AGFA CR 75.0 User manual

CR 75.0 Digitizer
Reference manual
CR 75.0
MUSICA

22241A EN 20040210
For more information on Agfa products and Agfa HealthCare products, please visit www.agfa.com, your
Point of Knowledge.
© Agfa-Gevaert N.V. 2004.
No parts of this document may be reproduced, copied, adapted or transmitted in any form or by any means
without the written permission of Agfa-Gevaert N.V.
Agfa-Gevaert N.V. makes no warranties or representation, expressed or implied, with respect to the
accuracy, completeness or usefulness of the information contained in this document and specifically
disclaims warranties of suitability for any particular purpose. Agfa-Gevaert N.V. shall under no
circumstances be liable for any damage arising from the use or inability to use any information, apparatus,
method or process disclosed in this document.
Agfa-Gevaert N.V. reserves the right to make changes to this document without prior notice.
Agfa-Gevaert N.V., Septestraat 27, B-2640 Mortsel, Belgium.
CR 75.0 is a trademark of Agfa-Gevaert N.V., Belgium.
Agfa and Agfa-Rhombus are trademarks of Agfa-Gevaert AG, Germany.

3
2241A EN 20040210
Table of contents
Chapter 1: Introducing the CR 75.0 ..................................................................5
CR 75.0 intended use.......................................................................................6
CR 75.0 features..............................................................................................7
Safety precautions ...........................................................................................8
Safety compliance..........................................................................................12
Operating modes ...........................................................................................13
The user interface..........................................................................................14
Switching on the CR 75.0...............................................................................23
Switching off the CR 75.0...............................................................................25
Resetting the Digitizer ....................................................................................26
Chapter 2: Basic operation (‘Operator mode’)...............................................27
Reading an image plate .................................................................................28
Reading an emergency image plate ...............................................................30
Re-erasing an image plate .............................................................................32
Chapter 3: Advanced operation (‘Key-operator mode’)................................37
Consulting the image transmission queue (‘Queue management’)..................38
Customizing the CR 75.0 (‘Digitizer set-up’)....................................................42
Setting the date and time................................................................................48
Sending test images.......................................................................................49
Consulting information on the CR 75.0 ...........................................................50
Installing a new software version....................................................................54
Installing a new language...............................................................................59
Installing new customer parameters ...............................................................64
Saving the configuration data on a diskette (backup)......................................70
Enabling/disabling fast preview ......................................................................73
Chapter 4: Preventive maintenance and troubleshooting............................75
Preventive maintenance.................................................................................76
General procedure in case of malfunction.......................................................77
Solving the ‘ERROR’ status............................................................................78
Clearing cassette jams...................................................................................80
Clearing image plate jams..............................................................................84
Appendix A: Equipment information sheet....................................................87
Specifications.................................................................................................88

42241A EN 20040210
Appendix B: ADC Compact cassette..............................................................91
Safety precautions .........................................................................................92
Description of the ADC Compact cassette......................................................93
Cleaning the image plate................................................................................95
Cleaning the cassettes...................................................................................97
Technical specifications of the cassettes.........................................................98
Technical specifications of the image plates .................................................100

Introducing the CR 75.0
This chapter draws attention to important safety
precautions and introduces the CR 75.0.
!CR 75.0 intended use
!CR 75.0 features
!Safety precautions
!Safety compliance
!Operating modes
!The user interface
!Switching on the CR 75.0
!Switching off the CR 75.0
!Resetting the Digitizer
Chapter
1

62241A EN 20040210
CR 75.0 intended use
This device must only be used to scan exposed X-ray cassettes, containing
an erasable image plate (IP). This device is part of a system, consisting of
X-ray cassettes with erasable phosphor image plates, an identification station
for the cassettes and a workstation where the resulting digital image
information is further processed and routed. It is intended that this device is
only operated in a radiological environment by qualified staff.

7
2241A EN 20040210
CR 75.0 features
The CR 75.0 scans the exposed ADC image plate, converts the information
into digital data and automatically transfers the image to the image
processing station for further processing and visualization.
The CR 75.0 requires but little manual interaction. All you have to do, after
exposure and identification of the cassette, is to place it in the input buffer of
the CR 75.0. You can deposit up to 10 cassettes of different sizes
simultaneously in the input buffer. The Digitizer takes in the cassettes one by
one. The Digitizer reads the demographic data and routing information from
the memory chip in the cassette, opens the cassette, removes the image
plate and scans the latent image by means of a sweeping laser beam.
Once the image is digitized, the cassette is returned to the output buffer to be
used for new exposures. After a full Digitizer cycle, the plate has turned 180°
in the cassette.
Depending on the X-ray intensity which has affected the phosphor during the
exposure, more or less light will be emitted during laser scanning. The light is
converted into an electrical signal. This signal is then converted into a digital
bit stream. Once converted into digital form, the digitized image is transferred
to the image processing station for further processing and visualization.
Further features of the CR 75.0 include:
"The CR 75.0 permits assigning the status ‘emergency’ to an image. An
emergency image will be given priority by the image processing station.
"The CR 75.0 permits re-erasing an image plate before re-using it. In specific
cases, this is necessary to prevent ghost images caused by previous
exposures or stray radiation from interfering with the image of interest. You
can erase a batch of up to 9 image plates.

82241A EN 20040210
Safety precautions
General safety instructions
•For software and other technical platforms, and/or in combination with any
consumable, which constitute, after installation, a system for the interpretation of
medical image data: such system is used by trained and qualified professionals.
It is the user's responsibility to ensure that image quality, display quality,
environmental lighting and other possible distractions are consistent with the
clinical application.
The user must be aware, that automatic collimation could possibly lead to
misinterpretation of the image.
•Make sure that the CR 75.0 is constantly monitored in order to avoid
inappropriate handling, especially by children.
•Only trained service personnel must make repairs. Only authorized service
personnel must make changes to the CR 75.0.
•If there is any visible damage to the machine casing, do not start nor use the
CR 75.0.
•If you want to connect the CR 75.0 with other devices, components or
assemblies and if the technical data do not permit determining whether the
combination with these devices, components or assemblies involves hazards,
you must consult the respective manufacturers to avoid danger for operating
personnel or the environment.
•Do not override or disconnect the integrated safety features.
•As is the case for all technical devices, the CR 75.0 must be operated, cared for
and serviced correctly.
•If you don’t operate the CR 75.0 correctly or if you don’t have it serviced
correctly, Agfa-Gevaert is not liable for resulting disturbances, damages or
injuries.
•When installing the CR 75.0, care must be taken to ensure that there is either a
mains plug or an all-cable disconnecting device in the internal installation fitted
near the CR 75.0 and that it is easily accessible.
•If you notice conspicuous noise or smoke, disconnect the CR 75.0 immediately.
•Check that the mains voltage is within the specified range of the self adapting
power supply of the machine.
Markings and labels
Always take into account the markings and labels provided on the inside and
outside of the machine. A brief overview of these markings and labels and
their meaning is given below.

9
2241A EN 20040210
Safety warning, indicating that the CR 75.0 manuals should be
consulted before making any connections to other equipment. The
use of accessory equipment not complying with the equivalent
safety requirements of this Digitizer may lead to a reduced level of
safety of the resulting system. Consideration relating to the choice
of accessory equipment shall include:
• Use of the accessory equipment in the patient vicinity,
• Evidence that the safety certification of the accessory equipment
has been performed in accordance with the appropriate IEC 601-
1 and IEC 601-1-1 harmonized national standard.
In addition all configurations must comply with the medical
electrical systems standard IEC 601-1-1. The party that makes the
connections acts as system configurator and is responsible for
complying with the systems standard.
If required contact your local service organization.
In order to reduce the risk of electric shock, do not remove any
covers.
Caution hot:
Keep hands clear from the erasure unit.
Type B equipment:
Indicates that the CR 75.0 complies with the limits for type B
equipment.
Supplementary protective earth connector:
Provides a connection between the CR 75.0 and the potential
equalization busbar of the electrical system as found in medical
environments. This plug should never be unplugged before the
power is turned off and the power plug has been removed.
Intergrounding connector:
Provides a connection between the Digitizer and other equipment
which might exhibit minor ground potential differences. These
differences may degrade the quality of communication between
different equipment. Never remove connections to this terminal.
Protective earth (ground):
Provides a connection between the Digitizer and the protective
earth of the mains. Do not remove this connection, because this
will have a negative influence on the leakage current.

10 2241A EN 20040210
•You can hurt your fingers if they are caught between the ADC Cassette and the
edge of the input slot. Insert the cassette in the input buffer as described in
‘Reading an image plate’ on page 28. At all times, keep your fingers clear of the
input slot. As soon as the CR 75.0 takes in the cassette, release it.
TÜV safety issues
Accessory equipment connected to the analog and digital interfaces must be
certified according to the respective IEC standards (e.g. IEC 950 for data
processing equipment and IEC 601-1 for medical equipment). Furthermore
all configurations shall comply with the valid version of the system standard
IEC 601-1-1. Everybody who connects additional equipment to the signal
input part or signal output part configures a medical system, and is therefore
responsible that the system complies with the requirements of the valid
version of the system standard IEC 601-1-1. If in doubt, consult your local
service organization.
Power On
Power Off
Note that the power cord has to be disconnected from the wall
outlet in order to disconnect the unit entirely from the mains.
Precautions for use in USA only:
Make sure that the circuit is single-phase center-tapped, if the
Digitizer is connected to a 240 V/60 Hz source instead of a 120 V/
60 Hz source.

11
2241A EN 20040210
Safety instructions for laser products
The CR 75.0 is a Class 1 Laser Product. It uses a 2x50 mW laser diode,
classification class IIIb.
Under normal operating conditions - when both doors are closed - there can
be no laser radiation outside the CR 75.0. It is nonetheless imperative that
the local radiation safety regulations regarding the protection of staff against
scattered radiation are complied with, if the CR 75.0 is located in the
immediate vicinity of an X-ray room.
Open the front left and right door only to solve cassette or image plate jams.
When you open either of the doors, the power supply of all critical
components is switched off automatically as a precaution.
Observe the Caution instructions on the Optical module label:
User interventions other than those described in this manual can be
hazardous with regard to laser radiation.
CAUTION!
CLASS 3B LASER RADIATION:
WHEN OPEN AVOID DIRECT
EXPOSURE TO THE BEAM!
CUIDADO!
RAYO LASER CLASE 3B:
EVITAR LA EXPOSICIÓN AL HAZ
CUANDO LA TAPA ESTÁ ABIERTA.
VORSICHT !
LASERSTRAHLUNG KLASSE 3B:
WENN ABDECKUNG GEÖFFNET
NICHT DEM STRAHL AUSSETZEN!
ATTENTION!
FAISCEAU LASER CLASSE 3B:
QUAND CAPOT OUVERT ÉVITER
DE S´EXPOSER AU RAYÓN!

12 2241A EN 20040210
Safety compliance
The CR 75.0 complies with:
•the general safety regulations:
EN 60601-1:1990+A1:1993+A2:1995,
IEC 601-1:1988+A1:1991+A2:1995,
IEC 601-1-1 / EN 60601-1-1,
EN 60601-1-2:1993,
UL 2601-1 Second Edition,
CAN/CSA 22.No.601.1-M909;
•the laser safety regulations:
EN 60825,
DHHS/FDA 21 CFR, Parts 1040.10 and 1040.11,
ANSI Z 136-1980.

13
2241A EN 20040210
Operating modes
The CR 75.0 can be operated in three modes: operator mode, key-operator
mode and service mode.
Operator mode
The operator mode groups all basic functions which are aimed at
radiographers:
•Reading an image plate;
•Reading an emergency image plate;
•Re-erasing an image plate.
A normal image plate is read automatically after it is placed in the CR 75.0
input buffer; the other functions of the operator mode can be accessed via the
keypad. All functions of the operator mode are described in Chapter 2, ‘Basic
operation (‘Operator mode’)’.
Key-operator mode
The key-operator mode groups advanced functions which are aimed at
technicians.
The key-operator mode can be accessed via the Key-operator key on the
keypad and is menu-driven. The key-operator functions are described in
Chapter 3, ‘Advanced operation (‘Key-operator mode’)’.
Service mode
The service mode functions are reserved for trained service personnel. They
are password protected.

14 2241A EN 20040210
The user interface
Main components of the Digitizer
The main components of the CR 75.0 are:
"Cassette input buffer
The cassette input buffer accepts up to 10 cassettes - even of different sizes -
for digitizing and up to 9 cassettes for erasure.
"Keypad
As the handling of the cassettes is fully automated, normal operation is a
zero-button operation. The keys on the keypad are only used to activate
special functions such as reading an emergency image plate or erasing an
image plate.
"Status indicator
A light indicates the status of the CR 75.0.
CR 75.0
MUSICA
Cassette
input buffer
Cassette
output buffer
Keypad
with display
Service connector Main switch
Status indicator

15
2241A EN 20040210
"Cassette output buffer
The cassette output buffer receives cassettes which have been handled by
the Digitizer.

16 2241A EN 20040210
The control panel
The control panel of the CR 75.0 consists of a backlit LCD display and 10
keys.
As the handling of the cassettes is fully automated, normal operation is a
zero-button operation. Only when you are performing special functions or in
the event of problems (e.g. a cassette or image plate jam), you will need the
keys.

17
2241A EN 20040210
The keypad
Special functions can be accessed via the keypad. The keypad features the
following keys:
Emergency
key To give an image the status ‘emergency’ when it is
sent to the image processing station.
Erase key
To erase images without digitizing them.
This must be done if:
• an image plate has not been used for more than
3 days;
• an image plate has been exposed to an
exceptionally high X-ray dose.
Key-operator
key To access advanced functions (‘key-operator
functions’).
Service key To access service-level functions.
Reserved for trained service personnel.
Escape key To quit the current function or exit a menu without
saving modifications.
Confirm key
In key-operator mode:
• to select a menu.
• to accept an entry in a menu and go back to
operator mode.
Up key
• To move the cursor to the previous entry field.
• To scroll upwards.
• To increment the number in a numeric entry
field.
Down key
• To move the cursor to the next entry field.
• To scroll downwards.
• To decrement the number in a numeric entry
field.

18 2241A EN 20040210
Left key
• To scroll backwards through multiple choices
within a field.
• To move the entry position in a numerical entry
field from right to left.
• To toggle between values in a field.
Right key
• To scroll forwards through multiple choices
within a field.
• To move the entry position in a numerical entry
field from left to right.
• To toggle between values in a field.

19
2241A EN 20040210
The display
The CR 75.0 control panel has a backlit LCD display with 8 lines of 40 char-
acters each. Its lay-out depends on the operating mode.
"In operator mode, the display has dedicated areas for specific information:
1Set-up of image processing station:
• [blank]: Default image processing station selected.
• Off line: Transmission to all image processing stations disabled.
• [process.station] not ready: Image processing station not available.
• [process.station] rerouted: Images rerouted to other image processing
station.
2Type of message
3Extra comment or action to take
4System status:
• READY: The CR 75.0 is ready for operation.
• BUSY: The CR 75.0 is busy with scanning or erasing.
• ERROR: An error has occurred. Refer to Chapter 4, ‘Preventive
maintenance and troubleshooting’.
• LOCKED: id.
• WARNING: id.
5Operation mode:
• [blank]: Normal operation mode.
• EMERGENCY: Emergency function for image plates with ID data.
• ERASURE: Re-erasure function.
6Error status: service code (SERVICE XXXXX) or error code (CODE
XXXXX)
7Station name of the CR 75.0
3
1
Set-up STATUS
OPERATION MODE
1ST MESSAGE
2nd Message
Patient_Last_Name Sub_Exam
Patient_Last_Name Sub_Exam
Patient_Last_Name Sub_Exam
Station Name ERROR
5
6
8.1
8.2
8.3
124
7

20 2241A EN 20040210
If the system has been idle for 5 minutes, the backlit LCD display dims. The
display lightens if:
•The display message changes, e.g. if the Digitizer receives a message from the
image processing station.
•You place a cassette in the input buffer.
•You press a key on the keypad.
"In key-operator mode, operation is menu driven. The menu displays the
key-operator functions, the active keys, and the service code.
Identifier of image plate being treated:
8.1
8.2
8.3
After image ID data is read;
During scanning of image plate and transmittal of image data;
During transmittal of image data to image processing station.
1Key-operator functions
2Active keys
3Service code
Queue management
Digitizer set-up
Date and Time
Send test image
System info
Install
Save configuration
Fast preview
: quit
: ok
: select
KEY-OPERATOR
MENU
SERVICE XXXXX
2
3
1
Table of contents
Other AGFA Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

GMC
GMC SIRAX TV 808-615 operating instructions

Environmental Express
Environmental Express OAKTON 200 Series instruction manual

Novus
Novus 6000 Series user manual

TEC
TEC MINNEAPOLIS BLOWER DOOR 3 Operation manual

koban
koban KTO AT-8 manual

Endress+Hauser
Endress+Hauser Proline Promass I 300 PROFIBUS DP operating instructions

Ludlum Measurements
Ludlum Measurements 3-98 manual

Rigol
Rigol MSO1000Z series installation guide

Baumer
Baumer OM20 operating manual

CLA-VAL
CLA-VAL X144D e-FlowMeter Quick Start Installation & Removal Instructions

Fixturlaser
Fixturlaser SMC user manual

DeFelsko
DeFelsko PosiTest HHD Verifier instruction manual