Atom Medical Corporation V-2200 Manual



CONTENTS
1. Definition of Symbols................................................................................................. 3
1-1. Symbols to indicate danger, warning or caution........................................... 3
1-2. Symbols to prohibit action ............................................................................. 3
1-3. Symbols to give instructions for action ......................................................... 3
1-4. Symbols of international standards (IEC) .................................................... 3
1-5. Other symbols................................................................................................. 4
2. General Precautions .......................................................................................... 7
2-1. DANGER................................................................................................. 7
2-2. WARNING............................................................................................... 7
2-3. CAUTION ............................................................................................... 7
3. Introduction................................................................................................................ 9
4. Parts Identification ................................................................................................... 10
4-1. Main Body..................................................................................................... 10
4-2. Display/Control Panel .................................................................................. 12
4-3. Screen Display.............................................................................................. 14
5. Maintenance Inspection........................................................................................... 20
5-1. Inspection before Use ................................................................................... 21
5-2. Three-month Inspection............................................................................... 29
5-3. Disinfection................................................................................................... 30
5-4. Periodical Replacement Parts...................................................................... 37
5-5. Inspection Checklist ..................................................................................... 41
6. How to Identify the Cause of Trouble....................................................................... 44
6-1. Parts Arrangement Plan .............................................................................. 44
6-2. Description of Functions (Description of Components).............................. 45
6-3. Description of Functions (Maintenance Screen)......................................... 49
6-4. Description of Functions (Setting the Pulse Oximeter) ............................. 55
6-5. Description of Functions (Prevention of a High Incubator
Air Temperature and a Low Incubator Air Temperature).......................... 56
6-6. Alarms........................................................................................................... 57
7. Repair ...................................................................................................................... 66
7-1. Jigs, Tools and Measuring Instruments to be Used ................................... 66
7-2. Flow of Parts Replacement .......................................................................... 68
7-3. Disassembling Procedure............................................................................. 69
7-4. Assembling Procedure ................................................................................ 107
7-5. Adjustment Procedure................................................................................ 145
7-6. Inspection Procedure.................................................................................. 160
7-7. Replacement Procedure.............................................................................. 180
8. Control Unit Block Diagram.................................................................................... 184
9. Wiring Diagram...................................................................................................... 186
9-1. Main Unit.................................................................................................... 186
9-2. HL-stand ..................................................................................................... 189
10. Replacement Parts List........................................................................................ 191
11. Technical Information ........................................................................................... 202
11-1. Technical Data .......................................................................................... 202
11-2. EMC Level and Classification.................................................................. 205


3
1. Definition of Symbols
1-1. Symbols to indicate danger, warning or caution
Symbol 《Title》and Indication
《General attention》
Indicates unspecified general danger, warning or caution.
《Caution: Hot surface》
Indicates that the surface can be dangerously hot under certain conditions.
1-2. Symbols to prohibit action
Symbol 《Title》and Indication
《General prohibition》
Indicates unspecified general prohibition.
《Prohibition of disassembly》
Indicates prohibition of disassembly of the unit where it may cause an electric shock or
other hazards.
《Prohibition of use of fire》
Indicates prohibition of use of fire where an external use of fire may cause the unit to
ignite under certain conditions.
1-3. Symbols to give instructions for action
Symbol 《Title》and Indication
《General instruction》
Indicates unspecified general action on the part of the user.
《Connect a ground wire》
Instructs the user to connect the ground wire without fail where the unit is provided with
a ground terminal.
《Remove the power plug from the power outlet》
Instructs the user to remove the power plug from the power outlet in the case of
malfunction or when there is a threat of lightning.
1-4. Symbols of international standards (IEC)
Symbol 《Title》and Indication
《Power on》
Indicates in relation to the power switch that the device is connected to the power source.
《Power off》
Indicates in relation to the power switch that the device is not connected to the power
source.
《Equipment partially on》
Indicates that a part of the device is “ON”.
《Equipment partially off》
Indicates that a part of the device is “OFF”.

4
Symbol 《Title》and Indication
《Type BF applied part》
Indicates that the device is classified as Type BF in terms of the degree of protection
against an electric shock.
《Bell》
Indicates a switch for controlling the bell function.
《Bell silenced》
Indicates a control switch that can switch off the bell, or the state of the bell function.
《Fan》
Indicates a control switch that controls the fan.
《Illuminance, Brightness》
Indicates a control switch for illuminance or brightness.
《Screen changeover》
Indicates a switch that changes over the screen.
《Caution》
Indicates that the user needs to see a relevant accompanying document before operating
the unit.
《Date of manufacture》
Indicates the date when the unit was manufactured in the factory.
《Manufacturer》
This symbol indicates the name and the address shown adjacent to the symbol is of the
manufacturer.
《WEEE symbol》
In the EC area, an electrical and electronic product falling in one of the categories
specified by “DIRECTIVE 2002/96/EC OF THE EUROPEAN PARLIAMENT AND OF
THE COUNCIL of 27 January 2003 on Waste Electrical and Electronic Equipment
(WEEE)” should be disposed of in a manner consistent with relevant laws and regulations.
This symbol indicates that the above-mentioned requirement applies to this product.
《Authorized representative in the European Community》
This symbol indicates the name and the address shown adjacent to the symbol is of the
authorized representative in the European Community.
1-5. Other symbols
Symbol 《Title》and Indication
《Setting》
Indicates that a setting is increased.
《Setting》
Indicates that a setting is decreased.
《Alarm》
Indicates a system failure alarm.
《Alarm》
Indicates a low voltage alarm.
《Alarm》
Indicates a set point alarm.

5
Symbol 《Title》and Indication
《Alarm》
Indicates that the rotation of the fan is abnormal or the fan has not been attached.
《Alarm》
Indicates a skin temperature probe alarm.
《Alarm》
Indicates a high temperature is above a set point.
《Heater》
Indicates the heater output.
《Manual control》
Indicates the manual control mode, manual control switch and incubator air temperature.
《Servo control》
Indicates the servo control mode, servo control switch and skin temperature.
《Override》
Indicates the override mode and override switch.
《Temperature unit》
Indicates the switch to select either °C or °F as the unit of the temperature to be displayed.
《Setting》
Indicates a set value.
《Humidity》
Indicates a function related to humidity.
《Humidity Sensor》
Indicates a function related to a humidity sensor.
《Humidity chamber off》
Indicates that the humidity chamber is not attached properly.
《Low water level》
Indicates that there is little water in the humidity chamber.
《Oxygen controller ON/OFF switch》
Indicates the oxygen controller ON/OFF switch.
《Oxygen》
Indicates a function related to oxygen.
《Oxygen flow rate》
Indicates a function related to an oxygen flow rate.
《Oxygen sensor》
Indicates a function related to an oxygen sensor.
《SpO2》
Indicates a detected current SpO2value.
《Pulse rate》
Indicates a detected pulse rate.

6
Symbol 《Title》and Indication
《Increased %SpO2》
Indicates that %SpO2has risen above the high limit.
《Decreased %SpO2》
Indicates that %SpO2has dropped below the low limit.
《Increased pulse rate》
Indicates that pulse rate has risen above the high limit.
《Decreased pulse rate》
Indicates that pulse rate has dropped below the low limit.
《Pulse oxisensor》
Indicates a function related to a sensor for pulse oximeter.
《Weight monitor》
Indicates the weight switch.
《Weight monitor》
Indicates the tare switch.
《Weight monitor》
Indicates the weight mode switch.
《Weight monitor calibration》
Indicates the switch to calibrate weight monitor.

7
2. General Precautions
2-1. DANGER
Death or serious injury, damage to equipment, or a fire will result if the instructions given below are
not followed.
Do not use or repair the unit where an explosive oxygen, nitrous oxide or a flammable
anesthetic gas is used.
2-2. WARNING
Death or serious injury due to a fire or an electric shock will result if the instructions given below
are not followed.
Do not operate or repair the unit with wet hands.
Connect the power cord of the unit to a power outlet including a ground terminal and meeting
the rated power requirements of the unit.
Be sure to disconnect the unit from the power source before handling the unit.
If a fingertip or a tool should touch a harness or a terminal in the unit, an electric shock
or a short circuit may result.
2-3. CAUTION
Injury or damage to ambient objects may result if the instructions given below are not followed.
Only qualified personnel authorized by Atom or a service representative properly trained in
the handling of the unit should inspect, repair, change the settings of, or disassemble and
assemble the unit.
Do not operate or repair the unit in electromagnetic environments where electrosurgical
knives or imaging devices such as magnetic resonance imaging (MRI) devices and
computerized tomography devices are used, or near a device equipped with an electromotor
that will consume a lot of electric currents.
The electronics of the unit may be affected by the radiation of various electromagnetic
waves in the hospital and may cause a malfunction of the unit.
Do not disinfect the unit by immersing it in a medical fluid. Avoid autoclaving or sterilization
with ethylene oxide gas (EOG).
Do not immerse the unit in water or a medical fluid.
If a portion of the water or the medical fluid should enter the unit, it may short an
electronic circuit inside.
After inspecting, repairing and disassembling and assembling, changing the settings of, or
disinfecting the unit, check that the unit operates normally.

8
The use of accessories and cables other than those sold by Atom as replacement parts for
internal components may result in increased emission or decreased immunity of the unit.
The unit or the system should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, observe the unit or the system to verify normal
operation in the configuration in which it will be used.
In replacing parts, make sure that the new parts you intend to use are those specified by
Atom.
Replacing parts with the ones not specified by Atom may compromise the safety of the
unit.

9
3. Introduction
This Service Manual describes procedures for inspecting, repairing, and changing the settings of
the Atom Incubator V-2200. It also describes precautions and troubleshooting methods during
alarm conditions. The procedures described in this Service Manual should be performed only by
service personnel trained in the operation and repair of medical electronic equipment.
Neither Atom Medical Corporation nor its local representative will be responsible for the quality or
the performance of the unit if it should be handled improperly contrary to Atom’s instructions or if
unauthorized parts should be used in repairing the unit. It should be noted that any responsibility
arising from inspecting, repairing, or changing the settings of the unit lies with the one who carried
it out.
Read this Service manual carefully and familiarize yourself thoroughly with its contents before
inspecting, repairing, or changing the settings of the unit. Keep this Manual where it is readily
accessible for reference in case of need.
For more detailed information on inspecting, repairing and changing the settings of the unit,
contact your local Atom representative.
Be sure to read the Operation Manual as well when using this Service Manual.

10
4. Parts Identification
4-1. Main Body
Note : ⑩Drawer (one-layer type and two-layer type) is optionally available.
Unlike the conventional incubator (V-2100G), the sensor module (⑬) does not include an alarm
lamp that comes on during an alarm condition. An alarm lamp is included in the display/control
panel(⑤).
No. Name No. Name
①Front admittance panel operating
knob
⑪Elevating pedal
⑫Caster
②Front admittance panel ⑬Sensor module
③Snap-open access port ⑭Iris access port
④Humidity chamber cover ⑮Tube introduction slit assembly
⑤Display/Control panel ⑯Relay box
⑥Pole arm ⑰MF rail
⑦Front admittance panel operating
lever
⑱Outer filter cover
⑲Hose cover
⑧Mattress
⑨Mattress platform tilting knob
⑩Drawer

11
No. Name No. Name
○
20 Filter cover ○
30 I/O port
○
21 Oxygen supply port 1 ○
31 Test switch
○
22
Oxygen supply port 2 (to use the internal
oxygen controller) ○
32 Mattress platform
○
33 Weight monitor assembly
○
23 Power switch ○
34 X-ray cassette tray
○
24 Weight monitor connector ○
35 Top board packing
○
25
Display output connector (for right side)
○
36 Top board
○
26 Hood stopper ○
37 Fan cover
○
27 Display output connector ○
38 Fan
○
28 Connector J2 ○
39 Conditioning chamber
○
29 Connector J1 ○
40 Nozzle

12
4-2. Display/Control Panel

13
Note : ⑬and ⑭are available only on the unit with
oxygen controller.
⑮is available only on the unit with built-in pulse
oximeter.
⑯, ⑰and ⑱are available only on the unit with
weight monitor.
Name
⑫Relative Humidity Setting Switch
Press this switch to set a relative humidity.
When this switch is pressed, the selection of
humidity message will be displayed. and a
relative humidity can be set.
Select a desired setting by pressing an
appropriate setting switch.
⑬ON/OFF Switch
Press this switch to turn the oxygen
controller on/off.
⑭O2Switch
Press this switch to set an oxygen
concentration. When this switch is pressed,
the selection of oxygen concentration
message will be displayed, and an oxygen
concentration can be set.
Select a desired setting by pressing an
appropriate setting switch.
⑮Alarm Limit Selector Switch
Press this switch to set an SpO2/pulse rate
alarm limit.
Every time this switch is pressed, selection
of a SpO2alarm message and the selection
of pulse rate alarm massage will be
displayed by turns, and an alarm limit for
the mode currently being displayed can be
set.
Select a desired setting by pressing an
appropriate setting switch.
⑯Weight Switch
Press this switch to weigh or store the
measurement value.
⑰Tare Switch
Press this switch to tare.
⑱Mode Switch
Press this switch to change the weight
measurement mode.
⑲Display
Displays an air temperature, a relative humidity
and an oxygen concentration in the incubator,
skin temperature, %SpO
2
, pulse rate and weight
of the infant etc. on the Numerical value screen,
Trend 1 screen and Trend 2 screen.
Name
① Alarm Indicator
Flashes or illuminates when an alarm
condition occurs.
② Power FailureAlarm Indicator
Illuminates when power supply is
interrupted due to power breakdown or a
disconnected power plug etc.
③ System Alarm Indicator
Illuminates when any system error is
detected by the self-diagnosis.
④ Screen Changeover Switch
Every time this switch is pressed, the
screen is changed.
⑤ Brightness Adjusting Switch
Press this switch to adjust the brightness
of the screen.
⑥ °C/°F Selector Switch
Press this switch to change the temperature
display unit from °C to °F and vice versa.
⑦ Setting Switch
Press this switch to set temperature,
relative humidity, SpO2/pulse rate alarm
limits, etc. to a desired level.
Every time switch is pressed, a
setting increases.
Every time switch is pressed, a
setting decreases.
⑧ Alarm Silence/ reset Switch
Press this switch to silence an audible
alarm or reset an alarm condition.
⑨ Manual Control Switch
Press this switch to operate the incubator
in the Manual Control mode.
⑩ Servo Control Switch
Press this switch to operate the incubator
in the Servo Control mode.
⑪ Override Switch
Press this switch while the Servo
indicator is flashing to select the override
mode; the “>37.5°C”indicator will flash,
and the high skin temperature range of
37.6-39.0°C will be available.
Press this switch while the Manual
indicator is flashing to select the override
mode; the “>37.0°C” indicator will flash,
and the high incubator air temperature
range of 37.1-39.0°C will be available.

14
4-3. Screen Display
Turn on the unit, and the start screen will appear on the display. It is then replaced by the
numerical value screen. Three different screens are available by pressing the (screen
changeover) switch: the numerical value screen, the trend 1 screen and the trend 2 screen.

15
4-3-1. Numerical Value Screen
The numerical value screen is the screen that is displayed unless otherwise selected. It consists of
three areas: the numerical value area, the message area and the time area.
●Numerical value area
This area displays all the numerical data including the selected temperature, the incubator air
temperature, the skin temperatures, the relative humidity, the oxygen concentration, %SpO2, the
pulse rate and the settings.

16
●Message area
(1) This area displays information on the weight monitor
most of the time.
(2) If a switch is operated, or an alarm should occur, a
corresponding message will be displayed.
*When the measurement section of the weight
monitor is not connected, “OFF” is normally
displayed. Messages, etc. will appear when an
operation or an alarm occurs.
●Time area
This area displays the current time and the alarm
silence indicator. Date is displayed in the
month/day/year format above. Time is displayed in
the hour/minute format below.
The alarm silence indicator appears when the
audible alarm is silenced.

17
4-3-2. Trend 1 Screen
The trend 1 screen displays the temperatures, the relative humidity and the oxygen concentration
graphically. It consists of four areas: the numerical value area, the temperature graph area, the
relative humidity & oxygen graph area, and the time area.
*Settings are not displayed on this screen.
●Numerical value area
This area displays the incubator air temperature,
the skin temperatures, the relative humidity, the
oxygen concentration, %SpO2and the pulse rate.
*Settings are not displayed in this area.
●Temperature graph area
This area displays the incubator air temperature,
the skin temperature 1, and the skin temperature 2
graphically. However, the skin temperature 1 and
the skin temperature 2 will not be displayed if the
respective skin temperature probes are not
connected. (See “5-5. Graphical Display of the
Incubator Air Temperature and the Skin
Temperature”.)
●Relative humidity & oxygen graph area
This area displays the relative humidity and the
oxygen concentration graphically. However, the
oxygen concentration will not be displayed if the
oxygen supply function is off. (See “6-2.Graphical
Display of the Relative Humidity and/or 8-3.Graphical
Display of Oxygen”.)
●Time area
This area displays the month, the day, the year, the
hour and the minute in this order.

18
4-3-3. Trend 2 Screen
The trend 2 screen displays %SpO2, the pulse rate and the weight graphically. It consists of four areas:
the numerical value area, %SpO2and pulse rate graph area, the weight graph area, and the time area.
*Settings are not displayed on this screen.
●Numerical value area
Like the trend 1 screen, this area displays the
incubator air temperature, the skin temperatures,
the relative humidity, the oxygen
concentration, %SpO2and the pulse rate.
*Settings are not displayed in this area.
●%SpO2& pulse rate area
This area displays the %SpO2and the pulse rate
graphically. The vertical axis on the left
shows %SpO2, and the vertical axis on the right shows
the pulse rate per minute. However, %SpO2and the
pulse rate will not be displayed if the probe is not
connected. However, no graph will be displayed if the
probe is not connected, or the main body is not
provided with a pulse oximeter. (See “7-3. Graphical
Display of the Pulse Oximeter”.)
●Weight graph area
This area displays weight readings and a weight graph.
The weight graph is shown in the left area, while the
current weighing mode and the latest reading are
shown in the right area. Recorded readings are
graphically displayed. The scale of the graph changes
depending on the weighing mode and the actual
readings. However, no graph will be displayed if the
measurement section of the weight monitor is not
connected, or the unit is not provided with a weight
monitor. (See “9-6-2. Graphical Display of Weight”.)
The data recorded for the weight graph remains stored even when power is turned off. Be
very careful when you are going to weigh another patient. When weight readings are
recorded, the graph will show the recorded readings. When no readings are recorded, the
graph will remain blank.
Table of contents
Other Atom Medical Corporation Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Otto Bock
Otto Bock 21A47 Instructions for use

Boston Scientific
Boston Scientific U125 Technical manual

INFIMED
INFIMED FENIX OT-01 user manual

Pari
Pari Easycare manual

Hydrogen For Health
Hydrogen For Health PEM+ instruction manual

Top shelf Orthopedics
Top shelf Orthopedics RecoveryWrap System Instructions for use