OPERATION
Please make sure that you have already read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model JM852
• Connect the black and red terminals to the Voltage AC range of your DMM or voltage measuring
instrument. Select the appropriate voltage range (2V AC). Clamp the probe around the conductor to be
tested. If the reading is less than 200 mV, select the lower range until you obtain the best resolution.
Read the value display on the DMM and multiply it by the probe ratio (3000/1 = 3000). (If reading =
0.592V AC, the current flowing through the probe is 0.592 x 3000 = 1776A).
• For best accuracy: carefully center the conductor inside the probe jaw, avoid if possible, the proximity of
other conductors which may create noise.
Tips for Making Precise Measurements
• When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
• Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE:
Warning
• For maintenance use only original replacement parts.
• To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
• To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents
into the probe.
Cleaning
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use a very fine sand paper (fine
600) to avoid scratching the jaw, then gently clean with a soft oiled cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA#
on the outside of the shipping container.
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive Dover, NH 03820 USA
(800) 945-2362 (Ext. 360) • (603) 749-6434 (Ext. 360) • [email protected]
(Or contact your authorized distributor)
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or
application of this instrument, please contact our technical hotline:
(800) 343-1391 • (508) 698-2115 • [email protected]
LIMITED WARRANTY
The current probe is warranted to the owner for a period of one year from the date of original purchase
against defects in manufacture. This limited warranty is given by AEMC®Instruments, not by the
distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused
or if the defect is related to service not performed by AEMC®Instruments.
Full warranty coverage and product registration is available on our website at:
www.aemc.com/warranty.html.
Please print the online Warranty Coverage Information for your records.
AC Current Probe
Model JM852
User Manual
DESCRIPTION
The JM852 (Catalog #2113.41) is designed for use in industrial environments. The “squared” jaws permit
multiple conductor or bus bar positioning. The voltage output makes it the perfect tool for measurement
with DMMs, recorders, power and harmonic meters. The Model JM852 is compatible with any voltmeter,
multimeter, or other voltage measuring instrument with an input impedance greater than 100K. To
achieve the stated accuracy, use the JM852 with the DMM having an accuracy of 0.75%.
WARNING
These safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument.
• Read the instruction manual completely and follow all the safety information before attempting
to use or service this instrument.
• Use caution on any circuit: Potentially high voltages and currents may be present and may
pose a shock hazard.
• Read the Safety Specifications section prior to using the current probe. Never exceed the
maximum voltage ratings given.
• Safety is the responsibility of the operator.
• ALWAYS connect the current probe to the display device before clamping the probe onto the
sample being tested.
• ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use.
Replace any defective parts immediately.
• NEVER use the current probe on electrical conductors rated above 600V in overvoltage
category III (CAT III). Use extreme caution when clamping around bare conductors or bus bars.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use
only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before
using the instrument.
This is a type A current sensor. This symbol signifies that application around and removal
from HAZARDOUS LIVE conductors is permitted.
DEFINITION OF MEASUREMENT CATEGORIES
Cat. I: For measurements on circuits not directly connected to the AC supply wall outlet such as
protected secondaries, signal level, and limited energy circuits.
Cat. II: For measurements performed on circuits directly connected to the electrical distribution system.
Examples are measurements on household appliances or portable tools.
Cat. III: For measurements performed in the building installation at the distribution level such as on
hardwired equipment in fixed installation and circuit breakers.
Cat. IV: For measurements performed at the primary electrical supply (<1000V) such as on primary
overcurrent protection devices, ripple control units, or meters.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately
with the carrier and notify your distributor at once, giving a detailed description of any damage.
ELECTRICAL SPECIFICATIONS
Current Range:
1 to 2400A AC, continuous cycle for
the full temperature range
(3000A if temperature < 35°C or 95°F)
Transformation Ratio:
3000:1
Output Signal:
0.333mV AC/A AC (1V at 3000A)
Accuracy and Phase Shift*:
Primary current 150A 600A 3000A
Accuracy % 1.5% 0.75% 0.5%
Phase shift 1.5° 0.75° 0.5°
(*Reference conditions: 23°C±3°K, 20 to 85%
RH, 48 to 65 Hz, external magnetic field < 40
A/m, no DC component, no external current
carrying conductor, test sample centered.)
Load impedance 1M.
Overload: 3600A during 5 mn
Accuracy: Per IEC 185-26-27
Class 0.5 from 48 to 1000 Hz
Frequency Range:
30 Hz to 5 kHz; current derating above
1 kHz for continuous use
Ampere Second Product: 90 AA
Load Impedance: ≥100K
Working Voltage: 600V AC
Common Mode Voltage: 600V AC
Influence of Adjacent Conductor:
0.005A/A AC
Influence of Conductor in Jaw Opening:
1% ± 0.1A of reading
MECHANICAL SPECIFICATIONS
Operating Temperature:
14° to 122°F (-10° to 50°C)
Storage Temperature:
-13° to 176°F (-25° to 80°C)
Influence of Temperature:
< 0.2% per 10°K
Jaw Opening: 3.54" (90mm)
Maximum Conductor Size:
Cable: 2.52" max. (64mm)
Bus bar: 1.97 x 5.31" (50 x 135mm)
2.52 x 3.94" (64 x 100mm)
Envelope Protection:
IP 20 (IEC 529)
Drop Test:
500mm (IEC 68-2-32)
Mechanical Shock:
100 g (IEC 68-2-27)
Vibration:
10/55/10Hz, 0.15 mm (IEC 68-2-6)
Polycarbonate Material:
Handles: 10% fiberglass charged
polycarbonate UL 94 V0
Jaws: ABS UL 94 V2
Dimensions:
4.72 x 12.40 x 1.89" (120 x 315 x 48 mm)
Weight:
2.65 lbs. (1200 g)
Colors:
Dark gray handles with red jaws
Output:
5 ft (1.5m) lead with safety 4mm banana plug
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation
between primary or secondary and outer case of
handle upon IEC 1010-2-32
Common Mode Voltage:
600V CAT III, Pollution: 2
ORDERING INFORMATION
Current Probe JM852 ...................Cat #2113.41
Accessories:
Banana plug adapter
(to non-recessed plug)....................Cat #1017.45
Black
S2
Red
S1
4.72"
(120 mm)
1.89"
(48 mm)
12.40"
(315 mm)
P1 P2
Ø =2.52" (64 mm)
Cable max.
1.97 x 5.31" (50 x 135 mm) or
2.52 x 3.94" (64 x 100 mm)
Bus Bar max.
ELECTRICAL SPECIFICATIONS
Current Range:
1 to 2400A AC, continuous cycle for
the full temperature range
(3000A if temperature < 35°C or 95°F)
Transformation Ratio:
3000:1
Output Signal:
0.333mV AC/A AC (1V at 3000A)
Accuracy and Phase Shift*:
Primary current 150A 600A 3000A
Accuracy % 1.5% 0.75% 0.5%
Phase shift 1.5° 0.75° 0.5°
(*Reference conditions: 23°C±3°K, 20 to 85%
RH, 48 to 65 Hz, external magnetic field < 40
A/m, no DC component, no external current
carrying conductor, test sample centered.)
Load impedance 1M.
Overload: 3600A during 5 mn
Accuracy: Per IEC 185-26-27
Class 0.5 from 48 to 1000 Hz
Frequency Range:
30 Hz to 5 kHz; current derating above
1 kHz for continuous use
Ampere Second Product: 90 AA
Load Impedance: ≥100K
Working Voltage: 600V AC
Common Mode Voltage: 600V AC
Influence of Adjacent Conductor:
0.005A/A AC
Influence of Conductor in Jaw Opening:
1% ± 0.1A of reading
MECHANICAL SPECIFICATIONS
Operating Temperature:
14° to 122°F (-10° to 50°C)
Storage Temperature:
-13° to 176°F (-25° to 80°C)
Influence of Temperature:
< 0.2% per 10°K
Jaw Opening: 3.54" (90mm)
Maximum Conductor Size:
Cable: 2.52" max. (64mm)
Bus bar: 1.97 x 5.31" (50 x 135mm)
2.52 x 3.94" (64 x 100mm)
Envelope Protection:
IP 20 (IEC 529)
Drop Test:
500mm (IEC 68-2-32)
Mechanical Shock:
100 g (IEC 68-2-27)
Vibration:
10/55/10Hz, 0.15 mm (IEC 68-2-6)
Polycarbonate Material:
Handles: 10% fiberglass charged
polycarbonate UL 94 V0
Jaws: ABS UL 94 V2
Dimensions:
4.72 x 12.40 x 1.89" (120 x 315 x 48 mm)
Weight:
2.65 lbs. (1200 g)
Colors:
Dark gray handles with red jaws
Output:
5 ft (1.5m) lead with safety 4mm banana plug
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation
between primary or secondary and outer case of
handle upon IEC 1010-2-32
Common Mode Voltage:
600V CAT III, Pollution: 2
ORDERING INFORMATION
Current Probe JM852 ...................Cat #2113.41
Accessories:
Banana plug adapter
(to non-recessed plug)....................Cat #1017.45
Black
S2
Red
S1
4.72"
(120 mm)
1.89"
(48 mm)
12.40"
(315 mm)
P1 P2
Ø =2.52" (64 mm)
Cable max.
1.97 x 5.31" (50 x 135 mm) or
2.52 x 3.94" (64 x 100 mm)
Bus Bar max.
99-MAN 100126.v4 07/13
OPERATION
Please make sure that you have already read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model JM852
• Connect the black and red terminals to the Voltage AC range of your DMM or voltage measuring
instrument. Select the appropriate voltage range (2V AC). Clamp the probe around the conductor to be
tested. If the reading is less than 200 mV, select the lower range until you obtain the best resolution.
Read the value display on the DMM and multiply it by the probe ratio (3000/1 = 3000). (If reading =
0.592V AC, the current flowing through the probe is 0.592 x 3000 = 1776A).
• For best accuracy: carefully center the conductor inside the probe jaw, avoid if possible, the proximity of
other conductors which may create noise.
Tips for Making Precise Measurements
• When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
• Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE:
Warning
• For maintenance use only original replacement parts.
• To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
• To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents
into the probe.
Cleaning
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use a very fine sand paper (fine
600) to avoid scratching the jaw, then gently clean with a soft oiled cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA#
on the outside of the shipping container.
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive Dover, NH 03820 USA
(800) 945-2362 (Ext. 360) • (603) 749-6434 (Ext. 360) • [email protected]
(Or contact your authorized distributor)
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or
application of this instrument, please contact our technical hotline:
(800) 343-1391 • (508) 698-2115 • [email protected]
LIMITED WARRANTY
The current probe is warranted to the owner for a period of one year from the date of original purchase
against defects in manufacture. This limited warranty is given by AEMC®Instruments, not by the
distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused
or if the defect is related to service not performed by AEMC®Instruments.
Full warranty coverage and product registration is available on our website at:
www.aemc.com/warranty.html.
Please print the online Warranty Coverage Information for your records.
AC Current Probe
Model JM852
User Manual
DESCRIPTION
The JM852 (Catalog #2113.41) is designed for use in industrial environments. The “squared” jaws permit
multiple conductor or bus bar positioning. The voltage output makes it the perfect tool for measurement
with DMMs, recorders, power and harmonic meters. The Model JM852 is compatible with any voltmeter,
multimeter, or other voltage measuring instrument with an input impedance greater than 100K. To
achieve the stated accuracy, use the JM852 with the DMM having an accuracy of 0.75%.
WARNING
These safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument.
• Read the instruction manual completely and follow all the safety information before attempting
to use or service this instrument.
• Use caution on any circuit: Potentially high voltages and currents may be present and may
pose a shock hazard.
• Read the Safety Specifications section prior to using the current probe. Never exceed the
maximum voltage ratings given.
• Safety is the responsibility of the operator.
• ALWAYS connect the current probe to the display device before clamping the probe onto the
sample being tested.
• ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use.
Replace any defective parts immediately.
• NEVER use the current probe on electrical conductors rated above 600V in overvoltage
category III (CAT III). Use extreme caution when clamping around bare conductors or bus bars.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use
only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before
using the instrument.
This is a type A current sensor. This symbol signifies that application around and removal
from HAZARDOUS LIVE conductors is permitted.
DEFINITION OF MEASUREMENT CATEGORIES
Cat. I: For measurements on circuits not directly connected to the AC supply wall outlet such as
protected secondaries, signal level, and limited energy circuits.
Cat. II: For measurements performed on circuits directly connected to the electrical distribution system.
Examples are measurements on household appliances or portable tools.
Cat. III: For measurements performed in the building installation at the distribution level such as on
hardwired equipment in fixed installation and circuit breakers.
Cat. IV: For measurements performed at the primary electrical supply (<1000V) such as on primary
overcurrent protection devices, ripple control units, or meters.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately
with the carrier and notify your distributor at once, giving a detailed description of any damage.

Other AEMC instruments Measuring Instrument manuals

AEMC instruments CA7027 User manual

AEMC instruments

AEMC instruments CA7027 User manual

AEMC instruments 5070 User manual

AEMC instruments

AEMC instruments 5070 User manual

AEMC instruments K100 User manual

AEMC instruments

AEMC instruments K100 User manual

AEMC instruments 407 User manual

AEMC instruments

AEMC instruments 407 User manual

AEMC instruments PowerPad IV 8345 User manual

AEMC instruments

AEMC instruments PowerPad IV 8345 User manual

AEMC instruments MN103 User manual

AEMC instruments

AEMC instruments MN103 User manual

AEMC instruments AmpFlex User manual

AEMC instruments

AEMC instruments AmpFlex User manual

AEMC instruments 6527 User manual

AEMC instruments

AEMC instruments 6527 User manual

AEMC instruments MD304 User manual

AEMC instruments

AEMC instruments MD304 User manual

AEMC instruments DIGITAL FLEXPROBE 400D-6 User manual

AEMC instruments

AEMC instruments DIGITAL FLEXPROBE 400D-6 User manual

AEMC instruments MN379T User manual

AEMC instruments

AEMC instruments MN379T User manual

AEMC instruments Clamp-On 3705 User manual

AEMC instruments

AEMC instruments Clamp-On 3705 User manual

AEMC instruments 5050 Operating and safety instructions

AEMC instruments

AEMC instruments 5050 Operating and safety instructions

AEMC instruments 6526 User manual

AEMC instruments

AEMC instruments 6526 User manual

AEMC instruments MD305 User manual

AEMC instruments

AEMC instruments MD305 User manual

AEMC instruments Chauvin Arnoux 6412 User manual

AEMC instruments

AEMC instruments Chauvin Arnoux 6412 User manual

AEMC instruments MN261 User manual

AEMC instruments

AEMC instruments MN261 User manual

AEMC instruments A196-24-BK User manual

AEMC instruments

AEMC instruments A196-24-BK User manual

AEMC instruments MD303 User manual

AEMC instruments

AEMC instruments MD303 User manual

AEMC instruments MN01 User manual

AEMC instruments

AEMC instruments MN01 User manual

AEMC instruments MR525 User manual

AEMC instruments

AEMC instruments MR525 User manual

AEMC instruments 6550 User manual

AEMC instruments

AEMC instruments 6550 User manual

AEMC instruments MN213 User manual

AEMC instruments

AEMC instruments MN213 User manual

AEMC instruments 6536 User manual

AEMC instruments

AEMC instruments 6536 User manual

Popular Measuring Instrument manuals by other brands

Hioki 8841 instruction manual

Hioki

Hioki 8841 instruction manual

Common CGR-01 Series manual

Common

Common CGR-01 Series manual

YOKOGAWA 96065 user manual

YOKOGAWA

YOKOGAWA 96065 user manual

PCB Piezotronics IMI SENSORS RTD602D11 Installation and operating manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS RTD602D11 Installation and operating manual

mercor MCRP 054 User instruction manual

mercor

mercor MCRP 054 User instruction manual

Mitutoyo LGF-110L-B manual

Mitutoyo

Mitutoyo LGF-110L-B manual

Hioki 3447-01 instruction manual

Hioki

Hioki 3447-01 instruction manual

Hioki 3661-20 instruction manual

Hioki

Hioki 3661-20 instruction manual

HEIDENHAIN LS 608 C Replacing Instructions

HEIDENHAIN

HEIDENHAIN LS 608 C Replacing Instructions

Kessel SonicControl Installation, operation and maintenance instructions

Kessel

Kessel SonicControl Installation, operation and maintenance instructions

Meriam meriGauge Plus user manual

Meriam

Meriam meriGauge Plus user manual

Bacharach Fyrite INSIGHT PLUS instruction manual

Bacharach

Bacharach Fyrite INSIGHT PLUS instruction manual

CS Instruments FA 410 instruction manual

CS Instruments

CS Instruments FA 410 instruction manual

PCB Piezotronics 353B77 Installation and operating manual

PCB Piezotronics

PCB Piezotronics 353B77 Installation and operating manual

Electro Industries/GaugeTech Shark 100T quick start guide

Electro Industries/GaugeTech

Electro Industries/GaugeTech Shark 100T quick start guide

Wentworth Technology Pegasus S200 Installation and operation manual

Wentworth Technology

Wentworth Technology Pegasus S200 Installation and operation manual

HORNER XL7 user manual

HORNER

HORNER XL7 user manual

Smart Caregiver TL-2015R2 Installation and use instructions

Smart Caregiver

Smart Caregiver TL-2015R2 Installation and use instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.