1. Ensure you are performing a calibration in a fresh air environment.
2. Turn piCO+™ON by holding the button down for three seconds.
3. Main Screen
a. Single click the button to move the pointer to
the calibration canister icon.
b. Next, double click the button to select the
calibration canister icon.
4. Calibration Mode
a. Double click button to select the canister icon.
b. Immediately turn on the gas by turning the valve.
No fine tuning is required because this regulator is fixed at 1 lpm.
c. The monitor’s reading will slowly climb from zero and
an hour glass will be shown on the screen as well.
d. When you receive a check mark, turn OFF the gas.
e. Double click the button to select return arrow and to go previous
menu.
f. Double click the button to select the return arrow and return to
main menu.
g. Calibration Complete
Note: If you receive and instead of a please contact coVita for
further assistance.