ATEQ D6 Series User manual

ATEQ D6 Class
Quick Start Guide

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
2/38
Table of contents
Safety advisory / Warranty
Good practices and safety instructions 4
Air quality requirements 5
Preamble
ATEQ D6 Class, a universal flow meter 6
Flow rate test 7
Principe of a cycle 8
Your ATEQ D6 Class
Front panel 9
Connectors on the back panel (with all options) 10
Power supply connectors 12
Digital links 13
Digital inputs/outputs 16
Pneumatic connectors 19
Pneumatics configuration 23
User interface
Overview 27
Keys 27
Display 28
Starting up
Power up 30
Preparing a program 30
Modifying a parameter 31
Selecting a program 32
Starting and stopping current cycle 32
User adjustments
Options of the menus 33
Specifications
Characteristics 38

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
3/38
ATEQ - Measurement Solution, Global Leader.
ATEQ
15, rue des Dames, Z.I.
78340 LES CLAYES-SOUS-BOIS
FRANCE
ateq.com
T.: +33 1 30 80 1020
F.: +33 1 30 54 1100
ATEQ K.K.
3 –41 ATEQ Building, Ikehata
Chiryu-city, Aichi-pref
JAPAN
ateq.co.jp
T.: +81 566-84-4670
F.: +81 566-84-4680
ATEQ China
98 Jian Peng Lu
Shanghai
CHINA
ateq.com.cn
T.: +86 21 6763 9508
F.: +86 21 6763 9528
ATEQ SYSTEMS ANALYSIS TAIWAN CO.,
LTD.
NO. 3, LAN 223, San Jia Dong Street
40642, TAICHUNG
TAIWAN
ateq.com.tw
T.: +886 4 2437 5278
F.: +886 4 2437 3675
ATEQ CORP.
35980 Industrial Road Suite L
Livonia MI 48150
UNITED STATES
atequsa.com
T.: +1 734-838-3100
F.: +1 734-838-0644
We continuously work on improving our products. This is why information contained in
this manual, the device and the technical specifications may be modified without prior
notification.
Pictures and figures in this manual are non-contractual.

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
4/38
Safety advisory / Warranty
GOOD PRACTICES AND SAFETY INSTRUCTIONS
Safety recommendations
If the device is supplied with 100 / 240 V AC, it is mandatory to connect it to the ground
with a good link to the ground, to protect against electric hazard or electrocution.
It is dangerous to change the status of the outputs.
They can control power actuators or other equipment (mechanical, pneumatic, hydraulic,
electrical or other) which can cause serious personal injury and damage to surrounding
material.
For safety and quality measurement reasons, it is important, before powering on the
device, to ensure that it is air supplied with a minimum operating pressure (0.6 MPa ±
15%).
Recommendations for the test environment
Keep the test area as clean as possible.
Recommendations for operators
ATEQ recommends that the operators who use the devices have training and a level of
qualification that correspond to the job to perform.
General recommendations
—Read the user manual before using the device.
—All electrical connections to the device must be equipped with safety systems (fuses,
circuit breakers, etc.) adapted to the needs and in accordance with the applicable
standards and rules.
—To avoid electromagnetic interference, electrical connections to the device must be
shorter than 2 meters.
—Power supply plug must be grounded.
—Disconnect the device from the mains before performing any maintenance work.
—Shut off the compressed air supply when working on the pneumatic assembly.
—Do not open a connected device.
—Avoid splashing water on the device.
ATEQ is at your disposal for any information concerning the use of the device under
maximum safety conditions.
We draw your attention to the fact that ATEQ cannot be held responsible for any accident
related to a misuse of the measuring instrument, the workstation or non-compliance of the
installation with safety rules.
In addition, ATEQ declines any responsibility for the calibration or the fitting of their
instruments that is not done by ATEQ.
ATEQ also declines any responsibility for any modification (program, mechanical or
electrical) of the device done without their written consent.

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
5/38
AIR QUALITY REQUIREMENTS
The air supplied into the device must be clean and dry. Even though the device is provided
with a filter, the presence of dust, oil or impurities may cause malfunction.
Air quality requirements according to ISO standard 8573
The air must be clean and dry.
The presence of impurities, oil or humidity in the air may cause deterioration
which will not be covered by the warranty.
When the instrument is working in vacuum conditions, impurities must be
prevented from being drawn into its internal components.
For this purpose, we strongly recommend that a suitable airtight filter is installed
between the part under test and the instrument.
ATEQ recommends the following characteristics for the air supplied into the device.
Recommended additional equipment
ATEQ recommends the installation of this additional equipment:
—Air dryer to provide dry air at less than - 40°C dew point
—25 microns and 1/100 microns double filter.
Air characteristics
ISO standard 8573 class
Grain size and concentration
0.1 μm and 0.1 mg/m3
Class 1
Dew point under pressure
- 40°C dew
Class 2
Maximum concentration of oil
0.01 mg/m3
Class 1

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
6/38
Preamble
ATEQ D6 Class, A UNIVERSAL FLOW METER
ATEQ D6 Class is a universal flow meter that measures flow rates through parts to test.
ATEQ D6 Class can memorize 128 different test programs.

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
7/38
FLOW RATE TEST
The ATEQ D6 series can do direct or indirect measurements. In both cases, the flow meter
can also work in vacuum (optional).
Measurement principle
Whenthefluid(gas)entersthedevice
1
,itmovesthroughacalibratedflowtube
2
which
causesadropinpressure.Thepressuredropismeasuredbyadifferentialpressuresensor
3
. The pressure ofthe part under test
5
ismeasured bythe sensor
4
.
1 Device
2 Calibrated flow tube
3 Differential pressure sensor
4 Pressure sensor
5 Part under test
Direct measurement
Once the part is filled, the device measures the flow rate through the part.
The test pressure is applied to the device 1 and then to the input of the part under test 2.
1 Device
2 Part under test

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
8/38
Indirect measurement
The indirect measurement (also called recovery mode or under bell) allows a considerable
gain in time as the flow which enters into the device is already stabilized.
The test pressure is first applied to the input of the part under test
1
and then to the
device 2. The output of the device 2 is vented to the atmosphere.
Thismethod allows totest thepartsatahigher pressure test (up to2MPa).
1 Part under test
2 Device
Depending on the part type, it may be possible to use a bell (ex: shower head where
it is impossible to recover the flow other than through the use of a bell). This method
can only be used if the recovery of the flow is easy.
PRINCIPLE OF A CYCLE
The measurement cycle is made of 4 main phases: fill, stabilization, test, dumping.
A Pressure auto zero additional phase 0can be placed at the start or at the end of a cycle,
depending on the requirement of the operator.
0 Pressure auto zero phase
1 Fill phase
2 Stabilization phase
3 Test
4 Dumping

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
9/38
Your ATEQ D6 Class
FRONT PANEL
The user interface is located on the front panel.
1 Display
2 Cycle keys
3 Navigation keys
4 USB connectors
For more information, refer to User interface.

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
10/38
CONNECTORS ON THE BACK PANEL (WITH ALL OPTIONS)

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
11/38
Ref
Name
Description
1
J1
Analog outputs - pressure and flow (option)
2
J5
Fieldbus connector
3
J8*
Extender (not operational)
4
J10
Program selection extension connector (option)
5
J12
Printer RS232 connector / Modbus (option) or Profibus
(option)
6
M1
External capillary option
7
J13
Temperature probe connector (option for external capillary)
8
P
External back pressure connector (option)
9
-
Input connector to the air filter (valves or regulator air
supply)
10
-
0.6 MPa valves air supply input (1 MPa range) or vacuum
input
11
TEST
Capillary connector to connect to the part under test
12
-
Regulator output (indirect mode or bypass option or
external capillary option)
13
-
Pneumatic output for external bypass option
14
ATM
Output capillary (indirect mode): has to be left to the
atmosphere
15
B
Pneumatic output for external shut off option or B
automatic connector option
16
ATM
Output to be left to the atmosphere
17
A
Pneumatic output for external dump option or A automatic
connector option
18
ATM
Output to be left to the atmosphere
19
-
Air supply energy information
20
ATEQ
Part number / Serial number
21
M2
External capillary connector (option)
22
-
Ground
23
J11
Relay board connector (digital inputs/outputs and 24 V DC
- 2 A
power supply)
24
J9
Outputs code board connector (digital inputs/outputs)
25
J7
Connector for 24 V DC - 2 A or 100 / 240 V AC power
supply
(according option provided)
26
J4*
USB (not operational)
27
J3*
Dry contact input for ERD test mode (option)
28
J6
Fieldbus connector
29
J2*
Network (not operational)
* These connectors are not operational. They are provided for future development
of our devices.

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
12/38
POWER SUPPLY CONNECTORS
The device can be connected to an external power supply (24 V DC –2A) or provided with
an internal power supply (100 / 240 V AC) (option).
External supply
24 V DC connector (J7)
The device can be connected to a 24 V DC –2A power supply through a M12 4 pins type
connector.
24 V DC on the relay board connector (J11) (option)
The device can also be connected to a 24 V DC - 2 A power supply through J11 connector
on the relay board.
Apply 24 V DC to the pin 2or 4.
Pin number
Signal
1
Not connected
2
+ 24 V DC
3
Not connected
4
Ground: 0 V
Pin number
Signal
2
+ 24 V DC
4
+ 24 V DC
16
Ground: 0 V

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
13/38
Internal supply only
100 / 240 V AC connector (J7) (option)
The device can be connected to a 100 / 240 V AC power supply (option).
This connector has an ON/OFF button.
It is mandatory to connect the device to the groundwith a good link to the ground, to protect
against electric hazard or electrocution.
1 ON
0 OFF
DIGITAL LINKS
PC USB connectors (on front face)
USB connectors can be used for connecting miscellaneous compatible USB devices.
The USB connectors are located under the rubber cover 1(see figure).
1 Rubber cover
2 USB connector to PC
3 USB connector to USB key
Do not connect two USB devices at the same time.
Do not use a cable longer than 2 m.
Push the rubber cover 1slightly forward for an easy access to USB connectors 2and 3.
Only use this connection for temporary communication. Connection to a PC cannot be used
permanently because the communication can be disconnected by the PC.

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
14/38
Printer RS232 connector / Modbus (option) or Profibus
(option)
(J12)
RS232 - SubD 9 pins male connector (printer)
RS232 for printer, bar code reader, PC
connection.
RS232 - SubD 9 pins female connector (Profibus) option
Profibus: SubD 9 pins female connector.
Devicenet connectors (J5) (J6) (option)
M12 type connector - 5 pins male connector (J5) (Devicenet input)
For connection to others ATEQ devices.
Pin number
Signal
1
Not used
2
RXD data input
3
TXD data input
4
Not used
5
Ground
6
Not used
7
RTS request to send
8
CTS clear to send
9
Not used
Pin number
Signal
1
PE (ground)
2
Not used
3
Data line A
4
CNTR - A (repeater control
signal)
5
DGND (logic ground)
6
VP (supply)
7
Not used
8
Data line B
9
Not used
Pin number
Signal
1
Drain
2
V+
3
V-
4
CAN_H
5
CAN_L

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
15/38
M12 type connector - 5 pins female connector (J6) (Devicenet
output)
For connection to other ATEQ devices.
Profinet connectors (J5 + J6) (option)
M12 D coded type connector - 4 pins female connector (J5 + J6)
Ethernet connector (J5 + J6) (option)
Standard connection Ethernet TCP / IP protocol.
One of these network protocols is available:
—Ethernet IP
—Profinet
—Ethercat
Pin number
Signal
1
Drain
2
V+
3
V-
4
CAN_H
5
CAN_L
Pin number
Signal
1
Ethernet Tx + (Transmit Data +)
2
Ethernet Rx + (Receive Data +)
3
Ethernet Tx - (Transmit Data -)
4
Ethernet Rx - (Receive Data -)

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
16/38
DIGITAL INPUTS/OUTPUTS
The 24V DC power supply for the digital inputs can be provided by 2 means:
—The internal power supply of the device (0.3 A max)
—An external power supply provided by the customer.
Inputs default mode is PNP. NPN mode is available on request.
Relay board connector (J11) (option)
Characteristics
—Inputs
• Activation: + 24 V DC
—Outputs
• Dry contacts
• 60 V AC / DC max - 200 mA max.
The device can be energized through the J11 connector of the relay board (except if
internal supply option):
0 V to the pin 16
24 V DC to the pin 2or 4
Pin number
Inputs / outputs
Description
1
Input 1
RESET
2
+ 24 V DC
Common
3
Input 2
START
4
+ 24 V DC
Common
5
Input 3
Program selection
6
Input 4
Program selection
7
Input 5
Program selection
8
Input 6
Program selection
9
Input 7
Program selection (programmable input)
10
Output
Common floating output
11
Output
Pass part
12
Output
High flow fail part
13
Output
Low flow fail part
14
Output
Warning
15
Output
End of cycle
16
0 V
Ground

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
17/38
Program selection extension connector (J10) (option)
The J10 connector is an extension of the J11 connector that enables the selection of 128
programs.
Characteristics
—Inputs
• Activation: + 24 V DC.
Program selection (J11 and J10)
The connectors J11 and J10 (option) enable you to select a program from digital inputs.
Combinations of connector pins to activate for program selection.
* X is equal to 0 or 1 in function of the program number.
Pin number
Inputs/outputs
Description
I8
Input 8
Program selection from 33 to 64 (programmable input)
I9
Input 9
Program selection from 65 to 128 (programmable input)
Program
number
J11
J10
Pin 5
(input 3)
Pin 6
(input 4)
Pin 7
(input 5)
Pin 8
(input 6)
Pin 9
(input 7)
Pin 1
(input 8)
Pin 2
(input 9)
1
0
0
0
0
0
0
0
2
1
0
0
0
0
0
0
3
0
1
0
0
0
0
0
4
1
1
0
0
0
0
0
5
0
0
1
0
0
0
0
6
1
0
1
0
0
0
0
7
0
1
1
0
0
0
0
8
1
1
1
0
0
0
0
9
0
0
0
1
0
0
0
10
1
0
0
1
0
0
0
11
0
1
0
1
0
0
0
12
1
1
0
1
0
0
0
13
0
0
1
1
0
0
0
14
1
0
1
1
0
0
0
15
0
1
1
1
0
0
0
16
1
1
1
1
0
0
0
17 to 32
X*
X
X
X
1
X
X
33 to 64
X
X
X
X
X
1
X
65 to 128
X
X
X
X
X
X
1

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
18/38
Valve codes and auxiliary outputs board connector (J9)
(option)
Characteristics
—Outputs:
• 24 V DC - 100 mA max per output.
—Inputs:
• Activation: + 24 V DC.
Pin number
Inputs / outputs
Description
1
+ 24 V DC
Common (outputs 1, 2,3)
2
Output 1
Open collector
3
Output 2
Open collector
4
Output 3
Open collector
5
+ 24 V DC
Common (outputs 4, 5, 6)
6
Output 4
Open collector
7
Output 5
Open collector
8
Output 6
Open collector
9
Input 1
Programmable input
10
Input 2
Programmable input
11
Input 3
Programmable input
12
Input 4
Programmable input
13
Input 5
Programmable input
14
0 V
Ground
15
Input 6
Programmable input
16
0 V
Ground

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
19/38
PNEUMATICS CONNECTORS
Pneumatic connectors used to connect the part under test are located on the back panel of
the device.
Pneumatic supply
The pneumatic supply has to meet specific requirements recommended by ATEQ. Refer to
Good practices and safety instructions section.
A specific filter may be necessary.
The air is supplied via the filter located on the back panel of the device.
Metal air filter
The metal filter is used for 1MPa range.
The maximum pressure admissible is 1.2
MPa.
Plastic air filter
The plastic filter is used for 0.5 MPa range
(direct and indirect modes) or 2MPa range
(indirect mode only).
The maximum pressure admissible is 690
kPa.

QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
20/38
Pneumatic outputs 0.6 MPa
DUMP: pneumatic output to connect to
an external valve for dump option.
SHUT OFF: pneumatic output to
connect to an external valve for shut off
option.
AandB:automaticconnectorsoption.
These connectors are used to drive
pneumatic caps on the part under test.
External capillary connectors
M1 and M2 connectors are pneumatic
inputs used for external capillary
connection (option).
External pressure monitoringconnector
The P connector is a pneumatic input
used for external pressure monitoring
connection (option).
Test connector
The test connector is used to connect the device to thepart to test for both direct and indirect
modes.
Metallic fitting diameter options (in mm):
—4/6
—6/8
—8/10
—15
—20
—25
Other manuals for D6 Series
2
Table of contents
Other ATEQ Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Wohler
Wohler VAMP AC-3/M Brochure & specs

Rice Lake
Rice Lake 1280 Enterprise Series installation manual

Zoom
Zoom Handy REcorder H1 Basic instructions

PPI
PPI ODPM-48 operating manual

Emerson
Emerson Rosemount Wireless Permasense WT210 quick start guide

Leaksafe Solutions
Leaksafe Solutions WaterComm Installation and user instructions