manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Axio MET
  6. •
  7. Multimeter
  8. •
  9. Axio MET AX-102 User manual

Axio MET AX-102 User manual

Other manuals for AX-102

1

Other Axio MET Multimeter manuals

Axio MET AX-573 User manual

Axio MET

Axio MET AX-573 User manual

Axio MET ax-585b User manual

Axio MET

Axio MET ax-585b User manual

Axio MET AX-190A User manual

Axio MET

Axio MET AX-190A User manual

Axio MET AX-18B User manual

Axio MET

Axio MET AX-18B User manual

Axio MET AX-160IP User manual

Axio MET

Axio MET AX-160IP User manual

Axio MET AX-582B User manual

Axio MET

Axio MET AX-582B User manual

Axio MET AX-588B User manual

Axio MET

Axio MET AX-588B User manual

Axio MET AX-594 User manual

Axio MET

Axio MET AX-594 User manual

Axio MET AX-MS811 User manual

Axio MET

Axio MET AX-MS811 User manual

Axio MET AX-201 User manual

Axio MET

Axio MET AX-201 User manual

Axio MET AX-101 User manual

Axio MET

Axio MET AX-101 User manual

Axio MET AX-174 User manual

Axio MET

Axio MET AX-174 User manual

Axio MET AX-TI220 User manual

Axio MET

Axio MET AX-TI220 User manual

Axio MET AX-MS8250 User manual

Axio MET

Axio MET AX-MS8250 User manual

Axio MET AX-588 User manual

Axio MET

Axio MET AX-588 User manual

Axio MET AX-176 User manual

Axio MET

Axio MET AX-176 User manual

Axio MET AX-T903 User manual

Axio MET

Axio MET AX-T903 User manual

Axio MET AX-582 User manual

Axio MET

Axio MET AX-582 User manual

Axio MET AX-101B User manual

Axio MET

Axio MET AX-101B User manual

Axio MET AX-588B User manual

Axio MET

Axio MET AX-588B User manual

Axio MET AX-105 User manual

Axio MET

Axio MET AX-105 User manual

Axio MET AX-T520 User manual

Axio MET

Axio MET AX-T520 User manual

Axio MET AX-102 User manual

Axio MET

Axio MET AX-102 User manual

Axio MET AX-C708 User manual

Axio MET

Axio MET AX-C708 User manual

Popular Multimeter manuals by other brands

Fluke 106 user manual

Fluke

Fluke 106 user manual

VAR DV-71600 manual

VAR

VAR DV-71600 manual

Agilent Technologies 34420A user guide

Agilent Technologies

Agilent Technologies 34420A user guide

Fluke 83 user manual

Fluke

Fluke 83 user manual

Hameg HM 8012 manual

Hameg

Hameg HM 8012 manual

Mastech MY70 Operation manual

Mastech

Mastech MY70 Operation manual

Steren MUL-300 user manual

Steren

Steren MUL-300 user manual

Somogyi Elektronic home SMA 92 instruction manual

Somogyi Elektronic

Somogyi Elektronic home SMA 92 instruction manual

Gossen MetraWatt METRAmax 6 operating instructions

Gossen MetraWatt

Gossen MetraWatt METRAmax 6 operating instructions

PeakTech 4000 Procedure of calibration

PeakTech

PeakTech 4000 Procedure of calibration

YOKOGAWA 90050B user manual

YOKOGAWA

YOKOGAWA 90050B user manual

Fluke 8846A Programmer's manual

Fluke

Fluke 8846A Programmer's manual

Tempo Communications MM200 instruction manual

Tempo Communications

Tempo Communications MM200 instruction manual

APPA 607 user manual

APPA

APPA 607 user manual

Elenco Electronics CM-1555 operating instructions

Elenco Electronics

Elenco Electronics CM-1555 operating instructions

Mastech MS8238C user manual

Mastech

Mastech MS8238C user manual

SDAR 830A, 830B, 830D, 830C user manual

SDAR

SDAR 830A, 830B, 830D, 830C user manual

RS PRO RS-9519BT instruction manual

RS PRO

RS PRO RS-9519BT instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

DIGITAL MULTIMETER
AX-102
OPERATION MANUAL
2
I.General
His is equipped with the LCD display of text height 18.9 mm, is a 3 ¾ digital multimeter which has
the merits of clear Reading, stable performance and high reliability. It could be used to measure DC
voltage, AC voltage, DC current, AC current, resistance , capacity, frequency/duty cycle, diode and
make on and off test. Meanwhile, it is available for unit symbol display, automatic/manual range
switching, automatic power off and alarm function. The complete machine takes an integrated
circuit which can directly drive LCD 8-bit microprocessor and double-integrating A/D switching and
a digital display drive offering high resolution and high precision. Due to its complete functions,
high measurement accuracy and convenient operation, the multimeter is the ideal tool in laboratory
and factory as well as for radio fans and family.
II.Open-package Inspection
Open the package box and take out the meter, check carefully if the following accessories are
absent or damaged. If there is any absence or damage, please contact the distributor immediately.
Digital Multimeter: 1PC
Instruction Manual: 1 copy
Test Leads: 1 pair
Temperature Probe (K-Thermocouple): 1 PC
III. Safety Considerations
The design of meter is in accordance with IEC1010 clause (the safety standard issued by
International Electrotechnical Commission). Prior to the operation of the instrument, please first
read the safety considerations.
1. When DC voltage above 30V, AC voltage above 25V, current above 10mA, AC power line with
inductive load or power line during electric fluctuation is measured, please beware of electric
shock.
2. Prior to measurement, check if the measurement function switch is at the correct position.
Check if the meter pen is contacted reliably, connected correctly, and grounded well and etc. in
order to avoid electric shock.
3. Only if the meter is used with the matched meter pen, can it meet the requirement of safety
standard. When the line of the meter pen is damaged, it is necessary to replace another one of the
same model or the same electrical specification.
4. Don’t use other unconfirmed or disapproved protector tube to replace the protector tube inside
the meter. Only the protector tube of the same model or same specifications can be replaced.
3
Before the replacement, the test leads must leave the measuring point and ensure there is no any
signal at the input terminal.
5. Don’t use other unconfirmed or disapproved battery to replace the battery inside the meter.
Only the battery of the same model or same electrical specification can be replaced. Before the
replacement, the test leads must leave the measuring point and ensure there is no any signal at the
input terminal.
6. When the electrical measurement is made, never let your body gets in touch with the ground
directly, and don’t touch uncovered metal terminal, output port, lead clamp and etc. where earth
potential may exist. Dry clothes, rubber shoes, rubber cushion and other insulating material are
usually used to keep your body insulated against the ground.
7. Don’t store and use it in the high-temperature, high-humidity, inflammable and strong magnetic
field environment.
8. It may do damage to the meter and endanger the operator’s safety if the voltage value beyond
the permitted ultimate voltage value is measured. The ultimate voltage value permitted for
measurement is marked on the instrument panel, and never measure the value exceeding the
standard. Don’t input the ultimate value out of regulation in order to avoid electric shock and the
damage to the meter.
9. When the test leads is inserted into the current socket, don’t measure any voltage for fear that
the meter should be damaged and operator’s safety be endangered.
10. Don’t try calibrating or repairing the meter. When necessary, only the qualified professional
personnel who have had special training or gained approval can make it.
11. During measurement, the requirement of measurement function must be in accordance with
LCD display. Please be sure to disconnect the line of the test leads with the measured object first
and ensure there is no any input signal. It is forbidden to switch the function/range selection switch
during measurement.
12. When “ ” is show on LCD display, please replace battery immediately to ensure the
measurement precision.
13. It is not allowed to insert the test leads into the current terminal to measure voltage!
14. Please don’t change the circuits of the meter freely for fear that the meter be damaged and the
safety be endangered.
15. Descriptions of safety symbols:
4
Warning!
DCA
High Voltage! Danger!
ACA
Ground
DCA & ACA
Double Insulation
In accordance wiyh the instructions
of European Trade Union
Battery Undervoltage
Fuse
IV. Instrument Panel & Button Function Description
1. Instrument Model Number
2. LCD Display: Display the measured data and unit.
3. Function Button
3.1. Hz/Duty (Frequency/Duty Cycle): Press this button to select the
frequency or duty cycle mode. The measurement mode of
voltage/frequency/duty cycle or current/frequency/duty cycle could
be selected by pressing this button in AC/DC voltage or AC/DC
current gear.
3.2. SELECT (Function Switch): Press this button, the function could
switch between DC/AC.
3.3. REL (Relative Value Measurement): The relative value
measurement of all functions could be conducted by pressing this
button except the Hz/Duty function
3.4. HOLD (Data Hold): Press this button, the reading is locked; Press it again, the lock is cleared
and enter normal measurement status.
4. Knob Switch: It could be used to change the measurement function and range.
5. Input Terminals
5.1. Current Voltage, Diode, Resistance, Capacity, Frequency, Buzzer, Temperature “-“ Input
terminal.
5.2. 10A “+”input terminal.
5
5.3. Voltage, Diode, Resistance, Capacity, Frequency, Buzzer, Temperature and “+” Input terminal
with current less than 200mA.
V. Other Functions
Automatic Power off
During measurement, the meter will automatically shut down (enter sleeping mode) to save power
if function buttons and knob switch are not operated in 15 minutes. In auto power off mode, press
any function buttons or rotate the knob switch, the instrument will get into the auto power on
mode (working mode); the auto power offer mode will be cancelled by pressing the HOLD button to
turn on the instrument.
VI. Property
1. General Property
1.1. Display: LCD
1.2. Max. Display: 3999 (3 ¾) counts automatic polarity display and unit display
1.3. Measuring Method: Dual integral A/D converter
1.4. Sampling Rage: Approx. 3 times/sec.
1.5. Over Range Indication: Display “OL”
1.6. Low Battery Indication: “ ” symbol appears.
1.7. Operation Environment: (0~40)°C, Relative Humidity: < 80%
1.8. Storage Environment: (0~50)°C, Relative Humidity: <80%
1.9. Power: 2 pcs 1.5V batteries (“AAA” 7#battery)
1.10. Dimension (size): 145x74x36 mm
1.11. Weight: Approx. 190g (Including 2 pcs
1.5V batteries)
1.12. Accessories: Instruction Manual (1pc),
Outer packing box (1pc),
10A test lead(1pair),
K-Thermocouple
6
2. Technical Properyty:
2.1. Accuracy: ±(% reading + digits), at (23±5)°C,
Relative humidity <75%,
One year calibration guarantee since the
time dispatched from the factory.
2.2. Technical Specifications:
2.2.1. DCV
1. Turn the knob switch to Range
2. The initiate state of the meter is in automatic range status, which shows “AUTO” symbol.
3. Make the test leads connected to the testing point. The voltage and polarity of the point where
the red test is contacted will be displayed on the screen.
Caution:
1.Don’t measure voltage over 600V. Otherwise, ther is a danger of meter being damaged.
2.When measuring high voltage, special attention should be given to personal safety and avoid your
body getting in touch with high voltage circuit.
Range
Accuracy
Resolution
400mV
±(0.5% + 4 d)
100µV
4V
1mV
40V
10mV
400V
100mV
600V
±(1.0% + 4 d)
1V
Input Impedence : 400m>40MΩ; 10MΩat other Ranges.
Overload Protection: 600V DC or 600V AC Peak Value.
2.2.2. ACV
1. Insert black test leads into the hole of “COM” and red one into
2. Rogate the function switch to gear, Press „SELECT” buton to selekt the AC measurement
mode.
3. The initiate state of the meter is in automatic range status, which shows “AUTO” symbol.
7
4. Make the test leads connected to the testing point. The voltage of the point where the red pen is
contacted will be displayed on the screen.
Caution:
1. Don’t measure voltage over 600V. Otherwise, there is a danger of meter being damaged.
2. When measuring high voltage, special attention should be given to personal safety and avoid your
body getting in touch with high voltage circuit.
Range
Accuracy
Resolution
4V
±(0.8% + 6 d)
1mV
40V
10mV
400V
100mV
600V
±(1.0% + 6 d)
1V
Input Impedance: >10MΩ
Overload protection: 600V DC or 600V AC Peak Value
Frequency Response: (50~200) Hz
Display: Average value response (RMS of sine wave)
2.2.3. DCA
1. Insert the black test lead into the “COM” input terminal and red one into the input
terminal. (Max 400mA), or 10A input terminal (Max 10A).
2. Rotate function switch to Current gear. The initiate state of the meter is in automatic range
status, which shows “DC” symbol. Then connect the test leads to the tested circuit in serial,
the tested current value and the current polarity of the point where the red one is contacted will
be displayed on the screen simultaneously.
Caution:
1. If “Ol” is displayed on LCD, it indicates the tested current value has exceeded the present range
limit, please select higher range to complete the measurement.
2. The Max input value is 400mA or 10A. (Depending on the terminal where the red test lead is
contacted).
8
Range
Accuracy
Resolutin
400µA
±(1.0% + 10 d)
0.1µA
4000µA
1µA
40mA
10µA
400mA
100µA
10A
±(1.2% + 10 d)
10mA
Max measurement voltage drop: Full Range mA is 0.4V, A is 100mV;
Max input current : 10A (less than 15 seconds);
Overload Protection: 0.4A/250V restorable fuse, 10A/250V fuse.
2.2.4. ACA
1. Insert the black meter pen into the “COM” input terminal and red one into the input
terminal. (Max 400mA), or 10A input terminal (Max 10A)
2. Rotate function switch to Current gear. Press “SELECT” button to select the AC measurement
mode. Then connect the test leads to the tested circuit in serial, the tested current value and the
current polarity of the point where the red one is connected will be displayed on the screen
simultaneously.
Caution:
1. If “OL” is displayed on LCD, it indicates the tested current value has exceeded the present range
limit, please select higher range to complete the measurement.
2. The Max input value is 400mA or 10A (Depending on the terminal where the red test lead is
contacted). The overrated current will lead to fuse melt or even damage the meter.
Range
Accuracy
Resolution
400uA
±(1.5% + 10 cyfr)
0.1uA
4000uA
1uA
40mA
10uA
400mA
100uA
10A
±(2.5% + 15 cyfr)
10mA
Max measurement voltage drop: Full Range mA is 0.4V, A is 100mV;
Max input current: 10A (less than 15 seconds)
9
Overload Protection: 0.4A/250V restorable fuse; 10A/250V fuse
Frequency Response: 10A Range (50~200) Hz
2.2.5. Resistance (Ω)
1. Insert the black meter pen into “COM” terminal and red one into terminal.
2. Rotate the Range to “Ω” gear. Cross connect the test leads to the tested resistor.
3. When measuring the low resistance, please short-circuit the meter pens at first to test the wire
resistance, and then deduct it from the actual resistance.
Caution:
1. If “OL” is displayed on LCD, it indicates the tested resistance value has exceeded the present
range limit, please select higher range to complete the measurement. When measuring the Resistor
higher than 1MΩ, the instrument will take several seconds to make the reading stable. It is normal
when measuring the high resistor.
2. When the input terminal is open circuit, it will display “OL”.
3. When measuring in-line resistor, be sure that the power is off and all capacitors are discharged
completely.
Range
Accuracy
Resolution
400Ω
±(0.8% + 5 d)
0.1Ω
4kΩ
±(0.8% + 4 d)
1Ω
40kΩ
10Ω
400kΩ
100Ω
4MΩ
1kΩ
40MΩ
±(1.2% + 10 d)
10kΩ
Open Voltage circuit: Less than 200mV
Overload Protection: 250 V DC or AC Peak Value
Note: When measuring at Range 400Ω, please short-circuit the meter pens at first to test wire
resistance, and then deduct it from the actual resistance.
10
2.2.6. Diode and Continuity Test
1. Insert the black meter pen to “COM” terminal and the red one to terminal. (The polarity
of red test lead is “+”)
2. Rotate the Range to “ ” gear. Press „SELECT” button to select the Diode
measurement mode.
3. Forward Measurement: Connect the red test lead to the diode positive polarity and the black one
to the diode negative polarity. The approximate value of diode forward voltage drop will show on
the display.
4. Backward Measurement: Connect the red test lead to the diode negative polarity and the black
one to the diode positive polarity. “OL” symbol will be displayed on the screen.
5. The complete diode testing includes forward and backward measurement, if the result does not
meet the above; it means the diode is bad.
6. Press “SELECT” button to select the Continuity measurement mode.
7. Connect the meter pens to two points of the tested circuit. If the built-in buzzer sounds,
the resistance between the two points is less than 50Ω.
Range
Display
Test Condition
Forward Voltage Drop of
Diode
Forward DC Current is approx. 0.5mA, Backward
Voltage is approx. 1.5V
Buzzer makes a long sound if
resistance is less tan 50Ω
Open circuit voltage is approx. 0.5V
Overload Protection: 250V DC or AC Peak Value.
CAUTION: DO NOT INPUT VOLTAGE AT THIS RANGE!
2.2.7. Capacity (C)
1. Rotate function switch to “ ”
2. Insert the black test lead to “COM” terminal and red one to terminal.
3. Connect the tested capacity by the test leads to “COM”, , input terminal, the screen will
show capacitance parameter. (The relative value measurement could be conducted by pressing
“REL” button)