12
Power Meter Set-up
With the Power Meter view displayed, pressing SET-UP shows the Power Meter set-up screen,
which gives access to the Setup Range soft-key. In Flicker mode, it also gives access to the
Flicker and Voltage Variations result pages. See the Flicker Meter chapter for details.
Set-up Range
Pressing the Setup Range soft-key (on either the Meter or Waveform Graph views) changes
the display to a graphical view of the current waveform with three further soft-keys: Up,Down
and Lock Range. Pressing any of these turns off auto-ranging; the Lock Range key keeps
the present measurement range, while the Up and Down keys change the range in ×2 and ÷2
steps respectively. The scaling of the waveform display changes to reflect the peak capability of
the new range and the maximum current is shown in the bottom left-hand corner of the display;
for example, 1500mA peak means that the top of the display represents +1500mA and the
bottom represents –1500mA with respect to the centre zero line.
The range can be set from 24mA peak to 400A peak in 2:1 steps. The waveform display on this
set-up screen is always updated every cycle, regardless of the mode set up for the Waveform
Graph view.
Important Note: To take compliance quality measurements the range must not change during
the test. The user must select a range that will accommodate the highest waveform peak that will
be encountered during the load’s operating cycle. It is therefore recommended that a preliminary
assessment of the load should be made, with the instrument in auto-range mode, to determine
the highest peak current operating condition. The instrument should then be locked on this range
to ensure that harmonics measurements are correct and that the derived values (Average, Max
Hold and Min Hold etc.) are meaningful.
If too low a range is chosen, no damage will be done, but all the measurement results will be
invalid and the message Overload will show in the status line at the top of the Meter View display.
It is more important to maintain adequate ‘headroom’for the measurement, i.e. to avoid overload,
than it is to maximise the vertical resolution on the Waveform Graph view; measurement
accuracy is maintained even when the vertical scaling of the display is well below optimum.
Once the range is locked, the Lock Range soft-key label becomes Auto Range;pressing
this key releases the lock and allows the instrument to auto-range on the present waveform.
Several other soft-keys 'toggle' between alternate choices in this way; it is only necessary to
remember that at any time the soft-key box shows what will happen when the key is pressed.
Having set the range (or selected auto-range), the previous view can be recalled by pressing the
VIEW key.
Auto-ranging is only available in Harmonics operating mode (see the Test Control section). Note
that auto-ranging is disabled while the LOAD switch is OFF to prevent the instrument auto-
ranging to the most sensitive range when there is no load current. At switch-on the default range
is 3A peak if the LOAD switch is OFF; at other times, switching the LOAD to OFF locks the range
at the setting currently in use.
In Flicker mode with voltage sensing, the current range can be manually changed, but should be
left at a setting that avoids overload at any time during the test (otherwise the test will be
needlessly aborted). When the current sensing method is selected, the range is fixed to suit the
requirements of the reference impedance calculations and cannot be changed.
On selecting Peak Inrush mode, the highest current range is initially set, but it can be manually
changed as required.
The Voltage range is determined by the Nominal Voltage value which must be declared on the
Test Control set-up menu to match the actual supply voltage.