12 ISDBT Field Tester MS8911A MM
Amplitude Reading at 1701 MHz:
26. Set the MG3692A output to 1701 MHz, with an RF output level of –30 dBm.
27. On the ISDBT Field Tester, press the Amplitude key, then press the Reference Level soft key.
28. Use the keypad to enter –27 and press the dBm soft key.
29. Press the Freq key and select the Center Freq soft key.
30. Use the keypad to enter 1701 and select the MHz soft key.
31. Press the Span key, use the keypad to enter 2, and select the MHz soft key.
32. Press the Span key and then press the Trace (5) key.
33. Press the # of Average soft key, use the keypad to enter 5, then press the Enter key.
34. Wait until the Trace Count displays 5/5.
35. Press the Marker key and select the Peak Search soft key.
36. Record the amplitude at 1701 MHz in the test records on page 44, Amplitude Reading at 1701 MHz.
37. Press the Freq key and select the Start Freq soft key.
38. Use the keypad to enter 26 and select the MHz soft key.
39. Press Stop Freq soft key, enter 28 and press the MHz soft key.
40. Press the Shift key and then press the Trace (5) key.
41. Press the # of Average soft key, use the keypad to enter 5, then press the Enter key.
42. Wait until the Trace Count displays 5/5.
43. Press the Marker key and select the Peak Search soft key.
44. Record the Marker 1 amplitude reading in the test records on page 44, Amplitude Reading at 1701 MHz.
45. Calculate the input related spurious level and verify that it is –50 dBm:
(Input Related Spurious = Marker 1 reading – amplitude reading for 1701 MHz).
Amplitude Reading at 2145 MHz:
46. Set the MG3692A output to 1701 MHz, with an RF output level of –30 dBm.
47. On the ISDBT Field Tester, press the Amplitude key, then press the Reference Level soft key.
48. Use the keypad to enter –27 and press the dBm soft key.
49. Press the Freq key and select the Center Freq soft key.
50. Use the keypad to enter 2145 and select the MHz soft key.
51. Press the Span key, use the keypad to enter 2, and select the MHz soft key.
52. Press the Shift key and then press the Trace (5) key.
53. Press the # of Average soft key, use the keypad to enter 5, then press the Enter key.
54. Wait until the Trace Count displays 5/5.
55. Press the Marker key and select the Peak Search soft key.
56. Record the amplitude at 2145 MHz in the test records on page 44, Amplitude Reading at 2145 MHz.
57. Press the Freq key and select the Start Freq soft key.
58. Use the keypad to enter 470 and select the MHz soft key.
59. Press Stop Freq soft key, enter 472 and press the MHz soft key.
60. Press the Shift key and then press the Trace (5) key.
61. Press the # of Average soft key, use the keypad to enter 5, then press the Enter key.
62. Wait until the Trace Count displays 5/5.
63. Press the Marker key and select the Peak Search soft key.
64. Record the Marker 1 amplitude reading in the test records on page 44, Amplitude Reading at 2145 MHz.
65. Calculate the input related spurious level and verify that it is –60 dBm:
(Input Related Spurious = Marker 1 reading – amplitude reading for 2145 MHz).