AGFA ADC Compact Plus User manual

ADC Compact Plus
Digitizer
User manual
A
D
C
C
O
M
P
A
C
T
P
L
U
S

22231C EN 20040224
For more information on Agfa products and Agfa HealthCare products, please visit www.agfa.com, your
Point of Knowledge.
© Agfa-Gevaert N.V. 2003.
No parts of this document may be reproduced, copied, adapted or transmitted in any form or by any means
without the written permission of Agfa-Gevaert N.V.
Agfa-Gevaert N.V. makes no warranties or representation, expressed or implied, with respect to the
accuracy, completeness or usefulness of the information contained in this document and specifically
disclaims warranties of suitability for any particular purpose. Agfa-Gevaert N.V. shall under no
circumstances be liable for any damage arising from the use or inability to use any information, apparatus,
method or process disclosed in this document.
Agfa-Gevaert N.V. reserves the right to make changes to this document without prior notice.
Agfa-Gevaert N.V., Septestraat 27, B-2640 Mortsel, Belgium.
ADC Compact Plus is a trademark of Agfa-Gevaert N.V., Belgium.
Agfa and Agfa-Rhombus are trademarks of Agfa-Gevaert AG, Germany.

3
2231C EN 20040224
Table of contents
Chapter 1: Introducing the ADC Compact Plus...............................................5
ADC Compact Plus, intended use ....................................................................6
ADC Compact Plus features.............................................................................7
Safety precautions ...........................................................................................8
Safety compliance.......................................................................................... 12
Operating modes ...........................................................................................13
The user interface..........................................................................................14
Switching on the ADC Compact Plus..............................................................21
Switching off the ADC Compact Plus..............................................................23
Resetting the ADC Compact Plus...................................................................24
Chapter 2: Basic operation (‘Operator mode’)...............................................25
Reading an image plate .................................................................................26
Reading an emergency image plate ...............................................................29
Re-erasing an image plate .............................................................................32
Chapter 3: Advanced operation (‘Key-operator mode’)................................37
Survey of advanced functions.........................................................................38
Preventive maintenance................................................................................. 39
General procedure in case of malfunction.......................................................41
Troubleshooting .............................................................................................42
Clearing cassette jams...................................................................................43
Clearing image plate jams..............................................................................47
Appendix A: Equipment information sheet....................................................51
Specifications.................................................................................................52
Appendix B: ADC Compact cassette..............................................................55
Safety precautions .........................................................................................56
Description of the ADC Compact cassette......................................................57
Cleaning the image plate................................................................................59
Cleaning the cassettes...................................................................................61
Technical specifications of the ADC Compact cassette...................................62
Technical specifications of the image plate .....................................................64

42231C EN 20040224

Introducing the
ADC Compact Plus
This chapter draws attention to important safety
precautions and introduces the ADC Compact Plus.
!ADC Compact Plus, intended use
!ADC Compact Plus features
!Safety precautions
!Safety compliance
!Operating modes
!The user interface
!Switching on the ADC Compact Plus
!Switching off the ADC Compact Plus
!Resetting the ADC Compact Plus
Chapter
1

62231C EN 20040224
ADC Compact Plus, intended use
This device must only be used to scan exposed X-ray cassettes, containing
an erasable image plate (IP). This device is part of a system, consisting of X-
ray cassettes with erasable phosphor image plates, an identification station
for the cassettes and a workstation where the resulting digital image
information is further processed and routed. It is intended that this device is
only operated in a radiological environment by qualified staff.

7
2231C EN 20040224
ADC Compact Plus features
The ADC Compact Plus scans the exposed ADC image plate, converts the
information into digital data and automatically transfers the image to the
image processing station for further processing and visualization.
The ADC Compact Plus requires but little manual interaction. All you have to
do, after exposure and identification of the cassette, is to place it in the input
buffer of the ADC Compact Plus. You can deposit up to 10 cassettes of
different sizes simultaneously in the input buffer. The Digitizer takes in the
cassettes one by one. The Digitizer reads the demographic data and routing
information from the memory chip in the cassette, opens the cassette,
removes the image plate and scans the latent image by means ofa sweeping
laser beam.
Once the image is digitized, the cassette is returned to the output buffer to be
used for new exposures. After a full Digitizer cycle, the plate has turned 180°
in the cassette.
Depending on the X-ray intensity which has affected the phosphor during the
exposure, more or less light will be emitted during laser scanning. The light is
converted into an electrical signal. This signal is then converted into a digital
bit stream. Once converted into digital form, the digitized image is transferred
to the image processing station for further processing and visualization.
Further features of the ADC Compact Plus include:
"The ADC Compact Plus permits assigning the status ‘emergency’ to an
image. An emergency image will be given priority by the image processing
station.
"The ADC Compact Plus permits re-erasing an image plate before re-using it.
In specific cases, this is necessary to prevent ghost images caused by
previous exposures or stray radiation from interfering with the image of
interest. You can erase a batch of up to 9 image plates.

82231C EN 20040224
Safety precautions
General safety instructions
•For software and other technical platforms, and/or in combination with any
consumable, which constitute, after installation, a system for the interpretation of
medical image data: such system is used by trained and qualified professionals.
It is the user's responsibility to ensure that image quality, display quality,
environmental lighting and other possible distractions are consistent with the
clinical application. The user must be aware, that automatic collimation could
possibly lead to misinterpretation of the image.
•Make sure that the ADC Compact Plus is constantly monitored in order to avoid
inappropriate handling, especially by children.
•Only trained service personnel must make repairs. Only authorized service
personnel must make changes to the ADC Compact Plus.
•If there is any visible damage to the machine casing, do not start nor use the
ADC Compact Plus.
•If you want to connect the ADC Compact Plus with other devices, components or
assemblies and if the technical data do not permit determining whether the
combination with these devices, components or assemblies involves hazards,
you must consult the respective manufacturers to avoid danger for operating
personnel or the environment.
•Do not override or disconnect the integrated safety features.
•As is the case for all technical devices, the ADC Compact Plus must be
operated, cared for and serviced correctly.
•If you don’t operate the ADC Compact Plus correctly or if you don’t have it
serviced correctly, Agfa-Gevaert is not liable for resulting disturbances, damages
or injuries.
•When installing the ADC Compact Plus, care must be taken to ensure that there
is either a mains plug or an all-cable disconnecting device in the internal
installation fitted near the ADC Compact Plus and that it is easily accessible.
•If you notice conspicuous noise or smoke, disconnect the ADC Compact Plus
immediately.
•Check that the mains voltage is within the specified range of the self adapting
power supply of the machine.

9
2231C EN 20040224
Markings and labels
Always take into account the markings and labels provided on the inside and
outside of the machine. A brief overview of these markings and labels and
their meaning is given below.
Safety warning, indicating that the ADC Compact Plus Manuals
should be consulted before making any connections to other
equipment. The use of accessory equipment not complying with
the equivalent safety requirements of this Digitizer may lead to a
reduced level of safety of the resulting system. Consideration
relating to the choice of accessory equipment shall include:
• Use of the accessory equipment in the patient vicinity,
• Evidence that the safety certification of the accessory equipment
has been performed in accordance with the appropriate IEC 601-
1 and IEC 601-1-1 harmonized national standard.
In addition all configurations must comply with the medical
electrical systems standard IEC 601-1-1. The party that makes the
connections acts as system configurator and is responsible for
complying with the systems standard.
If required contact your local service organization.
In order to reduce the risk of electric shock, do not remove any
covers.
Caution hot:
Keep hands clear from the erasure unit.
Type B equipment:
Indicates that the ADC Compact Plus complies with the limits for
type B equipment.

10 2231C EN 20040224
•You can hurt your fingers if they are caught between the ADC Cassette and the
edge of the input slot. Insert the cassette in the input buffer as described in
‘Reading an image plate’ on page 26. At all times, keep your fingers clear of the
input slot. As soon as the ADC Compact Plus takes in the cassette, release it.
Supplementary protective earth connector:
Provides a connection between the ADC Compact Plus and the
potential equalization busbar of the electrical system as found in
medical environments. This plug should never be unplugged
before the power is turned off and the power plug has been
removed.
Intergrounding connector:
Provides a connection between the Digitizer and other equipment
which might exhibit minor ground potential differences. These
differences may degrade the quality of communication between
different equipment. Never remove connections to this terminal.
Protective earth (ground):
Provides a connection between the Digitizer and the protective
earth of the mains. Do not remove this connection, because this
will have a negative influence on the leakage current.
Power on
Power off
Note that the power cord has to be disconnected from the wall
outlet in order to disconnect the unit entirely from the mains.
Precautions for use in USA only:
Make sure that the circuit is single-phase center-tapped, if the
Digitizer is connected to a 240 V/60 Hz source instead of a 120 V/
60 Hz source.

11
2231C EN 20040224
TÜV safety issues
Accessory equipment connected to the analog and digital interfaces must be
certified according to the respective IEC standards (e.g. IEC 950 for data
processing equipment and IEC 601-1 for medical equipment). Furthermore
all configurations shall comply with the valid version of the system standard
IEC 601-1-1. Everybody who connects additional equipment to the signal
input part or signal output part configures a medical system, and is therefore
responsible that the system complies with the requirements of the valid
version of the system standard IEC 601-1-1. If in doubt, consult your local
service organization.
Safety instructions for laser products
The ADC Compact Plus is a Class 1 Laser Product. It uses a 2x50 mW laser
diode, classification class IIIb.
Under normal operating conditions - when both doors are closed - there can
be no laser radiation outside the ADC Compact Plus. It is nonetheless
imperative that the local radiation safety regulations regarding the protection
of staff against scattered radiation are complied with, if the ADC Compact
Plus is located in the immediate vicinity of an X-ray room.
Open the front left and right door only to solve cassette or image plate jams.
When you open either of the doors, the power supply of all critical
components is switched off automatically as a precaution.
User interventions other than those described in this manual can be
hazardous with regard to laser radiation.

12 2231C EN 20040224
Safety compliance
The ADC Compact Plus complies with:
•the general safety regulations EN 60950, IEC 601-1-1:1992, UL 2601 and
CSA C22.2 No. 0 and IEC 601-1;
•the laser safety regulations EN 60825, DHHS/FDA 21 CFR, Parts 1040.10 and
1040.11 and ANSI Z 136-1980.

13
2231C EN 20040224
Operating modes
The ADC Compact Plus can be operated in three modes: operator mode,
key-operator mode and service mode.
Operator mode
The operator mode groups all basic functions which are aimed at
radiographers:
•Reading an image plate;
•Reading an emergency image plate;
•Re-erasing an image plate.
A normal image plate is read automatically after it is placed in the ADC
Compact Plus input buffer; the other functions of the operator mode can be
accessed via the keypad. All functions of the operator mode are described in
Chapter 2, ‘Basic operation (‘Operator mode’)’.
Key-operator mode
The key-operator mode groups advanced functions which are aimed at
technicians.
The key-operator mode can be accessed via the Key-operator key on the
keypad and is menu-driven. The key-operator functions are described in
Chapter 3, ‘Advanced operation (‘Key-operator mode’)’.
Service mode
The service mode functions are reserved for trained service personnel. They
are password protected.

14 2231C EN 20040224
The user interface
Main components of the Digitizer
The main components of the ADC Compact Plus are:
"Cassette input buffer
The cassette input buffer accepts up to 10 cassettes - even of different sizes -
for digitizing and up to 9 cassettes for erasure.
"Keypad
As the handling of the cassettes is fully automated, normal operation is a
zero-button operation. The keys on the keypad are only used to activate
special functions such as reading an emergency image plate or erasing an
image plate.
A
D
C
C
O
M
P
A
C
T
P
L
U
S
Cassette
input buffer
Cassette
output buffer
Keypad
with display
Service connector Main switch
Status indicator

15
2231C EN 20040224
"Status indicator
A light indicates the status of the ADC Compact Plus.
"Cassette output buffer
The cassette output buffer receives cassettes which have been handled by
the Digitizer.
The control panel
The control panel of the ADC Compact Plus consists of a backlit LCD display
and 10 keys.
As the handling of the cassettes is fully automated, normal operation is a
zero-button operation. Only when you are performing special functions or in
the event of problems (e.g. a cassette or image plate jam), will you need the
keys.

16 2231C EN 20040224
The keypad
Special functions can be accessed via the keypad. The keypad features the
following keys:
Emergency
key To give an image the status ‘emergency’ when it is
sent to the image processing station.
Erase key
To erase images without digitizing them.
This must be done if:
• an image plate has not been used for more than
3 days;
• an image plate has been exposed to an
exceptionally high X-ray dose.
Key-operator
key To access advanced functions (‘key-operator
functions’).
Service key To access service-level functions.
Reserved for trained service personnel.
Escape key To quit the current function or exit a menu without
saving modifications.
Confirm key
In key-operator mode:
• to select a menu.
• to accept an entry in a menu and go back to
operator mode.

17
2231C EN 20040224
Up key
• To move the cursor to the previous entry field.
• To scroll upwards.
• To increment the number in a numeric entry
field.
Down key
• To move the cursor to the next entry field.
• To scroll downwards.
• To decrement the number in a numeric entry
field.
Left key
• To scroll backwards through multiple choices
within a field.
• To move the entry position in a numerical entry
field from right to left.
• To toggle between values in a field.
Right key
• To scroll forwards through multiple choices
within a field.
• To move the entry position in a numerical entry
field from left to right.
• To toggle between values in a field.

18 2231C EN 20040224
The display
The ADC Compact Plus control panel has a backlit LCD display with 8 lines of
40 characters each. Its lay-out depends on the operating mode.
"In operator mode, the display has dedicated areas for specific information:
1Set-up of image processing station:
• [blank]: Default image processing station selected.
• Off line: Transmission to all image processing stations disabled.
• [process.station] not ready: Image processing station not available.
• [process.station] rerouted: Images rerouted to other image processing
station.
2Type of message
3Extra comment or action to take
4System status:
• READY: The ADC Compact Plus is ready for operation.
• BUSY: The ADC Compact Plus is busy with scanning or erasing.
• ERROR: An error has occurred.
• LOCKED: id.
• WARNING: id.
3
1
Set-up STATUS
OPERATION MODE
1ST MESSAGE
2nd Message
Patient_Last_Name Sub_Exam
Patient_Last_Name Sub_Exam
Patient_Last_Name Sub_Exam
Station Name ERROR
5
6
8.1
8.2
8.3
124
7

19
2231C EN 20040224
If the system has been idle for 5 minutes, the backlit LCD display dims. The
display lightens if:
•The display message changes, e.g. if the Digitizer receives a message from the
image processing station.
•You place a cassette in the input buffer.
•You press a key on the keypad.
"In key-operator mode, operation is menu driven. The menu displays the
key-operator functions, the active keys, and the service code.
5Operation mode:
• [blank]: Normal operation mode.
• EMERGENCY: Emergency function for image plates with ID data.
• ERASURE: Re-erasure function.
6Error status: service code (SERVICE XXXXX) or error code (CODE
XXXXX)
7Station name of the ADC Compact Plus
Identifier of image plate being treated:
8.1
8.2
8.3
After image ID data is read;
During scanning of image plate and transmittal of image data;
During transmittal of image data to image processing station.
1Key-operator functions
2Active keys
3Service code
Queue management
Digitizer set-up
Date and Time
Send test image
System info
Install
Save configuration
Fast preview
: quit
: ok
: select
KEY-OPERATOR
MENU
SERVICE XXXXX
2
3
1

20 2231C EN 20040224
The status indicator
The light at the top of the ADC Compact Plus indicates the status of the ADC
Compact Plus.
Audio signals
The ADC Compact Plus gives status information via beeps. The length of the
beep indicates the response of the system to a key command.
•A short beep means that ADC Compact Plus has accepted the key command
and is starting the operation.
•A long beep means that you have pressed a non-active key or that the ADC
Compact Plus has rejected the key command.
•An interval beep accompanies an error, locked or warning message.
Color Constant/
Flashing Status Action
Green
Constant Ready. Proceed.
Flashing Busy (treating image
plate). Proceed.
Red
Constant Error.
• Check display for messages.
• Refer to ‘General procedure
in case of malfunction’ on
page 41.
Flashing
• Locked or warning.
• Power on/self-test
in progress.
• Key-operator mode.
• Service mode.
• ADC Compact Plus
not connected to
image processing
device.
• Check display for messages.
• Refer to ‘General procedure
in case of malfunction’ on
page 41.
Other manuals for ADC Compact Plus
2
This manual suits for next models
1
Table of contents
Other AGFA Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

TELEDYNE API
TELEDYNE API T300 Manual addendum

Dinamica Generale
Dinamica Generale DG500 Quick user guide

Sonel
Sonel MIC-10 operating manual

Garmin
Garmin Approach Z80 Quick start manual

Cooper Atkins
Cooper Atkins DFP450W quick start guide

Rohde & Schwarz
Rohde & Schwarz Smart Instruments UP300 operating manual