
19
3 The optimum lockup pressure is 140±5 psig for the standard Con-
shelf XIV, and 125±5 psig for the Conshelf XIV Supreme, Supreme
EFA, Dry kit and 300 Bar DIN versions. To increase the MP, turn the
adjustment screw (30) standard & supreme, (34) supreme EFA and
(40) dr y kit “in” or clockwise; to decrease the MP, turn the adjustment
screw “out” or counterclockwise. Turn the adjustment screw using
a large at bladed screwdriver or 5/32 hex key adapter (supreme
EFA) . Turn the screw no more than 1/8 turns at a time, then purge
the second stage several times and observe the adjustment’s MP
increase. Repeat this process until the proper MP is set.
4 If the acceptable range cannot be achieved by varying the
adjustment screw (30) standard & supreme, (34) supreme EFA
and (40) dry kit, refer to the Troubleshooting Table 1 on page 24
for corrective guidelines ad specic procedures.
5 With the correct MP achieved, depress the purge button (2)
several times, release and observe the test bench MP gauge 5
seconds to 15 seconds after purging. The MP indicated must not
“creep” or slowly increase more than 5 psig within this time. If the
MP pressure will not stabilize at one pressure setting, refer to the
Troubleshooting Table 1 on page 24 for corrective guidelines and
specic procedures.
CAUTION: While adjusting the MP, be sure to “cycle” the regulator
by pushing the purge button (2) at least ten (10) times after turning
the adjustment screw (30) standard & supreme, (34) supreme
EFA and (40) dry kit. Failure to “cycle” the regulator can result
in a false reading of the rst stage MP “lock-up”.
First Stage Adjustment and Testing:
Test Bench Procedures
1 With the rst stage regulator mounted on the test bench HP valve
block adapter, attach a spare MP hose (26) to the open MP port
on the rst stage regulator. Attach the free end of this hose to the
male tting provided for it on the test bench.
NOTE: In conjunction with this subsection, the technician should refer
to Table 5 - Test Bench Specications, on page 29. In addition, all
tests described must rst be done at MP (500 psig) and then again at
service pressure (3000 psig).
2 While watching the MP gauge on the test bench, slowly pressurize
the bench and regulator to 500 psig. The MP gauge should indicate
a “lockup” pressure of 120-160 psig. If the gauge pressure increases
beyond 160 psig, turn the air source off immediately. This indicates a
failure of the rst stage regulator to achieve “lockup” pressure. Refer
to the Troubleshooting Table 1 on page 24 for corrective guidelines
and specic procedures. If the MP is not greater than 160 psig but
there is an audible leak or freeow from the second stage regulator(s),
adjustment of the lever (20) will be required before nal adjustment of
the rst stage MP can be completed. To adjust the lever see “Second
Stage Adjustment and Testing”, on p. 21 & 22.
Regulator Adjusting and Testing:
Manual Procedures
1 After attaching the rst and second stage (see Attachment of
First-stage to Second-stage Regulators, steps 1 through 4,
on page 18) unscrew one of the MP port plugs (20) and attach
a spare MP hose (26). Attach this hose to one of the open MP
ports on the rst stage body (16).
NOTE: If no test bench is available for testing the Conshelf XIV
regulator, the following adjustment and testing procedures should
be followed.
2 Thread the MP test pressure gauge (PN111610) onto the free
end of the MP hose. Make sure that the bleeder valve of the
test pressure gauge is open prior to pressurizing the regulator.
3Connect the rst stage regulator to a HP air source (partially lled
SCUBA cylinder - 500 psig). Slowly turn on the air source control
knob and pressurize the regulator. Once the regulator is pressurized,
slowly close gauge bleeder valve while watching the test gauge.
The gauge should indicate “lockup” pressure of 120-160 psig. If the
gauge pressure continues beyond 160 psig, immediately open the
gauge bleeder valve and close the air source control valve. Such a
steadily increasing gauge pressure beyond 160 psig indicates failure
of the rst stage regulator to achieve “lockup” pressure. Refer to the
Troubleshooting Table 1, on page 24 for corrective guidelines and
specic procedures.
4If the MP is not greater than 160 psig but there is an audible leak or
freeow from the second stage regulator(s), adjustment of the lever
(20) will be required before nal adjustment of the rst stage MP can be
completed. To adjust the lever see page 21 & 22 “Second Stage Adjust-
ment and Testing” for corrective guidelines and specic procedures.
5 The optimum lockup pressure is 140±5 psig for the Standard
Conshelf XIV, and 125±5 psig for the Conshelf XIV Supreme,
Supreme EFA , Dry kit and 300 Bar DIN versions. To increase
the MP, turn the adjustment screw (30) Standard & Supreme, (34)
Supreme EFA and (40) Dry Kit “in” or clockwise; to decrease the
MP, turn the adjustment screw “out” or counterclockwise. Turn the
adjustment screw using a large flat-bladed screwdriver or 5/32” hex
key adapter. Turn the screw no more than 1/8 turn at a time, then
purge the second stage several times and observe the adjustment
MP increase. Repeat this process until the proper MP is set.
6With the correct MP achieved, depress the purge button (2) sev-
eral times, release and observe the MP gauge 5 seconds to 15
seconds after purging. The MP indicated must not “creep” or slowly
increase more than 5 psig within this time. If the MP will not stabilize,
at one pressure setting, refer to the Troubleshooting Table 2, page
25, for corrective guidelines and specic procedures.
7 After stabilizing the MP repeat Steps 3 through 7 at 3000
psig, service pressure.
NOTE: For the Conshelf Supreme, Supreme EFA and Dry Kit:
See page 20 for reinstallation of environmental components on
the Supreme & Dry Kit Models.
CAUTION: If an adjustment of MP is required, be sure to
“cycle” the regulator (10 times) by pushing the purge button
(2) after turning the adjustment screw (30) standard & su-
preme, (34) supreme EFA and (40) dry kit). Failure to “cycle”
the regulator can result in a false reading of the rst stage
MP “lockup”.
Regulator Adjustment and Testing
8
Install the MP port plug (20) back into the rst stage body (16).
Using a 1/2” crow foot or socket attached to a in/lb torque
wrench, tighten the port plug to 15 in/lbs (1.7 Nm).