30 Washington Ave Ste D Haddonfield, NJ 08033
tel (800)707-5751 ●fax (800)721-2377 ●service@covita.net ●www.covita.net
1. Ensure you are performing a calibration in a fresh air environment.
2. Turn Micro+ ON by holding the on/off button down for three seconds.
3. Main Screen
a. On the main screen touch the gears icon.
4. Gears Screen
a. Touch the calibration icon to enter the calibration screen.
5. Calibration Mode
a. Touch the zero icon to zero the monitor. The monitor will
display an hour glass and then a check mark. Touch the
check mark to return back to the calibration screen.
b. Touch the calibration icon. (Note: the canister is no longer
crossed out)
c. Immediately turn on the gas by slowly turning the valve. Adjust the
flow to 1.0 lpm by balancing the silver ball on the middle line..
(Note: the valve is sensitive, so a small/delicate turn is all that is
necessary)
d. The monitor’s reading will slowly climb from zero and a spinning
hour glass will be shown on the screen as well.
e. When you receive a check mark, turn OFF the gas.
f. Touch the check mark to lock in calibration & to return
previous menu.
g. Touch the home icon or return arrow to return to the main
screen.
h. Calibration Complete
Note: If you receive and X instead of a check mark
please contact coVita for further assistance.