manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Avanti
  6. •
  7. Test Equipment
  8. •
  9. Avanti AVA-03 User manual

Avanti AVA-03 User manual

DIGITAL TRANSDUCER TESTER & PC-INTERFACE (AVA-03)
14-FEB-2022
PAGE 1 OF 12
© Avanti Elektronik Limited - (44) 01224-583132 - www.avanti.clara.net - [email protected]. k dqdtst_32.docx
1. GENERAL
The Avanti I²C Transducer Tester and PC-Interface (AVA-03) is a standalone device for
accessing ressure/tem erature transducers, which are equi ed with a two-wire serial I²C
interface. It was designed to be com atible with Quartzdyne's range of digital quartz
ressure transducers, including their V4.02 enhancements (checksum etc). U to 4
transducers can be connected simultaneously and their readings are dis layed on a backlit
LCD. There is also an RS232 ort rovided, which allows to access readings and
calibration coefficients from a PC.
2. HARDWARE DESCRIPTION
The AVA-03 I²C Transducer Tester is housed in an aluminium enclosure measuring 168 x
103 x 56mm.
Transducer Socket Pinout
Pin-1 A1
Pin-2 SCL
Pin-3 GND
Pin-4 SDA
Pin-5 +5V
Pin-6 A2
Pin-7/8/9 NC
RS-232 Socket Pinout
Pin-2 RX (of PC)
Pin-3 TX (of PC)
Pin-5 GND
Pin-
1/4/6/7/8/9
NC
There are four 9-way D-ty e sockets (A...D) for connecting u to four I²C transducers. The
connectors carry +5V, GND, SCL, SDA and two address lines A2 and A1. The address
lines are either o en (1) or grounded (0) and are coded differently at each socket: The first
socket sets A2=0/A1=0, the second socket sets A2=0/A1=1, the third socket sets
A2=1/A1=0 and the fourth socket sets A2=1/A1=1. This means that different I²C addresses
are allocated automatically for each socket, rovided the address ins are wired through to
the transducers, see the exam les shown in fig. 1 and fig. 2 below.
Figure 1: 6-wire connection, I²C address coded by the AVA-03 socket
DIGITAL TRANSDUCER TESTER & PC-INTERFACE (AVA-03)
14-FEB-2022
PAGE 2 OF 12
© Avanti Elektronik Limited - (44) 01224-583132 - www.avanti.clara.net - [email protected]. k dqdtst_32.docx
Figure 2: 4-wire connection, I²C address coded at the transducer
Important: Some versions of Quartzdyne transducers are terminated with a D-
type plug but with a different pin-out. Never plug directly into an V -03 socket!
Use a cross-over cable and double check for correct wiring.
vanti Elektronik Ltd assumes no responsibility and shall not be liable for any
damage to the transducer that might occur due to incorrect wiring.
The AVA-03 I²C Transducer Tester contains a 2 x 20 character backlit LCD for dis laying
readings and status information. A ush-button is rovided to ste through different
dis lay o tions. A further 9-way D-Ty e socket is located at the side for connecting to the
RS232 ort of a PC.
The I²C Transducer Tester runs on either the built-in 9V battery or on an external 12V
su ly. Note that the external su ly voltage must be higher than the battery voltage in
order to revent the battery from inadvertently sourcing current. Current consum tion is
a roximately 5mA when the backlight is off and a roximately 125mA when the
backlight is on.
3. LCD DISPLAY
After ower on, the AVA-03 I²C Transducer Tester dis lays its version number and
rom ts the user to ress the ush-button to ste through the available dis lay o tions.
After ressing the button, the tester dis lays the I²C address of the first transducer (the
transducer with the lowest address) together with its serial number and calibration date. A
single character (A...D) is shown as an identifier in the to right corner, according to the
current I²C address:
A for A2=0/A1=0, B for A2=0/A1=1, C for A2=1/A1=0 and D for A2=1/A1=1.
This identifier remains visible while ste ing through the following dis lays, making it
easy to kee track of multi le transducers.
After ressing the button, raw ressure and tem erature counts are dis layed as returned by
the transducer. The counts are shown as 32bit hexadecimal values. The transducers are
olled at a rate of once every 1.5 seconds. Note that the identifier letter in the to right
corner briefly blinks whenever the dis lay is u dated.
DIGITAL TRANSDUCER TESTER & PC-INTERFACE (AVA-03)
14-FEB-2022
PAGE 3 OF 12
© Avanti Elektronik Limited - (44) 01224-583132 - www.avanti.clara.net - [email protected]. k dqdtst_32.docx
Pressing the button once more advances to the frequency dis lay. Pressure and tem erature
frequencies are dis layed in Hz, under the assum tion that the transducer's internal
reference frequency is exactly 7.200MHz.
The next dis lay shows ressure in si (or bar) and tem erature in °C (or °F) using the
calibration coefficients held in the transducer's EEPROM memory.
Pressing the button again advances to the first dis lay of the next transducer, i.e. the
transducer with the next higher I²C address, showing again serial number and calibration
date. Re eatedly ressing the button allows ste ing through all available transducers in
sequence:
LCD Dis lay: Descri tion:
PRESS BUTTON TO STEP
THROUGH OPTIONS
1 Start-up Screen
A2A1 = 00 = XDUCER A
SN 000123 13-02-2022
2 Status Screen of the first transducer (lowest I²C address)
RAW-P = 0x00B60B61 A
RAW-T = 0x01C71C72
3 Raw Values
PF = 20000.000 Hz A
TF = 50000.000 Hz
4 Equivalent Frequencies
P = 11111.222 si A
T = 111.222 degC
5 Calculated P/T Values
A2A1 = 01 = XDUCER B
SN 000456 14-02-2022
6 Status Screen of next transducer, if connected Otherwise back to
Status Screen of first transducer
4. BACKLIGHT
The I²C Transducer Tester contains a backlit LCD. The backlight is ermanently on when
running on an external 12V su ly. However, when running on the internal battery, the
backlight is switched off after 30 seconds with no key ress, in order to reduce the current
consum tion. The backlight is switched on again by sim ly ressing the button. In this
case the dis lay o tion remains unchanged and a second key ress is required to advance
to the next o tion.
5. CONFIGURATION
The AVA-03 I²C Transducer Tester can be configured for different dis lay o tions. The
configuration is held in non-volatile memory. To change the configuration, the tester must
be switched off and then switched on again, while the ush-button is being ressed down.
This starts the tester in configuration mode.
There are 12 different o tions available, which can be selected in turn by re eatedly
ressing the ush-button:
DIGITAL TRANSDUCER TESTER & PC-INTERFACE (AVA-03)
14-FEB-2022
PAGE 4 OF 12
© Avanti Elektronik Limited - (44) 01224-583132 - www.avanti.clara.net - [email protected]. k dqdtst_32.docx
1: BAR-ABS/DEGC Absolute pressure is displayed in bar and temperature in °C
2: BAR-GAUGE/DEGC Gauge pressure (see paragraph 6 1) is displayed in bar and temperature in °C
3: PSI-ABS/DEGC Absolute pressure is displayed in psi and temperature in °C
This is the default setup
4: PSI-ABS/DEGF Absolute pressure is displayed in psi and temperature in °F
5: PSI-GAUGE/DEGC Gauge pressure (see paragraph 6 1) is displayed in psi and temperature in °C
6: PSI-GAUGE/DEGF Gauge pressure (see paragraph 6 1) is displayed in psi and temperature in °F
7: BAR-ABS/DEGF Absolute pressure is displayed in bar and temperature in °F
8: BAR-GAUGE/DEGF Gauge pressure (see paragraph 6 1) is displayed in bar and temperature in °F
9: PSI/DEGC/DUAL-P Absolute pressure is displayed in psi and temperature in °C Initial screen
shows pressure values of 2 transducers together (see paragraph 6 2)
10: PSI/DEGF/DUAL-P Absolute pressure is displayed in psi and temperature in °F Initial screen
shows pressure values of 2 transducers together (see paragraph 6 2)
11: BAR/DEGC/DUAL-P Absolute pressure is displayed in bar and temperature in °C Initial screen
shows pressure values of 2 transducers together (see paragraph 6 2)
12: BAR/DEGF/DUAL-P Absolute pressure is displayed in bar and temperature in °F Initial screen
shows pressure values of 2 transducers together (see paragraph 6 2)
Once the desired configuration is shown on the screen, the tester is sim ly switched off.
The tester remembers and a lies this configuration when owered u again in the normal
way.
6.1 ABSOLUTE / GAUGE PRESSURE
The AVA-03 I²C Transducer Tester can be configured to dis lay absolute or gauge
ressure (see aragra h 5 above). When gauge ressure is selected, the sequence of screens
is as follows:
LCD Dis lay: Descri tion:
PRESS BUTTON TO STEP
THROUGH OPTIONS
1 Start-up Screen
A2A1 = 00 = XDUCER A
SN 000123 04-12-2007
2 Status Screen
RAW-P = 0x00B60B61 A
RAW-T = 0x01C71C72
3 Raw Values
PF = 20000.000 Hz A
TF = 50000.000 Hz
4 Equivalent Frequencies
ATMOSPHERIC PRES: A
P = 14.500 si
5 Tare Screen: When the button is pressed again, the last pressure
value displayed is captured and applied as an atmospheric offset
until the unit is powered down
P = 0.000 sig A
T = 25.000 degC
6 Calculated pressure / temperature with atmospheric offset of
-14 5 psi applied
The tare screen is shown once only after ower-u . The atmos heric offset remains valid
until the tester is owered down. To change the atmos heric offset, the tester must be
switched off and on again. In case of multi le transducers, the transducer with the lowest
I²C address is used to determine the atmos heric offset. The same offset is then a lied to
all transducers.
DIGITAL TRANSDUCER TESTER & PC-INTERFACE (AVA-03)
14-FEB-2022
PAGE 5 OF 12
© Avanti Elektronik Limited - (44) 01224-583132 - www.avanti.clara.net - [email protected]. k dqdtst_32.docx
6.2 DUAL PRESSURE DISPLAY
The AVA-03 I²C Transducer Tester can be configured to dis lay two ressures together on
the same screen (see aragra h 5 above). When dual- ressure is selected, the sequence of
screens is as follows:
LCD Dis lay: Descri tion:
P1 = 11111.222 si A
P2 = 12222.333 si B
1 First screen after power up: Calculated pressures P1 and P2 of two
transducers (shown here with addresses A2/A1 = 00 = A and A2/A1
= 01 = B)
A2A1 = 00 = XDUCER A
SN 000123 13-02-2022
2 Status Screen of the first transducer
RAW-P = 0x00B60B61 A
RAW-T = 0x01C71C72
3 Raw P/T values
PF = 20000.000 Hz A
TF = 50000.000 Hz
4 Equivalent Frequencies
P = 11111.222 si A
T = 111.222 degC
5 Calculated P/T Values
... B
... B
... B
... B
6 as above for the second transducer
7
8
9
P1 = 11111.222 si A
P2 = 12222.333 si B
10 Dual pressure screen again
P1 is allocated to the transducer with the lowest I²C address and P2 to the transducer with
the second lowest I²C address. If more than two transducers are connected, the ones with
higher I²C addresses are ignored and not dis layed on the LCD. However their readings
remain available through the RS232 interface (see aragra h 7 below).
7. RS232 INTERFACE
The Avanti I²C Transducer Tester & PC-Interface contains an RS232 ort for direct
connection to a Personal Com uter via a 9-way D-ty e extension lead, where ins 2,3 and
5 are wired one-to-one. The tester communicates at 19200 baud with 1 start-bit, 8 data-bits
and 1-sto bit.
All commands and res onses are ASCII, so that sim le terminal emulation software (e.g.
Hy erterminal) can be used. All valid command characters are echoed. Invalid command
characters roduce a 0x07 (bee ) instead and the command is aborted. Note that the
command characters are case sensitive.
In order to revent any command characters from being lost due to inherent delays (for
instance when writing to the transducer's EEPROM), it is good ractice to send command
DIGITAL TRANSDUCER TESTER & PC-INTERFACE (AVA-03)
14-FEB-2022
PAGE 6 OF 12
© Avanti Elektronik Limited - (44) 01224-583132 - www.avanti.clara.net - [email protected]. k dqdtst_32.docx
characters one by one and to wait for the echo before sending the next character. This of
course is irrelevant when manually ty ing commands using terminal emulation software.
The following commands are im lemented:
HELP AND GENERAL INFO
PC SENDS TESTER
RESPONDS
DESCRIPTION BYTES
EXA PLE
?? Hel request 2 ??
?? Echo 2 ??
<CR> (0x0D) Carriage return 1 <CR>
<CR> (0x0D) Echo 1 <CR>
<LF> (0x0A) Line feed 1 <LF>
This command causes the I²C Transducer Tester to out ut a list of all im lemented RS232
commands, lus the FPGA identifier and status of all connected transducers.
GET A RAW PRESSURE READING (COUNTS)
PC SENDS TESTER
RESPONDS
DESCRIPTION BYTES
EXA PLE
P Pressure request 1 P
P Echo 1 P
A...D Transducer address 1 A
A...D Echo 1 A
<CR> (0x0D) Carriage return 1 <CR>
< > (0x20) S ace 1 < >
00000000...
FFFFFFFF
32bit ressure count as 8
digit hex value
8 016C16C1
<CR> (0x0D) Carriage return 1 <CR>
<LF> (0x0A) Line Feed 1 <LF>
GET A RAW TE PERATURE READING (COUNTS)
PC SENDS TESTER
RESPONDS
DESCRIPTION BYTES
EXA PLE
T Tem erature request 1 T
T Echo 1 T
A...D Transducer address 1 A
A...D Echo 1 A
<CR> (0x0D) Carriage return 1 <CR>
< > (0x20) S ace 1 < >
00000000...
FFFFFFFF
32bit tem erature count as
8 digit hex value
8 01C71C72
<CR> (0x0D) Carriage return 1 <CR>
<LF> (0x0A) Line Feed 1 <LF>
DIGITAL TRANSDUCER TESTER & PC-INTERFACE (AVA-03)
14-FEB-2022
PAGE 7 OF 12
© Avanti Elektronik Limited - (44) 01224-583132 - www.avanti.clara.net - [email protected]. k dqdtst_32.docx
GET A CALCULATED PRESSURE READING (PSI OR BAR)
PC SENDS TESTER
RESPONDS
DESCRIPTION BYTES
EXA PLE
Pressure request (lower
case )
1
Echo 1
A...D Transducer address 1 A
A...D Echo 1 A
<CR> (0x0D) Carriage return 1 <CR>
< > (0x20) S ace 1 < >
0.000 ...
99999.999
Pressure in si or bar as a
5.3 floating oint number
9 1234.567
<CR> (0x0D) Carriage return 1 <CR>
<LF> (0x0A) Line Feed 1 <LF>
GET A CALCULATED TE PERATURE READING (°C OR °F)
PC SENDS TESTER
RESPONDS
DESCRIPTION BYTES
EXA PLE
t Tem erature request
(lower case t)
1 t
t Echo 1 t
A...D Transducer address 1 A
A...D Echo 1 A
<CR> (0x0D) Carriage return 1 <CR>
< > (0x20) S ace 1 < >
0.000 ...
99999.999
Tem erature in °C or °F as
a 5.3 floating oint number
9 123.456
<CR> (0x0D) Carriage return 1 <CR>
<LF> (0x0A) Line Feed 1 <LF>
DIGITAL TRANSDUCER TESTER & PC-INTERFACE (AVA-03)
14-FEB-2022
PAGE 8 OF 12
© Avanti Elektronik Limited - (44) 01224-583132 - www.avanti.clara.net - [email protected]. k dqdtst_32.docx
READ FRO EEPRO
PC SENDS TESTER
RESPONDS
DESCRIPTION BYTES
EXA PLE
R Read EEPROM request 1 R
R Echo 1 R
A...D Transducer address 1 A
A...D Echo 1 A
0000...1FFF Start address 4 0100
0000...1FFF Echo 4 0100
00...FF Number of bytes to read
(00 = 256)
2 0A
00...FF Echo 2 0A
<CR> (0x0D) Carriage return 1 <CR>
< > (0x20) S ace 1 < >
00...FF First data byte 2 8A
... ... ... ...
00...FF Last data byte 2 E7
< > (0x20) S ace 1 < >
OK or NO Result of o eration 2 OK
< > (0x20) S ace 1 < >
0000...FFFF Checksum over all data
bytes
4 0171
<CR> (0x0D) Carriage return 1 <CR>
<LF> (0x0A) Line feed 1 <LF>
DIGITAL TRANSDUCER TESTER & PC-INTERFACE (AVA-03)
14-FEB-2022
PAGE OF 12
© Avanti Elektronik Limited - (44) 01224-583132 - www.avanti.clara.net - [email protected]. k dqdtst_32.docx
WRITE TO EEPRO
PC SENDS TESTER
RESPONDS
DESCRIPTION BYTES
EXA PLE
W Write to EEPROM request 1 W
W Echo 1 W
A...D Transducer address 1 A
A...D Echo 1 A
0000...1FFF Start address 4 0100
0000...1FFF Echo 4 0100
< > (0x20) S ace 1 < >
00...FF First data byte to write 2 8A
00...FF Echo 2 8A
... ... ... ... ... ...
00...FF Last data byte to write 2 E7
00...FF Echo 2 E7
<CR> (0x0D) Carriage return 1 <CR>
< > (0x20) S ace 1 < >
OK or NO Result of o eration 2 OK
< > (0x20) S ace 1 < >
0000...FFFF Checksum over all data
bytes
4 0171
<CR> (0x0D) Carriage return 1 <CR>
<LF> (0x0A) Line feed 1 <LF>
SET SERIAL NU BER AND CALIBRATION DATE
PC SENDS TESTER
RESPONDS
DESCRIPTION BYTES
EXA PLE
SN Set serial number request 2 SN
SN Echo 2 SN
<CR> (0x0D) Carriage return 1 <CR>
<CR> (0x0D) Echo 1 <CR>
<LF> (0x0A) Line feed 1 <LF>
This command rom ts the user to enter a new serial number and calibration date. The
user can choose to kee the contents of all other EEPROM locations or to fill them with
zeroes. Once acknowledged, the firmware writes to EEPROM and u dates the checksum.
The firmware writes the serial number and calibration date into 4 redundant locations, as
introduced by Quartzdyne in Jan 2010.
DIGITAL TRANSDUCER TESTER & PC-INTERFACE (AVA-03)
14-FEB-2022
PAGE 10 OF 12
© Avanti Elektronik Limited - (44) 01224-583132 - www.avanti.clara.net - [email protected]. k dqdtst_32.docx
COPY EEPRO CONTENTS
PC SENDS TESTER
RESPONDS
DESCRIPTION BYTES
EXA PLE
EC EEPROM co y request 2 EC
EC Echo 2 EC
<CR> (0x0D) Carriage return 1 <CR>
<CR> (0x0D) Echo 1 <CR>
<LF> (0x0A) Line feed 1 <LF>
This command reads the EEPROM contents of one transducer and writes it to a second
transducer. The user is rom ted to choose the source and the destination transducer. Once
acknowledged, the firmware co ies EEPROM locations from 0x0000 to 0x00FF into 4
redundant blocks (0x0000-0x00FF / 0x0100-0x1FF / 0x0200-0x2FF / 0x0300-0x3FF), as
introduced by Quartzdyne in Jan 2010.
START CONTINUOUS OUTPUT ODE
PC SENDS TESTER
RESPONDS
DESCRIPTION BYTES
EXA PLE
CM Continuous Mode request 2 CM
CM Echo 2 CM
<CR> (0x0D) Carriage return 1 <CR>
<CR> (0x0D) Echo 1 <CR>
<LF> (0x0A) Line feed 1 <LF>
This command rom ts the user to enter the required out ut rate in seconds, which
transducer addresses to consider (ABCD) and the data ty e required (raw, calculated or
both). The I²C Transducer Tester then starts sending data records together with ela sed
time at the rate s ecified, which can be logged to file using Hy erterminal or similar PC
software. Once started, continuous mode can only be sto ed by switching the I²C
Transducer Tester off and on again. While continuous mode is active, the rate at which
transducers are olled (and the LCD is u dated) is reduced from once er 1.5sec to once
er 2.0sec. There will also be a slight delay noticeable when ste ing through dis lay
o tions.
DIGITAL TRANSDUCER TESTER & PC-INTERFACE (AVA-03)
14-FEB-2022
PAGE 11 OF 12
© Avanti Elektronik Limited - (44) 01224-583132 - www.avanti.clara.net - [email protected]. k dqdtst_32.docx
8. THIRD PARTY LOGGING/CHARTING SOFTWARE
As described above, all RS232 commands and res onses contain only ASCII characters,
allowing full access with standard terminal software like Hy erterminal. For basic data
logging the AVA-03 can be ut into continuous out ut mode with the "CM" command.
If more com lex data logging or dis lay/charting functions are required, there is a low-cost
3rd- arty software available that runs on Windows PCs (see www.windmill.co.uk).
Windmill-7 data logging software includes a user configurable arser that can be set u to
suit the AVA-03 command set. Setu can be a bit tricky, but once configured the software
is easy to use. Please note however, that this is 3rd- arty software and beyond control of
Avanti Elektronik Ltd. No guarantee can be given regarding its suitability and no su ort
will be rovided.
Below are some screenshots obtained with Windmill-7 software.
Figure 3: Windmill Data Display Panels showing 2 Transducers (A and B)
Figure 4: Windmill Charting Module displaying 2 Pressures (A and B)
DIGITAL TRANSDUCER TESTER & PC-INTERFACE (AVA-03)
14-FEB-2022
PAGE 12 OF 12
© Avanti Elektronik Limited - (44) 01224-583132 - www.avanti.clara.net - [email protected]. k dqdtst_32.docx
Figure 5: Windmill Data Logging Module writing raw and engineering values to file
9. ACKNOWLEDGE ENTS AND FURTHER INFOR ATION
Quartzdyne is a trademark of Quartzdyne Inc.
Information on their range of digital ressure transducers can be downloaded from their
web-site at htt ://www.quartzdyne.com
The following documents are of articular interest:
DigitalTransS ec. df - Digital Quartz Pressure Transducer S ecifications
DigitalTransProg. df - Digital Transducer Programming Manual
I²C is a trademark of NXP Semiconductors (formerly Phili s).
S ecifications and a lication notes can be downloaded from their web-site at
htt ://www.nx .com
For I²C bus s ecifications refer to the following document:
i2c.bus.s ecification. df (I²C S ecification and User Manual Rev.03 / 2007)
Windmill-7 is co yright of Windmill Software Ltd (www.windmill.co.uk)
odel No. AVA-03
Serial Number
Firmware Revision
Date Tested
Signature

Popular Test Equipment manuals by other brands

multicomp pro MP Dual Channel Series user manual

multicomp pro

multicomp pro MP Dual Channel Series user manual

PDK PDPG-H user manual

PDK

PDK PDPG-H user manual

Bacharach Informant 2 Specifications

Bacharach

Bacharach Informant 2 Specifications

Elma Sagab Volt Stick Wide manual

Elma

Elma Sagab Volt Stick Wide manual

Tektronix TBS1000C Series Service manual

Tektronix

Tektronix TBS1000C Series Service manual

Testifire XTR2 quick start guide

Testifire

Testifire XTR2 quick start guide

Keysight Technologies 16380V Operation and service manual

Keysight Technologies

Keysight Technologies 16380V Operation and service manual

Forch 5439 172 manual

Forch

Forch 5439 172 manual

Cascade Microtech IZI Probe Quick reference guide

Cascade Microtech

Cascade Microtech IZI Probe Quick reference guide

Hillstone HAC415-120PX operating manual

Hillstone

Hillstone HAC415-120PX operating manual

Telwin DTS700 manual

Telwin

Telwin DTS700 manual

AEMC MN113 user manual

AEMC

AEMC MN113 user manual

EKF Diagnostics Hemo Control user manual

EKF Diagnostics

EKF Diagnostics Hemo Control user manual

Dwyer Instruments TDS Series Specifications - Operating Instructions

Dwyer Instruments

Dwyer Instruments TDS Series Specifications - Operating Instructions

Rohde & Schwarz R&S CMW-KM300 user manual

Rohde & Schwarz

Rohde & Schwarz R&S CMW-KM300 user manual

Huazheng HZJZ-200V manual

Huazheng

Huazheng HZJZ-200V manual

Benning DUSPOL digital plus user manual

Benning

Benning DUSPOL digital plus user manual

Wuhan Huatian Electric Power Automation HTDT-10A manual

Wuhan Huatian Electric Power Automation

Wuhan Huatian Electric Power Automation HTDT-10A manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.