
5061-084-REV-1LIGHTWAVE 3 & 3+ SPECTROPHOTOMETERS USER’S MANUAL
5
SAFETY INFORMATION
Hazards and Warnings
This section describes potential hazards which may
exist in the operation of these units. A number of
warning labels and symbols are axed to your
instrument. These symbols are used to inform you of
potential dangers which may exist or where caution
is required. Before installing your new unit, please
take time to familiarise yourself with these warnings
and symbols.
This instrument is subject to the following
identied hazards:
This unit uses a Xenon lamp. The lamp energy is
mainly conned within the unit but traverses the
cell holder when a measurement is being taken.
Although the energy present is low and intermit-
tent you are advised not to stare into the beam
or attempt to deect the beam as prolonged
exposure could result in permanent eye damage.
High voltages exist within the power supply unit
and the Xenon lamp housing. Repair and main-
tenance should only be carried out by individuals
trained to work on these instruments.
There are no biohazardous materials within
the unit, however, this unit may be exposed to
biohazardous samples during normal laboratory
use. To protect users against these hazards we
recommend the following decontamination
procedures:
• Wipe the exterior casework with disinfectant
cleaning wipes.
• Remove cuvettes and cuvette holders.
• Wash with disinfectant appropriate
for the biohazard in question.
• Rinse with distilled water.
• Allow to dry thoroughly before reuse.
To further reduce the possibility of biohazards:
• Include an appropriate decontamination
certicate for equipment returned for repair.
• Ensure that the operator of the equipment is
provided with a safe working environment.
• Use, store and dispose of any chemicals in
accordance with manufacturer’s guidelines
and local safety regulations.
• Provide suitable ventilation when working
with volatile solvents or toxic substances.
• Dispose of solvents and chemicals that may
be classed as hazardous waste in accor-
dance with local regulatory practice.
• Determine if personal protective equipment
(PPE) is required for handling laboratory
samples.
All models can be connected to and operat-
ed from a PC. To preserve the integrity of the
measuring equipment it is essential that the
attached PC itself conforms to basic safety
and EMC standards and is set up in accor-
dance with the manufacturers’ instructions.
If in doubt, consult the information that came
with your PC.
The following safety precautions should be
observed when operating a PC:
• To reduce the chance of eye strain, set up
the PC display with the correct viewing
position, free from glare and with appropriate
brightness and contrast settings.
• To reduce the chance of cross contamination
from biological samples, use appropriate
personnel protection measures and
disinfectant wipes on keyboard and mouse.
In the event of contamination, malfunction
or hazard occurring, the operator should
disconnect the unit, by removing the power
cord, and isolate for decontamination
and/or repair.