
12
Input A Configuration Options
The default configuration options for Input A at power-on are: AC coupling, 1MΩinput
impedance, 1:1 attenuation, rising edge polarity and no filter; with the Threshold control set to
mid-position a measurement should be possible with the majority of waveforms. Changes to the
configurations will, however, be necessary for certain waveforms, e.g. DC coupling and low pass
filter in circuit will improve measurement of low frequencies.
Input Coupling: AC coupling is the default and can be used with either input impedance setting.
Select DC coupling for very low frequencies (<30Hz) or if the waveform duty cycle is very low.
DC coupling should normally be used with the input impedance set to 1MΩ; selection of 50Ωis
allowed but, because a 50kΩprotection resistor is fitted in parallel with the coupling capacitor,
the actual impedance will be much higher than 50Ωuntil the input frequency is greater than
approximately 300kHz. This configuration can be useful to avoid charging up the coupling
capacitor on asymmetrical waveforms.
When AC coupling is selected the instrument will assume there is no signal and set the display to
0.0 after about 1 second if no transition occurs. When DC coupling is selected it will allow for
very slow signals by waiting forever for an input transition; the display will continue to show the
last value.
Input Impedance:1MΩis the default and can be used with both AC and DC coupling. It can be
used directly or in conjunction with x1, x10 or x100 oscilloscope probes as appropriate to the
signal amplitude. Select 50Ωfor higher frequencies and where the signal source impedance is
50Ωto minimise spurious counting caused by reflections.
Input Attenuation: 1:1 (no attenuation) is the default. Select 5:1 for larger signals, particularly if
noise is significant. When measuring standard logic signals, use 1:1 attenuation for 1.8V (or
lower) CMOS and 5:1 for CMOS at 2.5V (or higher), or TTL. Additional attenuation can be
achieved by attenuating the signal externally before being presented to the counter; a x10
oscilloscope probe can be used with the 1MΩinput impedance or a 50Ω attenuator can be used
with the 50Ω input impedance to preserve matching.
Input Polarity: Rising edge (pulse High) is the default setting; with this setting Frequency and
Period measurements start and finish on the rising edge and Count is the total number of rising
edge occurrences. The Width measurement is from rising edge to falling edge which, together
with the Period measurement, yields the calculated Ratio (High:Low time) and Duty (High time as
a percentage of Period) measurements.
If the Polarity is changed to falling edge (pulse Low), Frequency and Period measurements will
start and finish on the falling edge and Count will totalise the occurrences of falling edges. If the
waveform being measured has a slow rising edge but fast falling edge, setting the Polarity to
falling edge might be advantageous in reducing measurement jitter. Changing the Polarity for
Width measurements, however, will change the interpretation of Ratio and Duty and should only
be used with care.
Low Pass Filter: The default setting is for no filter. If Filter In is selected the FILT annunciator
will show in the display; the nominal cut-off frequency is 50kHz. The filter is particularly useful for
low frequency measurements but, with an adequate input signal, it can be helpful at frequencies
up to 200kHz or more.
Trigger Level Threshold Adjustment: The trigger level control is associated with two yellow
LED lamps which indicate the signal balance at the output of the A input amplifier. Their
brightness varies from bright to dim depending on the relationship between the trigger threshold
and the average value of the input signal. When the threshold setting matches the average value
of the input signal they are of equal brightness. If a signal is applied and the instrument is not
counting, move the threshold control towards the dimmer of the two lamps. Note that the smaller
the input signal level, the more critical this setting becomes.
When AC coupling is selected (the default configuration) a threshold feedback mechanism is
engaged, with the threshold control providing a small offset above or below the average signal
level. Normally the control should be set with the marker at the midway position marked AC.