HoldPeak 870N User manual

1
870N
6000 DIGITS DUAL DISPLAY
AC/DC AUTO CAL CLAMP MULTIMETER
OPERATION MANUAL
This LCD Auto Range & Auto Power off Digital AC/DC AUTO
CAL clamp multimeter is a portable, 6000 digit multimeter. It is
ideally suited for field, laboratory, shop and home applications.
1. SAFETY INFORMATION
The following safety information must be observed to insure
maximum personal safety during the operation at this meter.
1) When measuring voltage ensure that instrument is not
switched to the current range, resistance range, diode and
continuity range, capacitance range or temperature range.
2) Use extreme care when measuring voltage above 50V.
especially from sources where high energy is existed.
3) Avoid making connections to "live" circuits whenever possible.
4) Before making resistance measurements, diode or continuity
test, capacitance test or temperature test, ensure that the
circuit under test is de-energized.
5) Always ensure that the correct function and range is selected.
6) Extreme care should be taken when using the instrument to
conjunction with a current transformer connected to the
terminals if an open circuit occurs.
7) Ensure that the test leads and probes are in good condition
with no damage to the insulation.
8) Take care not to exceed the over-load limits as given in the
specifications.
9) Before opening the cover of the battery cabinet to replace
batteries. disconnect the test leads from any external circuit,
set the selector switch to "OFF" position.
10) Keep the fingers after the protection ring when measuring
through the instrument lead.
11) Change the battery when the symbol “ ” appears to avoid
incorrect data.
2. Panel Layout
1) Rotary Switch: use this switch to select functions and ranges.
2) D.HOLD/ Back Light key:In any range, push the key, the
present display value will be locked and the " H " symbol will
appear, push it again to exit HOLD and the " H"symbol
disappear. Press “D.HOLD” button more than 2 seconds, the
back light will light, press it more than 2 seconds again, the
back light will light off or After 15 seconds Auto light out.
3) SELECT key:This key work on the "CAP Ω" range , Push
the key to choose resistance, diode ,continuity or capacitance
test, on the voltag or current range,change to DC/AC, and ℃/
℉range , change to ℃/℉
4) MAX/MIN key:Push the key to select MAX mode, push it
again to change MIN mode, push once again to change
max-min,press the key for more than 2 seconds to go back
auto range mode. And Push the key to change manual range
mode .But in Hz/Duty and Capacitance measurement, it can
not use.
5) Hz/DUTY key:In “ACV/ACA”or “Hz”range, push the
key, you can measure the Hz ,push again, can measure the
duty.
6)REL key:Push the key, the present display value will be stored
in memory, then the new display value is the difference
between input value and stored data. In Hz/Duty measurement,
it can not work. press the key for more than 2 seconds to
measure inrush current at ACA/DCA range
7) LCD Display:LCD Dual Display, facilitates reads the data.
8)T+V Ω Input Jack 、COM Input Jack、T- Input Jack.
2. SPECIFICATIONS
2.1 GENERAL SPECIFICATIONS
Display: 6000 digit LCD with a max. reading of 6000.
Range control: Auto range control.
Polarity: Automatic negative polarity indication.
Zero adjustment: Automalic.
Overrange indication: Only the "OL" display.
Lowbattery: The" " is displaywhenthe battery voltage is below6.2V.
Auto Power Off: 15 minutes after stopping the switch or no push
button, the meter automatically enter to power off
mode.Push SELECT button, auto power off disable.
Safety Standards: The meter is up to the standards of IEC1010
Double Insulation, Pollution Degree 2, Overvoltage
Category III 1000V.
Clamp opening size: 45mm.
Operating Environment: Temperature 32~104°F(0~40℃),
humidity<80%RH.
Storage Environment: Temperature -4~140°F(-20~60℃),
humidity<90%RH.
Power supply: 9V Zinc-carbon battery.
Dimension: 225(H)×97(W)×40(D)mm.
Weight: Approx. 350g (including batteries).
2.2 ELECTRICAL SPECIFICATIONS
Accuracies are ±(% of reading +numder in last digit) at 23±5℃,
≤70%RH.
2.2.1 DC Voltage
Range
Accuracy
Resolution
600mV
1.0% of rdg+5digits
0.1mV
6V
1mV
60V
10mV
600V
100mV
1000V
1.5% of rdg+5 digits
1V
Overload protection: 1000V DC/750Vrms AC
Impedance: 10MΩ,More than 100MΩon 400mV scale
2.2.2 AC Voltage(RMS)
Range
Accuracy
Resolution
Frequency
600mV
3.0% of rdg+15 digits
0.1mV
40~100Hz
6V
1.5% of rdg+5 digits
1mV
40~400Hz
60V
10mV
600V
100mV
750V
2.5% of rdg+5 digits
1V
40~100Hz
Average sensing, calibrated to rms of sine wave
Overload protection: 1000V DC/750Vrms AC
Impedance: 10MΩ.
2.2.3 DC Current
Range
Accuracy
Resolution
600A
3.0% of rdg+10 digits
0.1A
1000A
3.5% of rdg+10 digits
1A
Overload protection: 1000Arms within 60 seconds
2.2.4 AC Current
Range
Accuracy
Resolution
Frequency
600A
3.0% of rdg+10 digits
0.1A
50~60Hz
1000A
3.5% of rdg+10 digits
1A
Average sensing, calibrated to rms of sine wave
Overload protection: 1000Arms within 60 seconds
2.2.5 Resistance
Range
Accuracy
Resolution
600Ω
1.0% of rdg+8 digits
0.1Ω
6kΩ
1Ω
60kΩ
10Ω
400kΩ
100Ω
6MΩ
1kΩ
60MΩ
2.5% of rdg+15 digits
10kΩ
Overload protection: 250V DC/250Vrms AC
2.2.6 Capacitance
Range
Accuracy
Resolution
9.999nF
2.5% of rdg+15 digits
1PF
99.99nF
10PF
999.9nF
2.5% of rdg+10 digits
100PF
9.999μF
1nF
99.99μF
10nF
999.9μF
3.0% of rdg+20 digits
100nF
9.999mF
5.0% of rdg+25digits
1uF
99.99mF
10uF
Overload protection: 250V DC/250Vrms AC
TRUERMS
AUTOCAL
HP-870N
T+ T-
REL
INRUSH
Hz/Duty
MAX/MIN
SELECT
F
600A
600V
CAT.IV
2
2.2.7 Frequency AND Duty
Range
Accuracy
Resolution
Sensitivity
9.999Hz
0.5% of rdg+15 digits
0.001Hz
MIN of input
voltage:0.1V~1
.5V AC
99.99Hz
0.01Hz
999.9Hz
0.1Hz
9.999kHz
1Hz
99.99kHz
10Hz
999.9kHz
100Hz
9.999MHz
1kHz
Duty cycle: 1%~99%Accuracy: ±0.5
Overload protection: 250V DC/250Vrms AC
2.2.8 Temperature(NiCr-NiSi sensor)
Range
Accuracy
Resolution
℃
-20~1000℃
2.0% of rdg+3 digits
1℃
℉
-4~1832℉
2.5% of rdg+5 digits
Overload protection: 36V DC/36Vrms AC
2.2.9 Diode and Audible continuity test
Overload protection: 250V DC/250Vrms AC
3. MEASURING INSTRUCTION
3.1 DC/AC Voltage
1) Connect the black test lead to "T-COM" socket and red test
lead to the " T+VΩHz " socket.
2) Set the Rotary switch to "V" position Push the SELECT
key to desired DC /AC test
3) connect the probes across the source or load under
measurement.
4) Read the result from the LCD panel.
3.2 Resistance、Diode and Audible continuity
Measurement
1) Connect the black test lead to " T-COM " socket and red test
lead to the " T+VΩHz " socket.
2) Set the Rotary switch to desired "Ω" position, the
present function is resistance measurement, if it is other
function,push theSELECTKeytoselect.Diode and Audible
continuity
3) Push "SELECT" to select diode test.
4) Connect the black and red test probe to the cathode (-)
and anode (+) ends of diode to be tested
repectively, read the forward voltage drop (junction) value
from the display. If reverse connected the probes to diode,
display shows over-load.
5) Push "SELECT" to select Audible continuty test.
6) Connect the probes across circuit to be tested, the beeper
sounds continuously if the resistance is less than approx. 50Ω
7) Connect the probes across circuit to be tested.
8) Read the result from the LCD panel.
Caution: Ensure that the circuit to be tested is "dead".
Max.input over-load: 250V rms<10sec
3.3 Capacitance Measurement
1) Connect the black test lead to "T-COM" socket and red test
lead to the "T+VΩHz" socket.
2) Set the Rotary switch to desired "CAP " position.
3) Connect the probes to the capacitance to be tested.
4) Read the result from the LCD panel.
Caution:
a) Capacitors should be discharged before being tested.
b) When testing large capacitance, it will take longer time before
the final indication(For 1uF~99.99mF range, it will take about
4~7 seconds).
c) When testing small capacitance(≤1uF), to assur the
measurement accuracy, first press "REL", then go on
measureing.
Max.input over-load: 250V rms<10sec
3.4 Frequency Measurement
1) Connect the black test lead to "T-COM" socket and red test
lead to the " T+VΩHz " socket.
2)Set the “Hz/Duty”key to desired Hz/Duty test
3) connect the probes across the source or load under
measurement.
4) Read the result from the LCD panel.
3.5 Temperature Measurement
1) Set the Rotary switch to "℃/℉" position.
2) Connect the black test lead of the sensor to "T-COM"
socket and the red test lead to the " T+ " socket.
3) Put the sensor probe into the temperasure field under
measurement.
4) Read the result from the LCD panel.
Max.input over-load: 250V rms<10sec
A. The temperature function shows the random number at ordinary
times, must insert the thermocouple in temperature test hole while
examining temperature.
B. This meter inclosure WRNM-010 type contact thermocouple
limit temperature is 250 ℃(300 ℃shortly ) ;
C. Please don't change the thermocouple at will , otherwise we can't
guarantee to measure accuracy ;
Please don’t importing the voltage in the temperature function.
D. Please use special probe for test high temperature.
3.6 DC/AC Current Measurement
1) Set the Rotary switch to desired “600A ”or “1000A ”
position.
2) Set the SELECT key to desired DC or AC test
3) Press the "REL" Key, the display show "0", ACA auto Zero.
4) Open the clamp by pressing the jaw-opening handle and insert
the cable to be measured into the jaw.
5) Close the clamp and get the reading from the LCD panel.
Note:
Before this measurement, disconnect the test lead with the
meter for safety.
4. CARE AND MAINTENANCE
4.1 CARING FOR YOUR MULTIMETER
Your Digital Multimeter is an example of superior design and
craftsmanship. The following suggestions will help you care for
the multimeter so you can enjoy it for years.
1) Keep the multimeter dry. If it gets wet, wipe it dry immediately.
Liquids can contain minerals that can corrode electronic circuits.
2) Use and store the multimeter only in normal temperature
environments. Temperature extremes can shorten the life of
electronic devices, damage batteries and distort or melt plastic parts.
3) Handle the multimeter gently and carefully. Dropping it can
damage the circuit boards and cause and can accuse the
multimeter to work improperly.
4) When take current measurement, keep the cable at the center
of the clamp will get more accurate test result.
5) Keep the multimeter away from dust and dirt, which can cause
premature wear of parts.
6) Wipe the mutimeter with a damp cloth occasionally to keep it
looking new. Do not use harsh chemicals, cleaning solvents, or
strong detergents to clean the multimeter.
7) Use only fresh batteries of the required size and type. Always
remove old or weak batteries. They can leak chemicals that
destroy electronic circuits.
8) Please take out the battery when not using for a long time.
4.2 9Volt battery replacement
1) Ensure the instrument is not connected to any
extemal circuit. Set the Rotary switch to "OFF" position and
remove the test leads from the terminals.
2) Open the cover of the battery cabinet by a screwdriver.
3) Replace the old batteries with the same type batteries.
4) Close the battery cabinet cover and fasten the screw.
Above picture and content just for your reference. Please be
subject to the actual products if anything different or updated.
Please pardon for not informing in advance.
Range
Description
Test condition
“
Display read approx.
Forward voltage of
diode.
Forward DC current
approx. 1.0mA
Reversed DC voltage
approx. 3.0V
”
Built-in buzzer sounds
if resistance is less
than 50Ωapprox.
Open circuit voltage
approx. 1.0V

Other manuals for 870N

1

Other HoldPeak Multimeter manuals

HoldPeak 870N User manual

HoldPeak

HoldPeak 870N User manual

HoldPeak 770HD User manual

HoldPeak

HoldPeak 770HD User manual

HoldPeak 36 k User manual

HoldPeak

HoldPeak 36 k User manual

HoldPeak 90B User manual

HoldPeak

HoldPeak 90B User manual

HoldPeak 6688D User manual

HoldPeak

HoldPeak 6688D User manual

HoldPeak HP-760E User manual

HoldPeak

HoldPeak HP-760E User manual

HoldPeak 860N User manual

HoldPeak

HoldPeak 860N User manual

HoldPeak 6300 User manual

HoldPeak

HoldPeak 6300 User manual

HoldPeak 770D User manual

HoldPeak

HoldPeak 770D User manual

HoldPeak 760A Installation manual

HoldPeak

HoldPeak 760A Installation manual

HoldPeak 860D User manual

HoldPeak

HoldPeak 860D User manual

HoldPeak 6208 User manual

HoldPeak

HoldPeak 6208 User manual

HoldPeak HP-37A User manual

HoldPeak

HoldPeak HP-37A User manual

HoldPeak 39B User manual

HoldPeak

HoldPeak 39B User manual

HoldPeak 870E Installation manual

HoldPeak

HoldPeak 870E Installation manual

HoldPeak HP-90E User manual

HoldPeak

HoldPeak HP-90E User manual

HoldPeak HP-770H User manual

HoldPeak

HoldPeak HP-770H User manual

HoldPeak HP-90F User manual

HoldPeak

HoldPeak HP-90F User manual

HoldPeak 760J Installation manual

HoldPeak

HoldPeak 760J Installation manual

HoldPeak 870CR User manual

HoldPeak

HoldPeak 870CR User manual

HoldPeak HP-36C User manual

HoldPeak

HoldPeak HP-36C User manual

HoldPeak 770G User manual

HoldPeak

HoldPeak 770G User manual

HoldPeak 39B User manual

HoldPeak

HoldPeak 39B User manual

HoldPeak 770A User manual

HoldPeak

HoldPeak 770A User manual

Popular Multimeter manuals by other brands

Maxwell 25201 user manual

Maxwell

Maxwell 25201 user manual

isweek M300 Operator's instruction manual

isweek

isweek M300 Operator's instruction manual

A.W. Sperry Instruments DSA-2002P operating instructions

A.W. Sperry Instruments

A.W. Sperry Instruments DSA-2002P operating instructions

Gossen MetraWatt METRAHIT Base operating instructions

Gossen MetraWatt

Gossen MetraWatt METRAHIT Base operating instructions

Greenlee DM-40 instruction manual

Greenlee

Greenlee DM-40 instruction manual

Kewtech KT62 instruction manual

Kewtech

Kewtech KT62 instruction manual

Extech Instruments ExStik EC505 user manual

Extech Instruments

Extech Instruments ExStik EC505 user manual

Minipa ET-1100 instruction manual

Minipa

Minipa ET-1100 instruction manual

Tektronix DM 502 instruction manual

Tektronix

Tektronix DM 502 instruction manual

Tektronix DM 502A instruction manual

Tektronix

Tektronix DM 502A instruction manual

Tektronix TEKTOOLS DMM155 instructions

Tektronix

Tektronix TEKTOOLS DMM155 instructions

Agilent Technologies 34450A Application note

Agilent Technologies

Agilent Technologies 34450A Application note

KNOVA KN 8059 manual

KNOVA

KNOVA KN 8059 manual

Micro Seven VM10C Operator's manual

Micro Seven

Micro Seven VM10C Operator's manual

Rigol DM3068 manual

Rigol

Rigol DM3068 manual

Fluke 106 user manual

Fluke

Fluke 106 user manual

Hameg HM 8012 manual

Hameg

Hameg HM 8012 manual

Zico ZI-849 Operating	 instruction

Zico

Zico ZI-849 Operating instruction

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.