Arroyo Instruments TECSource 5400 Series User manual


Page 2 · 5400 Series TECSource User’s Manual
Table of Contents
Introduction.................................................................................................................... 3
Safety Terms and Symbols ........................................................................................... 4
Quick Start ..................................................................................................................... 6
Installation...................................................................................................................... 7
Power and Cable Connections...................................................................................... 9
Connecting to the TECSource..................................................................................... 12
Front Panel Operation ................................................................................................. 12
Settings and Menus..................................................................................................... 16
Remote Mode Operation ............................................................................................. 24
Installing the USB Drivers............................................................................................ 25
Saving and Restoring Configurations.......................................................................... 25
Using the Function Keys.............................................................................................. 26
Understanding the Sensor Inputs ............................................................................... 27
Using the Auxiliary Interface ........................................................................................ 28
Auxiliary Sensors..................................................................................................... 28
Digital Inputs........................................................................................................... 28
Digital Outputs ........................................................................................................ 29
Relay ....................................................................................................................... 29
Interlock .................................................................................................................. 30
Using the Interlock....................................................................................................... 30
Selecting the Fixture .................................................................................................... 30
Determining Temperature Stability.............................................................................. 31
Out Off Menu ............................................................................................................... 32
Applying Sensor Limits................................................................................................ 33
Working With Thermistors ........................................................................................... 34
Working With RTDs...................................................................................................... 35
2-Wire versus 4-Wire Measurements ...................................................................... 36
Working With AD590s and LM335s............................................................................. 37
Gain Control and the PID Loop ................................................................................... 37
Using the AutoTune Function...................................................................................... 38
Using the AutoTune Function Remotely...................................................................... 40
Controlling the Temperature Rate of Change ............................................................. 40
Changing the Set Point Step Size ............................................................................... 41
Resistive Heaters and Heat/Cool Only Modes ............................................................ 41
Using the Current Limit................................................................................................ 41
Using the Voltage Limit................................................................................................ 42
Correcting the TEC Voltage Measurement.................................................................. 43
Software Compensation.......................................................................................... 43
4-Wire Measurement............................................................................................... 44
Analog Output.............................................................................................................. 44
External Fan Control.................................................................................................... 46
User Calibration ........................................................................................................... 47
Linking Instruments ..................................................................................................... 48
Internal Power Management ....................................................................................... 48
Specifications .............................................................................................................. 51
Error Messages
............................................................................................................ 53
Maintenance, Calibration and Warranty ...................................................................... 57
European Community Declaration of Conformity ....................................................... 59

5400 Series TECSource User’s Manual · Page 3
Introduction
Thank you for choosing the 5400 Series TECSource from Arroyo Instruments.
Your TECSource is a combination of leading edge technology combined with
years of experience in the field of temperature control.
With a crystal clear 4-line VFD display, very high power output, multi-sensor
support, both RS232 and USB computer interfaces, and auxiliary I/O, the
TECSource will fit into almost any temperature control application.
Unlike other temperature controllers in its class which use inexpensive 7-
segment LED displays, the TECSource takes advantage of its large VFD display
to simultaneously display set point temperature, actual temperature, current, and
voltage.
The user interface of the TECSource is engineered to make using the
instrument straightforward. With its text-based menus, there is never any
confusion over which setting is being changed, and parameters are displayed in
clear English (no cryptic numbers or LEDs to decode).
The TECSource offers all the features you would expect from a modern
temperature controller, including:
AutoTune for automatic PID parameter calculation
0.004°C temperature stability
0.001°C resolution for both set point and actual temperatures
What’s in the Box
Along with the TECSource itself, you will find a printed copy of the manual, a 2-
pin interlock header, and a CD with electronic copies of this manual, the
Computer Interfacing Manual, and USB drivers are included. For USA
customers, a power cord is included. For non-USA customers, an IEC-60320-
C13 rated AC power cord must be provided.
Accessories
Arroyo Instruments also sells several accessories designed to work with the
TECSource. These include:
TECSource Cable, 30A, 2m (p/n 1264)
This cable has 17W2 male/female connectors for interfacing to the
LaserMount or other connectorized fixtures, includes wiring for the fan
interface, and supports up to 30A of TEC current. A pigtailed version of
this cable is available as p/n 1265.

Page 4 · 5400 Series TECSource User’s Manual
TECSource Cable, 17W2 to DB15, 15A, 2m (p/n 1264-DB15)
This cable has a 17W2 male connector for the instrument end and a
DB15 female connector for the mount end, and is used to connect the
5400 to mounts that utilize the DB15 TEC connector, such as the 274
TECMount. The 1264-DB15 supports up to 15A of TEC current.
4400/5400 Series 2U Rack Mount Kit (p/n 1403-RM)
For installing your 5400 Series TECSource into a standard 19” rack.
10kΩThermistor (p/n 1600-10K)
100ΩPlatinum RTD (p/n 1600-100RTD)
RS-232 NULL Cable, 3m (p/n 1200-NULL)
USB Cable, 3m (p/n 1201)
Safety Terms and Symbols
The following safety-related terms are used in this manual:
Warnings (noted by the WARNING heading) explain dangers that
could result in physical injury or death;
Cautions (noted by the CAUTION heading) explain conditions that
could result in damage to the instrument, other equipment, or your
device.
Notes (noted by the NOTES heading) are not safety-related, and are
intended simply to point out important information.
If, at any time, any of the following conditions exist, or are suspected of existing,
discontinue use of the unit until it can be inspected by qualified service
personnel:
Visible damage to the unit, including damage or stress caused during
product shipment;
Storage of the unit outside the standard storage temperature or
humidity rating, or prolonged storage under harsh conditions;
Failure to operate properly.
If needed, contact your distributor or Arroyo Instruments for service or repair to
ensure the safety of the product is maintained.

5400 Series TECSource User’s Manual · Page 5
Symbols
Power Off Power On
Caution, refer to manual Earth ground
Caution, risk of electric shock
General Warnings
CAUTION
There are no user-serviceable parts inside. All service
and repair work shall be done by Arroyo Instruments or
personnel authorized by Arroyo Instruments.
Modifications done by non-authorized personnel will
void the warranty. Please see the Service section later in
this manual for instructions on how to obtain service for
this instrument.
WARNING
To avoid electrical shock, ensure a 3-prong power cord
is used, and is plugged into an earth-grounded
receptacle. Failure to do so can result in severe injury or
death.
WARNING
This instrument is intended for use by qualified
personnel who understand the shock and laser hazards
and are familiar with safety procedures required to avoid
injury. Read this manual completely before attempting to
use this product.

Page 6 · 5400 Series TECSource User’s Manual
Quick Start
The 5400 TECSource was designed with ease of use in mind, and you will likely
have little need for this manual for almost all of the features the unit offers. This
section will show how you can quickly get the unit up and running in almost no
time.
The TECSource has a universal input power supply, accepting 90 to 240 VAC,
and 50 or 60 Hz. This covers all conventional power worldwide, but ensure your
AC power meets these requirements.
Plug the AC cord into the unit and into the wall outlet. Turn on the power switch
located on the IPC, and the unit will power up, displaying the model information
and firmware version number.
Press the Menu button to enter the menu, and using the knob, turn to the right
until the I Lim setting is displayed. Press the knob to edit the setting, and adjust
the limit as appropriate to your Peltier. Press the knob again to save the value.
Make the same adjustments to the high and low temperature limits (T-High Lim
and T-Low Lim), as appropriate for your application.
The unit comes pre-programmed for the BetaTHERM 10K3A1 thermistor. If the
TECSource is being connected to a LaserMount, no changes need to be
made, as this is the thermistor used in the mount. However, if you are using
manufacturer’s thermistor or a different BetaTHERM thermistor, you will need to
verify the Steinhart-Hart coefficients are set correctly. Navigate to the Main
Sensor Menu, pressing the knob to enter the sub-menu. Adjust the constants to
reflect the values for your thermistor.
Once you have made all your adjustments, press the Menu button to exit the
menu (in the future, if you’re only changing one value, you do not need to press
the knob --- pressing the Menu button while changing a value will save the value
and exit).
Next, connect the cable between your LaserMount or other fixture and the
Output connector of the TECSource. We recommend using our cables as they
have been designed to work well with the TECSource. If using your own cables,
ensure they have been properly wired according to the pin-out of the
TECSource and your fixture.
Finally, set the set point to an appropriate temperature and press the Output
button. The output will turn on and you will see the voltage and current begin
driving the fixture to the set point you have chosen. Depending on the thermal
size of your fixture, it may take seconds or several minutes to reach the set
point.

5400 Series TECSource User’s Manual · Page 7
If you notice the temperature is oscillating around the set point and not
stabilizing, you may need to adjust the Gain setting in the menu. You can use
the AutoTune PID feature to automatically calculate the best PID values, or
select from a set of eight factory preset values that typically cover most
applications. To use the factory gains, if the temperature is quickly jumping up
and down, the Gain will typically need to be reduced. If the temperature is slowly
moving up and down, try a higher Gain. You may need to experiment with
several gain settings to find the ideal value, and for even finer control, you can
set the Gain to PID and directly set the PID control values.
For more detailed operating and installation instructions, read on.
Installation
Installation of the 5400 TECSource is very straightforward, as the quick start
section above illustrated. This section will provide additional details and
considerations for installing your TECSource.
After unpacking the unit, make sure all packing materials have been removed
and nothing obscures the ventilation ports on the back and bottom of the unit.
The TECSource has a universal input power supply, accepting 90 to 240 VAC,
and 50 or 60 Hz. This covers all conventional power worldwide, but ensure your
AC power meets these requirements.
Powering Up the Unit
Once the correct voltage selection has been made, connect the AC power cord
to the unit.
Turn the power switch, located on the front of the unit, into the on (|) position.
The unit will display the current firmware revision, go through a quick power-up
self-test, and return to the last known operating state.
CAUTION
Do not exceed 250VAC on the line input.
It is critical to select the proper voltage selection prior to
applying power to the unit. If the actual voltage exceeds
the voltage selection by +/-10%, damage to the unit
may occur.

Page 8 · 5400 Series TECSource User’s Manual
Ventilation
The TECSource has vent holes on the rear and left side of the unit. You must
not block these vent holes, or overheating may occur, causing damage to the
unit.
Rack Mounting
The rack mounting kit (p/n 1403-RM) for standard 19” racks are available for the
TECSource. Because the unit draws air from the side, and therefore inside the
rack housing, be sure that the internal rack ambient temperature (which will
typically be several degrees higher than room ambient) does not exceed the
unit’s operating temperature.
Warm-up and Environmental Considerations
In order to achieve the highest level of accuracy, the TECSource should be
powered on for at least one hour prior to taking measurements. In addition,
ensure that the unit is not operating outside the ambient temperature range or
humidity conditions.
NOTE
As the unit nears maximum load conditions, it is normal
for the internal fan to significantly increase in speed.
CAUTION
Do not operate the unit above +40°C ambient, and
ensure the instrument is properly ventilated, or the unit
may overheat and possible damage to the instrument
may occur.

5400 Series TECSource User’s Manual · Page 9
Power and Cable Connections
In addition to the input power connector described above, there are five
connectors on the rear panel of the TECSource: Output connector, Auxiliary
connector, Interlock connector, USB connector, and RS232 connector.
TECSource Rear Panel
TEC Output Connector
The Output connector is a female 17W2, and has the following pin-out:
Pin Description Pin Description
A1 TE (+) A2 TE (–)
1 TEC On LED+ 8 TEC On LED– (Signal Ground)
2 Remote TEC Voltage Sense+ 9 Remote TEC Voltage Sense–
3 Sensor 1– 10 Sensor 1 Remote Sense–
4 Sensor 1+ 11 Sensor 1 Remote Sense+
5 Sensor 3 (AD590) – 12 Sensor 3 (AD590)+
6 Sensor 2– 13 Earth Ground
7 Fan+ 14 Sensor 2+
15 Fan– (Signal Ground)
Output Connector (17W2 Female)
Inverting the TE Output
In the event the polarity of the TE pins is backwards, it is possible to reverse the
direction of the TEC current by setting the Advanced » Invert ITE to Yes. While
not recommended as it can lead to confusion when the controller is used later is
properly wired setups, this allows for correct operation of system when the TE
pins are wired backwards and rewiring is not possible or feasible.
Using Remote Sensor
The Remote Sensor 1 pins (10 & 11) are used in RTD 4-wire mode, and provide
a remote measurement of the RTD voltage to eliminate voltage measurement

Page 10 · 5400 Series TECSource User’s Manual
errors. See Working with RTDs below for more information. Only Sensor 1 has 4-
wire remote sense capability.
Using Remote TEC Voltage Sense
The Remote TEC Voltage Sense pins (2 & 9) are used to remotely measure the
TEC voltage. See Using Remote TEC Voltage Sense below for more information.
Using the TEC On LED
Pins 1 & 8 on the Output connector can be used to drive an LED to indicate
when the TEC is on. The signal is driven from a +5V buffer with a 249Ωresistor
in series with the output, limiting the current to 20mA. This is suitable for the
direct connection of most green and red LEDs. Add additional series resistance
for lower current devices, such as blue LEDs.
TEC On LED Circuit Diagram
Auxiliary Connector
The Auxiliary connector is a female DB-25, and has the following pin-out:
Pin Description Pin Description
1 Digital Input 1 14 Digital I/O Ground
2 Digital Input 2 15 Digital I/O Ground
3 Digital Output 1 16 Digital I/O Ground
4 Digital Output 2 17 Digital I/O Ground
5 +5V (100mA MAX) 18 Digital I/O Ground
6 Relay N/C Contact 19 Relay Common
7 Relay N/O Contact 20
Reserved, no connection
8
Reserved, no connection
21
Reserved, no connection
9
Reserved, no connection
22 Sensor Ground
10 Auxiliary Sensor 1 23 Sensor Ground
11 Auxiliary Sensor 2 24 Sensor Ground
12 Auxiliary Sensor 3 25 Sensor Ground
13 Auxiliary Sensor 4
Auxiliary Connector (DB-25 Female)
See Using the Auxiliary Interface below for additional information.

5400 Series TECSource User’s Manual · Page 11
Interlock Connector
The Interlock connector is a 2-pin Phoenix socket with the following pin-out:
Pin Description
1 Interlock
2 Interlock Ground
Interlock Connector (2-Pin Phoenix)
See Using the Interlock below for additional information.
USB Connector
The USB connector is a standard Type B female connector, and can be plugged
into any USB 1.1 or USB 2.0 port. For more information on using the USB
interface, see the Computer Interfacing Manual which is included on the CD that
accompanied this product.
RS232 Connector
The RS232 connection is male DB-9 connector wired in a NULL modem
configuration.
Pin
Description
2
Receive
3
Transmit
5
Ground
1,4,6
Commoned together
7,8
Commoned together
9
No connection
Shell
Earth ground
RS232 Connector (DB-9 Male)
For more information on using the RS232 interface, see the Computer Interfacing
Manual which is included on the CD that accompanied this product.

Page 12 · 5400 Series TECSource User’s Manual
Connecting to the TECSource
Arroyo Instruments carries two cable assemblies specifically designed for
connecting the TECSource to temperature controlled fixtures or devices. The
1264 is a two meter cable, capable of up to 30A of TEC current, designed for
use with Arroyo Instruments mounts, and has 17W2 connectors on both ends,
one male and one female. The 1265 is identical to the 1264, except that it has
bare wires on the device end, allowing you to terminate the cable with your own
connector or mount.
See the manual for your fixture for additional safety and operational information.
Front Panel Operation
Operation of the TECSource is very straightforward. The sections below will
help familiarize you with the front panel, the display, and the menu structure.
The front panel is designed for simplicity in operation. There are six buttons on
the front panel: Output, Menu, Display, Func A, Func B, and the adjustment
knob button.
The Output button is used to turn the output on and off. Whenever the output is
on, the blue Output On LED will be lit.
The Menu button is used to enter the TECSource menu. When in remote mode,
it acts as a Local button, returning the instrument to local mode.
NOTE
Connections to the TECSource and the fixture must be
secure. Tighten any screws on the connectors, and
make sure all connections are in good condition.

5400 Series TECSource User’s Manual · Page 13
The Display button is used to cycle through the display modes of the controller.
Because the controller monitors more information than can be displayed at
once, the Display button allows you to cycle between the TEC information,
auxiliary temperature measurements, and digital I/O status.
The Func A and Func B buttons are used to execute user-defined macros,
recall saved settings, or other actions as defined in the Function Keys Menu.
The large adjustment knob located on the right hand side of the unit is used to
change the set point or parameters in the menu. It also acts as a push button,
primarily as an enter button, when making changes in the menu.
There are ten LEDs:
Color Function When lit, indicates:
Blue Output On TECSource is actively controlling the temperature
Red Error Error has occurred
Yellow Remote Unit is being controlled via a computer
Green Cool Instrument is cooling
Green Heat Instrument is heating
Green Stable Temperature is stable
Green ITE Limit Controller is in current limit
Green VTE Limit Controller is in voltage limit
Yellow T Limit Controller is in temperature limit
Green Link Controller is linked to another instrument
Whenever an error is generated, the red Error LED will light, and the error will be
displayed on the VFD display. There may be one or more errors, but only the
first error will be displayed. To display the next error, press Menu button. To
clear all errors, press the knob. A list of error codes can be found in the Error
Messages section below.
When the unit is in remote mode, the yellow Remote LED will be lit. More
information about how the instrument behaves in remote mode can be found in
the Remote Mode Operation section below.
Control Modes
The TECSource offers three control modes: constant temperature mode (T
Mode), and constant resistance mode (R Mode), and constant current mode (Ite
Mode).

Page 14 · 5400 Series TECSource User’s Manual
Changing the control mode is done through the menu by changing the Mode
parameter to T Mode, R Mode, or Ite Mode.
Constant temperature mode uses the sensor constants to calculate the
resistance of the thermistor at the desired or actual temperature.
While most users will only need the temperature control mode, the constant
resistance mode allows you to bypass the sensor equation and directly select
the sensor set point. This can be useful when only the desired sensor value is
known, or when the sensor-to-temperature conversion values are not available
for your sensor.
Main Display Screen
The top of the main display will be the temperature of the active sensor, except
when adjusting the set point. The bottom half of the display can display one of
three sets of data:
1. Set point and TEC current and voltage
2. Sensor readings for the non-active primary sensors
3. Sensor readings for the auxiliary sensors
4. Digital input and output status
Example displays are shown below:
Main Display: Set Point and TEC Current and Voltage
Main Display: Additional Primary Sensors
T=25.000°c
Sens 2= 25.000°c
Sens 3= Disabled
T=25.000°c
TSET= 25.000°c
I= 3.25AV= 6.72V

5400 Series TECSource User’s Manual · Page 15
Main Display: Auxiliary Sensors
Main Display: Digital I/O State
The set point will depend on the control mode you have selected. When in T
Mode, the set point will be temperature. When in R Mode, the set point will be in
the active sensor’s units (ohms for thermistor and RTD, uA for AD590, and mV
for LM335). When in Ite Mode, the set point will be in amps.
Measurements
The second line of the display contains the measurements appropriate for the
control mode. In T Mode, the actual temperature, current, and voltage can be
displayed, depending on the display mode. In R Mode, the actual sensor value
is displayed instead of the actual temperature. In Ite Mode, the actual
temperature is displayed.
No Set Point Error
Because the TECSource uses a digital PID loop, it will always stabilize at the set
point, and you will never have a temperature offset often found with analog
controllers.
Using Limits
The TECSource supports four different types of limits: temperature, sensor,
current, and voltage. There is both a temperature high and temperature low limit,
and if the actual temperature exceeds either of these limits, the output will be
shut off. Likewise, both high and low limits exist for the sensor (R limits), and the
output will be shut off if the sensor measurement exceeds either of these limits,
as determined by settings in the Output Off Menu.
T=25.000°c
Din 1/2/I:On / Off / Off
Dout 1/2/R:Off / Off / Off
T=25.000°c
Ax1=25.00 Ax3= Dis
Ax2=25.00 Ax4= Dis

Page 16 · 5400 Series TECSource User’s Manual
Settings and Menus
All parameters of the TECSource can be viewed and changed within the menu
(except for the user calibration adjustments, which must be done via the
computer interface). The menus are constructed with the most used parameters
first. To change any setting, press Menu to enter the menu then rotate the knob
to select the parameter to change. Press the knob to begin changing the value.
As a visual indication that you are in edit mode, you will see an asterisk appear
next to the value. Once you have made your change, press the knob or Menu
button to store the value. Pressing the Menu button will store and exit the menu,
while pressing the knob will store the value but leave you in the menu to make
additional changes.
Some settings are contained inside a sub menu, such as communications
settings. To access the sub menu, simply press the knob to enter the sub menu
when its name is displayed.
Below is a complete list of available settings:
Menu Description Factory
Default
Root Menu
Mount
Specify the mount connected to the
TECSource. If using a LaserMount, select
the appropriate model. Otherwise, select
User Defined.
User
Defined
Sensor
Sets the temperature measurement sensor
for the active sensor. Can be Therm 100uA,
Therm 10uA, LM335, AD590, RTD, or RTD
(4-wire).
Therm
100uA
Gain
Gain controls the response of the
temperature controller. A higher gain value
will cause the controller to respond more
quickly to the difference between the set
point and the actual temperature, while a
lower value will cause it to respond more
slowly. Select PID for direct access to the
PID parameters. See Gain Control and the
PID Loop section below.
10
PID P The proportional term of the PID loop. Will
only be available if Gain is set to PID. 1

5400 Series TECSource User’s Manual · Page 17
Menu Description Factory
Default
PID I The integral term of the PID loop. Will only
be available if Gain is set to PID. 0.01
PID D The derivative term of the PID loop. Will
only be available if Gain is set to PID.0
I Lim
I Lim sets the current limit of the
temperature controller. The limit should be
set to a value that is suitable for your Peltier
device. See Using the Current Limit below
for more details.
3A
V Lim
V Lim sets the voltage limit of the
temperature controller. See Using the
Voltage Limit below for more details.
Instrument
Rating
+ 2 Volts
T-Lo Lim
Only displayed when operating in T & Ite
Modes, T-Lo Lim is the lower temperature
limit of the active sensor. It is a duplicate of
the active sensor T Lo Lim in the Sensor
Limits menu, copied here for ease of
access.
-99°C
T-Hi Lim
Only displayed when operating in T & Ite
Modes, T-Hi Lim is the upper temperature
limit of the active sensor. It is a duplicate of
active sensor T Hi Lim in the Sensor Limits
menu, copied here for ease of access.
125°C
R-Lo Lim
Only displayed when operating in R Mode,
R-Lo Lim is the lower sensor limit of the
active sensor. It is a duplicate of the R-Lo
Lim in the Sensor Limits menu, copied here
for ease of access.
0.01kΩ
R-Hi Lim
Only displayed when operating in R Mode,
R-Hi Lim is the upper sensor limit of the
active sensor. It is a duplicate of the R-Hi
Lim in the Sensor Limits menu, copied here
for ease of access.
45kΩ
Mode
This sets the operating mode (T Mode, R
Mode, or Ite Mode) of the temperature
controller.
T Mode

Page 18 · 5400 Series TECSource User’s Manual
Menu Description Factory
Default
Main Sensor Menu
Active Sensor Selects which primary sensor to be used as
feedback sensor for the control loop. 1
Sens 1 Sets the Primary 1 temperature sensor
type.
Therm
100uA
Sens 2 Sets the Primary 2 temperature sensor
type. Disabled
Sens 3 Sets the Primary 3 temperature sensor
type. Disabled
ThermA The A term in the Steinhart-Hart equation
for use with primary sensors 1 & 2. 1.12924E-03
ThermB The B term in the Steinhart-Hart equation
for use with primary sensors 1 & 2. 2.34108E-04
ThermC The C term in the Steinhart-Hart equation
for use with primary sensors 1 & 2. 0.87755E-07
RTD A The A term in the RTD equation for use
with primary sensors 1 & 2. 3.98480E-03
RTD B The B term in the RTD equation for use
with primary sensors 1 & 2. -0.58700E-06
RTD C The C term in the RTD equation for use
with primary sensors 1 & 2. 4.00000E-12
RTD R0The R0term (in Ω) in the RTD equation
for use with primary sensors 1 & 2. 100.00
AD590 M The slope term in the AD590 correction
equation for use with primary sensor 3. 1.00000
AD590 B
The offset term, in °C, in the AD590
correction equation for use with primary
sensor 3.
0.000
LM335 M
The slope term in the LM335 correction
equation for use with primary sensors 1
& 2.
1.00000
LM335 B
The offset term, in °C, in the LM335
correction equation for use with primary
sensors 1 & 2.
0.000

5400 Series TECSource User’s Manual · Page 19
Menu Description Factory
Default
Tol Time
Tolerance time is the amount of time, in
seconds, that the actual temperature must
be within the set point temperature +/- the
Tol Temp value for the unit to be
considered in tolerance.
5 seconds
Tol Temp
Tolerance temperature is a temperature
band (in °C) around the set point
temperature. When the actual temperature
is within this band for longer than the Tol
Time setting, then the unit is considered to
be in tolerance.
0.1°C
Aux Sensor Menu
Aux1 Sets the Auxiliary 1 temperature sensor
type. Disabled
Aux2 Sets the Auxiliary 2 temperature sensor
type. Disabled
Aux3 Sets the Auxiliary 3 temperature sensor
type. Disabled
Aux4 Sets the Auxiliary 4 temperature sensor
type. Disabled
ThermA The A term in the Steinhart-Hart equation
for use with auxiliary sensors. 1.12924E-03
ThermB The B term in the Steinhart-Hart equation
for use with auxiliary sensors. 2.34108E-04
ThermC The C term in the Steinhart-Hart equation
for use with auxiliary sensors. 0.87755E-07
Sensor Limits Menu
T1 Lo Lim Low temperature limit for Primary Sensor 1 -99°C
T1 Hi Lim High temperature limit for Primary Sensor 1 125°C
T2 Lo Lim Low temperature limit for Primary Sensor 2 -99°C
T2 Hi Lim High temperature limit for Primary Sensor 2 125°C
T3 Lo Lim Low temperature limit for Primary Sensor 3 -99°C
T3 Hi Lim High temperature limit for Primary Sensor 3 125°C
Ax1 Lo Lim Low temperature limit for Aux Sensor 1 -99°C

Page 20 · 5400 Series TECSource User’s Manual
Menu Description Factory
Default
Ax1 Hi Lim High temperature limit for Aux Sensor 1 125°C
Ax2 Lo Lim Low temperature limit for Aux Sensor 2 -99°C
Ax2 Hi Lim High temperature limit for Aux Sensor 2 125°C
Ax3 Lo Lim Low temperature limit for Aux Sensor 3 -99°C
Ax3 Hi Lim High temperature limit for Aux Sensor 3 125°C
Ax4 Lo Lim Low temperature limit for Aux Sensor 4 -99°C
Ax4 Hi Lim High temperature limit for Aux Sensor 4 125°C
R-Lo Lim Sensor low limit. Only applies to the active
Primary sensor. 0.01kΩ
R-Hi Lim Sensor high limit. Only applies to the active
Primary sensor. 45kΩ
Analog Out Menu
Function Selects the output function of the Analog
Output. Disabled
Analog Set When Function is set to Direct, sets the
output voltage of the Analog Out BNC. 0.000V
T Error
When Function is set to T Error, selects the
transfer function for generating the output
voltage as a function of the temperature
error.
0.01°C/V
T Range
When Function is set to Temperature,
selects the transfer function for generating
the output voltage as a function of
temperature.
10°C/V
Digital I/O Menu
Di1 Func Selects the action to perform when Digital
Input 1 is true. None
Di1 Invert Inverts the active state of Digital Input 1. No
Di2 Func Selects the action to perform when Digital
Input 2 is true. None
Di2 Invert Inverts the active state of Digital Input 2. No
Do1 Func Selects what status is used to control
Digital Output 1. Off
This manual suits for next models
3
Table of contents
Other Arroyo Instruments Temperature Controllers manuals
Popular Temperature Controllers manuals by other brands

Ropex
Ropex RESISTRON RES-5012 operating instructions

Store It Cold
Store It Cold CoolBot instruction manual

MESCOLI
MESCOLI LK100 instruction manual

Honeywell
Honeywell C7189U installation instructions

Greystone Energy Systems
Greystone Energy Systems HATSOB Series installation instructions

Siemens
Siemens RDF300 Series Documentation