As prompted, aach the regulator to the
piCO™ monitor and follow the onscreen
instrucons to run the calibraon
Open the gas regulator
to perform the test
Close the regulator when the calibraon is nished
To perform a calibraon test, rstly turn on the piCO™ monitor Press the sengs icon to open the sengs menu
Press the calibraon icon 5 mes to unlock the opon then
press the icon once more to start the calibraon
Allow the piCO™ monitor to zero. If the monitor is too cold or hot
please wait for it to come to ambient temperature before connuing
If the the piCO™ failed its calibraon a red cross will show,
press the retry icon to try again. If unsuccessful consult the troubleshoong
secon of the manual or contact your local distributor for help
If the piCO™ successfully completed its calibraon a
green ck will show.
The piCO™ should be wiped clean
with the alcohol-free anbacterial wipes provided
with the monitor before and aer each test
A new SteriBreath™ mouthpiece should be
used for each paent. It is best pracce for
the paent to insert their own mouthpiece
Follow the instrucons
on the piCO™ to complete
a breath test
The paent will be
prompted to inhale and
hold their breath
There will be a 15 second
countdown
The paent will be prompted to exhale
slowly into the monitor, and should aim
to completely empty their lungs
The results will rise and stabilise
onscreen
Allow the paent to throw away their
used mouthpiece
Results can be interpreted using the
Smokerlyzer® CO Chart
The piCO™ should be wiped clean with the
alcohol-free anbacterial wipes provided
with the monitor before and aer each test
Bedfont® Scienc Ltd
Staon Road, Harrietsham, Maidstone, Kent,
ME17 1JA, England
Tel: +44 (0)1622 851122, Fax: +44 (0)1622 854860
Email: ask@bedfont.com, Web: www.bedfont.com
Issue 1 - November 2020, Part No: MKT528
Bedfont® Scienc Limited reserve the right to change or update this literature without
prior noce. Registered in: England and Wales. Registered No: 1289798
The external surfaces of the D-piece™
lter should also be wiped down before and
aer each test
D-piece™ should be replaced every 30 days
The external surfaces of the D-piece™ lter should
also be wiped down before and aer each test
D-piece™ should be replaced every 30 days
Bedfont® Scienc Ltd. has put together some advice on how to safely perform breath tests during the pandemic, however,
rst and foremost, it is recommended that you follow your local guidelines on breath tesng procedures and necessary PPE
How to perform a breath test:
How to perform a calibraon:
Bedfont® GmbH
Mierbachweg 18, 5081, Anif, Austria
Tel: +43 662 635050 Fax: +43 662 635050-50
Email: bedfont-gmbh@bedfont.com