Age The Normal Temperature Clinical Deviation
0-2 years 36.4℃~38.0℃≤±0.3℃
3 to 10 years 36.1℃~37.8℃≤±0.2℃
11-65years 35.9℃~37.5℃≤±0.2℃
Under the age of 65, 35.8℃~37.4℃≤±0.3℃
Note: It is not unusual for temperature values to vary according to skin color,
thickness, body parts, and ambient temperature.The reason is that the more
exposed the body parts are more affected by the environmental temperature.
Statement:
The clinical accuracy or clinical deviation of the thermometer has been clinically verified, and its safety and
effectiveness are in line with the relevant national regulations. The verification method of the claimed
performance of the product can be obtained by contacting the customer service department of the company.
Contraindications, precautions, warnings and suggestive instructions
1. The Contraindications
(1) Thereis no standard body temperature. Consult your doctor if you have a fever.
(2) patients with coma and terminal stage.
2. Precautions
(1) Thisproduct is a professional medical infrared forehead thermometer for measuring human body temperature.
According to the difference of human skin, the measured temperature will be different.
(2) Whenthe room temperature changes greatly, the medical infrared forehead thermometer should be placed in
the room for 20 minutes before use.
(3) When the tested person comes from a place where there is a big difference with the measured ambient
temperature, he should stay in the test environment for at least 5 minutes, and then be measured after the
temperature is consistent with the ambient temperature.
(4) Thesurrounding environment of the tested person should be stable, and the test should not be conducted in
places with large airflow such as fans and air vents of air conditioners.
(5) The instrument should not be used outdoors or in places with strong sunlight.
(6) Donot hold the front end of the thermometer by hand when measuring.
(7) Themeasuring area shall not be covered by hair.If there are sweat stains, please clean them with a dry towel
and measure them. Otherwise, the measurement results will be affected.
(8) Fever patients with cold forehead, sweating and other cooling measures will make the measurement results
low, should be avoided in this case.
(9) Whenusing the product, please keep away from the high-voltage equipment and signal transmitter to avoid the
danger of interference and wrong action.
(10) This thermometer has sensitive electronic components, so do not in the environment with strong
electromagnetic interference (such as: mobile power, microwave, etc.) directly use the thermometer, because it is
likely to lead to accuracy distortion.
(11) Do not replace the original parts with parts not provided by the manufacturer. If the original parts are replaced
with parts not provided by the manufacturer or the machine is dismantled by itself, measurement errors and
accuracy may be caused. If necessary, please contact the supplier