Infrared thermometer
Instruction manual
A. Introduction
G. Maintenence
H. Specification
B. WARNING& CAUTIONS
213
564
D. EMISSIVITY
F. LCD display & buttons
A
B
CDEFG
H
IJK
1
SET
MODE
STO
CAL
E. Operation
2.
(figure 2)
Diagram description
7
3
4
6
2
5
5
C. Distance to spot size
38mm@
300mm
75mm@
900mm
132mm@
1500mm
1.5"@
12"
3.0"@
36"
5.3"@
60"
D:S=12:1
This infrared thermometer is used for measuring the tem-
perature of the object's surface, which is applicable for va-
rious hot, hazardous or hard-toreach objects without contact
safely and quickly.
This unit consist of Optics, Temperature Sensor Signal
amplifier, Processing circuit and LCD Display. The Optics
collected the infrared energy emitted by object and focus onto
the Sensor. Then the sensor translates the energy into an
electricity signal. This signal will be turned out to be digital
shown on the LCD after the signal amplifier and processing
circuit.
1. Warning:
To avoid the potential situation may cause harm or dam-
age to people, please pay attention to the following items:
1) . Do not point laser directly at eye or indirectly off reflec-
tive surfaces.
2) . The unit cannot measure through transparent surfaces
such as glass or plastic. It will measure the surface tem-
perature of these materials instead.
3) . Steam, dust, smoke, or other particles can prevent acc-
urate measurement by obstructing by the units optics.
2. Cautions:
Infrared thermometer shouldbe protected forthe
following:
1). EMF (electro-magneticfields) from arcwelders,
induction heaters.
2). Thermal shock(cause by largeor abrupt ambient
temperature changes allow 30 minutes for unit to
stabilize before use).
3). Do notleave the uniton or nearobjects of high
temperature.
Emissivity: Most organicmaterials and paintedor
oxidized surfaces havean emissivity of0.95(preset in
the unit). Inaccuratereadings will resultfrom measuring
shiny or polishedmetal surfaces. To compensate forthis,
adjust the unitsemissivity reading orcover the surfaceto
be measured withmasking tape orflat black paint.
Measure the tapeor painted surfacewhen the tapeor
painted reach thesame temperature asthe material
underneath.
1. When take measurement, pay attention to the Distance
to Spot Size. As the Distance (D) from the target surface
increases, the spot size (S) of the area measured by the
unit becomes larger.
The Distance to Spot size of the unit is 12:1.
This unit is equipped with a laser, which is used for aiming.
2. Field of view:
Make sure the target is larger than the unit's spot size.
The smaller the target the closer measure distance.
When accuracy is critical, make sure the target is at least
twice as large as the spot size.
0.30
0.95
0.95
0.70
0.50
0.90
0.85
0.95
0.95
0.95
0.94
0.90
0.93
0.85
0.98
0.70
0.50
0.98
0.94
0.93
0.95
0.95
0.95
0.90
0.98
0.90
0.80
0.94
0.93
0.94
Emissivity
Emissivity
Marterial
Marterial
Aluminum
Asbestos
Asphalt
Basalt
Brass
Brick
Carbon
Ceramic
Concrete
Copper
Dirt
Frozen food
Hot food
Glass(plate)
Ice
Iron
Lead
Limestone
Oil
Paint
Paper
Plastic
Rubber
Sand
Skin
Snow
Steel
Textiles
Water
Wood
1. Operating the unit
1). Open the battery door and insert a 9V battery properly.
2). Pull the trigger to turn on the unit;
3). Aim at the target surface and pull the trigger, then temp-
erature will be shown on the LCD.
This unit is equipped with a laser, which is only used for
aiming.
2. Locating a Hot Spot
To find a hot spot, aim the thermometer outside of interest,
then scan across with an up and down motion until you
locate the hot spot. (Figure 1)
Figure 1
1.LCD display:
A: measuring reading
B: measuring unit
C: laser on icon
D: back light on icon
E: battery power icon
F: scanning icon
G: data hold icon
H: mode/emissivity indicator
I: data storge / read icon
J: low temperature alarm icon
K: high temperature alarm icon
Figure 2
(1) Trigger:When turn onLCD display VERXXsoftware
version for 1 sec. And turn to display reading with
SCAN icon. Release the trigger, display reading with
HOLD icon. Built in auto power off in 30sec.
(2) Laser /back light button:when back lightturn on ,
any operations will remain back light for 10 sec. LCD
indicate on/off status.
(3) (6) key functions:press 3 key, LCD subdisplay
blinks MAX-MIN-DIF-AVG-HAL-LAL-STO-EMS seg-
ment(only main displaymeans normal measuring
mode) press 4key to enter.
a. MAX: measuringmaximum temperature
b. MIN: measuringminimum temperature
c. DIF: Basicon the readingbefore press 4key,
compute the differenceof current reading.
d. AVG: measuring average temperature
e. HAL: hightemperature alarm--when selectedHAL,
press 5 keysto set hightemperature alarm trigger
and confirmed bypressing 4 key. When readingover
trigger, LCDdisplay HI iconwith BiBi audiosounds.
f. LAL: lowtemperature alarm--when selectedLAL,
press 5 keysto set lowtemperature alarm triggerand
confirmed by pressing4 key.
When reading overtrigger, LCDdisplay LOW icon
with BiBi audiosounds
g. STO: datastorage--when selected STO,lock & DATA
& 1---indicator willshown when press4 key.After
temperature read outpress 6 keyto store, then2---
memory unit willbe shown. There12 groups memory
unit available. To recall the stored data innormal
measuring mode bypressing 6 key, remove alldata
by pressing 6keys for 3seconds.
H. EMS: Emissivitysetup-- press 5key for emissivity
settings, press 4key to savesetup and backto
normal status.
(7) Clesius /Fahrenheit switch: Pleaseopen battery
and push theslide switch forconvertsion.
1. Lens Cleaning:
Blow off loose particles using clean compressed
air. Gently brush remaining debris away with a
moist cotton swab. The swab may be moistened
with water.
2. Case cleaning: Clean the case with a damp spo-
nge/cloth and mild soap.
Note:
1) Do not use solvent to clean plastic lens.
2) Do not submerge the unit in water.
Accuracy
0~700 (32~1292 ) : 1.5 ( 2.7 )
or 1.5%
-50~0 (-58~32 ): 3 ( 5 )
Whichever is greater
Repeatability 1% of reading or 1
Response time 500 mSec, 95% response
8-14 um
Emissivity 0.10~1.00 Adjustable (0.95 Preset)
0 ~40 (32 ~ 104 )
-20 ~ 60 (-4~140 )
Power 9V Alkaline or NiCd battery
Non-laser mode: 22 hrs;
Laser Models:12 hrs
12:1
Spectral response
Operating Humidity
Storage Temperature
Typical battery life
(Alkaline)
Distance to Spot size
Weight
Dimension
-50~700 (-58~1292 )
Temperature range
111*50*172mm
222g
Operating Temperature
10~95%RH non-condensing
up to 30 (86 )
Resolution 0.1 or 0.1
MADE IN CHINA