Baker Hughes Panametrics HygroPro User manual

HygroPro
User’s manual
916-099 Rev. D

ii

HygroPro
Moisture transmitter
User’s manual
916-099 Rev. D
November 2012
iii

[no content intended for this page - proceed to next page]
iv

Contents
Product registration .................................................................................vii
Services .............................................................................................vii
Terms and conditions................................................................................vii
Typographical conventions ..........................................................................vii
Safety issues ........................................................................................vii
Auxiliary equipment .................................................................................viii
Environmental compliance ..........................................................................viii
Chapter 1: Installation....................................................................1
1.1 Introduction ......................................................................................1
1.2 Sample system guidelines .......................................................................2
1.3 Mounting the transmitter ........................................................................3
1.4 Wiring the transmitter ...........................................................................4
1.4.1 Standard wiring connections .............................................................................4
1.4.2 Hazardous area wiring connections ......................................................................6
Chapter 2: Operation ...................................................................9
2.1 Powering up and programming ..................................................................9
2.1.1 Menu Map................................................................................................10
2.1.2 Keypad ...................................................................................................10
2.2 Basic setup ......................................................................................11
2.2.1 Unlocking the display .....................................................................................11
2.2.3 Selecting measurement parameters .....................................................................11
2.2.4 Setting up the display....................................................................................12
2.2.5 Setting up the analog output ............................................................................12
2.2.6 Entering the node ID......................................................................................13
2.2.7 Setting up the RS485 output .............................................................................13
2.3 Advanced setup ................................................................................ 14
2.3.1 Setting up the pressure/temperature displays...........................................................14
2.3.2 Entering sensor calibration data .........................................................................14
2.3.3 Locking and unlocking the keypad.......................................................................15
2.3.4 Locking and unlocking the menus .......................................................................15
Chapter 3: Service and maintenance .................................................. 16
3.1 The service menus .............................................................................. 16
3.2 Moisture probe error conditions................................................................. 16
3.3 Cleaning the moisture probe.....................................................................17
3.3.1 Preparing to clean the probe ............................................................................17
3.3.2 Replacing the RTE ........................................................................................17
3.3.4 Removing the probe from the transmitter ...............................................................17
3.3.5 Cleaning the sensor and the shield ......................................................................18
3.3.6 Installing the probe in the transmitter ...................................................................18
v

3.3.7 Evaluating the cleaned probe............................................................................18
Chapter 4: Specifications .............................................................. 19
4.1 General ......................................................................................... 19
4.2 Electrical........................................................................................ 19
4.3 Mechanical ..................................................................................... 19
4.4 Moisture sensor................................................................................. 19
4.5 Built-in temperature sensor.....................................................................20
4.6 Built-in pressure sensor.........................................................................20
4.7 Certifications ...................................................................................20
vi

Product registration
Thank you for purchasing a HygroPro moisture transmitter
from Panametrics. Please register your product at https://
info.bakerhughesds.com/New-product-registration-LP.
htmlproduct support such as the latest software/firmware
upgrades, product information and special promotions.
Services
Panametrics provides customers with an experienced
staff of customer support personnel ready to respond to
technical inquiries, as well as other remote and on-site
support needs. To complement our broad portfolio of
industry-leading solutions, we offer several types of flexible
and scalable support services including: training, Product
repairs, service Agreements and more. Please visit change
to: https://www.bakerhughesds.com/panametrics/services
for more details.
Terms and conditions
Panametrics’ sales terms and conditions for your recent
purchase of a Panametrics product, including the
applicable product warranty, can be found on our website
at the following link: https://www.bakerhughesds.com/sales-
terms-conditions.
Typographical conventions
Note: These paragraphs provide information that provides
a deeper understanding of the situation, but is not essential
to the proper completion of the instructions.
IMPORTANT:
These paragraphs provide information that emphasizes
instructions that are essential to proper setup of the
equipment. Failure to follow these instructions carefully
may cause unreliable performance.
CAUTION!
This symbol indicates a risk of potential minor
personal injury and/or severe damage to
the equipment, unless these instructions are
followed carefully.
WARNING!
This symbol indicates a risk of potential
serious personal injury, unless these
instructions are followed carefully.
Safety issues
WARNING!
It is the responsibility of the user to make sure
all local, county, state and national codes,
regulations, rules and laws related to safety
and safe operating conditions are met for
each installation.
Attention European customers!
To meet CE marking requirements for all
units intended for use in the EU, all electrical
cables must be installed as described in this
manual.
vii

Auxiliary equipment
Local safety standards
The user must make sure that he operates all auxiliary
equipment in accordance with local codes, standards,
regulations, or laws applicable to safety.
Working area
WARNING!
Auxiliary equipment may have both manual
and automatic modes of operation. As
equipment can move suddenly and without
warning, do not enter the work cell of this
equipment during automatic operation,
and do not enter the work envelope of this
equipment during manual operation. If you
do, serious injury can result.
WARNING!
Make sure that power to the auxiliary
equipment is turned OFF and locked out
before you perform maintenance procedures
on this equipment.
Qualification of personnel
Make sure that all personnel have manufacturer-approved
training applicable to the auxiliary equipment.
Personal safety equipment
Make sure that operators and maintenance personnel have
all safety equipment applicable to the auxiliary equipment.
Examples include safety glasses, protective headgear,
safety shoes, etc.
Unauthorized operation
Make sure that unauthorized personnel cannot gain access
to the operation of the equipment.
Environmental compliance
RoHS
The HygroPro moisture transmitter fully complies with RoHS
regulations (Directive 2011/65/EU).
Waste Electrical and Electronic Equipment (WEEE) directive
Panametrics is an active participant in Europe’s Waste
Electrical and Electronic Equipment (WEEE) take-back
initiative (Directive 2012/19/EU).
The equipment that you bought has required the extraction
and use of natural resources for its production. It may
contain hazardous substances that could impact health
and the environment.
In order to avoid the dissemination of those substances in
our environment and to diminish the pressure on the natural
resources, we encourage you to use the appropriate take-
back systems. Those systems will reuse or recycle most of
the materials of your end life equipment in a sound way.
The crossed-out wheeled bin symbol invites you to use
those systems.
If you need more information on the collection, reuse and
recycling systems, please contact your local or regional
waste administration.
Please visit https://www.bakerhughesds.com/health-
safetyand-environment-hse for take-back instructions and
more information about this initiative.
viii

Chapter 1: Installation
1.1 Introduction
The HygroPro moisture transmitter is a compact,
intrinsically-safe, loop-powered, 4-20 mA transmitter
that provides accurate dew/frost point measurements
over a range of –110° to 20°C (-166° to 68°F). It features an
integrated display and a six-button keypad, and it is
housed in an IP67/Type 4X enclosure.
The HygroPro transmitter uses an aluminum oxide
moisture sensor, and it includes a temperature thermistor
and a pressure transducer on a common mount for
calculation of parameters such as:
• ppmvin gases
• ppmwin liquids
• pounds per million standard cubic feet
in natural gas
• process relative humidity.
IMPORTANT
To install the HygroPro in a hazardous (classified) area,
see "Hazardous area wiring connections" on page 6.
1

1.2 Sample system guidelines
The HygroPro transmitter can be installed in a sample
system or directly in the process line. However, Panametrics
recommends that the unit be installed in a sample
system to protect the probe from potentially damaging
components in the process stream. Figure 1 below
shows a typical sample system.
HygroPro
Probe
Needle
valve 1
Needle
valve 2
Flow cell
Flow
See note below
Intake near the center of the
pipe is recommended.
Figure 1: A typical sample system
Note: At least 5 ft (1.5 m) of 1/4” (6 mm) tubing vented
to atmosphere will ensure an accurate process sample
and avoid diffusion of ambient air moisture back into
the process.
In the sample system shown in Figure 1 on page 2,
fully open valve 1 and use valve 2 to regulate the sample
flow for measurements at process system pressure.
For measurements at atmospheric pressure, fully open
valve 2 and use valve 1 to regulate the sample flow.
Before constructing a sample system, consult a Panametrics
application engineer and observe the following guidelines:
• The sample system should be very simple and should
contain as few components as possible. All or most of
those components should be located downstream
of the measurement point.
• Sample system components must not affect moisture
readings. Most common filters and pressure regulators
are not suitable for sample systems because the wetted
parts adsorb moisture or release moisture into the sample
system. They may also allow ambient contamination to
enter the sample system. If possible, use stainless steel
for all wetted parts.
• The HygroPro probe should be oriented perpendicular
to the sample system inlet. For dimensions and
other sample system requirements see "Mechanical
specifications" on page 19.
• Sample systems should be tested for leaks prior to
operation, using a Snoop leak detector, to verify the
integrity of connections, components and fittings.
IMPORTANT
When pressurizing or depressurizing the sample system,
be careful to avoid shock damage to the moisture sensor.
2

1.3 Mounting the transmitter
CAUTION!
If the HygroPro will be installed directly
into the process line, consult Panametrics
for proper installation instructions and
precautions before proceeding.
Refer to Figure 2 below and complete the steps on the next
page to install the HygroPro transmitter.
Probe hex nut
3/4-16 UNF-2A
Shield
Figure 2: HygroPro installation
1. Make sure the sintered or stainless-steel shield is in
place over the sensor. This shield protects the aluminum
oxide sensor from damage during operation.
2. Using the integral 3/4-16 straight male thread, screw the
probe end of the transmitter into the process or sample
system fitting. Make sure not to damage the threads.
Note: A 3/4-16 to G 1/2 thread adapter is available from
Panametrics.
3. Using a 1-1/8” wrench on the probe hex nut, tighten the
probe securely into the process or sample system fitting.
CAUTION!
Do not apply torque to the transmitter
electronic module to tighten the unit into its
fitting.
Note: If there is insufficient space to rotate the HygroPro
during installation, remove the replaceable transducer
element (RTE) from the transmitter, install it in the fitting,
then re-install the the transmitter onto the RTE.
3

1.4 Wiring the transmitter
Proceed to the appropriate section for instructions on:
• Standard wiring connections (below)
• "Hazardous area wiring connections" on page 6
1.4.1 Standard wiring connections
The HygroPro is a 4-20 mA loop-powered device that can
use the same two wires for its measurement signal lines
and its power supply lines. Follow the instructions in this
section only for units not requiring hazardous (classified)
area protection.
IMPORTANT
To install the HygroPro in a hazardous (classified) area, see
"Hazardous area wiring connections" on page 6
The transmitter must be wired using the factory-supplied,
2-meter cable. If a different cable length is required, please
contact the factory for assistance.
Note: If a longer cable is required, you may splice an
extension onto the standard cable. Be sure to connect
leads of the same color together. For example, connect blue
lead to blue lead, brown lead to brown lead, etc.
The HygroPro may be connected to either of the following
types of external systems:
• an external device that can provide the loop power
to the HygroPro and can also receive and display the
4-20 mA measurement output from the HygroPro
(proceed to next page)
• an external power supply to provide power to the
HygroPro plus a personal computer (PC) running
Panametrics’ PanaView™ interface software
(proceed to page 13)
1.4.1a Standard connections - without a PC
IMPORTANT
To install the HygroPro in a hazardous (classified) area,
see "Hazardous area wiring connections" on page 6.
Refer to "Figure 3" on page 5 and Table 1 below, and
complete the following steps to wire the transmitter.
Table 1: Cable leads - without a PC
Lead Connection description
Blue Power supply (+) [12-28 VDC]
Brown Power supply (–) [return]
Shield Earth ground [recommended]
1. Push the female connector end of the factory-supplied
cable into the mating male connector on the transmitter
module. Make sure the pins are properly aligned. Then,
secure the connectors together by sliding the metal
sleeve on the cable over the connectors and turning
it clockwise until it is tight.
2. Using the flying leads at the other end of the factory-
supplied cable, connect the transmitter to the external
system.
Note: The blue and brown leads also carry the measurement
signal current output of 4-20 mA.
3. Trim any unused leads even with the outer cable
jacket to remove any bare wire and prevent accidental
short circuits.
4

Shield
+
Power and
measurement circuits
-
Brown
12-28 VDC
Return
Blue
Earth
ground
Figure 3: Standard connections - without a PC
1.4.1b Standard connections - with a PC
IMPORTANT
To install the HygroPro in a hazardous (classified) area, see
"Hazardous area wiring connections" on page 6.
Refer to "Figure 4" on page 5 and Table 2 below, and
complete the following steps to wire the transmitter.
Table 2: Cable leads - with a PC
Lead Connection description
Blue Power supply (+) [12-28 VDC]
Brown Power supply (–) [return]
White RS485-RS232 converter (+) [positive]
Black RS485-RS232 converter (–) [negative]
Ground Earth ground
1. Connect an RS485-RS232 converter (customer-supplied)
to an available serial port on the PC.
2. Push the female connector end of the factory-supplied
cable into the mating male connector on the transmitter
module. Make sure the pins are properly aligned. Then,
secure the connectors together by sliding the metal
sleeve on the cable over the connectors and turning
it clockwise until it is tight.
3. Using the flying leads at the other end of the factory-
supplied cable, connect the transmitter to the external
system.
4. Trim any unused leads even with the outer cable
jacket to remove any bare wire and prevent accidental
short circuits.
Power
supply
24 VDC
RS485 ? RS232
converter box
PanaView
RS232
+-
Blue
Brown
+ White
? Black
Earth
Ground
Figure 4: Standard connections with a PC
5

1.4.2 Hazardous area wiring connections
Before installing and using the HygroPro in a hazardous
(classified) area, be sure to read and understand all
applicable reference materials. This includes:
• all EU or North American Standards and Directives
(see Table 3 on page 7 and Table 4 on page 7)
• all local safety procedures and practices
• the FM schematic drawing (see Figure 5 on page 6)
• this user’s manual
Note: It is the installer’s responsibility to follow all applicable
standards and procedures.
WARNING!
The procedures in this section must be
performed only by trained technicians who
have the necessary skills and qualifications.
Associated apparatus
Note 3 RS485 comm
Note 4
Associated apparatus
Note 3 Loop-power device
Note 4
Notes:
1. Installation shall be in accordance with the National Electrical
Code (ANSI/NFPA 70) or part one of the Canadian Electrical Code
(C221), as applicable.
2. The subject equipment is FM approved, and FM approved for
Canada for the following hazardous (classified) locations and
protection methods:
Intrinsically safe, class I, II, III, division 1, group ABCDEFG, T4
T Ambient: -40°C to 60°C
Enclosure: IP 67
3. Galvanically isolated FM approved (US) or Canadian certified
(for Canada) associated apparatus installed in accordance
with the manufacturer’s instructions.
4. Equipment to which associated apparatus is connected shall not
use or generate in excess of 250V.
5. Maintain the following relationship between intrinsic aafety (entity)
parameters of HygroPro and associated apparatus:
HygroPro Vmax ≥ associated apparatus Voc
HygroPro Imax ≥ associated apparatus Isc
HygroPro Pi ≥ associated apparatus Po
HygroPro Ci + Ccable ≤ associated apparatus Ca
HygroPro Li + Lcable ≤ associated apparatus La
Hazardous or non-hazardous location
class I, division 1, group ABCD
class II, division 1, group EFG
class III, division 1
Non-hazardous location
HygroPro
Loop terminal
Safety entity parameters:
Vmax = 28 V
Imax = 93.3 mA
Pi = 0.653 W
Ci = 0
Li = 62 µH
RS485 comm terminal safety
entity parameters:
Vmax = 3.72 V
Imax = 228 mA
Pi = 212 mW
Ci = 0
Li = 62 µH
blue
brown
black
white
Figure 5: FM schematic (ref. dwg #752-262)
6

1.4.2a Applicable standards and directives
When the HygroPro is installed in hazardous areas with
potentially explosive atmospheres, it complies with the ATEX
directive 2014/34/EU, the EU standards listed in Table 3 below,
and the North American FM/CSA and IEC standards listed in
Table 4 below.
Table 3: European Union (EU) standards
Title Number
General requirements IEC/EN 60079-0
Intrinsic safety “i” IEC/EN 60079-11
Group II, category 1G,
zone 0 equipment IEC/EN 60079-14
Table 4: North American standards
Title Number
Electrical Equipment for Use in
Hazardous (Classified) Locations
General Requirements
Class
No. 3600
Intrinsically Safe Apparatus and
Associated Apparatus for use in
Class I, II, and III, Division 1, and
Class I, Zone 0 and 1 Hazardous
(Classified) Locations
Class
No. 3610
Electrical Equipment for
Measurement, Control and
Laboratory Use
Class
No. 3810
Intrinsically Safe and
Non-incendive Electrical for
Use in Hazardous Locations
CSA-C22.2
No. 60079-11
Safety Requirements for
Electrical Equipment for
Measurement, Control,
and Laboratory Use - Part 1:
General Requirements
(Adopted IEC 61010-1:2001, MOD)
(Tri-National standard, with
UL 61010-1 and ISA 82.02.01)
CSA-C22.2
No. 61010-1
Degrees of Protection Provided
by Enclosures (IP Code) ANSI/IEC 60529
Degrees of Protection Provided
by Enclosures (IP Code)
CSA-C22.2
No. 60529
1.4.2b Compliance requirements
The HygroPro installation must comply with EN 60079-14
in Europe and with the National Electrical Code (ANSI/NFPA
70) or part one of the Canadian Electrical Code (C22.1), as
applicable, in North America. In other regions, additional
local codes may also apply.
1.4.2c Electrical connections
The HygroPro transmitter is certified intrinsically-safe
for use in zone 0. However, the external power must be
provided by one of the following methods:
• an isolated, intrinsically-safe, 24 VDC power supply
mounted in the safe area
• an isolated zener barrier mounted in the safe area
and installed between a standard 24 VDC power
supply and the HygroPro
Refer to "Figure 6" on page 8 for a typical hazardous
(classified) area HygroPro wiring diagram. This installation
uses an MTL706 zener barrier, in accordance with the
second option listed above.
If using serial communications with a PC, an isolated
RS232-RS485 converter mounted in the safe area between
the computer and HygroPro must be used. The converter is
typically powered by its own standard 24 VDC power supply.
WARNING!
Do NOT power an RS232-RS485 converter
from the same intrinsically-safe 24 VDC
power supply used to power the HygroPro.
7

Bus bar
Brown
Blue
Red
3
41
2
+24V
24V return
–+
OCB(*)
Green
Black
Black
IS ground
R1, 250 ohms, ±5%, 1/4 Watt
MTL706 Power supply
4 - 20 mA
Analog input device
Figure 6: Hazardous area connections
1.4.2d Requirements for external devices
When connecting the HygroPro to external devices, the
allowable total load capacitance and inductance for
those devices are listed in the manufacturer’s datasheets.
The entity parameters of the external devices (e.g. voltage,
current and power) must be equal to or lower than the
same specifications for the HygroPro.
The entity parameters for the HygroPro are listed in
Table 5 below.
Table 5: HygroPro entity parameters
Loop power supply
Ui= 28 V Pi= 0.653 W Li= 62 µH
Ii= 93.3 mA Ci= 0
RS485 digital output
Ui= 3.72 V Pi= 212 mW Li= 62 µH
Ii= 228 mA Ci= 67 µF at 5.36 V
1.4.2e Special conditions for safe operation
The “X” at the end of the HygroPro ATEX certificate number,
Bas06ATEX0019X, indicates that special conditions are
required for safe operation in Europe. These conditions are:
1. The equipment must be protected against impact or
friction with ferrous metals.
2. The bare ends of the connecting cable must be
terminated in such a way that the terminations are
afforded a degree of protection of not less than IP20
ingress protection.
3. The HygroPro is incapable of withstanding the
500 VAC test between all inputs and frame for one
minute. Therefore, the HygroPro must be earthed by
connecting the outer grounding screw on the HygroPro
with the equipotential bonding system. Follow the
local standards and electrical codes regarding the
equipotential bonding system.
WARNING!
Never connect or disconnect the HygroPro in
the hazardous area when supply power or the
communication circuit is energized. Isolate
the supply lines in the non-hazardous area
first.
Following the same special conditions listed above is
recommended for regions outside of Europe as well.
8

Chapter 2: Operation
2.1 Powering up and programming
After the HygroPro has been installed as described in
chapter 1, installation, power may be applied to the unit.
The transmitter requires up to 60 seconds to initialize and
begin normal operation. The unit will meet its specified
accuracy within 3 minutes.
Figure 7 below shows a close-up view of the HygroPro
display and keypad, and "Figure 8" on page 10 a
complete menu map of the HygroPro setup program.
Figure 7: HygroPro display and keypad
9

2.1.1 Menu Map
# of views
1
2
3
Normal
Reverse
Contrast
None
Even
Odd
1
2
7
8
2400
4800
9600
Versions
Upgrade
Default analyzer
4-20mA
NAMUR
Special
Base trim
Span trim
%
Mode
Zero
Span
Zero
Span
Test
Normal
Hygro °C
Pressure FP
Measure
Type
Range
Trim
Display Display
Passcode Lock keypad
Startup screen
Lock menus Passcode Lock keypad Lock menus
[escape] - [enter] - [escape]
unlock - [enter]
* Factory-level passcode is required
to access the Service menu.
[22 options]
Calibrate
Live
Constant
enter value
Hygro Curve
Press curve
CS table
Const pressure
Live
Constant
enter value
Const
temperature Node ID
enter value
Baud rate
Parity
Stop bits
Data Bits
Save Table?
RS485 Service*
Zero mV
Zero PSIg
Span mV
Span PSIg
Temp °C 6
CS 6
Temp °C 1
CS 1
Row 16
Units 16
Row 1
Units 1
Figure 8: Programming menu map
2.1.2 Keypad
After entering the setup program, the keys on the HygroPro keypad (see "Figure 7" on page 9) perform the
following functions:
• enter - confirm a selection or move to the next screen
• escape - cancel a selection or move to the
previous screen
• up - scroll upward through a list of options or
increase the value of a selected character
• down - scroll downward through a list of options
or decrease the value of a selected character
• left - move the cursor to the next character to the left
• right - move the cursor to the next character to the right
10

2.2 Basic setup
The HygroPro transmitter is easily programmed to meet the user’s requirements by referring to the menu map in "Figure 8" on
page 10 and following the instructions in this section.
2.2.1 Unlocking the display
Upon startup, the following screen appears on the
HygroPro display:
The symbol in the upper
right corner indicates that
the screen is locked.
To unlock the screen, press
escape, enter, escape.
2.2.2 Entering the setup program
To enter the setup program, complete the following steps:
Using the arrow keys,
highlight the opened lock
and press enter.
Select passcode and
press enter.
Note: The default operator-level passcode is 2719.
Enter your passcode using
the up and down keys to
change the value of each
passcode digit, and the
left and right keys to move
from one digit to the next.
When the passcode is
correct, press enter.
Note: To enter the service menu, the factory-level passcode
must be entered.
2.2.3 Selecting measurement parameters
To select the measurement parameters, complete the
following steps:
To select the number
of measurements to be
displayed on each screen,
select # of views and
press enter.
Use the up and down keys
to select the number of
views desired and press
enter. Then, press enter
again to return to the
previous menu.
Press escape to see the
view setup. To change a
measurement parameter,
use the arrow keys to
highlight the parameter
name and then press enter.
Use the up and down
arrow keys to select a
measurement parameter
and then press enter
twice.
Note: As an example, dew point has been selected as the
measurement parameter to be changed.
Use the up and down
arrow keys to select a unit
of measure and press
enter twice.
Repeat the previous steps to make any other desired
measurement parameter changes. Then, continue
as follows:
• If the operator-level passcode was used, highlight
the lock symbol and press enter again to return to the
DISPLAY menu.
• If the service-level passcode was used, continue the
programming steps below.
11

If the number of decimal
places is acceptable,
press escape. To change
the number of decimal
places, press enter.
Use the up and down keys
to change the number
of decimal places and
press enter.
The programming sequence is now complete and you are
returned to the view menu.
2.2.4 Setting up the display
To set up the display, complete the following steps:
Use the up and down keys
to select display and press
enter. Then, enter your
passcode and press enter.
Use the up and down keys
to select display and
press enter.
If the display type is
acceptable, press escape
to return to the previous
menu. To change the
display type, use the up
and down keys to select
normal or reverse and
press enter.
If you wish to change the display contrast, at the above
screen, highlight that choice and press enter.
Use the arrow keys to
change the contrast value
and press enter. Then
press escape twice to
return to the main menu.
2.2.5 Setting up the analog output
To set up the analog output, complete the following steps:
After entering your
password, use the right
arrow key to scroll to
output and press enter.
Select the measure option
and press enter.
Use the arrow keys to
scroll to the desired
output parameter and
press enter.
In the output menu, select type and press enter. The
following screen appears:
Select the desired output
type and press enter.
If special was chosen at the previous prompt, the following
screen appears:
Select zero and press
enter.
Use the arrow keys to
enter the zero value for
the special output and
press enter.
Repeat the above two steps to enter the span value for the
special output. In the output menu, select range and press
enter. The following screen appears:
12
Table of contents
Other Baker Hughes Transmitter manuals

Baker Hughes
Baker Hughes XMTC User manual

Baker Hughes
Baker Hughes Masoneilan 12400 Series User manual

Baker Hughes
Baker Hughes Panametrics oxy.IQ User manual

Baker Hughes
Baker Hughes DigitalFlow XGF868i Owner's manual

Baker Hughes
Baker Hughes DigitalFlow XGF868i User manual

Baker Hughes
Baker Hughes Panametrics DigitalFlow XGM868i User guide