Adixen ASM 142 D User manual

1
CONDENSED MANUAL
• References refer to a specific chapter of the
User’s Manual.
• For further information, please refer to User’s
Manual supplied with your unit.
© AXESS
ASM 142/142 D/GRAPH/GRAPH D Detector connections................ 2
ASM 142/142 D Operator interface .... 3
ASM GRAPH/GRAPH D Operator interface 4
Start-up......................... 5
User interface level ................ 5
Test cycles ....................... 6
Helium signal analog display.......... 7
Audio alarm...................... 7
Zero function ..................... 8
Air inlet vent ..................... 9
Calibration....................... 10
Assistance to the test................ 11
P/N: 107768 Ed.03
Insert your
Condensed Manual
on the unit cover

2
DETECTOR CONNECTIONS
Main power
cord
Main power
switch “I”
Neutral gas purge (ASM 142 D and
ASM GRAPH D only)
Left side Right side
Printer interface
/headphone/
loudspeaker (option)
RS232 interface
(option)
I/O
interface
(option)
Quick connector
exhaust
Long distance sniffer
quick disconnect

3
OPERATOR INTERFACE ASM 142/142 D 1 Inlet port pressure analog display
2 Control and menu selection indicators (ON when activated)
3 Auto-calibration START/ABORT control key
4 Sniffing mode ON/OFF control key
5 Auto-zero ON/OFF control key
6 Cycle START/STOP control key
7 Control keys (4 keys)
8 Standby ON/OFF indicator
9 Evacuation ON/OFF indicator
10 Test ON/OFF indicator
11 Helium signal analogic display
12 Helium signal analogic scale ON/OFF indicator
13 Helium signal Zero scale ON/OFF indicator
14 Correction factor COR indicator (applied to digital display)
15 Units ON/OFF indicator
16 Helium signal digital display
17 Alphanumeric display (4 lines x 20 characters)
18 Parameter function keys (1 key per display line)
19 Modification access keys (4 keys)
20 NEXT : next display/parameter circular function
21/22 Plus or minus value adjustment, parameter selection, audio volume
adjustment keys
23 RESET of previously displayed values (cancels temporary inputs)
24 Menu selection access key (4 keys)
25 SET POINT menu selection key
26 SPECTRO calibration and analyzer cell configuration menu selection
key
27 MAINTENANCE menu selection key
28 OTHER menus selection key (test mode selection, inlet VENT
selection, date/time)
29 Remote control connection
AUTO CAL ZERO
NEXT
-
+
RESET
CYCLE
SNIFFER
SET POINTS SPECTRO
MAINTENANCE
OTHER
F1
F3
F2
F4
10
mbar
hPa
Torr
-3
10
-12
10
-11
10
-10
10
-9
10
-8
10
-7
10
-6
10
-5
10
-4
10
-3
10
-2
10
-2
10
-1
110
1
10
2
10
3
COR
mbar.l/s
STDBY
INLET PRESSURE
LEAK RATE
EVAC TEST
Pa.m
3
/s
Torr.l/s
13
12
1098
2
3
1
11
16
18 17
20 21 22 23
25
24
19
26 27 28
29
14
15
4 5
7
6
Setting and maintenance part Operation part

OPERATOR INTERFACE ASM GRAPH/GRAPH D 1 Inlet port pressure analog display
2 Control and menu selection indicators (ON when activated)
3 Auto-calibration START/ABORT control key
4 Sniffing mode ON/OFF control key
5 Auto-zero ON/OFF control key
6 Cycle START/STOP control key
7 Control keys (4 keys)
8 Standby ON/OFF indicator
9 Evacuation ON/OFF indicator
10 Test ON/OFF indicator
11 Helium signal analogic display
12 Helium signal analogic scale ON/OFF indicator
13 Helium signal Zero scale ON/OFF indicator
14 Correction factor COR indicator (applied to digital display)
15 Units of measurement selection
16 Helium signal digital display
17 Menu display (4 lines)
18 Spectro pressure analog display
19 Modification access keys (4 keys)
20 RESET: next display/parameter circular function
21/22 Plus or minus value adjustment, parameter selection, audio volume
adjustment keys
23 NEXT of previously displayed values (cancels temporary inputs)
24 Menu selection access key (4 keys)
25 SET POINT menu selection key
26 SPECTRO calibration and analyzer cell configuration menu selection
key
27 MAINTENANCE menu selection key
28 OTHER menus selection key (test mode selection, inlet VENT
selection, date/time)
29 Remote control connection: connect if before switching on
the detector
30 Graphic interface selection key
31 Color touch screen
4
AUTO CAL ZERO
CYCLE
SNIFFER
SET POINTS SPECTRO
MAINTENANCE
OTHER
10
-12
10
-11
10
-10
10
-9
10
-8
10
-7
10
-6
10
-5
10
-4
10
-3
10
-2
COR
mbar.l/s
STDBY
LEAK RATE
EVAC TEST
Pa.m
3
/s
Torr.l/s
13
12
1098
11
16
17
25 26 27
29
14
15
1
18
3 4 5
7
6
2
20
22 2321
19
24
28
30
31
Setting and maintenance part Operation part

The detector offers 4 user interface levels for this section to
accomodate any application requirements.
START-UP USER INTERFACE LEVEL
After unpacking the unit, please fill
up the roughing pump with oil, as
indicated in the User’s Manual
E 750
1 Connect the main cable from the detector to the
proper power outlet.
2Depress the main switch to position “I”. On the
control panel, the indicator lights flash.
3 The following screens are shown on the LCD.
4 When the detection pump
reaches its nominal speed,
the unit autocalibrates
itself.
5 When calibration is
completed, the unit is
ready to start a cycle.
STDBYSTDBY
F1
F3 F2
F4
PLEASE WAIT…
ALCATEL ASM...
UNIT : mbar.l/s
SOFT : Lxx Vx.x rxx
F1
F3 F2
F4
PLEASE WAIT…
DETECTION PUMP: 75%
F1
F3 F2
F4
AUTOCAL IN PROGRESS
ELECTR. ZERO ADJUST.
80
F1
F3 F2
F4
READY FOR CYCLE
INLET VENT : off
To know your user interface level and to change it C 120
Setting and maintenance part User part
LEVEL
11
This level has very limited information on the alphanumeric display (LCD). This level is generally selected for production
types of applications.
No access to control keys (Cycle key
included).
LEVEL
22
This level allows the operator to visualize some parameters without the possibility of making any
changes.
Same as Level
11
, this level is usually selected for production types of applications.
Access to all the control keys.
LEVEL
33
Same as level
22
but with possibility to set some parameters. This level is generally selected for
maintenance applications.
LEVEL
44
This level allows access to all the parameters and is generally used for settings all the parameters.
Note: When switching from level
44
to any other level, the switch can be performed without using the
password. This level is generally selected for R&D applications.
ASM 142/GRAPH
5

6
TEST CYCLES
HARD VACUUM TEST MODE
Leak detector in stand-by mode; connect the part or assembly
to be tested to the detector.
SNIFFING TEST MODE
Leak detector in stand-by mode; connect the long distance sniffer
probe to the quick connector.
STDBYSTDBY
F1
F3 F2
F4
READY FOR CYCLE
INLET VENT : off
STDBYSTDBY
F1
F3 F2
F4
READY FOR CYCLE
INLET VENT : off
CYCLE
STDBYSTDBY
EVACEVAC
EVACEVAC
TESTTEST
SNIFFER
SNIFFER
SNIFFER
SNIFFER
CYCLE
TESTTEST
STDBYSTDBY
Starting a cycle
Ending a cycle
Starting Sniffing test mode
Ending Sniffing test mode
(1) As soon as the inlet pressure reaches 10 mbar (7.5 torr),
the unit goes in gross leak test mode, or when the pressure
has reached 5.10-1 mbar (0.37 torr), the unit goes in fine
leak test mode.
F1
F3 F2
F4
READY FOR CYCLE
INLET VENT : off
F1
F3 F2
F4
SNIFFING MODE
PLEASE WAIT…
F1
F3 F2
F4
SNIFFING MODE
Selecting a test mode C 210
(1) Read leak
value

7
He SIGNAL ANALOG SCALE DISPLAY AUDIO ALARM
Leak detector in hard vacuum or sniffing
test mode and zero function not activated.
How to read the He signal analog scale?
reject point is visualized by a blinking led.
if the leak value exceeds the reject point, the leds will turned red (the blinking led will turn orange).
if the leak value remains under the reject point, the leds will remain green.
Example : Reject point = 1 x 10-7 mbar.l/s
The audio alarm offers 2 modes of operation. They are both linked to the zero function.
zero function not activated: the audio alarm starts when the He signal exceeds a fixed set point:
this set point is programmable.
zero function activated: the audio alarm is modulated with respect to the position of the helium
background.
To active/deactive audio alarm C 520
Blinking led = reject point
AUTO CAL ZERO
NEXT
-
+
RESET
CYCLE
SNIFFER
SET POINTS SPECTRO
MAINTENANCE
OTHER
F1
F3
F2
F4
10
mbar
hPa
Torr
-3
10
-12
10
-11
10
-10
10
-9
10
-8
10
-6
10
-5
10
-4
10
-3
10
-2
10
-2
10
-1
110
1
10
2
10
3
COR
mbar.l/s
STDBY
INLET PRESSURE
LEAK RATE
EVAC TEST
Pa.m
3
/s
Torr.l/s
10
-7

Purpose: the zero function offers the operator the possibility to detect small leaks that are smaller than the
helium background.
Activation of zero function: connect the part or installation to be tested.
Deactivation of zero function:
the digital display shows the standard He signal.
Use Helium signal analog scale.
Analog display:
• When zero function is activated, use the Helium signal zero scale.
• The He signal zero scale displays 2 leds signal centered around the zero value.
Example:
Operator could find an example in the User’s Manual C 540 8
ZERO FUNCTION
CYCLE
on the digital display, the detector He background displays.
ZERO
the digital display becomes 0.0E-00. On and after this time, it will display only Helium variation.
ZERO
2 leds flash
1x10-7 1x10-8 1.1x10-7 9x10-8 2x10-7
AUTO CAL ZERO
NEXT
-
+
RESET
CYCLE
SNIFFER
SET POINTS SPECTRO
MAINTENANCE
OTHER
F1
F3
F2
F4
10
mbar
hPa
Torr
-3
10
-12
10
-11
10
-10
10
-9
10
-8
10
-6
10
-5
10
-4
10
-3
10
-2
10
-2
10
-1
110
1
10
2
10
3
COR
mbar.l/s
STDBY
INLET PRESSURE
LEAK RATE
EVAC TEST
Pa.m
3
/s
Torr.l/s
10
-7

9
AIR INLET VENT
Purpose: the air inlet valve vents the inlet of the detector back to atmosphere at the end of the test.
• The indicator “inlet valve = off” indicates that the venting valve is not activated (= closed)
at the end of cycle.
• The setting by default is «off» (= valve closed).
F1
F3 F2
F4
READY FOR CYCLE
INLET VENT : off
F3F3
(opening) /
(closing)
F3F3
Opening / closing air inlet valve
(user level
22
,
33
or
44
)
To open/close air inlet valve with user level
11
C 500

10
CALIBRATION
Internal: The internal autocalibration is automatically activated during
the start-up process. It doesn’t require any operator action.
Thanks to the initial autocalibration, the leak detector can be
immediately operational.
Internal autocalibration on request: it can be started by the operator
whenever needed (the unit has to be off-cycle).
The result of the autocalibration process is displayed.
External: The external autocalibration allows direct readout in cases of
operation with an auxiliary pumping system.
F1
F3 F2
F4
READY FOR CYCLE
INLET VENT : off
F1
F3 F2
F4
READY FOR CYCLE
INLET VENT : off
F1
F3 F2
F4
AUTOCAL IN PROGRESS
ELECTR. ZERO ADJUST.
80
To perform an external calibration C 303
AUTOCAL

11
ASSISTANCE TO THE TEST The leak detector offers to the user 4 interesting functions in
order to improve test.
MEMO FUNCTION. . . . . . Memorization of the latest He signal measured after depressing the CYCLE key at
the end of the cycle.
CYCLE END . . . . . . . . . . . . Automatic control of the roughing and measure timers.
BARGRAPH ZOOM . . . . Display a greater resolution of the He signal around the reject point.
ON THE REJECT POINT
HELIUM POLLUTION . . . Device that prevents the unit from getting polluted with Helium.
PREVENTION
Memo function
Cycle end
Bargraph zoom on the reject point
Helium pollution prevention
To active/deactive this function and adjust display time of the leak value
C 550
To active/deactive this function and adjust roughing and measure times C 530
To active/deactive this function C 510
To active/deactive this function C 560
This manual suits for next models
3
Table of contents
Other Adixen Security Sensor manuals
Popular Security Sensor manuals by other brands

C.P. Electronics
C.P. Electronics MWS1A-C-PRM-LV installation guide

Alarmcom
Alarmcom Flex IO LED Patterns and Troubleshooting

Timeguard
Timeguard WFPIRB Assembly/Installation/Operating Instructions

C.P. Electronics
C.P. Electronics MWS3 Product Data and Installation Sheet

C.P. Electronics
C.P. Electronics EBDSPIR-PRM-IP-LT30 installation guide

C.P. Electronics
C.P. Electronics EBDSPIR-PRM-VFC installation guide