Anderson DA2 User manual

Anderson Instrument Co. Inc.
156 Auriesville Road
Fultonville, NY 12072
1-800-833-0081
Fax 518-922-8997
www.andinst.com
Instruction
Manual
Instrument Model Number _____________________________
Instrument Serial Number _________________________________
Form Number AIC2034
© June 2003 revised: 2/13
Model DA2 ELECTRODLESSModel DA2 ELECTRODLESS
Model DA2 ELECTRODLESSModel DA2 ELECTRODLESS
Model DA2 ELECTRODLESS
CONDUCTIVITY TRANSMITTERCONDUCTIVITY TRANSMITTER
CONDUCTIVITY TRANSMITTERCONDUCTIVITY TRANSMITTER
CONDUCTIVITY TRANSMITTER
(for conductivity, % concentration and TDS measurement)

PAGE 2
TT
TT
Table of Contentsable of Contents
able of Contentsable of Contents
able of Contents
Section 1 - General Information ...................................4
1.1 Capability Highlights ..................................................................................4
1.2 Transmitter Safety ......................................................................................4
1.3 Retained Configuration Values .................................................................5
1.4 Transmitter Serial Number ........................................................................5
1.5 EMC Conformance .....................................................................................5
Section 2 - Specifications ............................................6
2.1 Operational .................................................................................................6
2.2 Transmitter Performance (Electrical, Analog Outputs) ..........................8
2.3 Mechanical ..................................................................................................8
Section 3 - Physical Installation ..................................9
3.1 Location ......................................................................................................9
3.2 Panel Mounting (DA201A) .......................................................................10
3.3 Wall and Pipe Mounting (DA202A) ..........................................................11
3.4 Integral Sensor Mounting (DA202A with HC1012000 only) ..................12
Section 4 - Electrical Connections ............................13
4.1 HC1 Series Electrodless Conductivity Sensor Wiring ..........................14
4.2 Two-Wire Hookup .....................................................................................15
4.3 Three-Wire Hookup ..................................................................................16
4.4 Four-Wire Hookup ....................................................................................17
Section 5 - Operation ..................................................18
5.1 Display ......................................................................................................18
5.2 Keypad ......................................................................................................18
5.3 MEASURE Screen (normal display mode) .............................................20
Section 6 - Menu Structure ........................................ 21
6.1 Displaying Main Branch Selection Screens............................................21
6.2 Displaying Top-Level Menu Screens ......................................................22
6.3 Displaying Submenu Screens .................................................................23
6.4 Adjusting Edit / Selection Screen Values ..............................................23
6.5 Entering (Storing) Edit/Selection Screen Values/Choices ....................23

PAGE 3
Section 7 - Transmitter Configuration ......................24
7.1 Selecting Language to Operate Transmitter ..........................................24
7.2 Configuring Sensor Characteristics .......................................................24
7.3 Set °C OR °F (temperature display format) ............................................34
7.4 Configuring Analog Output .....................................................................34
7.5 SET Passcode (feature enabled or disabled) ........................................36
7.6 Configuration Setting Summary .............................................................38
Section 8 - Transmitter Calibration ...........................40
8.1 Important Information ..............................................................................40
8.2 Zero Procedure (first-time sensor calibration only) ..............................41
8.3 Conductivity Calibration ..........................................................................41
8.4 % Concentration Calibration ...................................................................45
8.4 TDS Calibration ........................................................................................47
8.6 Analog Output Calibration .......................................................................49
Section 9 - Testing & Maintenance ............................50
9.1 STATUS Check (transmitter and sensor) ...............................................50
9.2 HOLD OUTPUT .........................................................................................51
9.3 OUTPUT Test Signal ................................................................................51
9.4 Firmware (EPROM VERSION) Check ......................................................51
9.5 SELECT SIM Measurment ........................................................................52
9.6 SIM SENSOR Setting ................................................................................52
9.7 RESET CONFIGURE Values to Factory Defaults ...................................53
9.8 RESET CALIBRATE Values to Factory Defaults ....................................53
Section 10 - Troubleshooting ....................................54
10.1 Keeping the Transmitter Calibrated .....................................................54
10.2 Avoiding Electrical Interference ...........................................................54
10.3 Checking Electrical Connections .........................................................54
10.4 Verifying Sensor Operation ...................................................................54
10.5 Verifying Transmitter Operation ...........................................................54
10.6 Verifying Interconnect Cable Integrity .................................................55
Section 11 - Ordering Matrix ......................................55
Section 12 - Warranty and Return Statement ...........56

PAGE 4
Section 1 - General InformationSection 1 - General Information
Section 1 - General InformationSection 1 - General Information
Section 1 - General Information
1.1 Capability Highlights1.1 Capability Highlights
1.1 Capability Highlights1.1 Capability Highlights
1.1 Capability Highlights
Sensor Input
The transmitter can be used with the Model HC1-Series electrodeless conductivity sensor.
These sensors have a built-in Pt 1000 RTD temperature compensator element.
MEASURE Screen
The measure screen (normal display mode) can provide different readouts of measured data.
With the MEASURE screen displayed, press ÕÕ
ÕÕ
Õor ÖÖ
ÖÖ
Ö to show:
•Measured conductivity, % concentration or TDS
•Measured temperature (°C or °F)
•Measured conductivity, % concentration or TDS and temperature
•Measured analog output value (mA)
• Uncompensated conductivity corresponding to concentration readout (only shown
when transmitter is set to measure concentration)
Password-Protected Access
For security, you can enable a passcode feature to restrict access to configuration and
calibration settings to authorized personnel only. See Section 7.5 for details.
Calibration Methods
Because each sensor has a unique zero point and span, always zero the sensor in air when
calibrating it for the first time (Section 8.2). Depending on the configured measurement
(conductivity, % concentration or TDS), different methods are available for calibrating sensor
span (see Section 8.3, 8.4 or 8.5 respectively). The analog output loop can also be calibrated
(Section 8.6).
Analog Output
The transmitter’s isolated 4-20 mA analog output can be assigned to represent one of these:
•Measured conductivity, % concentration or TDS
•Measured temperature.
Parameter values can be entered to define the endpoints at which the 4 mA and 20 mA analog
output values are desired (range expand). For analog output setup details, see Section 7.4.
NOTE: During calibration, the analog output is automatically held at the last measured value
and, upon completion, returned to its active state.
1.2 T1.2 T
1.2 T1.2 T
1.2 Transmitter Safetyransmitter Safety
ransmitter Safetyransmitter Safety
ransmitter Safety
The transmitter is completely safe to handle. Only low DC voltage is present.

PAGE 5
1.3 R1.3 R
1.3 R1.3 R
1.3 Retained Configuration Vetained Configuration V
etained Configuration Vetained Configuration V
etained Configuration Valuesalues
aluesalues
alues
All user-entered configuration values are retained indefinitely, even if power is lost or turned off.
The non-volatile transmitter memory does not require battery backup.
1.4 T1.4 T
1.4 T1.4 T
1.4 Transmitter Serial Numberransmitter Serial Number
ransmitter Serial Numberransmitter Serial Number
ransmitter Serial Number
A label with the transmitter model number, serial number, and build date is located between the
terminal blocks.
1.5 EMC Conformance1.5 EMC Conformance
1.5 EMC Conformance1.5 EMC Conformance
1.5 EMC Conformance
The transmitter is designed to provide protection from most normally encountered
electromagnetic interference. This protection exceeds U.S. standards and meets European IEC
1000 (EN 61000) series testing for electromagnetic and radio frequency emissions and
immunity. Refer to Figure 1-1 and the specifications in Section 2.1 for more information.
Figure 1-1 EMC Diagram
EMISSIONS IMMUNITY

PAGE 6
Section 2 - SpecificationsSection 2 - Specifications
Section 2 - SpecificationsSection 2 - Specifications
Section 2 - Specifications
2.1 Operational2.1 Operational
2.1 Operational2.1 Operational
2.1 Operational
Display Two-line by 16 character LCD
NOTE: The measured value (conductivity, % concentration or TDS) and temperature can be
displayed separately or shown together on a single screen. The corresponding 4-20
mA analog output value can also be shown. (When measuring concentration, the
transmitter can also show a corresponding readout of uncompensated conductivity.)
Measurement Ranges
Conductivity µS/cm: 0-200.0 or 0-2000
mS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000
S/cm: 0-2.000
% Concentration 0-99.99% or 0-200.0%
TDS 0-9999 ppm
Temperature -4.0 to 347.0°F or -20.0 to +175.0°C
Analog Output 4.00-20.00 mA
Ambient Conditions:
Operation -4 to +140°F (-20 to +60°C); 0-95% relative humidity, non-
condensing
Storage -22 to +158°F (-30 to +70°C); 0-95% relative humidity, non-
condensing
Temperature Compensation Automatic from 14.0 to 347.0°F (-10.0 to +175.0°C) with
selection for Pt 1000 ohm RTD temperature element, or
manually fixed at a user-set temperature
NOTE: The selected measurement (conductivity, % conc. or TDS) determines which of the
following temperature compensation methods are available:
Linear % per °C slope, built-in natural water temperature properties table, user-
entered temperature table, or no compensation
Sensor-to-Transmitter Distance Maximum cable length is a function of the measuring range
and allowable non-linearity. The following schedule is
recommended:
Full-scale Range Max. Length
200 to 2000 µS/cm 200 ft. (61 m)
2000 to 2,000,000 µS/cm 300 ft. (91 m)
NOTE: When measuring % concentration, convert the transmitter full-scale value to
conductivity to determine the maximum distance.

PAGE 7
Power Requirements (Class 2 Power Supply):
Two-wire Hookup 16-30 VDC
Three-wire Hookup 14-30 VDC
Four-wire Hookup 12-30 VDC
Calibration Methods:
Sensor Zero With the dry sensor in air, press keys to initiate
(all measurements) automatic system zeroing
Conductivity Measurement:
COND CAL Enter compensation reference temperature, and reference
solution’s known linear % per °C slope and value
SAMPLE CAL Enter one sample value (determined by laboratory analysis
or a comparison reading)
ELECTRONIC CAL This method requires that the sensor be removed from the
process and utilizes a decade resistance box to simulate
conductivity rather than an actual solution.
Concentration Measurement:
CONC CAL Enter one sample value (determined by laboratory analysis
or a comparison reading)
COND CAL Enter compensation reference temperature, and reference
solution’s known linear % per °C slope and value
TDS Measurement:
TDS CAL Enter one sample value (determined by laboratory analysis
or a comparison reading)
Analog Output Isolated 4-20 mA output with 0.004 mA (12-bit) resolution
NOTE: The output can be assigned to represent the measured value (conductivity, %
concentration or TDS) or measured temperature. Parameter values can be entered to
define the endpoints at which the 4 mA and 20 mA output values are desired (range
expand). During calibration, the output is automatically held at the last measured
value and, upon completion, returned to its active state.
Maximum Loop Load Dependent on power supply voltage, transmitter hookup
arrangement, and wire resistance (see load resistance
charts for respective hookup diagrams in Section 4.2, 4.3 or
4 .4)
Memory (non-volatile) All user settings are retained indefinitely without battery
backup
Electrical Certifications UL, C-UL Pending

PAGE 8
2.2 T2.2 T
2.2 T2.2 T
2.2 Transmitter Pransmitter P
ransmitter Pransmitter P
ransmitter Performance (Electrical, Analog Outputs)erformance (Electrical, Analog Outputs)
erformance (Electrical, Analog Outputs)erformance (Electrical, Analog Outputs)
erformance (Electrical, Analog Outputs)
Accuracy* ± 0.1% of span
Sensitivity* ± 0.05% of span
Repeatability* ± 0.05% of span
Temperature Drift* Zero and Span: ± 0.02% of span per °C
Response Time 1-60 seconds to 90% of value upon step change (with
sensor filter setting of zero)
*These typical performance specifications are:
1. Based on 25°C with conductivity of 500 µS/cm and higher. Consult factory for applications in
which conductivities are less than 500 µS/cm.
2. Derated above 100°C to the maximum displayed temperature of 200°C. Consult factory for details.
2.3 Mechanical2.3 Mechanical
2.3 Mechanical2.3 Mechanical
2.3 Mechanical
Enclosure Polycarbonate; NEMA 4X general purpose; choice of
included mounting hardware
Mounting Configurations Panel, wall, pipe or integral sensor mounting
Dimensions With Back Cover:
3.75 in. W x 3.75 in. H x 2.32 in. D
(95 mm W x 95 mm H x 60 mm D)
Without Back Cover for Panel Mount:
3.75 in. W x 3.75 in. H x 0.75 in. D
(95 mm W x 95 mm H x 19 mm D)
Net Weight 10 oz. (280 g) approximately

PAGE 9
Section 3 - Physical InstallationSection 3 - Physical Installation
Section 3 - Physical InstallationSection 3 - Physical Installation
Section 3 - Physical Installation
Unpack and examine the equipment even if you do not use it immediately. If there is evidence of
damage, notify the transit carrier immediately.
Recommendation: Save the shipping carton and packing materials in case the instrument must
be stored or re-shipped.
3.1 Location3.1 Location
3.1 Location3.1 Location
3.1 Location
1. It is recommended to locate the transmitter as close as possible to the installed sensor.
The maximum allowable distance between an installed sensor and the transmitter
depends upon the full-scale value you set for the transmitter measuring range:
200-2000 µS/cm
Full-scale Value
2000-2,000,000 µS/cm
Full-scale Value
200 feet (61 m) max. 300 feet (91 m) max.
NOTE: When measuring % concentration, convert the transmitter full-scale value to
conductivity to determine the maximum distance.
2. Mount the transmitter in a location that is:
• Clean and dry where there is little or no vibration.
• Protected from corrosive fluids.
• Within ambient temperature limits (-4 to +140°F or -20 to +60°C).
CAUTION:
EXPOSING THE TRANSMITTER TO DIRECT SUNLIGHT MAY INCREASE
THE OPERATING TEMPERATURE ABOVE ITS SPECIFIED LIMIT, AND
DECREASE DISPLAY VISIBILITY.

PAGE 10
3.2 P3.2 P
3.2 P3.2 P
3.2 Panel Mounting (DA201A)anel Mounting (DA201A)
anel Mounting (DA201A)anel Mounting (DA201A)
anel Mounting (DA201A)
Figure 3-1 illustrates how to panel mount the transmitter using the supplied panel mount
hardware kit.
1. Cut a 3.30-inch (84 mm) square cutout hole in panel.
2. Position panel-mount gasket over cutout in front of panel, and place retainer plate
behind panel with its four threaded inserts facing away from back of panel.
3. Attach transmitter to retainer plate using its four captive screws.
NOTE: If panel is too thick, remove captive screws from transmitter, and use longer screws
provided in hardware kit.
FIGURE 3-1 Panel Mounting Details (DA201A)

PAGE 11
3.3 W3.3 W
3.3 W3.3 W
3.3 Wall and Pipe Mounting (DA202A)all and Pipe Mounting (DA202A)
all and Pipe Mounting (DA202A)all and Pipe Mounting (DA202A)
all and Pipe Mounting (DA202A)
Figure 3-2 illustrates how to wall or pipe mount the transmitter using the supplied hardware kit.
Determine the mounting method, and attach the hardware as shown.
1. Fasten the wall/pipe adapter to the wall or pipe.
2. Using a blunt tool, open both cable entry knockout holes in the back cover.
3. Insert-and-twist the back cover onto the installed wall/pipe adapter, and tighten its two
screws to lock back cover onto the adapter.
4. Attach transmitter to back cover using its four captive screws.
FIGURE 3-2 Wall and Pipe Mounting (DA202A)

PAGE 12
3.4 Integral Sensor Mounting (DA202A with HC1012000 only)3.4 Integral Sensor Mounting (DA202A with HC1012000 only)
3.4 Integral Sensor Mounting (DA202A with HC1012000 only)3.4 Integral Sensor Mounting (DA202A with HC1012000 only)
3.4 Integral Sensor Mounting (DA202A with HC1012000 only)
Figure 3-3 illustrates how to integrally mount the transmitter onto a sensor using the supplied
mounting hardware kit.
1. Using a blunt tool, open knockout hole in bottom of swivel ball for routing the sensor
cable.
2. Attach swivel-mount assembly onto back end of sensor using a stainless steel 1/2”
NPT female close coupling that you provide. Be sure to tape the NPT threads to
prevent moisure ingression
3. Insert-and-twist the back cover onto the installed swivel-mount assembly. Tighten its
two screws to lock the back cover onto the swivel-mount assembly.
NOTE: To change mounting angle, loosen swivel-mount assembly by lifting tab on bottom of
swivel nut. Position to desired angle and re-tighten swivel nut.
4. Attach transmitter to back cover using its four captive screws.
FIGURE 3-3 Integral Sensor Mounting (DA202A)

PAGE 13
Section 4 - Electrical ConnectionsSection 4 - Electrical Connections
Section 4 - Electrical ConnectionsSection 4 - Electrical Connections
Section 4 - Electrical Connections
Figure 4-1 shows the terminal block arrangement and terminal designations for the transmitter.
NOTE: All terminals are suitable for single wires up to 14 AWG (2.5 mm2).
Wiring Tip! To comply with European Community (CE) electromagnetic
compatibility requirements, follow these general wiring guidelines:
1. Locate transmitter as far as possible from motors and other non-CE certified devices
with excessive electromagnetic emissions.
2. Use specified ferrites and cables. Failure to do so may eliminate compliance. Locate all
ferrites as close as possible to the transmitter.
•DC Power Supply Cable: Connect cable shield to earth ground at the supply end. Loop
cable 2-1/2 times through ferrite (Steward #28B0686-200, Fair-Rite Corp. #2643665702
or equivalent).
•Sensor Cable: Keep cable shields as short as possible. At the transmitter end, connect
the outer shield to earth ground, and the inner shield to the SHIELD terminal. Clamp
ferrite (Steward #28A2025-OAO, Fair-Rite Corp. #0431164281, or equiv-alent) on
sensor cable.
•Analog mA Output Cable (four-wire hookup only): Connect cable shield to earth ground
at the supply end. Loop cable 2-1/2 times through ferrite (Steward #28B0686-200, Fair-
Rite Corp. #2643665702, or equivalent).
FIGURE 4-1 Transmitter Terminal Designations

PAGE 14
4.1 HC1 Series Electrodeless Conductivity Sensor Wiring
Depending on how transmitter is mounted, route the sensor (or interconnect) cable into the
transmitter as follows:
• Wall/Pipe-mounted Transmitter: Route cable through left side cable entry knockout hole
in the back cover.
• Panel-mounted Transmitter: Route cable behind panel to the exposed TB2 terminal
strip.
• Integral Sensor-mounted Transmitter: Route cable through a 1/2” NPT female close
coupler and then through the swivel ball knockout hole and center hole in back cover.
(Do not open left side cable entry knockout hole in back cover.)
Wiring Tip! Route the sensor cable in 1/2-inch, grounded metal conduit to
protect it from moisture, electrical noise, and mechanical damage.
For installations where the distance between sensor and transmitter exceeds
the sensor cable length, indirectly connect the sensor to the transmitter using a
junction box and interconnect cable.
NOTE: Do not route the sensor cable in any conduit containing AC or DC power wiring
(“electrical noise” may interfere with the sensor signal). Also, always re-calibrate the
system when the cable length between sensor and transmitter changes.
Refer to Figure 4-2 and connect the sensor (or interconnect) cable wires as shown, matching
colors as indicated. (Terminal 6 is unused.)
NOTE: For systems not requiring CE compliance and lacking an earth ground, connect the
outer shield to Terminal 3 on TB2.
FIGURE 4-2 Connecting HC1 Series Electrodless Conductivity Sensor
1 White (Drive)
2 Blue (Drive)
3 Shield (Ground)
4 Red (Temp)
5 Yellow (Ground)
6
7Green (Signal)
Green
Yellow
Inner Shield
White
Blue
Red
Outer Shield
To Earth Ground

PAGE 15
4.2 T4.2 T
4.2 T4.2 T
4.2 Twowo
wowo
wo--
--
-Wire HookupWire Hookup
Wire HookupWire Hookup
Wire Hookup
In a two-wire hookup, at least 16 VDC is required for operation. A load device can be connected
in the current loop (see Figure 4-3 for details).
Depending on how the transmitter is mounted, route the DC power/analog output wiring into the
transmitter as follows:
•Wall/Pipe-mounted Transmitter: Route cable through right side cable entry knockout
hole in the back cover.
•Panel-mounted Transmitter: Route cable behind panel to the exposed TB1 terminal
strip.
•Integral Sensor-mounted Transmitter: Route cable through right side cable entry
knockout hole in the back cover. (Do not open left side cable entry knockout hole in
cover).
Wiring Tip! Use high quality, shielded instrumentation cable.
FIGURE 4-3 Two-Wire Hookup
5
7
6
1
3
4
2
-
Loop Power
16-30VDC
Class 2
Loop Resistance
(See nomograph)
+
Jumper
Required
16
0
600
300
RESISTANCE
IN OHMS
100
200
400
500
700
800
30
DC VOLTAGE
18 20 22 24 26 28
50 Ohms/Volt
LOOP LOAD

PAGE 16
4.3 Three4.3 Three
4.3 Three4.3 Three
4.3 Three--
--
-Wire HookupWire Hookup
Wire HookupWire Hookup
Wire Hookup
In a three-wire hookup, the transmitter can be wired two ways depending on load “sinking” or
“sourcing”. At least 14 VDC is required for operation.
Depending on how the transmitter is mounted, route the DC power, analog output, and wiring
into the transmitter as follows:
•Wall/Pipe-mounted Transmitter: Route cable through right side cable entry knockout
hole in the back cover.
•Panel-mounted Transmitter: Route cable behind panel to the exposed TB1 terminal
strip.
•Integral Sensor-mounted Transmitter: Route cable through right side cable entry
knockout hole in the back cover. (Do not open left side cable entry knockout hole in
cover).
Wiring Tip! Use high quality, shielded instrumentation cable.
FIGURE 4-4 Three-Wire Hookup - Load Sinking
+
1
2
3
4
5
6
7
-
Loop Resistance
(See nomograph)
Jumper
Required
200
0
14 16 18
RESISTANCE
IN OHMS
800
400
600
1000
1200
1400
DC VOLTAGE
20 22 24 26 28 30
50 Ohms/Volt
LOOP LOAD
Loop Power
14-30VDC
Class 2
FIGURE 4-5 Three-Wire Hookup - Load Sourcing
+
1
2
3
4
5
6
7
-
Loop Resistance
(See nomograph)
Jumper Required
200
0
14 16 18
RESISTANCE
IN OHMS
800
400
600
1000
1200
1400
DC VOLTAGE
20 22 24 26 28 30
50 Ohms/Volt
LOOP LOAD
Loop Power
14-30VDC
Class 2

PAGE 17
4.4 Four4.4 Four
4.4 Four4.4 Four
4.4 Four--
--
-Wire HookupWire Hookup
Wire HookupWire Hookup
Wire Hookup
In a four-wire hookup, at least 12 VDC is required for operation.
Depending on how the transmitter is mounted, route the DC power, analog output wiring into the
transmitter as follows:
•Wall/Pipe-mounted Transmitter: Route cable through right side cable entry knockout
hole in the back cover.
•Panel-mounted Transmitter: Route cable behind panel to the exposed TB1 terminal
strip.
•Integral Sensor-mounted Transmitter: Route cable through right side cable entry
knockout hole in the back cover. (Do not open left side cable entry knockout hole in
cover).
Wiring Tip! Use high quality, shielded instrumentation cable
FIGURE 4-6 Four-Wire Hookup
5
7
6
1
3
4
2
-
Loop Resistance
(See nomograph)
+
-
+
1000
RESISTANCE
IN OHMS
161412
400
200
0
800
600
1400
1200
LOOP LOAD
50 Ohms/Volt
2624222018
DC VOLTAGE
3028
Loop Power
12-30VDC
Class 2
External Power
12-30VDC
Class 2

PAGE 18
Section 5 - OperationSection 5 - Operation
Section 5 - OperationSection 5 - Operation
Section 5 - Operation
The user interface consists of a two-line LCD display and a keypad with MENU, ENTER, ESC,
ÕÕ
ÕÕ
Õ, ÖÖ
ÖÖ
Ö, ××
××
×, and ØØ
ØØ
Økeys.
5.1 Display5.1 Display
5.1 Display5.1 Display
5.1 Display
By using the keypad, you can display three types of screens:
1. Measure Screens: The normal display mode shows the measured value (conductivity,
% concentration or TDS). Pressing the ÖÖ
ÖÖ
Ö key sequentially scrolls through these other
measurement readouts:
• Measured process temperature
• Measured value and temperature
• Measured analog output mA value
• Uncompensated conductivity corresponding to % concentration readout (only
shown when transmitter is set to measure concentration)
2. MENU Screens: These top-level and lower-level (submenu) screens within the three
main branches of the menu tree are used to access edit/selection screens for configu-
ration. (EXIT screens at the end of each menu branch enable you to move up one level
in the menu tree by pressing the ENTER key. This is functionally the same as pressing
the ESC key.)
3. Edit/Selection Screens: These screens enter values / choices to calibrate, configure,
and test the transmitter.
5.2 K5.2 K
5.2 K5.2 K
5.2 Keypadeypad
eypadeypad
eypad
The keypad enables you to move throughout the transmitter menu tree. The keys and their
related functions are:
1. MENU key: Pressing this key with the MEASURE screen displayed shows the “MAIN
MENU CALIBRATE” screen. To display the configure or test/maint top-level main
branch screen, press the ØØ
ØØ
Ø key. Pressing the MENU key with a menu screen dis-
played always shows the top-level screen in that branch. (Pressing the MENU key also
“aborts” the procedure to change values or selections.)
2. ENTER key: Pressing this key does two things; it displays submenu and edit/selection
screens, and it enters (saves) configuration values/selections.
3. ESC key: Pressing this key always takes the display up one level in the menu tree.
(Example: With any “MAIN MENU” screen displayed, pressing the ESC key once takes
the display up one level to the measure screen.) The ESC key can also “abort” the
procedure to change a value or selection.
4. ÕÕ
ÕÕ
Õ and ÖÖ
ÖÖ
Ökeys: Depending on the type of displayed screen, these keys do the
following:
• MEASURE Screen: Changes readout (in continuous loop sequence) to show
different measurements.
• Menu Screens: These keys are non-functional.
• Edit/Selection Screens: Moves cursor left or right to select digit for adjustment with
××
××
× and ØØ
ØØ
Ø keys.
5. ××
××
×and ØØ
ØØ
Økeys: Depending on the type of displayed screen, these keys do the
following:

PAGE 19
• MEASURE Screen: These keys are non-functional.
• Menu Screens: Moves up or down respectively between other same-level menu
screens.
• Edit/Selection Screens: Adjusts selected digit value up or down, or moves up or
down between choices.
FIGURE 5-1 Transmitter Keypad

PAGE 20
5.3 MEASURE Screen (normal display mode)5.3 MEASURE Screen (normal display mode)
5.3 MEASURE Screen (normal display mode)5.3 MEASURE Screen (normal display mode)
5.3 MEASURE Screen (normal display mode)
The MEASURE screen is normally displayed. Pressing the MENU key temporarily replaces the
MEASURE screen with the top-level “MAIN MENU CALIBRATE” branch selection screen.
Using the keypad, you can then display other screens to calibrate, configure or test the
transmitter. If the keypad is not used within 30 minutes, except during calibration or while
using specific transmitter test/ maintenance functions, the display will automatically
return to the MEASURE screen. To display the measure screen at any time, press the MENU
key once and then press the ESC key once.
The MEASURE screen can show four different readout versions. To select between them, in
continuous loop sequence, press the ÕÕ
ÕÕ
Õ or ÖÖ
ÖÖ
Ö key. These are examples of the different
versions:
ÖÖ
ÖÖ
Ö ÖÖ
ÖÖ
Ö
ÖÖ
ÖÖ
Ö ÖÖ
ÖÖ
Ö
When set to measure concentration, the transmitter can also show an uncompensated
conductivity reading corresponding to the measured concentration, as illustrated by this
example:
NOTE: When the transmitter returns to its normal MEASURE screen mode, the appearing
readout is always the version last selected.
Note that three MEASURE screen readout examples show the factory-default “cond”
notation on their top lines, illustrating the transmitter notation feature. To create your
own notation, refer to Section 7.2, subheading “ENTER NOTE (top line of MEASURE
screen).”
When the measured value is beyond the transmitter measuring range, a series of “ + “ or “ - “
screen symbols appear, respectively indicating that the value is above or below range.
Table of contents
Other Anderson Transmitter manuals
Popular Transmitter manuals by other brands

JUMO
JUMO JUMO dTRANS p20 DELTA operating manual

Emerson
Emerson CSI 9420 Reference manual

ASCOM
ASCOM T942C installation guide

King gates
King gates Myo quick start guide

Ramsey Electronics
Ramsey Electronics FM-100 Assembly and instruction manual

Lee Technology
Lee Technology LTK-1400AH User instruction manual