Bedfont Gastrolyzer User manual

User manual
0086
Our family, innovang health, for yours.
GastroCH4ECK
®
Gastrolyzer®
GastroCH4ECK Gastrolyzer®®

!
!
1!
!
!
Our$family,$innovating$health,$for$yours.$
!"#$%#$&!
Definitions!.................................................................................................................................................!2!
Intended!use!.............................................................................................................................................!2!
Introduction!..............................................................................................................................................!2!
Safety!notes!...............................................................................................................................................!3!
Contraindications!......................................................................................................................................!3!
Safety!during!installation!and/or!maintenance!........................................................................................!3!
Electrical!safety!.........................................................................................................................................!4!
Safety!against!risk!of!fire!...........................................................................................................................!4!
Mechanical!safety!.....................................................................................................................................!4!
Environment!..............................................................................................................................................!5!
Safety!information!and!device!symbols!....................................................................................................!6!
Monitor!installation!..................................................................................................................................!7!
Monitor!set-up!..........................................................................................................................................!7!
Temperature!requirements!and!device!warm!up!.....................................................................................!8!
Performing!a!breath!test!...........................................................................................................................!8!
Breath!bag!sample!collection!....................................................................................................................!9!
Analysing!a!breath!bag!test!.....................................................................................................................!10!
Instrument!layout!...................................................................................................................................!11!
Standard!pack!contents!...........................................................................................................................!11!
Technical!specification!............................................................................................................................!11!
Display!symbols!.......................................................................................................................................!12!
Warnings!and!maintenance!....................................................................................................................!16!
Cleaning!...................................................................................................................................................!18!
Routine!maintenance!..............................................................................................................................!18!
Calibration!...............................................................................................................................................!18!
GastroCHART™!software!–!connecting!to!the!PC!....................................................................................!19!
Troubleshooting!procedure!....................................................................................................................!20!
Returns!procedure!..................................................................................................................................!20!
Warranty!.................................................................................................................................................!21!
Testing!protocol!thresholds!....................................................................................................................!22!
References!...............................................................................................................................................!22!
Spares!......................................................................................................................................................!22!

!
!
2!
!
!
Our$family,$innovating$health,$for$yours.$
!
'%()#)$)"#&*
!+,-./0,*indicates!a!potentially!hazardous!situation,!which,!if!not!avoided,!may!result!in!minor!or!
moderate!injury.!It!may!be!used!to!alert!against!unsafe!practices.!It!may!also!be!used!to!indicate!the!
possibility!of!erroneous!data!that!could!result!in!an!incorrect!diagnosis!(does!not!apply!to!all!
products).!!
.12/3-+0-,!is!used!for!comments!that!add!value!to!the!step!or!procedure!being!performed.!
Following!the!.12/3-+0-!advice!adds!benefit!to!the!performance!of!the!equipment!or!to!a!process.!
0/-4,!is!used!to!call!attention!to!notable!information!that!should!be!followed!during!installation,!
use,!or!servicing!of!the!equipment.!
!
.#$%#5%5*6&%*
The!GastroCH4ECK®!Gastrolyzer®!is!a!portable!desktop!monitor!which!measures!hydrogen!(H2),!
methane!(CH4)!and!oxygen!(O2)!levels!in!expired!breath!samples!in!response!to!appropriate!
substrates.!It!is!intended!to!be!for!multi-patient!use!on!children!and!adults!and!used!by!healthcare!
professionals!primarily!in!a!hospital,!or!physician’s!surgery.!
The!expired!breath!can!be!delivered!direct!to!the!monitor!for!immediate!analysis!via!a!mouthpiece!
or!a!sample!can!be!taken!remotely!via!a!breath-bag!for!subsequent!analysis.!
The!GastroCH4ECK®!Gastrolyzer®!will!be!used!by!a!single!patient!at!a!time,!but!will!interface!with!a!
PC!to!allow!a!multi-patient!sampling!protocol!to!be!performed.!
The!GastroCH4ECK®!Gastrolyzer®!can!be!used!as!an!aid!to!diagnose!the!following!disorders:!
•Carbohydrate!breakdown!deficiency!
•Carbohydrate!malabsorption!
•Lactose!intolerance!
•Bacterial!overgrowth!
•Determination!of!time!of!passage!through!gut!
Specific!diagnosis!is!not!possible!with!this!device;!further!specific!testing!would!need!to!be!carried!
out!to!diagnose!a!patient’s!condition.!
!
.#$7"568$)"#*
H2!and!CH4!are!generated!in!the!intestinal!lumen!by!bacterial!action!on!carbohydrates!in!the!small!or!
large!intestine.!Once!the!resultant!H2!and!CH4!is!diffused!into!the!bloodstream!it!is!transported!to!
the!alveoli!and!can!then!be!detected!in!expiratory!air.!Levitt1!demonstrated!the!correlation!between!
intestinal!lumen!H2!and!CH4!production!and!H2!and!CH4!excretion!in!expiratory!air.!Accurate!

!
!
3!
!
!
Our$family,$innovating$health,$for$yours.$
measurement!of!H2!and!CH4!in!parts!per!million!(ppm)!in!expiratory!air!reveals!intolerance!and/or!
malabsorption!of!carbohydrates;!or!bacterial!overgrowth.!
The!GastroCH4ECK®!!Gastrolyzer®!also!measures!O2!levels!in!the!breath!sample!in!order!to!correct!
for!any!tests!that!do!not!achieve!an!end-tidal!breath!sample.!The!end-tidal!breath!sample!is!that!last!
portion!of!expired!air!and!contains!the!most!accurate!levels!of!H2!and!CH4.!End-tidal!breath!contains!
approx.!14-16%!O2!and!therefore!corrections!to!the!final!result!can!be!made!if!required!by!applying!a!
correction!factor!if!more!than!14%!O2!is!recorded!in!the!sample.!!
Operation!of!the!monitor!is!straightforward.!A!sampling!system!enables!end-expired!breath!to!be!
sampled!easily!and!hygienically,!using!single-use!disposable!mouthpieces!and!single-use!disposable!
face!masks.!!
*
9:(%$;*#"$%&**
To!avoid!electric!shock,!this!equipment!must!be!connected!to!a!mains!supply!with!protective!earth-
grounding.!!
The!operator!should!not!touch!the!patient!whilst!in!contact!with!either!the!computer!or!the!USB!
connector!parts.!!
The!GastroCH4ECK®!Gastrolyzer®!should!only!be!connected!to!a!computer!that!is!manufactured!in!
accordance!with!EN!60950!–!Safety!of!information!technology!equipment.!!
Please!do!not!attempt!to!modify!the!equipment!in!any!way!or!use!accessories!not!specified!for!use!
by!the!manufacturer.!Any!attempt!to!do!so!will!invalidate!the!warranty!and!may!affect!the!safety!of!
the!device.!!
Any!modification!to!this!equipment!during!the!actual!service!life!requires!evaluation!to!the!
requirement!of!EN!60601-1.!
*
!"#$7:)#5)8:$)"#&**
If!exhaling!gently!to!produce!a!breath!sample!is!physically!demanding!then!it!may!cause!
contraindications!in!patients!who!have!had!a!recent!myocardial!infarction.!Also!extensive!exhalation!
might!lead!to!syncope.!
!
9:(%$;*567)#<*)#&$:==:$)"#*:#5>"7*?:)#$%#:#8%**
The!GastroCH4ECK®!Gastrolyzer®!can!be!installed!by!a!trained!user!as!long!as!the!instruction!manual!
has!been!read.!The!GastroCH4ECK®!Gastrolyzer®!weighs!8kg.!!
Do!not!open!the!GastroCH4ECK®!Gastrolyzer®.!This!could!result!in!a!void!in!warranty.!Servicing!
should!only!be!carried!out!by!a!Bedfont®!trained!representative.!

!
!
4!
!
!
Our$family,$innovating$health,$for$yours.$
4=%8$7)8:=*&:(%$;**
To!reduce!the!risk!of!electrical!shock,!this!equipment!uses!a!three-wire!electrical!cord!and!plug!to!
connect!the!GastroCH4ECK®!Gastrolyzer®!to!earth-ground.!To!preserve!this!safety!feature:!!
•Make!sure!that!the!matching!wall!outlet!receptacle!is!properly!wired!and!earth-grounded.!
Check!that!the!line!voltage!agrees!with!the!voltage!listed!on!the!rear!label!affixed!to!the!
GastroCH4ECK®!Gastrolyzer®.!
•Do!not!install!the!GastroCH4ECK®!Gastrolyzer®!on!a!ground!fault-protected!power!source.!!
•Do!not!place!containers!holding!liquid!on!or!near!the!GastroCH4ECK®.!If!they!spill,!liquid!may!
enter!the!GastroCH4ECK®!Gastrolyzer®!and!damage!electrical!or!mechanical!components.!
•Only!use!properly!rated!safety!fuses!as!described!on!the!rear!of!the!GastroCH4ECK®!(T!3.15!
AH!250V).!
!
9:(%$;*:<:)#&$*7)&@*"(*()7%**
Fuses!protect!certain!electrical!circuits!within!the!GastroCH4ECK®!Gastrolyzer®!against!overcurrent!
conditions.!For!continued!protection!against!the!risk!of!fire,!replace!only!with!the!same!type!and!
rating!specified.!!
Only!fuses!with!the!following!ratings!should!be!applied:!250Vac,!3.15AH!T,!1500A!breaking!capacity.!
The!GastroCH4ECK®!Gastrolyzer®!is!not!designed!for!use!with!materials!capable!of!developing!
flammable!or!explosive!vapours.!Do!not!use/handle/store!such!materials!(such!as!chloroform!or!
ethyl!alcohol)!on!or!within!the!required!30cm!(1ft)!area!surrounding!the!GastroCH4ECK®!
Gastrolyzer®.!!
!
1%8A:#)8:=*&:(%$;**
For!safe!operation!of!the!equipment,!observe!the!following:!!
Use!only!accessories!designed!for!use!with!the!GastroCH4ECK®!Gastrolyzer®.!!
Do!not!lift!or!move!the!GastroCH4ECK®!Gastrolyzer®!while!it!is!running.!This!could!damage!the!
methane!bench.!!!
!
!
!
!

!
!
5!
!
!
Our$family,$innovating$health,$for$yours.$
4#B)7"#?%#$**
The!GastroCH4ECK®!Gastrolyzer®!complies!with!the!directive!EN!60601-1-2!electromagnetic!
compatibility!but!can!be!affected!by!cellular!phones!and!by!electromagnetic!interference!exceeding!
the!levels!specified!in!EN!50082-1.!
Guidance!and!manufacturer’s!declaration:!Electromagnetic!Immunity!(IEC!60601-1-2)!
The!GastroCH4ECK®!Gastrolyzer®!is!intended!for!the!use!in!the!electromagnetic!environment!specified!
below.!The!customer!or!the!user!of!the!GastroCH4ECK®!Gastrolyzer®!should!assure!that!it!is!used!in!
such!an!environment.!
Immunity!test!!
!
IEC!60601!Test!level!!
!
Compliance!level!!
!
Electromagnetic!environment!–!
guidance!!
!
Conducted!immunity!
!
!
!
!
!
Radiated!immunity!
3!V!rms!150!khz!~!
80mhz!outside!ISM!
bands*!!
3V!(1khz!80%)!
150khz!–!80mhz!
!
!
3!V/m!80!mhz!~!2.5!
ghz!!
!
!
3V!(1khz!80%)!
150khz!–!80mhz!
!
!
!
!
!
!
!
!
!
80!mhz!to!2.5!ghz!
3!V/m!
!
!
Portable!and!mobile!RF!
communications!equipment!should!
be!used!no!closer!to!any!part!of!the!
GastroCH4ECK®!Gastrolyzer®!
including!cables,!than!the!
recommended!separation!distance!
calculated!from!the!equation!
appropriate!to!the!frequency!of!the!
transmitter.!!
!
!
Interference!may!occur!in!the!
vicinity!of!equipment!marked!with!
the!following!symbol:!!
!
!
!
0/-4*CD!At!80!mhz!and!800!mhz,!the!higher!frequency!range!applies.!!
0/-4*ED!These!guidelines!may!not!apply!in!all!situations.!Electromagnetic!propagation!is!affected!by!
absorption!and!reflection!from!structures,!objects,!and!people.!
The!ISM!(industrial,!scientific,!and!medical)!bands!between!150!khz!and!80!mhz!are!6.765!mhz!to!6.795!
mhz;!13.553!mhz!to!13.567!mhz;!26.957!mhz!to!27.283!mhz;!and!40.66!mhz!to!40.70!mhz.!!
The!compliance!levels!in!the!ISM!frequency!bands!between!150!khz!and!80!mhz!and!in!the!frequency!
range!80!mhz!to!2.5!ghz!are!intended!to!decrease!the!likelihood!that!mobile/portable!communications!
equipment!could!cause!interference!if!it!is!inadvertently!brought!into!patient!areas.!For!this!reason,!an!
additional!factor!of!10/3!is!used!in!calculating!the!recommended!separation!distance!for!transmitters!in!
these!frequency!ranges.!!
Field!strengths!from!fixed!transmitters,!such!as!base!stations!for!radio!(cellular/cordless)!telephones!
and!land!mobile!radios,!amateur!radio,!AM!and!FM!radio!broadcast!and!TV!broadcast!cannot!be!
predicted!theoretically!with!accuracy.!To!assess!the!electromagnetic!environment!due!to!fixed!RF!
transmitters,!an!electromagnetic!site!survey!should!be!considered.!If!the!measured!field!strength!in!the!
location!in!which!the!GastroCH4ECK®!Gastrolyzer®!is!used!exceeds!the!applicable!RF!compliance!level!
above,!the!GastroCH4ECK®!Gastrolyzer®!should!be!observed!to!verify!normal!operation.!If!abnormal!
performance!is!observed,!additional!measures!may!be!necessary,!such!as!reorienting!or!relocating!the!
GastroCH4ECK®!Gastrolyzer®.!!
Over!the!frequency!ranges!150!khz!to!80!mhz!field!strength!should!be!less!than!3!V/m.!

!
!
6!
!
!
Our$family,$innovating$health,$for$yours.$
*
!+,-./0D*If!the!GastroCH4ECK®!Gastrolyzer®!is!used!in!a!manner!other!than!that!specified!in!this!
manual,!the!safety!and!performance!of!this!equipment!could!be!impaired.!Further,!the!use!of!any!
equipment!other!than!that!recommended!by!Bedfont®!has!not!been!evaluated!for!safety.!Use!of!any!
equipment!not!specifically!recommended!in!this!manual!is!the!sole!responsibility!of!the!user.!*
!+,-./0D*Degree!of!safety!of!application!in!the!presence!of!a!flammable!anaesthetic!mixture!with!
air,!oxygen!or!nitrous!oxide:!Equipment!not!suitable!for!use!in!the!presence!of!flammable!mixtures.*
!+,-./0D*Do!not!place!the!GastroCH4ECK®!Gastrolyzer®!near!areas!containing!flammable!reagents!
or!combustible!fluids.!Vapours!from!these!materials!could!enter!the!GastroCH4ECK®!Gastrolyzer®!air!
system!and!be!ignited.!!
!+,-./0D*Do!not!apply!pressures!greater!than!4!bar!to!the!direct!sample!line!connector,!at!a!
maximum!flow!rate!of!3!LPM.!!Doing!so!could!result!in!the!disconnection!of!internal!tubing!to!
protect!against!sensor!damage.!
!
9:(%$;*)#("7?:$)"#*:#5*5%B)8%*&;?F"=&**
'%<7%%*"(*G7"$%8$)"#*:<:)#&$*%=%8$7)8*&A"8@*
Type!BF!applied!part!
-;G%*"(*G7"$%8$)"#*:<:)#&$*%=%8$7)8*&A"8@*
Class!I!equipment;!(earthed)!
'%<7%%*"(*G7"$%8$)"#*:<:)#&$*)#<7%&&*"(*=)H6)5*
IPX0!–!not!protected!against!water!
ingress!
'%<7%%*"(*&:(%$;*:GG=)8:$)"#*)#*$A%*G7%&%#8%*"(*:*
(=:??:F=%*:#:%&$A%$)8*?)I$67%*J)$A*:)7K*"I;<%#*"7*
#)$7"6&*"I)5%*
Equipment!not!suitable!for!use!in!the!
presence!of!flammable!mixtures.!
!"#&6=$*)#&$768$)"#&*("7*6&%**
!
+=$%7#:$)#<*8677%#$*
!
!4*?:7@**
!
-;G%*LM*:GG=)%5*G:7$*
!
')&G"&%*"(*:88"75)#<*$"*N444*
!
9%7):=*#6?F%7*
!
1:#6(:8$67%5*F;*:#5*5:$%*
!
L%5("#$O*="<"*
!
!
!

!
!
7!
!
!
Our$family,$innovating$health,$for$yours.$
1"#)$"7*)#&$:==:$)"#*
Maintain!a!30cm!(1ft)!clearance!envelope!above!the!GastroCH4ECK®!Gastrolyzer®!while!it!is!running.!
The!equipment!should!not!be!positioned!in!such!a!way!that!it!is!difficult!to!operate!the!
disconnection!device!(mains!plug).!If!it!is!necessary!to!move!the!GastroCH4ECK®!Gastrolyzer®,!then!
the!following!conditions!need!to!be!maintained:!
1. Select!a!location!away!from!heat-producing!laboratory!equipment,!with!sufficient!ventilation!
to!allow!heat!dissipation.!!
2. Position!the!GastroCH4ECK®!Gastrolyzer®!on!a!level!surface,!such!as!a!sturdy!table!or!
laboratory!bench!that!can!support!the!weight!of!the!GastroCH4ECK®!Gastrolyzer®,!8!kg,!and!
resist!vibration.!
3. Place!the!GastroCH4ECK®!Gastrolyzer®!at!least!5!cm!(2!in)!from!the!front!edge!of!the!surface.!!
4. In!addition!to!space!for!the!GastroCH4ECK®!Gastrolyzer®,!allow!15cm!(6in)!clearances!at!the!
sides!and!back!to!ensure!sufficient!air!circulation.!The!GastroCH4ECK®!Gastrolyzer®!must!
have!adequate!air!ventilation!to!ensure!compliance!to!local!requirements!for!vapours!
produced!during!operation.!!
5. Relative!humidity!should!not!exceed!75%!(non-condensing).!
1"#)$"7*&%$P6G*
1. Connect!the!power!lead!to!the!GastroCH4ECK®!Gastrolyzer®!via!the!socket!on!the!rear!of!the!
panel!and!connect!to!the!electricity!supply:!!
!
2. Switch!the!monitor!on!using!the!on/off!switch!on!the!back!of!the!unit.!The!screen!will!light!
up!and!the!unit!will!proceed!to!the!warm!up!screen.!!If!any!warning!screens!are!displayed,!
these!should!be!checked!in!the!warnings!and!errors!section!of!the!manual.!The!
GastroCH4ECK®!Gastrolyzer®!monitor!warm!up!takes!30!minutes,!see!‘temperature!
requirements’,!and!a!countdown!timer!will!be!shown!onscreen!for!the!duration!of!the!warm!
up!as!shown:!
0/-4D!The!warm-up!time!can!be!shortened!if!the!unit!is!already!warmed-up.!

!
!
8!
!
!
Our$family,$innovating$health,$for$yours.$
3. 5!minutes!before!the!completion!of!the!warm!up,!the!GastroCH4ECK®!Gastrolyzer®!will!begin!
a!purge!cycle!to!clear!any!residual!gas!in!the!tubing.!
4. Connect!the!sample!line!to!the!direct!sample!line!connector!for!breath!samples,!ensuring!the!
connecter!‘clicks’!fully!into!place!or!push!the!moisture!filter!onto!the!bag!sample!connector!
for!breath!bag!sampling.!!
!
-%?G%7:$67%*7%H6)7%?%#$&*:#5*5%B)8%*J:7?*6G*
Temperature!should!not!exceed!those!stated!in!the!specifications!section!of!this!document.!The!
monitor!should!be!calibrated!at!the!same!temperature!at!which!it!would!be!used,!ideally!23°c!±3°c!
(69.8°F!±!5.4°F).!!
0/-4D*At!25°c!ambient!temperature!the!expected!lifetime!of!the!unit!is!at!least!30,000!hrs.!For!every!
10°c!rise!in!temperature,!the!lifetime!is!reduced!by!half.!
The!unit!requires!a!warm!up!time!of!30!minutes!before!it!can!be!used.!After!2!minutes,!the!unit!
periodically!does!a!self-check!in!order!to!know!if!it!is!ready!to!be!used.!In!the!case!it!is!detected!to!be!
ready,!a!bypass!option!is!provided.This!bypass!should!only!be!used!if!the!unit!is!accidently!switched!
off!or!needs!to!be!moved!after!it!has!already!warmed!up.!The!bypass!must!be!done!within!2!minutes!
of!the!unit!being!switched!off,!as!using!the!device!without!the!correct!warm-up!time!will!
compromise!accuracy.!The!bypass!is!carried!out!by!pressing! !on!the!warm-up!screen.!You!will!be!
asked!to!confirm!this!decision!by!again!pressing! !on!the!next!screen;!this!will!take!you!to!the!
device’s!home!screen.!!
0/-4D*The!instrument!should!be!calibrated!prior!to!first!use,!after!transportation!and!every!4!weeks.*
!
2%7("7?)#<*:*F7%:$A*$%&$*
1. Once!the!warm!up!has!been!completed,!attach!the!sample!line!and!a!breath!sampling!
mouthpiece.!!
2. Press !to!begin.!
3. A!timer! !will!be!shown!onscreen,!whilst!the!sensors!are!zeroing.!This!may!take!up!to!75!
seconds!if!a!test!has!just!been!carried!out.!
4. If!you!are!using!GastroCHART™!to!record!the!readings,!make!sure!you!are!on!the!patient’s!
profile!and!click!‘capture!reading’!now.!
5. Inhale!when!prompted!by!and!hold!breath!for!the!pre-set!15!second!countdown,!as!
shown!on!screen.!!
6. A!beep!will!sound!during!the!last!three!seconds!of!the!countdown .!!

!
!
9!
!
!
Our$family,$innovating$health,$for$yours.$
7. Blow!slowly!into!mouthpiece!when!prompted!by! .!
8. An!indicator!on-screen!will!illustrate!how!hard!to!blow ;!keep!the!arrow!pointing!in!
the!green!section!of!this!indicator!throughout!the!test.!!
9. The!arrow!will!change!colour!as!the!O2!level!in!the!breath!sample!reduces!to!the!target!15%,!
at!which!point!it!will!turn!green!and!the!test!will!automatically!stop!after!3!seconds!
.!
10. If!the!patient!feels!they!are!unable!to!exhale!for!any!longer!and!the!test!has!not!finished!
they!can!end!the!test!at!any!time!by!pressing! .!
11. The!final!results!will!be!shown!on!screen! .!
12. Once!the!test!is!complete,!the!final!O2!concentration!can!be!observed!by!pressing!! .!This!
will!show!the!O2!concentration!in!%!and!the!correction!factor!applied!to!the!reading!if!it!was!
required.!A!visual!indicator!will!help!interpretation!of!the!result:!
a. !Good!test,!little!or!no!correction!factor!applied!
b. Satisfactory!test,!correction!factor!applied!
c. Poor!test,!high!correction!factor!applied!and!another!test!is!recommended!
13. Press! !to!return!to!the!results!screen.!
14. When!you!are!finished,!press! !to!return!to!the!home!screen.!0/-4D!this!will!lose!the!
reading;!please!ensure!you!have!downloaded!the!results!to!GastroCHART™!or!recorded!
them!manually.!!
15. Remove!the!mouthpiece!and!sample!line!between!tests!to!purge!sensors!with!fresh!air.!
!
L7%:$A*F:<*&:?G=%*8"==%8$)"#**
1. Place!two!bag!clamps!over!the!inlet!and!outlet!tubes!of!the!breath!bag,!but!keep!them!open!
!
2. Insert!a!breath!bag!mouthpiece!into!the!bag!inlet!tube!of!the!breath!bag!
!
3. With!both!bag!clamps!open,!exhale!gently!through!the!bag!at!a!steady!rate!
4. Continue!exhaling!until!you!feel!you!have!nearly!emptied!your!lungs!–!this!clears!the!dead!
air!in!the!upper!airspace!of!the!lungs!in!order!to!achieve!a!good!end-tidal!sample!

!
!
10!
!
!
Our$family,$innovating$health,$for$yours.$
5. As!soon!as!you!feel!your!lungs!are!nearly!empty,!and!end-tidal!breath!has!almost!been!
reached,!engage/close!the!clamp!at!the!outlet!of!the!bag!
!
6. Whilst!still!maintaining!a!steady!flow,!fill!up!the!breath!bag!COMPLETELY!with!your!end!tidal!
breath,!ensuring!you!do!not!inhale!whilst!doing!this!
!
7. Engage!the!clamp!at!the!inlet!of!the!bag.!
8. Insert!the!blue!plastic!plugs!in!each!end!of!the!breath!bag!to!ensure!no!breath!is!lost!from!
the!sample.!!
Following$these$steps$should$ensure$that$a$valid$sample$has$been$collected.$
$
+#:=;&)#<*:*F7%:$A*F:<*$%&$*
1. Once!warm!up!has!been!completed,!attach!the!moisture!filter!to!the!device.!
2. Press! to!start!the!bag!test!procedure.!
3. The!device!will!then!illustrate!to!attach!the!breath!bag.!Ensure!to!carefully!remove!the!blue!
plug!on!one!end!only!and!attach!the!bag!to!the!moisture!filter,!then!open!the!clamp!on!that!
end!only!to!preserve!the!sample.!
4. Press! !to!begin.!!
5. A!timer! !will!be!shown!onscreen,!whilst!the!sensors!are!zeroing.!This!may!take!up!to!75!
seconds!if!a!test!has!just!been!carried!out.!
6. If!you!are!using!GastroCHART™!to!record!the!readings,!make!sure!you!are!on!the!patient’s!
profile!and!click!‘capture!reading’!now.!
7. Once!zeroed,!breath!bag!sampling!will!begin!automatically!–!make!sure!that!the!clamp!is!
open.!At!this!point,!the!device!needs!no!further!intervention!until!a!reading!is!recorded.!
8. The!device!will!draw!the!breath!sample!for!analysis;!this!will!take!45!seconds,!as!indicated!on!
the!screen.!
9. The!final!results!will!be!shown!on!screen .!
10. Once!the!test!is!complete,!the!final!O2!concentration!can!be!observed!by!pressing!! .!This!
will!show!the!O2!concentration!in!%!and!the!correction!factor!applied!to!the!reading!if!it!was!
required.!A!visual!indicator!will!help!interpretation!of!the!result:!
a. !Good!test,!little!or!no!correction!factor!applied!

!
!
11!
!
!
Our$family,$innovating$health,$for$yours.$
b. Satisfactory!test,!correction!factor!applied!
c. Poor!test,!high!correction!factor!applied!and!another!test!is!recommended!
11. Press! !to!return!to!the!results!screen.!
12. When!you!are!finished,!press! !to!return!to!the!home!screen.!0/-4D!this!will!lose!the!
reading;!please!ensure!you!have!downloaded!the!results!to!GastroCHART™!or!recorded!
them!manually.!
!
.#&$76?%#$*=:;"6$*
!
!
!
9$:#5:75*G:8@*8"#$%#$&*
1. GastroCH4ECK®!Gastrolyzer®!monitor!
7. Direct!sample!line!
2. Operating!manual!
8. Quick!start!guide!
3. Mains!cable!
9. Patient!preparation!guide!
4. GastroCHART™!software!on!USB!!
10. Interpretation!wheel!
5. USB!cable!
11. Ledochowski!book!
6. Moisture!filter!
!
!
-%8A#)8:=*&G%8)()8:$)"#*
!
Q:&$7"!RS4!TO*Q:&$7"=;U%7O*

!
!
12!
!
!
Our$family,$innovating$health,$for$yours.$
!"#8%#$7:$)"#*7:#<%**
!RS*
0-200!ppm!
RE*
0-200!ppm!!
/E*
14-23%!
2"J%7*)#G6$**
230V/100V,!50Hz!–!60!Hz,!0.5!–!1.0!A!
M6&%*
T!3.15!AH!
N:7?P6G*$)?%*
30!minutes!
!:=)F7:$)"#*(7%H6%#8;**
Every!4!weeks!!
')&G=:;*
Full!colour!touchscreen!
'%$%8$)"#*G7)#8)G=%*
Electrochemical!sensor!(O2!&!H2)!
Infrared!(CH4)!
*
*
*
*
+8867:8;**
*
!RS*
3%&"=6$)"#*
1!ppm!
+8867:8;*
±!10%!of!reading!!
3%G%:$:F)=)$;*
<5%!difference!on!consecutive!readings!
*
RE*
3%&"=6$)"#*
1!ppm!
+8867:8;*
±!10%!of!reading!!
3%G%:$:F)=)$;*
<5%!difference!on!consecutive!readings!
*
/E*
3%&"=6$)"#*
0.1%!
+8867:8;*
±!2.5%!of!reading!!
3%G%:$:F)=)$;*
<5%!difference!on!consecutive!readings!
!:7F"#*?"#"I)5%*87"&&P&%#&)$)B)$;*VRE*"#=;W*
<1%!
-%?G%7:$67%*7:#<%*
/G%7:$)#<*
15-35°C!(59°-!95°F)!
9$"7:<%*
0-40°C!(32°-!104°F)!
27%&&67%*7:#<%**
/G%7:$)#<*
900-1100mbar!
9$"7:<%*
800-1200mbar!
R6?)5)$;*7:#<%*
/G%7:$)#<*
30-75%!non-condensing!
9$"7:<%*
15-90%!non-condensing!
*
9%#&"7*"G%7:$)#<*=)(%*
!RS*
5!years!
RE*
2!years!
/E*
2!years!
')?%#&)"#&*
Approx.!474!x!310!x!167!mm!
N%)<A$*
Approx.!8kg!
1"#)$"7*8"#&$768$)"#*
Case:!aluminium!
!=:&&)()8:$)"#*
Class!I!ME!equipment:!(externally!powered)!!
Type!BF!applied!part!!
Method!of!sterilization!(not!suitable!for!
sterilization)!!
Not!suitable!for!use!in!an!oxygen!rich!
environment!!
Intended!for!continuous!use!
3%&G"#&%*$)?%**
≤45!seconds!
N:77:#$;*
Q:&$7"!RS4!TO*J)$A"6$*
&%#&"7&*
2!years!
!RSK*RE*:#5*/E*&%#&"7&*
1!year!
')&G=:;*&;?F"=&*
'%&87)G$)"#*
Q:&$7"!RS4!TO*Q:&$7"=;U%7O*

!
!
13!
!
!
Our$family,$innovating$health,$for$yours.$
L7%:$A*$%&$**
!
.#A:=%*
!
R"=5*F7%:$A*
!
!"6#$5"J#*VF7%:$A*A"=5W*
!
4IA:=%*
!
GG?*7%:5)#<**
!
/E*7%:5)#<**
!
!"77%8$)"#*(:8$"7*
!
!A%8@*$%&$*H6:=)$;*
!
9%$$)#<&*
!
R"?%**
!
4IA:=:$)"#*&G%%5*<:6<%*
!
Q""5*$%&$*V=)$$=%*"7*#"*8"77%8$)"#*
(:8$"7*:GG=)%5W*
!
9:$)&(:8$"7;*$%&$*V8"77%8$)"#*(:8$"7*
:GG=)%5W*
!
2""7*$%&$*VA)<A*8"77%8$)"#*(:8$"7*
:GG=)%5*:#"$A%7*$%&$*)&*
7%8"??%#5%5W*
!
L7%:$A*F:<*$%&$*
!

!
!
14!
!
!
Our$family,$innovating$health,$for$yours.$
L:<*&:?G=%*8"6#$5"J#*
!
0%I$*&$%G*
!
27%B)"6&*&$%G*
!
9%#&"7*8A:#<%*56%*&""#*
!
9%#&"7*8A:#<%*"B%756%*
!
'%B)8%*8:=)F7:$)"#*56%*
!
3%&%$*8:=)F7:$)"#*
!
+7%*;"6*&67%*;"6*J:#$*$"*7%&%$X*
!
!"#()7?*7%&%$*
!
9$:7$*8:=)F7:$)"#*G7"8%567%*
!
'%B)8%*U%7")#<*&%#&"7&**
!
!A%8@*8:=)F7:$)"#*<:&*
7%H6)7%?%#$&**
!

!
!
15!
!
!
Our$family,$innovating$health,$for$yours.$
+$$:8A*7%<6=:$"7*$"*<:&*8:#)&$%7*
!
+$$:8A*8:=)F7:$)"#*:5:G$"7*$"*
5%B)8%*:#5*$67#*"#*<:&*(="J*
!
9;&$%?*G7"8%&&)#<*
!
!:=)F7:$)"#*&688%&&(6=*
!
!:=)F7:$)"#*(:)=%5*
!
3%$7;*8:=)F7:$)"#>U%7"*
!
9%7B)8%*:7%:*VG)#*G7"$%8$%5W*
!
9%#&"7*&%7B)8)#<**
!
-7"6F=%&A""$)#<**
!
'):<#"&$)8&**
!
!:=)F7:$)"#*&%7B)8)#<**
!
45)$*5:$%>$)?%*
!
':;P?"#$AP;%:7*
!
1"#$AP5:;P;%:7*
!
ES*A"67*("7?:$*
!
CE*A"67*("7?:$*
!

!
!
16!
!
!
Our$family,$innovating$health,$for$yours.$
RE*&%#&"7**
!
/E*&%#&"7*
!
!AS*&%#&"7*
!
RE*&%#&"7*U%7"*(:)=67%*
!
/E*&%#&"7*U%7"*(:)=67%*
!
!AS*&%#&"7*U%7"*(:)=67%*
!
!AS*&%#&"7*(:)=67%**
!
9%#&"7*)#&$:==:$)"#*5:$%*
!
9%#&"7*7%#%J:=*5:$%*
!
9%#&"7*7%#%J:=*5:$%*"B%756%*
!
9:B%*&%$$)#<&*
!
M)7?J:7%*B%7&)"#*
!
9%7):=*#6?F%7*
!
!
N:7#)#<&*:#5*?:)#$%#:#8%*
1. Mouthpieces!should!be!replaced!after!every!patient.!!
2. Hands!should!be!washed!regularly!in!accordance!with!infection!control!practice.!!
3. Please!do!not!attempt!to!modify!the!equipment!in!any!way!or!use!accessories!not!specified!
by!the!manufacturer.!Any!attempt!to!do!so!will!invalidate!the!warranty!and!may!
compromise!the!safety!of!the!device.!!
4. Bedfont®!will!make!available,!upon!request,!service!training!to!appropriately!qualified!
persons.!
V2.23

!
!
17!
!
!
Our$family,$innovating$health,$for$yours.$
5. Pressing!the!reset!button! !will!prompt!the!device!to!calibrate!without!taking!into!
account!the!sensor’s!safety!windows.!You!will!be!asked!to!confirm!your!choice;!click!to!
confirm.!If!completed,!new!safety!windows!and!calibration!coefficients!are!saved.!The!next!
calibration!needed!is!in!4!weeks.!To!abort,!click! .!Aborting!the!reset!calibration!
before!the!gas!is!sampled!will!revert!the!calibration!back!to!the!previous!settings.!!
! !

!
!
18!
!
!
Our$family,$innovating$health,$for$yours.$
0/-4:!Only!available!if!the!unit!has!been!on!for!1!hour,!from!switch!on.!If!not,!the!reset!
button!is!in!grey! !
6. The!GastroCH4ECK®!should!we!switched!off!once!every!24!hours.!
!
!=%:#)#<*
1. Bedfont®!recommends!wiping!the!instrument’s!external!surfaces!with!a!product!specifically!
developed!for!this!purpose.!Bedfont®!provides!instrument!cleaning!wipes.!The!device!or!
consumables!cannot!be!sterilised.!
2. 04Y43!use!alcohol!or!cleaning!agents!containing!alcohol!or!other!organic!solvents!as!long!
term!exposure!to!these!vapours!will!damage!the!sensors!inside.!!
3. Under!no!circumstances!should!the!instrument!be!immersed!in!liquid!or!splashed!with!
liquid.!
!
3"6$)#%*?:)#$%#:#8%*
1. The!device!must!be!calibrated!once!every!4!weeks!with!a!Bedfont®!calibration!kit!and!
associated!gas,!see!‘calibration’.!
2. Bedfont®!recommends!an!annual!service!in!order!to!check!sensor!and!component!parts!
performance.!
3. The!H2!sensor!should!be!replaced!every!two!years.!
4. The!O2!sensor!should!be!replaced!every!two!years.!
5. The!CH4!sensor!should!be!replaced!every!five!years.!
6. The!monitor!will!provide!daily!reminders,! ,!via!the!reminder!screen!after!turning!on!the!
GastroCH₄ECK®!Gastrolyzer®!for!1!month!leading!up!to!the!sensor!change!due!date.!!
7. Once!this!date!has!elapsed!the!date!will!turn!red!to!indicate!that!sensor!replacement!is!
overdue.!
.12/3-+0-D!This!may!compromise!result!accuracy.!!
8. Additional!technical!information!can!be!made!available!on!request;!please!contact!Bedfont®!
or!its!distributor.!
!:=)F7:$)"#*
The!GastroCH4ECK®*Gastrolyzer®!is!calibrated!at!21°c!+3°c!(69.8°F!+5.4°F)!before!leaving!Bedfont®.!
The!instrument!should!be!calibrated!at!21°c!+3°c!(69.8°F!+5.4°F)!as!this!is!the!temperature!at!which!
we!recommend!it!is!to!be!used.!The!GastroCH4ECK®*Gastrolyzer®!should!be!calibrated!every!4!weeks!
and!will!give!a!reminder!during!start-up!when!calibration!is!due.!The!calibration!gas!required!is!
100ppm!H2,!100ppm!CH4!and!20.9%!O2.!!

!
!
19!
!
!
Our$family,$innovating$health,$for$yours.$
1. Make!sure!the!monitor!is!plugged!in!at!the!mains!and!switched!on!at!the!back.*0/-4D!
calibration!is!only!possible!if!the!unit!has!been!on!for!1!hour,!from!switch!on.!If!not,!a!red!
crossed!circle!is!displayed!on!top!of!the!image! .!
2. Press! .!
3. Start!the!calibration!process!by!pressing! .!
4. First!the!monitor!must!be!zeroed!and!this!will!happen!automatically,!showing!this!symbol!
.!Once!the!calibration!process!has!started,!do!not!connect!the!gas!at!this!stage.!
5. Once!this!has!been!successfully!carried!out,!the!first!step!of!the!calibration!process!will!be!
shown!onscreen!indicating!for!you!to!use!the!correct!gas!mixture! .!Then!select!the!
arrow!at!the!bottom!right!of!the!screen!to!move!to!the!next!instruction.!!
6. The!next!instruction!on!screen!will!ask!you!to!attach!the!regulator!to!the!gas!cylinder,!as!
shown! .!Then!select!the!arrow!at!the!bottom!right!of!the!screen!to!move!to!the!next!
instruction.!
7. The!next!instruction!onscreen!will!indicate!for!you!to!attach!the!calibration!adaptor!to!the!
end!of!the!direct!sample!line!and!then!open!the!gas!regulator .!
Select!the!arrow!at!the!bottom!right!hand!side!of!the!screen!to!start!the!calibration!process!!
Z[ Q:&*(="J*&A"6=5*F%*&%$*$"*C*=)$7%*G%7*?)#6$%[**
9. A!successful!calibration!will!be!indicated!by!the ;!press! !to!return!to!the!home!screen.!
10. A!failed!calibration!will!be!indicated!by ;!press! !to!attempt!calibration!again.!If!the!
problem!persists!see!‘troubleshooting’!or!call!your!local!supplier!of!Bedfont®!products.!!
Return!to!the!home!screen!by!pressing! .!
Q:&$7"!R+3-\*&"($J:7%*]*8"##%8$)#<*$"*$A%*2!*
The!GastroCHART™!software!can!be!downloaded!and!installed!from!the!USB!that!comes!with!the!
GastroCH4ECK®!Gastrolyzer®.!Place!one!end!of!the!connection!lead!into!the!USB!socket!at!the!rear!of!
the!GastroCH4ECK®*Gastrolyzer®,!connecting!the!other!end!to!the!USB!port!on!the!PC.!Before!
starting!the!software,!ensure!that!the!GastroCH4ECK®*Gastrolyzer®!is!connected!to!the!PC!and!
switched!on.!Double!click!the!GastroCHART™!icon!on!the!PC!to!start!the!programme.!Refer!to!the!
supplied!documentation!for!how!to!operate!GastroCHART™.!(GastroCHART™!is!also!available!to!
download!for!M344!at!www.bedfont.com)!