NOTES

Agilent 53131A/132A/181A 225 MHz
Counter
Agilent 53131A/132A/181A
Assembly-Level Service Guide
This guide describes how to service the Agilent 53131A and Agilent 53132A
Universal Counters, and the Agilent 53181A Frequency Counter.
深圳德标仪器 135-1095-0799

Agilent Technologies, Inc. 7.NC.NL.A.11.03.97.R1.P.CW6FC
900 South Taft Avenue
Loveland, Colorado 80537 U.S.A.
Copyright 1994 - 2009
Agilent Technologies, Inc.
All Rights Reserved. Reproduction,
adaptation, or translations without
prior written permission is
prohibited, except as allowed under
the copyright laws.
March 2009
Printed in Malaysia
Manual part number
53131-90039
Certification and Warranty
Certification
Agilent Technologies certifies that
this product met its published
specification at the time of shipment
from the factory. Agilent
Technologies further certifies that its
calibration measurements are
traceable to the United States
National Institute of Standards and
Technology (formerly National
Bureau of Standards), to the extent
allowed by the Institute’s calibration
facility, and to the calibration
facilities of other International
Standards Organization members.
Warranty
Agilent warrants Agilent hardware,
accessories and supplies against
defects in materials and
workmanship for a period of one year
from date of shipment. If Agilent
receives notice of such defects during
the warranty period, Agilent will, at
its option, either repair or replace
products which prove to be defective.
Replacement products may be either
new or like-new.
Agilent warrants that Agilent
software will not fail to execute its
programming instructions, for the
period specified above, due to
defects in material and workmanship
when properly installed and used. If
Agilent receives notice of such
defects during the warranty period,
Agilent will replace software media
which does not execute its
programming instructions due to
such defects.
For detailed warranty information,
see back matter.
Safety Considerations
General
This product and related
documentation must be reviewed for
familiarization with this safety
markings and instructions before
operation.
Before Cleaning
Disconnect the product from
operating power before cleaning.
Warning Symbols That May Be
Used In This Book
Instruction manual symbol; the
product will be marked with this
symbol when it is necessary for the
user to refer to the instruction
manual.
Indicates hazardous voltages.
Indicates earth (ground) terminal.
or
Indicates terminal is connected to
chassis when such connection is not
apparent.
Indicates Alternating current.
Indicates Direct current.
Safety Considerations (contd)
W
ARNING
BODILY INJURY OR DEATH
MAY RESULT FROM
FAILURE TO HEED A
WARNING. DO NOT
PROCEED BEYOND A
WARNING UNTIL THE
INDICATED CONDITIONS
ARE FULLY UNDERSTOOD
AND MET.
C
AUTION
Damage to equipment, or incorrect
measurement data, may result
from failure to heed a caution. Do
not proceed beyond a CAUTION
until the indicated conditions are
fully understood and met.
Safety Earth Ground
An uninterruptible safety earth
ground must be maintained from the
mains power source to the product’s
ground circuitry.
W
ARNING
WHEN MEASURING POWER
LINE SIGNALS, BE
EXTREMELY CAREFUL AND
ALWAYS USE A
STEP-DOWN ISOLATION
TRANSFORMER WHICH
OUTPUT IS COMPATIBLE
WITH THE INPUT
MEASUREMENT
CAPABILITIES OF THIS
PRODUCT. THIS PRODUCT’S
FRONT AND REAR PANELS
ARE TYPCIALLY AT EARTH
GROUND. THUS, NEVER TRY
TO MEASURE AC POWER
LINE SIGNALS WITHOUT AN
ISOLATION TRANSFORMER.
For additional safety and acoustic
noise information, see back matter.
深圳德标仪器 135-1095-0799

Continued from front matter. . .
7.NL.A.11.03.97.R1.P.CW6BC
Warranty (contd)
Agilent does not warrant that the
operation of Agilent products will be
uninterrupted or error free. If Agilent
is unable, within a reasonable time,
to repair or replace any product to a
condition as warranted, customer
will be entitled to a refund of the
purchase price upon prompt return of
the product.
Agilent products may contain
remanufactured parts equivalent to
new in performance or may have
been subjected to incidental use.
The warranty period begins on the
date of delivery or on the date of
installation if installed by Agilent. If
customer schedules or delays Agilent
installation more than 30 days after
delivery, warranty begins on the 31st
day from delivery.
Warranty does not apply to defects
resulting from (a) improper or
inadequate maintenance or
calibration, (b) software, interfacing,
parts or supplies not supplied by
Agilent, (c) unauthorized
modification or misuse, (d) operation
outside of the published
environmental specifications for the
product, or (e) improper site
preparation or maintenance.
TO THE EXTENT ALLOWED BY
LOCAL LAW, THE ABOVE
WARRANTIES ARE EXCLUSIVE
AND NO OTHER WARRANTY OR
CONDITION, WHETHER
WRITTEN OR ORAL, IS
EXPRESSED OR IMPLIED AND
Agilent SPECIFICALLY
DISCLAIMS ANY IMPLIED
WARRANTIES OR CONDITIONS
OF MERCHANTABILITY,
SATISFACTORY QUALITY, AND
FITNESS FOR A PARTICULAR
PURPOSE.
Agilent will be liable for damage to
tangible property per incident up to
the greater of $300,000 or the actual
amount paid for the product that is
the subject of the claim, and for
damages for bodily injury or death,
to the extent that all such damages
are determined by a court of
competent jurisdiction to have been
directly caused by a defective
Agilent product.
TO THE EXTENT ALLOWED BY
LOCAL LAW, THE REMEDIES IN
THIS WARRANTY STATEMENT
ARE CUSTOMER’S SOLE AND
EXCLUSIVE REMEDIES.
EXCEPT AS INDICATED ABOVE,
IN NO EVENT WILL Agilent OR
ITS SUPPLIERS BE LIABLE FOR
LOSS OF DATA OR FOR DIRECT,
SPECIAL, INCIDENTAL,
CONSEQUENTIAL (INCLUDING
LOST PROFIT OR DATA), OR
OTHER DAMAGE, WHETHER
BASED IN CONTRACT, TORT, OR
OTHERWISE.
For consumer transactions in
Australia and New Zealand: the
warranty terms contained in this
statement, except to the extent
lawfully permitted, do not exclude,
restrict or modify and are in addition
to the mandatory statutory rights
applicable to the sale of this product
to you.
Assistance
Product maintenance agreements and
customer assistance agreements are
available for Agilent Technologies
products.
For any assistance, contact your
nearest Agilent Technologies Sales
and Service Office.
Safety Considerations (contd)
WARNING
INSTRUCTIONS FOR
ADJUSTMENTS WHILE
COVERS ARE REMOVED
AND FOR SERVICING ARE
FOR USE BY SERVICE-
TRAINED PERSONNEL
ONLY. TO AVOID
DANGEROUS ELECTRIC
SHOCK, DO NOT PERFORM
SUCH ADJUSTMENTS OR
SERVICING UNLESS
QUALIFIED TO DO SO.
WARNING
ANY INTERRUPTION OF THE
PROTECTIVE GROUNDING
CONDUCTOR (INSIDE OR
OUTSIDE THE PRODUCT'S
CIRCUITRY) OR
DISCONNECTING THE
PROTECTIVE EARTH
TERMINAL WILL CAUSE
A POTENTIAL SHOCK
HAZARD THAT COULD
RESULT IN PERSONAL
INJURY. (GROUNDING ONE
CONDUCTOR OF A TWO
CONDUCTOR OUTLET IS
NOT SUFFICIENT
PROTECTION.)
Whenever it is likely that the
protection has been impaired, the
instrument must be made inoperative
and be secured against any
unintended operation.
If this instrument is to be energized
via an autotransformer (for voltage
reduction) make sure the common
terminal is connected to the earthed
pole terminal (neutral) of the power
source.
Instructions for adjustments while
covers are removed and for servicing
are for use by trained-personnel only.
To avoid dangerous electric shock,
do not perform such adjustments or
servicing unless qualified to do so.
For continued protection against fire,
replace the line fuse(s) of the same
current rating and type (for example,
normal blow, time delay). Do not use
repaired fuses or short circuited
fuseholders.
Acoustic Noise Emissions
LpA<47 dB at operator position, at
normal operation, tested per
EN 27779. All data are the results
from type test.
Geräuschemission
LpA<47 dB am Arbeits
platz, normaler Betrieb, geprüft nach
EN 27779.
Die Angagen beruhen auf
Ergebnissen von Typenprüfungen.
深圳德标仪器 135-1095-0799

For further information, please contact your local Agilent Technologies sales office, agent or distributor.
Authorized EU-representative: Agilent Technologies Deutschland GmbH, Herrenberger Stra
β
e 130, D 71034 Böblingen, Germany
I
II
I DECLARATION OF CONFORMITY
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturer’s Name: Agilent Technologies, Incorporated
Manufacturer’s Address: Santa Clara Site
5301 Stevens Creek Blvd
Santa Clara, California 95051
Declares, that the product
Product Name: Universal Counter Frequency Counter
Model Number: 53131A, 53132A 53181A
Product Options: This declaration covers all options of the above product.
Conforms with the following European Directives:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC
(including 93/68/EEC) and carries the CE Marking accordingly.
EMC Standard
IEC 61326-1:1997+A1:1998 / EN 61326-1:1997+A1:1998
CISPR 11:1990 / EN 55011:1991
IEC 61000-4-2:1995+A1:1998 / EN 61000-4-2:1995
IEC 61000-4-3:1995 / EN 61000-4-3:1995
IEC 61000-4-4:1995 / EN 61000-4-4:1995
IEC 61000-4-5:1995 / EN 61000-4-5:1995
IEC 61000-4-6:1996 / EN 61000-4-6:1996
IEC 61000-4-11:1994 / EN 61000-4-11:1994
Canada: ICES-001:1998
Australia/New Zealand: AS/NZS 2064.1
Limit
Group 1 Class A [1]
4kV CD, 8kV AD
3 V/m, 80-1000 MHz
0.5kV signal lines, 1kV power lines
0.5 kV line-line, 1 kV line-ground
3V, 0.15-80 MHz I cycle, 100%
Safety IEC 61010-1:1990+A1:1992+A2:1995 / EN 61010-1:1993+A2:1995
Canada: CSA C22.2 No. 1010.1:1992
Supplemental Information:
[1] The product was tested in a typical configuration with Agilent Technologies test systems.
July 31, 2001
Date Art Nanawa, Product Regulations Manager
深圳德标仪器 135-1095-0799

Contents
Assembly-Level Service Guide iii
Preface
How to Use This Guide xiii
Repair Strategy xiii
Instrument Identification xiii
Instruments Covered by this Guide xiv
Assembly-Level Service Guide Organization xiv
How to Order Guides xv
Description of the 225 MHz Counters xvi
Options xviii
Hardware xviii
Support xix
Accessories Supplied and Available xix
Accessories Supplied xix
Accessories Available xix
1 Performance Tests
Introduction 1-2
Operational Verification 1-2
Complete Performance Tests 1-3
Optional GPIB Verification 1-3
Recommended Calibration Cycle 1-3
Test Record 1-3
Equipment Required 1-4
Agilent 53131A/132A/181A Operational Verification 1-5
Power-On Self Tests 1-5
Run Self Test 1-8
Termination Check 1-8
External Arm Test (Agilent 53131A/132A Only) 1-10
Equipment 1-10
深圳德标仪器 135-1095-0799

Contents
iv Assembly-Level Service Guide
Counter Setup 1-10
Procedure 1-11
External Timebase Tests 1-12
Equipment 1-12
1 MHz External Timebase Input (Agilent 53131A and Agilent
53181A Only) 1-13
5 MHz External Timebase Input (Agilent 53131A and Agilent
53181A Only) 1-14
10 MHz External Timebase Input (Agilent 53131A and Agilent
53181A Only) 1-14
10 MHz External Timebase Input for the Agilent 53132A 1-15
Agilent 53131A/132A Complete Performance Tests 1-16
Test 1: Time Interval (Agilent 53131A/132A Only) 1-17
Equipment 1-17
Counter Setup 1-17
Procedure 1-18
Test 2: Trigger Level (Agilent 53131A/132A Only) 1-19
Equipment 1-19
Counter Setup 1-19
Procedure 1-20
Test 3: Channels 1 and 2 Frequency Sensitivity (Agilent 53131A/132A
Only) 1-22
Equipment 1-22
Counter Setup 1-22
100 kHz to 100 MHz Sensitivity for Agilent 53131A/132A 1-23
100 MHz to 200 MHz Sensitivity for Agilent 53131A/132A 1-25
200 MHz to 225 MHz Sensitivity for Agilent 53131A/132A 1-26
Test 4: Channels 1 and 2 Frequency Accuracy (Agilent 53131A/132A
Only) 1-27
Equipment 1-27
Counter Setup 1-27
Procedure 1-28
深圳德标仪器 135-1095-0799

Contents
Assembly-Level Service Guide v
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (Agilent
53131A/132A Only) 1-30
Equipment 1-30
Counter Setup 1-30
100 MHz to 2.5 GHz Sensitivity for Agilent 53131A/132A Option
030 only 1-31
2.7 to 3.0 GHz Sensitivity for Option 030 only 1-32
200 MHz to 5.0 GHz Sensitivity for Agilent 53131A/132A (Option
050 only) 1-33
200 MHz to 12.4 GHz Sensitivity for Agilent 53131A/132A (Option
124 only) 1-34
Test 6: Option 030/050/124 Channel 3 Frequency Accuracy
(Agilent 53131A/132A Only) 1-35
Equipment 1-35
Counter Setup 1-35
Procedure 1-35
Test 7: Peak Volts, Channels 1 and 2 (Agilent 53131A/132A
Only) 1-37
Equipment 1-37
Counter Setup 1-37
Vo l t Pe a k 1 1- 3 8
Vo l t Pe a k 2 1 -3 9
Agilent 53131A/132A GPIB Verification Program (Optional) 1-40
Agilent 53131A/132A Performance Test Record 1-41
深圳德标仪器 135-1095-0799

Contents
vi Assembly-Level Service Guide
Agilent 53181A Complete Performance Tests 1-47
Test 1: Trigger Level (Agilent 53181A Only) 1-48
Equipment 1-48
Counter Setup 1-48
Procedure 1-49
Test 2: Channel 1 Frequency Sensitivity (Agilent 53181A
Only) 1-51
Equipment 1-51
Counter Setup 1-51
100 kHz to 100 MHz Sensitivity for Agilent 53181A 1-52
100 MHz to 200 MHz Sensitivity for Agilent 53181A 1-54
200 MHz to 225 MHz Sensitivity for Agilent 53181A 1-54
Test 3: Channel 1 Frequency Accuracy (Agilent 53181A Only) 1-55
Equipment 1-55
Counter Setup 1-55
Procedure 1-56
Test 4: Option 015/030/050/124 Channel 2 Frequency
Sensitivity 1-58
Equipment 1-58
Counter Setup 1-58
100 MHz to 1.5 GHz Sensitivity for Option 015 Only 1-59
100 MHz to 2.7 GHz Sensitivity for Option 030 Only 1-60
2.8 to 3.0 GHz Sensitivity for Option 030 Only 1-61
200 MHz to 5.0 GHz Sensitivity for Option 050 only 1-61
200 MHz to 12.4 GHz Sensitivity for Option 124 only 1-62
Test 5: Option 015/030/050/124 Channel 2 Frequency Accuracy
(Agilent 53181A Only) 1-63
100 MHz to 1.5 GHz Accuracy Test for Option 015 1-63
Equipment 1-63
Counter Setup 1-63
Procedure 1-64
Minimum to Maximum Accuracy Test for Options 030/050/
124 1-65
深圳德标仪器 135-1095-0799

Contents
Assembly-Level Service Guide vii
Equipment 1-65
Counter Setup 1-65
Procedure 1-66
Test 6: Peak Volts, Channel 1 (Agilent 53181A Only) 1-68
Equipment 1-68
Counter Setup 1-68
Procedure 1-69
Agilent 53181A GPIB Verification Program (Optional) 1-70
Agilent 53181A Performance Test Record 1-71
2 Service
Introduction 2-2
Returning the Instrument to Agilent Technologies for Service 2-3
To Provide Repair Information 2-3
To Pack in the Original Packaging Materials 2-4
To Pack in the Commercially Available Materials 2-5
About the Agilent 53131A/132A Calibration Menu 2-6
Overview of the Agilent 53131A/132A Calibration Menu 2-6
The Agilent 53131A/132A Calibration Menu Tree 2-8
To View the Calibration Menu and Security Status 2-9
To Unsecure for Calibration 2-9
To Initiate the Calibration Routines 2-9
To Secure Against Calibration 2-13
To Change to a New Security Code 2-13
To View the Calibration Count 2-14
To Get Help with the Calibration Menu 2-14
The Agilent 53131A/132A Calibration Procedures 2-15
First Determine the Counter Firmware Revision 2-15
To Calibrate the Offset for Channels 1 and 2 2-15
To Calibrate the Gain for Channels 1 and 2 2-18
To Calibrate Time Interval 2-19
深圳德标仪器 135-1095-0799

Contents
viii Assembly-Level Service Guide
CAL: TI QUIK? Calibration 2-20
Equipment 2-20
CAL: TI FINE? Calibration 2-21
Equipment 2-21
To Calibrate the Standard Timebase 2-23
To Calibrate the High Stability Timebase Option (Medium, High, or
Ultra-High) 2-24
About the Agilent 53181A Calibration Menu 2-25
Overview of the Agilent 53181A Calibration Menu 2-25
The Agilent 53181A Calibration Menu Tree 2-27
To View the Calibration Menu and Security Status 2-28
To Unsecure for Calibration 2-28
To Initiate the Calibration Routines 2-29
To Secure Against Calibration 2-29
To Change to New the Security Code 2-30
To View the Calibration Count 2-30
To Get Help with the Calibration Menu 2-30
The Agilent 53181A Calibration Procedures 2-31
First Determine the Counter Firmware Revision 2-31
To Calibrate the Offset for Channel 1 2-31
To Calibrate the Gain for Channel 1 2-33
To Calibrate the Standard Timebase 2-34
To Calibrate the High Stability Timebase Option (Medium, High, or
Ultra-High) 2-35
Pre-Troubleshooting Information 2-36
Safety Considerations 2-36
Recommended Test Equipment 2-37
Repair Considerations 2-38
Electrostatic Discharge 2-38
Surface Mount Repair 2-38
Disassembly and Reassembly Specifics 2-38
After Service Considerations 2-39
Product Safety Checks 2-39
深圳德标仪器 135-1095-0799

Contents
Assembly-Level Service Guide ix
Product Performance Checks 2-39
Assembly Identification and Location 2-40
Troubleshooting the Counter 2-45
Power Supply Check 2-45
Overview of the Self-Test Routines 2-47
Diagnosing the Faulty Assembly by Using the Self Tests 2-49
To Run the Test-All Self Test 2-49
To Run the Individual Self Tests 2-50
3 Replacing Assemblies
Introduction 3-2
Tools Required 3-3
Do This First 3-3
To Remove the Cover 3-4
To Remove the Front Bezel 3-6
To Remove A1 Motherboard Assembly 3-9
To Remove the A2 Display Board, Keypads, and Window 3-11
To Remove A3 1.5/3.0/5.0/12.4 GHz Channel Assembly (Option 015/
030/050/124) 3-13
To Remove A4 AC Power Supply Assembly 3-15
To Remove A5 DC Power Input Assembly (Option 002) 3-17
To Remove A6 High Stability Timebase Assembly (Options 001, 010,
and 012) 3-19
To Remove the Rear Terminals (Option 060) 3-21
4 Retrofitting Options
Introduction 4-2
Tools Required 4-2
Do This First 4-2
To Retrofit A3 1.5/3.0/5.0/12.4 GHz Channel Assembly (Option 015/
030/050/124) 4-3
To Retrofit A5 DC Power Input Assembly (Option 002) 4-4
Option 002 DC Power Input Assembly Parts 4-4
深圳德标仪器 135-1095-0799

Contents
x Assembly-Level Service Guide
Preliminary Procedure 4-4
Retrofitting Procedure 4-5
To Retrofit A6 High Stability Timebase Assembly (Options 001, 010,
and 012) 4-11
Option 001 Medium Stability Timebase Assembly Parts 4-11
Option 010 High Stability Timebase Assembly Parts 4-11
*Option 012 Ultra-High Stability Timebase Assembly Parts 4-11
Procedure 4-12
To Retrofit the Rear Terminals (Option 060) 4-14
Option 060 Rear Terminals Parts 4-14
5 Replaceable Parts
Introduction 5-2
Exchange Assemblies 5-2
Reference Designations 5-3
Replaceable Parts 5-3
How To Order A Part 5-4
Parts Identification 5-4
Contacting Agilent Technologies 5-5
Cabinet Parts and Hardware 5-6
6 Backdating
Introduction 6-2
Manual Changes 6-2
Older Instruments 6-2
Backdating Hardware 6-3
Backdating Firmware 6-8
Backdating Specifications 6-16
7 Agilent 53131A/132A Specifications
Introduction 7-2
Instrument Inputs 7-2
Time Base 7-4
深圳德标仪器 135-1095-0799

Contents
Assembly-Level Service Guide xi
Measurement Specifications 7-5
Measurement Definitions 7-12
Measurement Arming and Processing 7-14
General Information 7-16
8 Agilent 53181A Specifications
Introduction 8-2
Instrument Inputs 8-2
Time Base 8-4
Measurement Specifications 8-5
Measurement Definitions 8-9
Measurement Arming and Processing 8-10
General Information 8-11
Index
深圳德标仪器 135-1095-0799

Contents
xii Assembly-Level Service Guide
深圳德标仪器 135-1095-0799

Assembly-Level Service Guide xiii
Preface
This guide provides assembly-level service information for the Agilent 53131A and
Agilent 53132A Universal Counters, and the Agilent 53181A Frequency Counter.
How to Use This Guide
Repair Strategy
This service guide is designed to isolate failures to the assembly level only.
The Agilent 53131A/132A/181A Counter can be returned to Agilent Technologies for
all service work, including troubleshooting, and verifying specifications. Contact
your nearest Agilent Technologies Sales and Service Office for more details.
Note: there is an Express Repair/Performance Calibration Service for USA customers
if downtime is critical. You can receive your repaired Counter via overnight shipment.
Just call 800-403-0801 and ask for Express Repair/Performance Calibration Service.
When your Counter is repaired, it will be returned via overnight shipment at no extra
charge.
If you decide to service the Counter yourself, use the troubleshooting procedures in
Chapter 2 (Service) and the disassembly and reassembly procedures in Chapter 3
(Replacing Assemblies). Then use the calibration instructions in Chapter 2 to
calibrate the Counter for peak-performance operation, and finally perform the
complete performance tests in Chapter 1 to verify the Counter's specifications.
Instrument Identification
Instrument identification is made from the serial number on the rear panel of the
Counter. Agilent uses a two-part serial number with the first part (prefix) identifying a
series of instruments and the second part (suffix) identifying a particular instrument
within a series. An Agilent-assigned alpha character between the prefix and suffix
identifies the country in which the instrument was manufactured.
深圳德标仪器 135-1095-0799

Preface
xiv Assembly-Level Service Guide
Instruments Covered by this Guide
This guide applies directly to Agilent 53131A, Agilent 53132A, and Agilent 53181A
Counters that have the same serial number prefix(es) shown on the title page. If the
serial number prefix of your Counter differs from that listed on the title page of this
guide, then there may be differences between this guide and your instrument.
Instruments having a higher serial prefix are covered when required by one or more
manual-change sheets included with this guide. If a required change sheet is missing,
contact your nearest Agilent Technologies Sales Office listed at the back of this guide.
Instruments having a serial prefix lower than that listed on the title page are covered
in Chapter 6, “Backdating.”
Assembly-Level Service Guide Organization
This assembly-level service guide consists of a table of contents, preface, eight
chapters and an index. The page running headers identify the chapters and sections of
this manual. The chapter contents are summarized as follows:
Chapter 1, “Performance Tests,” provides procedures that verify the Counter operates
properly and meets the Agilent 53131A/132A or Agilent 53181A specifications given
in Chapter 7, “Agilent 53131A/132A Specifications,” or Chapter 8, “Agilent 53181A
Specifications,” in this guide.
Chapter 2, “Service,” is divided into seven main sections that provide instructions for
returning the Counter to Agilent Technologies for service, calibrating the Counter,
and troubleshooting the assemblies in the Counter.
Chapter 3, “Replacing Assemblies,” provides procedures for replacing defective
assemblies and/or modules in the Counter.
Chapter 4, “Retrofitting Options,” provides field-installation procedures that help you
install options into the Counter.
Chapter 5, “Replaceable Parts,” lists the replaceable parts contained in the Counter,
and explains how to order replacement parts for your Counter.
Chapter 6, “Backdating,” contains information required to adapt this manual for older
instruments.
深圳德标仪器 135-1095-0799

Preface
Assembly-Level Service Guide xv
Chapter 7, “Agilent 53131A/132A Specifications,” lists all the specifications and
operating characteristics for the Agilent 53131A/132A Universal Counter.
Chapter 8, “Agilent 53181A Specifications,” lists all the specifications and operating
characteristics for the Agilent 53181A Frequency Counter.
How to Order Guides
The part number for this guide is listed in the Certification and Warranty page
(located on the back of the title page) and on the rear cover of this guide. Preface will
always follow table of contents. Page numbers will be in roman lower case.
深圳德标仪器 135-1095-0799

Preface
xvi Assembly-Level Service Guide
Description of the 225 MHz Counters
The Agilent 53131A and Agilent 53132A are universal counters capable of
measuring frequencies to 225 MHz on Channels 1 and 2. With an optional Channel 3,
this frequency is extended to 3.0 GHz (Option 030); 5.0 GHz (Option 050) or 12.4
GHz (Option 124).”
For the Agilent 53131A, frequency and time interval resolutions are 10 digits in one
second and 500 picoseconds, respectively. The Agilent 53131A provides users with
a GPIB measuring speed of 200 measurements per second, and is suitable for bench-
top operation and lower-volume ATE operation. The frequency and time interval
resolutions for the Agilent 53132A are up to 12 digits in one second and 150
picoseconds, respectively. The Agilent 53132A provides users with exceptional
resolution, and is ideal for ATE systems operation.
The Agilent 53131A/132A basic measurement functions include Frequency, Period,
Pulse Width, Duty Cycle, Rise/Fall Time, Time Interval, Frequency Ratio, Totalize,
Phase, and Peak Voltage.
The Agilent 53181A is a frequency counter capable of measuring frequencies to 225
MHz on Channel 1. Depending on which optional Channel 2 the counter contains,
this capability is extended to 1.5 GHz (Option 015) or 3.0 GHz (Option 030). The
Agilent 53181A has a frequency resolution of 10 digits in one second.
The Agilent 53181A provides users with a GPIB measuring speed of
200 measurements per second, and is suitable for bench-top and ATE operation.
The Agilent 53181A basic measurement functions include Frequency, Period, Peak
Voltage, and Frequency Ratio (if Channel 2 is installed).
The Agilent 53131A/132A/181A include additional measurement functions and
features that are designed specifically for manufacturing and service applications:
• 1, 5, 10 MHz external reference capability—to match customer’s house standard
(however, the Agilent 53132A’s external reference capability is 10 MHz only),
• optional ultra-high stability , high stability, or medium stability oven oscillators
for high accuracy needs,
深圳德标仪器 135-1095-0799
This manual suits for next models
2
Table of contents
Other Agilent Technologies Cash Counter manuals

Agilent Technologies
Agilent Technologies 53181A User manual

Agilent Technologies
Agilent Technologies 53210A User manual

Agilent Technologies
Agilent Technologies 5352B User manual

Agilent Technologies
Agilent Technologies 53220A User manual

Agilent Technologies
Agilent Technologies 53151A User manual

Agilent Technologies
Agilent Technologies 5385A User manual