Bedfont Micro Plus Smokerlyzer User manual

One CO monitor for all your smoking cessaon needs.
Infecon control plays an important role in the medical device industry, but now more than ever, we must take extra
precauon to ensure the safety of healthcare professionals and paents alike when performing CO breath tests to help
people stop smoking.
This guide has been put together to help you safely resume CO monitoring.
Contents
How to perform a bump test with a non-smoker: 2
How to perform a bump test with veried gas: 3
How to perform a calibraon: 4
How to perform a breath test: 5
Follow the instrucons
on the Micro⁺™ to complete
a breath test
The praconer will be
prompted to inhale and
hold their breath
There will be a 15 second
countdown
The praconer will be prompted to
exhale slowly into the monitor, and
should aim to completely empty their
lungs
Prior to undertaking your rst CO test, please check the Micro⁺™ Smokerlyzer® is reading correctly by performing a Bump Test.
This can be done by performing a breath test yourself or with a known non-smoker, or by using veried gas.
Press the breath test icon to begin
If the monitor is reading correctly, a non-smoker reading will be shown
onscreen; this result should be green and under 6ppm.
If the reading comes back a lile high, it could be due to environmental
CO; please verify the Smokerlyzer® with a second non-smoker.
If the reading is too high, you might need to calibrate your machine.
Aach the regulator to the Micro⁺™ monitor
If it’s been a while since you’ve used your Micro⁺™,
please check the baeries have charge.
Replace them if necessary
As roune maintenance, a bump test should always be performed aer
long periods of no use to check the device is reading correctly. You can
perform a bump test using either a non-smoker or cered gas to ensure
the device is reading correctly.
Once the countdown has nished and the screen prompts
to exhale, open the gas regulator to perform the bump test
Start the breath test on the Micro⁺™ monitor
Once the reading has stabilised on screen
close the regulator to turn o the gas
Make sure the Micro⁺™ monitor is reading
between 15 and 22 ppm
If the Micro⁺™ is not reading between this range,
please calibrate the monitor
To perform a bump test using gas, rstly aach the gas canister
to the gas regulator by screwing the canister into the regulator
How to perform a bump test with veried gas:How to perform a bump test with a non-smoker:
Start the breath test on the piCO™ monitor
1
2
3
21
6
9
7
5
4
3
9
As prompted, aach the regulator to the
Micro⁺™ monitor and follow the onscreen
instrucons to run the calibraon
Open the gas regulator
to perform the test
Close the regulator when the calibraon is nished
To perform a calibraon test, rstly turn on the Micro⁺™ monitor Press the sengs icon to open the sengs menu
Press the calibraon icon 5 mes to unlock the opon then
press the icon once more to start the calibraon
Allow the Micro⁺™ monitor to zero. If the monitor is too cold or hot
please wait for it to come to ambient temperature before connuing
If the the Micro⁺™ failed its calibraon a red cross will show,
press the retry icon to try again. If unsuccessful consult the troubleshoong
secon of the manual or contact your local distributor for help
If the Micro⁺™ successfully completed its calibraon a
green ck will show.
The Micro⁺™ should be wiped clean
with the alcohol-free anbacterial wipes provided
with the monitor before and aer each test
A new SteriBreath™ mouthpiece should be
used for each paent. It is best pracce for
the paent to insert their own mouthpiece
Follow the instrucons
on the Micro⁺™ to complete
a breath test
The paent will be
prompted to inhale and
hold their breath
There will be a 15 second
countdown
The paent will be prompted to exhale
slowly into the monitor, and should aim
to completely empty their lungs
The results will rise and stabilise onscreen Allow the paent to throw away their used mouthpiece
Results can be interpreted using the
Smokerlyzer® CO Chart
The Micro⁺™ should be wiped clean with the
alcohol-free anbacterial wipes provided
with the monitor before and aer each test
The external surfaces of the D-piece™
lter should also be wiped down before and
aer each test
D-piece™ should be replaced every 30 days
The external surfaces of the D-piece™ lter should
also be wiped down before and aer each test
D-piece™ should be replaced every 30 days
Bedfont® Scienc Ltd. has put together some advice on how to safely perform breath tests during the pandemic, however,
rst and foremost, it is recommended that you follow your local guidelines on breath tesng procedures and necessary PPE
How to perform a breath test:
How to perform a calibraon:
Bedfont® Scienc Ltd
Staon Road, Harrietsham, Maidstone, Kent,
ME17 1JA, England
Tel: +44 (0)1622 851122, Fax: +44 (0)1622 854860
Email: ask@bedfont.com, Web: www.bedfont.com
Issue 2 - February 2021, Part No: MKT529
Bedfont® Scienc Limited reserve the right to change or update this literature without
prior noce. Registered in: England and Wales. Registered No: 1289798
Bedfont® GmbH
Mierbachweg 18, 5081, Anif, Austria
Tel: +43 662 635050 Fax: +43 662 635050-50
Email: bedfont-gmbh@bedfont.com
1 2
3 4
576
98
23
1
4
6
9
8
7
5

Other manuals for Micro Plus Smokerlyzer

1

Other Bedfont Test Equipment manuals

Bedfont ToxCO User manual

Bedfont

Bedfont ToxCO User manual

Bedfont Smokerlyzer Range User manual

Bedfont

Bedfont Smokerlyzer Range User manual

Bedfont GastroCH4ECK Gastrolyzer User manual

Bedfont

Bedfont GastroCH4ECK Gastrolyzer User manual

Bedfont GastroCH4ECK Gastrolyzer User manual

Bedfont

Bedfont GastroCH4ECK Gastrolyzer User manual

Bedfont Smokerlyzer COmpact User manual

Bedfont

Bedfont Smokerlyzer COmpact User manual

Bedfont NOBreath User manual

Bedfont

Bedfont NOBreath User manual

Bedfont NOBreath User manual

Bedfont

Bedfont NOBreath User manual

Popular Test Equipment manuals by other brands

Sourcetronic ST9010 Series user manual

Sourcetronic

Sourcetronic ST9010 Series user manual

Aventics Essential-Test-Box operating manual

Aventics

Aventics Essential-Test-Box operating manual

sauter HMM instruction manual

sauter

sauter HMM instruction manual

Artos PT-5 owner's manual

Artos

Artos PT-5 owner's manual

Bibby Sterilin TECHNE Tecal 140F Operator's manual

Bibby Sterilin

Bibby Sterilin TECHNE Tecal 140F Operator's manual

Beijing TIME 5100 instruction manual

Beijing

Beijing TIME 5100 instruction manual

Fluke 707 instruction sheet

Fluke

Fluke 707 instruction sheet

Time Electronics 1068 user manual

Time Electronics

Time Electronics 1068 user manual

PCE Instruments PCE-SC 42 manual

PCE Instruments

PCE Instruments PCE-SC 42 manual

Integrity NexPore CF 7.2 instruction manual

Integrity

Integrity NexPore CF 7.2 instruction manual

Martel LC100 manual

Martel

Martel LC100 manual

Proceq Zonotip Service manual

Proceq

Proceq Zonotip Service manual

AMS SCRAM Quick reference guide

AMS

AMS SCRAM Quick reference guide

inTest Temptronic ThermoStream TP04300 Series Operator's manual

inTest

inTest Temptronic ThermoStream TP04300 Series Operator's manual

Owon PDS5022S user manual

Owon

Owon PDS5022S user manual

Dymond RS Milzar Prestige II Series owner's manual

Dymond

Dymond RS Milzar Prestige II Series owner's manual

Fluke CableIQ user manual

Fluke

Fluke CableIQ user manual

AEMC 6418 user manual

AEMC

AEMC 6418 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.