ADCMT 5350 User manual

MANUAL NUMBER
C
Printed in Japan
ADC CORPORATION
All rights reserved.
5350
Digital Electrometer
Operation Manual
FOE-00000151A00
First printing Apr 9, 2019
2019
Cover


Safety-1
Safety Summary
FOE-ANZENA01
To ensure thorough understanding of all functions and to ensure efficient use of this instrument, please read the
manual carefully before using. Note that ADC Corporation (hereafter referred to as ADC) bears absolutely no re-
sponsibility for the result of operations caused due to incorrect or inappropriate use of this instrument.
If the equipment is used in a manner not specified by ADC, the protection provided by the equipment may be im-
paired.
• Warning Labels
Warning labels are applied to ADC products in locations where specific dangers exist. Pay care-
ful attention to these labels during handling. Do not remove or tear these labels. If you have any
questions regarding warning labels, please ask your nearest ADC dealer. Our address and phone
number are listed at the end of this manual.
Symbols of those warning labels are shown below together with their meaning.
DANGER: Indicates an imminently hazardous situation which will result in death or serious
personal injury.
WARNING: Indicates a potentially hazardous situation which will result in death or serious
personal injury.
CAUTION: Indicates a potentially hazardous situation which will result in personal injury or
a damage to property including the product.
• Basic Precautions
Please observe the following precautions to prevent fire, burn, electric shock, and personal inju-
ry.
• Use a power cable rated for the voltage in question. Be sure however to use a power cable
conforming to safety standards of your nation when using a product overseas.
• When inserting the plug into the electrical outlet, first turn the power switch OFF and then
insert the plug as far as it will go.
• When removing the plug from the electrical outlet, first turn the power switch OFF and then
pull it out by gripping the plug. Do not pull on the power cable itself. Make sure your hands
are dry at this time.
• Before turning on the power, be sure to check that the supply voltage matches the voltage
requirements of the instrument.
• Connect the power cable to a power outlet that is connected to a protected ground terminal.
Grounding will be defeated if you use an extension cord which does not include a protective
conductor terminal.
• Be sure to use fuses rated for the voltage in question.
• Do not use this instrument with the case open.
• If there is any abnormality such as abnormal heart, smoke, smell or noise, immediately stop
using the instrument, turn off the power and disconnect the plug from the outlet.
Safety Summary

Safety Summary
Safety-2
FOE-ANZENA01
• Do not place anything on the product and do not apply excessive pressure to the product. Al-
so, do not place flower pots or other containers containing liquid such as chemicals near this
product.
• When the product has ventilation outlets, do not stick or drop metal or easily flammable ob-
jects into the ventilation outlets.
• When using the product on a cart, fix it with belts to avoid its drop.
• When connecting the product to peripheral equipment, turn the power off.
• Caution Symbols Used Within this Manual
Symbols indicating items requiring caution which are used in this manual are shown below to-
gether with their meaning.
DANGER: Indicates an item where there is a danger of serious personal injury (death or seri-
ous injury).
WARNING: Indicates an item relating to personal safety or health.
CAUTION: Indicates an item relating to possible damage to the product or instrument or relat-
ing to a restriction on operation.
• Safety Marks on the Product
The following safety marks can be found on ADC products.
: ATTENTION - Refer to manual.
: Protective ground (earth) terminal.
: DANGER - High voltage.
: CAUTION - Risk of electric shock.
• Replacing Parts with Limited Life
The following parts used in the instrument are main parts with limited life.
Replace the parts listed below before their expected lifespan has expired to maintain the perfor-
mance and function of the instrument.
Note that the estimated lifespan for the parts listed below may be shortened by factors such as
the environment where the instrument is stored or used, and how often the instrument is used.
The parts inside are not user-replaceable. For a part replacement, please contact the ADC sales
office for servicing.
Each product may use parts with limited life.
For more information, refer to the section in this document where the parts with limited life are
described.

Safety Summary
Safety-3
FOE-ANZENA01
Main Parts with Limited Life
• Hard Disk Mounted Products
The operational warnings are listed below.
• Do not move, shock and vibrate the product while the power is turned on.
Reading or writing data in the hard disk unit is performed with the memory disk turning at a
high speed. It is a very delicate process.
• Store and operate the products under the following environmental conditions.
An area with no sudden temperature changes.
An area away from shock or vibrations.
An area free from moisture, dirt, or dust.
An area away from magnets or an instrument which generates a magnetic field.
• Make back-ups of important data.
The data stored in the disk may become damaged if the product is mishandled. The hard disc
has a limited life span which depends on the operational conditions. Note that there is no
guarantee for any loss of data.
• Precautions when Disposing of this Instrument
When disposing of harmful substances, be sure dispose of them properly with abiding by the
state-provided law.
Harmful substances: (1) PCB (polycarbon biphenyl)
(2) Mercury
(3) Ni-Cd (nickel cadmium)
(4) Other
Items possessing cyan, organic phosphorous and hexadic chromium
and items which may leak cadmium or arsenic (excluding lead in sol-
der).
Example: fluorescent tubes, batteries
Part name Life
Unit power supply 5 years
Fan motor 5 years
Electrolytic capacitor 5 years
LCD display 6 years
LCD backlight 2.5 years
Memory backup battery 5 years

Environmental Conditions
Safety-4
FOE-ANZENA01
This instrument should be only be used in an indoor area which satisfies the following conditions:
• An area free from corrosive gas
• An area away from direct sunlight
• A dust-free area
• An area free from vibrations
• Altitude of up to 2000 m
Figure-1 Environmental Conditions
• Operating position
Figure-2 Operating Position
• Storage position
Figure-3 Storage Position
• The classification of the transient over-voltage, which exists typically in the main power supply, and
the pollution degree is defined by IEC61010-1 and described below.
Impulse withstand voltage (over-voltage) category II defined by IEC60364-4-443
Pollution Degree 2
Vibration
Direct sunlight
Corrosive
gas
Dust
Front
The instrument must be used in a hor-
izontal position.
To prevent the internal temperature
from rising, cooling fans are installed
in some instruments.
The air vents on the case must be un-
blocked.
A clear space of 10 centimeters or more
must be kept around the air vents.
Front
This instrument should be stored in a horizontal
position.
When placed in a vertical (upright) position for
storage or transportation, ensure the instrument is
stable and secure.
-Ensure the instrument is stable.
-Pay special attention not to fall.

Safety-5
Types of Power Cable
FOE-ANZENA01
Replace any references to the power cable type, according to the following table, with the appropriate power cable
type for your country.
Plug configuration Standards Rating, color
and length
Model number
(Option number)
PSE: Japan
Electrical Appliance and
Material Safety Law
125 V at 7 A
Black
2 m (6 ft)
Straight: A01402
Angled: A01412
UL: United States of America
CSA: Canada
125 V at 7 A
Black
2 m (6 ft)
Straight: A01403
(Option 95)
Angled: A01413
CEE: Europe
DEMKO: Denmark
NEMKO: Norway
VDE: Germany
KEMA: The Netherlands
CEBEC: Belgium
OVE: Austria
FIMKO: Finland
SEMKO: Sweden
250 V at 6 A
Gray
2 m (6 ft)
Straight: A01404
(Option 96)
Angled: A01414
SEV: Switzerland 250 V at 6 A
Gray
2 m (6 ft)
Straight: A01405
(Option 97)
Angled: A01415
SAA: Australia, New Zealand 250 V at 6 A
Gray
2 m (6 ft)
Straight: A01406
(Option 98)
Angled: ---------
BS: United Kingdom 250 V at 6 A
Black
2 m (6 ft)
Straight: A01407
(Option 99)
Angled: A01417
CCC:China 250 V at 10 A
Black
2 m (6 ft)
Straight: A114009
(Option 94)
Angled: A114109


5350 Digital Electrometer Operation Manual
C-1
TABLE OF CONTENTS
1. PREFACE ...................................................................................................... 1-1
1.1 Product Overview .............................................................................................. 1-1
1.2 Supplied Accessories ........................................................................................ 1-2
1.3 Input Cables ...................................................................................................... 1-3
1.4 Optional Accessories ......................................................................................... 1-4
1.5 Operating Environment ..................................................................................... 1-5
1.6 Unpacking Inspection ........................................................................................ 1-6
1.7 Cleaning, Storage, and Transport Methods ...................................................... 1-7
1.7.1 Cleaning ...................................................................................................... 1-7
1.7.2 Storage ........................................................................................................ 1-7
1.7.3 Transport ..................................................................................................... 1-7
1.8 Life Limited Parts ............................................................................................. 1-8
1.9 Product Disposal and Recycling ....................................................................... 1-9
2. FRONT PANEL AND REAR PANEL .................................................. 2-1
2.1 Front Panel ........................................................................................................ 2-1
2.1.1 Display ........................................................................................................ 2-1
2.1.2 Control Panel .............................................................................................. 2-3
2.1.3 Input Terminal Section ............................................................................... 2-6
2.2 Rear Panel ......................................................................................................... 2-7
2.3 Displayed Messages .......................................................................................... 2-8
3. BASIC MEASUREMENTS ...................................................................... 3-1
3.1 Power Supply .................................................................................................... 3-1
3.1.1 Power Specifications .................................................................................. 3-1
3.1.2 Changing Power Voltage, and Checking and Replacing Power Fuse ........ 3-2
3.1.3 Connecting Power Cable ........................................................................... 3-4
3.1.4 Power ON and Self Test ............................................................................. 3-5
3.1.5 Setting Line Frequency ............................................................................... 3-6
3.1.6 Warm Up .................................................................................................... 3-6
3.2 DC Voltage Measurement ................................................................................ 3-7
3.2.1 Connection for DC Voltage Measurement ................................................. 3-8
3.2.2 Precautions for DC Voltage Measurement ................................................. 3-11
3.3 DC Current Measurement ................................................................................ 3-12
3.3.1 Connection for DC Current Measurement ................................................ 3-13
3.3.2 Precautions for DC Current Measurement ................................................. 3-16
4. MENU ............................................................................................................. 4-1
4.1 Menu Structure .................................................................................................. 4-1
4.2 Menu Operation Overview ................................................................................ 4-3
4.2.1 Parameters to Be Selected .......................................................................... 4-4
4.2.2 Parameters to Be Numerically Input .......................................................... 4-5
4.2.3 Parameters to Be Executed ......................................................................... 4-7
TABLE OF CONTENTS

5350 Digital Electrometer Operation Manual
Table of Contents
C-2
5. FUNCTIONAL REFERENCE ................................................................. 5-1
5.1 Sampling Operation .......................................................................................... 5-1
5.1.1 Sampling FREE RUN/HOLD .................................................................... 5-2
5.1.2 Sampling Rate ............................................................................................ 5-2
5.1.3 Trigger Delay .............................................................................................. 5-3
5.1.4 Sampling Count .......................................................................................... 5-4
5.1.5 Sampling Interval ....................................................................................... 5-4
5.2 Integration Time ................................................................................................ 5-6
5.3 Measurement Range ......................................................................................... 5-7
5.3.1 Measurement Auto Range .......................................................................... 5-8
5.3.2 Measurement Range Upper Limit/Lower Limit ......................................... 5-8
5.3.3 Measurement Auto Range Level ............................................................... 5-9
5.4 Auto Zero ......................................................................................................... 5-10
5.5 Driving Guard ................................................................................................... 5-11
5.6 Zero Check ....................................................................................................... 5-13
5.7 Measurement Data Display .............................................................................. 5-14
5.7.1 Measurement Data Display OFF ................................................................ 5-14
5.7.2 Measurement Data Display Digits .............................................................. 5-14
5.8 Math Operations ................................................................................................ 5-15
5.8.1 NULL Operation ........................................................................................ 5-15
5.8.2 Smoothing Operation .................................................................................. 5-16
5.8.3 Averaging Operation ................................................................................. 5-18
5.8.4 Totalizing Operation ................................................................................... 5-19
5.8.5 Scaling Operation ....................................................................................... 5-21
5.9 Display OFF ...................................................................................................... 5-22
5.10 Setting Parameters ............................................................................................ 5-23
5.10.1 Setting Parameters at Power ON ................................................................ 5-23
5.10.2 Setting Parameter Saving ........................................................................... 5-23
5.10.3 Setting Parameter Loading ........................................................................ 5-24
5.10.4 Setting Parameter Initialization ................................................................. 5-24
5.11 Measurement Data Memory .............................................................................. 5-28
5.11.1 Measurement Memory Store ...................................................................... 5-28
5.11.2 Measurement Data Memory Recall ............................................................ 5-29
5.12 System Functions ............................................................................................. 5-31
5.12.1 Buzzer ......................................................................................................... 5-31
5.12.2 Self Test ..................................................................................................... 5-32
5.12.3 Key Test ...................................................................................................... 5-32
5.12.4 Display Test ................................................................................................ 5-33
5.12.5 Error Log .................................................................................................... 5-33
5.12.6 Information Display .................................................................................... 5-34
5.12.7 Panel Lock .................................................................................................. 5-35
5.12.8 Password Setting ........................................................................................ 5-37
6. REMOTE CONTROL ................................................................................ 6-1
6.1 Interface Overview ............................................................................................ 6-1
6.1.1 Interface Selection ...................................................................................... 6-1
6.1.2 Compatibility with Former Model .............................................................. 6-3
6.2 GPIB .................................................................................................................. 6-5

5350 Digital Electrometer Operation Manual
Table of Contents
C-3
6.2.1 GPIB Overview .......................................................................................... 6-5
6.2.2 Precautions when Using GPIB ................................................................... 6-6
6.2.3 GPIB Setting ............................................................................................... 6-6
6.3 USB ................................................................................................................... 6-8
6.3.1 USB Overview ............................................................................................ 6-8
6.3.2 Connecting with Personal Computer .......................................................... 6-8
6.3.3 USB Setting ................................................................................................ 6-8
6.4 LAN ................................................................................................................... 6-10
6.4.1 LAN Overview ........................................................................................... 6-10
6.4.2 LAN Setting ................................................................................................ 6-10
6.4.3 Communication Setting on Computer ........................................................ 6-10
6.4.4 Network Connection ................................................................................... 6-11
6.4.5 IP Connection Check .................................................................................. 6-11
6.4.6 Communication Method ............................................................................. 6-12
6.5 Status Register ................................................................................................... 6-13
6.5.1 Event Register ............................................................................................ 6-14
6.5.2 Enable Register ........................................................................................... 6-21
6.5.3 Status Registers in 8240-Compatible Mode ............................................... 6-22
6.6 Measurement Data Format (Talker Format) ..................................................... 6-26
6.6.1 Measurement Data ...................................................................................... 6-26
6.6.2 Memory Recall Data ................................................................................... 6-29
6.7 Remote Commands ........................................................................................... 6-30
6.7.1 Command Syntax ....................................................................................... 6-30
6.7.2 Data Format ................................................................................................ 6-32
6.7.3 Remote Command Index ............................................................................ 6-33
6.7.4 Remote Command List ............................................................................... 6-35
6.7.5 Sample Program ......................................................................................... 6-45
6.8 External Control Signals ................................................................................... 6-46
6.8.1 TRIGGER IN (External Trigger Input) ...................................................... 6-46
6.8.2 COMPLETE OUT (Measurement End Output) ......................................... 6-48
6.8.3 Selecting Pulse Width of Measurement End Output .................................. 6-50
6.8.4 Selecting Output Mode of Measurement End Output ................................ 6-51
6.8.5 D/A OUT (D/A Output) ............................................................................. 6-52
6.8.6 Preamplifier Output (AMP OUT) .............................................................. 6-55
7. PERFORMANCE TEST ............................................................................ 7-1
7.1 Performance test ................................................................................................ 7-1
7.2 Connections ....................................................................................................... 7-1
7.3 Test Method ...................................................................................................... 7-1
8. CALIBRATION ........................................................................................... 8-1
8.1 Instruments and Cable Necessary for Calibration ............................................. 8-2
8.2 Precautions ........................................................................................................ 8-3
8.3 Connections ....................................................................................................... 8-4
8.4 Calibration Points and Tolerance Ranges ......................................................... 8-6
8.5 Remote Calibration ........................................................................................... 8-8
8.5.1 Remote Calibration Flowchart ................................................................... 8-8
8.5.2 Remote Calibration Data Saving ................................................................ 8-11

5350 Digital Electrometer Operation Manual
Table of Contents
C-4
8.5.3 Remote Calibration Data Initialization ....................................................... 8-11
8.6 Manual Calibration ............................................................................................ 8-12
8.6.1 Voltage Measurement Calibration (DCV) .................................................. 8-12
8.6.2 Current Measurement Calibration (DCI) .................................................... 8-14
8.6.3 D/A OUT Calibration (DAOUT) ............................................................... 8-16
8.6.4 Manual Calibration Data Saving and Ending ............................................. 8-17
8.6.5 Manual Calibration Data Initialization ....................................................... 8-18
9. SPECIFICATIONS ...................................................................................... 9-1
9.1 DC Voltage Measurement ................................................................................. 9-1
9.2 DC Current Measurement ................................................................................. 9-2
9.3 Measurement Time and Display Digits ............................................................. 9-4
9.4 Math Operations ................................................................................................ 9-5
9.5 Advanced Setting Functions ............................................................................. 9-6
9.6 Interface Function ............................................................................................. 9-7
9.7 General Specifications ...................................................................................... 9-9
APPENDIX ................................................................................................................. A-1
A.1 When Problems Occur (Before Requesting Repairs) ........................................ A-1
A.2 Error Message List ............................................................................................ A-3
A.3 Execution Time ................................................................................................. A-5
DIMENSIONAL OUTLINE DRAWING........................................................... EXT-1
ALPHABETICAL INDEX ..................................................................................... I-1

F-1
5350 Digital Electrometer Operation Manual
LIST OF ILLUSTRATIONS
No. Title Page
2-1 Screen Annotations ........................................................................................................... 2-1
2-2 Keys on Control Panel ......................................................................................................2-3
2-3 Input Terminal Section .....................................................................................................2-6
2-4 Rear Panel ......................................................................................................................... 2-7
3-1 Connecting Power Cable .................................................................................................. 3-4
4-1 Menu Operation Overview ............................................................................................... 4-3
5-1 Sampling Operation Overview ......................................................................................... 5-1
5-2 Relationship between Sampling Interval and Measurement Time ................................... 5-5
5-3 Driving Guard Principles .................................................................................................. 5-12
5-4 Math Operation Diagram .................................................................................................. 5-15
5-5 Smoothing Operation ........................................................................................................5-16
5-6 Averaging Operation ......................................................................................................... 5-18
5-7 Totalizing Operation ......................................................................................................... 5-19
6-1 Status Register Overview .................................................................................................. 6-13
6-2 Status Register Structure ................................................................................................... 6-15
6-3 Status Byte Register Structure .......................................................................................... 6-16
6-4 Status Register Structure in 8240-Compatible Mode ....................................................... 6-22
6-5 TRIGGER IN Equivalent Circuit ..................................................................................... 6-47
6-6 Operation Timing of TRIGGER IN and COMPLETE OUT ............................................ 6-47
6-7 COMPLETE OUT Equivalent Circuit .............................................................................. 6-48
6-8 Operation Timing of COMPLETE OUT (Single Output) ................................................ 6-49
6-9 Relationship between Measured Data and D/A Output Voltage ...................................... 6-52
6-10 Internal Connection of AMP OUT Terminal .................................................................... 6-56
8-1 Connection for Voltage Measurement Calibration at Zero Position (Red-Blue Shorted) 8-4
8-2 Connection for Voltage Measurement Calibration at Full Scale in 200 mV Range ......... 8-4
8-3 Connection for Voltage Measurement Calibration at Full Scale in 2000 mV/20 V Range 8-4
8-4 Connection for Current Measurement Calibration at Zero Position (Input Open) ........... 8-4
8-5 Connection for Current Measurement Calibration at Full Scale in 200 pA to 200 A Range 8-5
8-6 Connection for Current Measurement Calibration at Full Scale in 2000 A to 20 mA Range 8-5
8-7 Connection for D/A OUT ................................................................................................. 8-5
8-8 Calibration Flow ............................................................................................................... 8-8
8-9 Voltage Measurement Calibration Flow ........................................................................... 8-9
8-10 Current Measurement Calibration Flow ........................................................................... 8-10
8-11 D/A OUT Calibration Flow .............................................................................................. 8-11


T-1
5350 Digital Electrometer Operation Manual
LIST OF TABLES
No. Title Page
1-1 Supplied Accessory List ................................................................................................... 1-2
1-2 Optional Accessory List ....................................................................................................1-4
1-3 Life Limited Parts ............................................................................................................. 1-8
1-4 Substances and Components to Be Sorted ........................................................................ 1-9
2-1 Displayed Messages .......................................................................................................... 2-8
3-1 Power Specification .......................................................................................................... 3-1
3-2 Maximum Allowable Input Voltage for DC Voltage Measurement ................................ 3-11
3-3 Maximum Allowable Input Voltage for DC Current Measurement ................................. 3-16
3-4 Voltage between Input Terminals at Overload ................................................................. 3-17
4-1 Menu Structure ................................................................................................................. 4-1
5-1 Measurement Function and Measurement Range ............................................................. 5-7
5-2 Driving Guard and Input Cable ........................................................................................ 5-11
5-3 Setting Parameters Default Values ................................................................................... 5-25
5-4 List of Events to Activate Buzzer ..................................................................................... 5-31
6-1 Interface Setting Items and Factory Default ..................................................................... 6-1
6-2 Interface Function ............................................................................................................. 6-5
6-3 Standard Bus Cable ........................................................................................................... 6-6
6-4 LAN Setting Items ............................................................................................................ 6-10
6-5 Communication Settings when Using LAN ..................................................................... 6-10
6-6 Status Byte Register (STB) ............................................................................................... 6-17
6-7 Standard Event Status Register (SESR) ............................................................................ 6-18
6-8 Device Event Status Register (DESR) .............................................................................. 6-19
6-9 Error Register (ERR) ........................................................................................................ 6-20
6-10 Status Byte Register in 8240-Compatible Mode .............................................................. 6-23
6-11 Standard Event Status Register in 8240-Compatible Mode ............................................. 6-24
6-12 Error Register in 8240-Compatible Mode ........................................................................ 6-25
6-13 Main Header ..................................................................................................................... 6-26
6-14 Sub Header ........................................................................................................................ 6-26
6-15 Mantissa Part and Exponent Part ...................................................................................... 6-27
6-16 Mantissa Part and Exponent Part at Error Occurrence ..................................................... 6-28
6-17 Block Delimiter List ......................................................................................................... 6-28
6-18 Remote Command List ..................................................................................................... 6-35
6-19 Sample Program List ........................................................................................................ 6-45
6-20 List of External Control Signals ....................................................................................... 6-46
6-21 TRIGGER IN Signal Level ............................................................................................... 6-46
6-22 COMPLETE OUT Signal Level ....................................................................................... 6-48
6-23 D/A OUT Specifications ................................................................................................... 6-52
6-24 D/A Output Setting List .................................................................................................... 6-53
6-25 Output Voltage of AMP OUT Terminal ........................................................................... 6-55
6-26 Maximum Allowable Input Voltage of AMP OUT Terminal .......................................... 6-55

T-2
5350 Digital Electrometer Operation Manual
List of Tables
No. Title Page
8-1 Instruments Necessary for Calibration ............................................................................. 8-2
8-2 Cables Necessary for Calibration ..................................................................................... 8-2
8-3 Voltage Measurement Calibration Ranges ....................................................................... 8-6
8-4 Current Measurement Calibration Ranges ........................................................................ 8-6
8-5 D/A OUT Calibration Ranges ........................................................................................... 8-7
A-1 Items to be Inspected before Requesting Repair .............................................................. A-1
A-2 Error Message List ............................................................................................................ A-3

5350 Digital Electrometer Operation Manual
1. PREFACE
1-1
1. PREFACE
1.1 Product Overview
The 5350 is a 51/2-digit Digital Electrometer that features a DC voltage measurement function and a DC cur-
rent measurement function.
• DC voltage measurement with high-input impedance of 1013 ranging from 1 V to 19.9999 V
• DC current measurement with a wide range from 1 fA to 19.9999 mA
• Both high-speed measurement and low-noise measurement possible by setting the integration time
from 0.5 ms to 3200.0 ms with a resolution of 0.1 ms
• GPIB and USB as standard interface, and LAN as factory option
Remote commands compatible with those of the former model 8240
• Measurement data memory to store up to 100,000 data
• Various math operations: NULL, smoothing, averaging, totalizing and scaling operations
• D/A out function to D/A convert measurement data

5350 Digital Electrometer Operation Manual
1.2 Supplied Accessories
1-2
1.2 Supplied Accessories
You must fully examine the supplied accessories listed below when you receive the 5350. If any accessories
are missing or damaged, contact an ADC CORPORATION sales representative. Specify the part number
when ordering.
Table 1-1 Supplied Accessory List
Name Part number Quantity Remarks
Power cable *1 A01402 1
Input cable A01010 1
Power fuse *2 DFT-AAR5A-1 1 100 VAC/120 VAC
DFT-AAR315A-1 220 VAC/240 VAC
Quick Manual E5350(Q) 1 English
Operation Manual (CD-ROM) E5350 1 English
*1: The power cable can be changed by specifying the option at the time of order.
For more information, refer to "Safety Summary."
*2: Either of fuses is included depending on the power supply option.

5350 Digital Electrometer Operation Manual
1.3 Input Cables
1-3
1.3 Input Cables
The supplied input cable is a double coaxial cable.
100 cm
Red
Black
Blue
Rated voltage: 300 Vrms AC Rated current: 2 A
1. Connector (TXA104-P) 3. Alligator clip (middle)
2. Triaxial cable
Internal connection
Red
Black
Blue

5350 Digital Electrometer Operation Manual
1.4 Optional Accessories
1-4
1.4 Optional Accessories
The 5350 optional accessories are listed below. Specify the part number when ordering.
Table 1-2 Optional Accessory List
Name Part number Remarks
Input cable (TRIAX-TRIAX) A01009
Input cable (TRIAX-BNC) A01011
Connector (TRIAXJ-TRIAXJ) A04201 Relay adaptor
Connector (TRIAXJ-BNCP) A04202
A04203
Connector (BNCJ-MP) A04207
Connector (BNCJ-TRIAXP) CC015006
Connector (TRIAXJ receptacle) A04208
BNC-alligator clip MI-03
Input output cable (BNC-BNC 1.5 m) A01036-1500
Safety cable A01044 Red, black 1 of each
Banana tip adapter A08531 Red, black 1 of each
Alligator clip adapter A08532 Red, black 1 of each
Voltage divider probe 12602
Test lead 12603
Test fixture 12701A, 12706A, 12702A/B,
12704A, 12708, 42, 12707
Standard resistance 15045 series
JIS rack mount set A02263
JIS rack mount set (twin) A02264
EIA rack mount set A02463
EIA rack mount set (twin) A02464
Panel mount set A02039
Panel mount set (twin) A02040
Table of contents
Other ADCMT Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Evident
Evident Vanta GX user manual

GW Instek
GW Instek GLA-1016 user manual

Precision Digital Corporation
Precision Digital Corporation ProtEX FarVu PD6870 instruction manual

Hioki
Hioki PW3365-20 quick start guide

Sperry instruments
Sperry instruments SPR-300 PLUS A operating instructions

BiOptic
BiOptic Qsep100 Operation Quick Start Guide