.
Troubleshooting
If the AV400 detects that it cannot operate properly, it turns o the vein dis-
play light and displays an alert or fault screen (See Non-Operating Display un-
der User Screens). When the AV400 is not operating properly turn the device o
using the round Power on/o button.
A yellow header on the screen indicates an alert. Alerts appear if you oper-
ate the device in an environment that is too cold (less than 4°C/39°F) or too
hot (more than 33°C/90°F). Should a temperature alert occur, simply bring the
AV400 back within its operating temperature range.
Other alerts indicate when the battery needs to be charged or replaced.
The fault screen has a red header and is picture above. If the fault screen ap-
pears, you should stop using the AV400 immediately. Turn o the device and
contact AccuVein support via telephone at (816) 997-9400 option 3.
Warranty Summary
AccuVein warrants the AV400, when purchased new from an authorized sell-
er, is of sound materials and workmanship and conforms to the manufacturer’s
specications for a period of one year from the date of purchase from AccuVein
or its authorized distributors or agents. The warranty period for accessories is
as follows: Charging Cradle – 1 year; Power Supply – 1 year; Battery – 90 days.
AccuVein will repair or replace any product that does not satisfy these warran-
ties within this time limit at no cost to the customer. You must obtain a Re-
turn of Material Authorization (RMA) number from AccuVein support. The re-
paired unit has a 90-day warranty from the date it is received by the customer.
Contact AccuVein support at (816) 997-9400 option 3 for service or addition-
al information.
These warranties do not cover any misuse or abuse. These warranties are also
void if the instrument is repaired by other than AccuVein or an authorized
agent. There are no other warranties, either express or implied, other than
those set forth in AccuVein’s written warranty.
This is only a summary of the AccuVein warranty coverage. Contact AccuVein
or see www.accuvein.com/legal/warranty for full details of warranty coverage.
Limitation of Liability
In no event shall either AccuVein or the customer be liable to the other for any
incidental, indirect, special or consequential damages (including without lim-
itation lost prots or lost business opportunity) that the other party may incur
by reason of its having entered into or relied on a purchase or lease agreement,
or arising out of the performance or breach of such an agreement, even if the
party against whom a claim is made was advised or knew of the possibility of
such damages. The foregoing limitation shall apply regardless of the form of
the claim in which such liability may be asserted, including breach of contract,
tort (including negligence) or otherwise.
Warranty and Liability
16
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
15
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • right page
-AV400_RevB_compiled.indd 15 5/22/13 7:48 AM

Production 1P after translator
AV400_ES_RevB.indd • May 22, 2013 7:48 AM • right page keith@typefl ownyc.com
917.370.8219
AccuVein® es una marca registrada de AccuVein Inc. El contenido de este manual, así
como todas sus fotografías y diagramas, es información propiedad de AccuVein Inc.
Última modicación de este manual: Abril de 2013
Advertencias y precauciones
Lea todas las instrucciones, precauciones y advertencias antes de usar el dispositivo.
Únicamente profesionales médicos calicados deben utilizar este producto. El
dispositivo AV400 no debe emplearse como único método de localización de venas y
no sustituye el buen criterio médico ni la localización visual y táctil de las venas.
Antes de utilizar el dispositivo AV400 en un paciente, los profesionales médicos
calicados deben leer y comprender este manual del usuario. Antes de emplear el
dispositivo por primera vez, los usuarios deben comparar la detección de las venas
que realiza el dispositivo AV400 con técnicas de detección visual y palpación.
Manual del usuario de AV400
© AccuVein Inc. 2012
AccuVein Inc.
40 Goose Hill Road
Teléfono: +1 (816) 997-9400
Cold Spring Harbor, NY 11724
Estados Unidos de América
www.accuvein.com
El dispositivo AV400 se fabrica para:
AccuVein Inc.
40 Goose Hill Road
Cold Spring Harbor, NY 11724
Estados Unidos de América
El dispositivo cumple
cabalmente con la Directiva
Europea 2002/364/CE.
Conformité Européenne
(Conformidad europea)
Para clientes de Estados Unidos de
América: IEC/EN 60825-1 Seguridad
de productos láser: Cumple con los
artículos 1040.10 y 1040.11 del título
21 del Código de Normas Federales
(21 CFR 1040.10 y 1040.11), con
la excepción de las desviaciones
conforme al aviso relativo a productos
láser n.º 50, con fecha del 24 de junio
de 2007.
EMERGO EUROPE
Molenstraat 15
2513 BH, La Haya
Países Bajos
Teléfono: +31.70.345.8570
Fax: +31.70.346.7299
Puede descargar este manual del usuario de AV400 e información
adicional de www.accuvein.com.
Longitud de onda nm/ o nm
Potencia máxima mW / mW
Energía de pulso nJ / nJ
Longitud de pulso . ns
IEC/EN-(/)
RADIACIÓN LÁSER
VISIBLE E INVISIBLE.
NO MIRE HACIA EL
HAZ DE LUZ. PRODUCTO
LÁSER CLASE .
Manual del usuario de
AccuVein AV400
Guía del profesional médico para el
uso y funcionamiento de
AccuVein AV400
Español
AV400
Product Specication
Weight 275 grams (9.7 oz.)
Size 5 x 6 x 20 cm (2” x 2.4” x 7.9”)
Battery AccuVein® BA400 user
replaceable Li-Ion 3.6V.
BA400 – 3,100 mAh
Continuous (vein light on)
run time on full charge:
Typically 180 minutes
Number of standard viewing
procedures that can be performed
using a fully charged battery
90
Charge time (full) Typically 3 hours
Product labels
Information Label with Laser Product Notice
(behind battery cover)
Laser Safety Label with Laser Classication Notice
Operating Temperature 4°C to 33°C
(39°F to 90°F)
Humidity 5% to 85% RH
non-condensing
Pressure 75kPa to 106kPa
Transport Temperature -20°C to 50°C
(-4°F to 122°F)
Humidity 5% to 85% RH
non-condensing
Storage Temperature -20°C to 50°C
(-4°F to 122°F)
Humidity 5% to 85% RH
non-condensing
S p e c i c a t i o n s
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
16
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • left page
-AV400_RevB_compiled.indd 16 5/22/13 7:48 AM

Production 1P after translator
AV400_ES_RevB.indd • May 22, 2013 7:48 AM • right page keith@typefl ownyc.com
917.370.8219
AccuVein® es una marca registrada de AccuVein Inc. El contenido de este manual, así
como todas sus fotografías y diagramas, es información propiedad de AccuVein Inc.
Última modicación de este manual: Abril de 2013
Advertencias y precauciones
Lea todas las instrucciones, precauciones y advertencias antes de usar el dispositivo.
Únicamente profesionales médicos calicados deben utilizar este producto. El
dispositivo AV400 no debe emplearse como único método de localización de venas y
no sustituye el buen criterio médico ni la localización visual y táctil de las venas.
Antes de utilizar el dispositivo AV400 en un paciente, los profesionales médicos
calicados deben leer y comprender este manual del usuario. Antes de emplear el
dispositivo por primera vez, los usuarios deben comparar la detección de las venas
que realiza el dispositivo AV400 con técnicas de detección visual y palpación.
Manual del usuario de AV400
© AccuVein Inc. 2012
AccuVein Inc.
40 Goose Hill Road
Teléfono: +1 (816) 997-9400
Cold Spring Harbor, NY 11724
Estados Unidos de América
www.accuvein.com
El dispositivo AV400 se fabrica para:
AccuVein Inc.
40 Goose Hill Road
Cold Spring Harbor, NY 11724
Estados Unidos de América
El dispositivo cumple
cabalmente con la Directiva
Europea 2002/364/CE.
Conformité Européenne
(Conformidad europea)
Para clientes de Estados Unidos de
América: IEC/EN 60825-1 Seguridad
de productos láser: Cumple con los
artículos 1040.10 y 1040.11 del título
21 del Código de Normas Federales
(21 CFR 1040.10 y 1040.11), con
la excepción de las desviaciones
conforme al aviso relativo a productos
láser n.º 50, con fecha del 24 de junio
de 2007.
EMERGO EUROPE
Molenstraat 15
2513 BH, La Haya
Países Bajos
Teléfono: +31.70.345.8570
Fax: +31.70.346.7299
Puede descargar este manual del usuario de AV400 e información
adicional de www.accuvein.com.
Longitud de onda nm/ o nm
Potencia máxima mW / mW
Energía de pulso nJ / nJ
Longitud de pulso . ns
IEC/EN-(/)
RADIACIÓN LÁSER
VISIBLE E INVISIBLE.
NO MIRE HACIA EL
HAZ DE LUZ. PRODUCTO
LÁSER CLASE .
Manual del usuario de
AccuVein AV400
Guía del profesional médico para el
uso y funcionamiento de
AccuVein AV400
Español
AV400
Product Specication
Weight 275 grams (9.7 oz.)
Size 5 x 6 x 20 cm (2” x 2.4” x 7.9”)
Battery AccuVein® BA400 user
replaceable Li-Ion 3.6V.
BA400 – 3,100 mAh
Continuous (vein light on)
run time on full charge:
Typically 180 minutes
Number of standard viewing
procedures that can be performed
using a fully charged battery
90
Charge time (full) Typically 3 hours
Product labels
Information Label with Laser Product Notice
(behind battery cover)
Laser Safety Label with Laser Classication Notice
Operating Temperature 4°C to 33°C
(39°F to 90°F)
Humidity 5% to 85% RH
non-condensing
Pressure 75kPa to 106kPa
Transport Temperature -20°C to 50°C
(-4°F to 122°F)
Humidity 5% to 85% RH
non-condensing
Storage Temperature -20°C to 50°C
(-4°F to 122°F)
Humidity 5% to 85% RH
non-condensing
S p e c i c a t i o n s
keith@typefl ownyc.com
917. 370 .8 219
Production 1Pc after translations
AV400_EN_RevB.indd • May 22, 2013 7:48 AM • left page
17
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • right page
-AV400_RevB_compiled.indd 17 5/22/13 7:48 AM

Conceptos básicos de uso y funcionamiento Conceptos básicos de uso y funcionamiento Índice
Conceptos básicos
de uso y funcionamiento 6
Conguración
Indicaciones
Descripción del producto
Controles y pantallas del usuario
Uso del dispositivo AV400
Carga, limpieza
y solución de problemas13
Carga del dispositivo AV400
Limpieza y desinfección
Solución de problemas
Garantía y
responsabilidad 16
Resumen de la garantía
Limitación de responsabilidad
Especicaciones del
producto 17
Especicaciones
Etiquetas del producto
Manual del usuario de AccuVein AV400
Guía del profesional médico para el uso y funcionamiento
de AccuVein AV400
Español
AV400
5
Production 1P after translator
AV400_ES_RevB.indd • May 22, 2013 7:48 AM • right page keith@typefl ownyc.com
917.370.8219
Conceptos básicos de uso y funcionamiento
Conceptos básicos de uso y funcionamiento Índice
4
Production 1P after translator
AV400_ES_RevB.indd • May 22, 2013 7:48 AM • left page
keith@typefl ownyc.com
917.370.8219
keith@typeownyc.com
917. 370 .8 219
-AV400_RevB_compiled • May 21, 2013 5:41 PM • left page
-AV400_RevB_compiled.indd 18 5/22/13 7:48 AM
Other manuals for AV400
2
Table of contents
Other AccuVein Medical Equipment manuals