BODY COMPOSITION ANALYZER
Install Manual
1. If you use the InBody230 on a carpet, static electricity may cause serious damage to the
electronic components. If the InBody230 is to be installed on a carpet, use antistatic mats
below the InBody230. Use InBody tissue to wipe feet and hands before the measurement
starts.
2. Do not wipe the base and hand electrode with cleaning liquids. If cleaning liquids enter
into InBody230, it may cause corrosion or electric shock by short circuit. For cleaning of
the InBody230, please use InBody tissue, which also has antibacterial functionality.
3.Do not dismantle InBody230. Internal parts are not for customer use and may cause
electric shock. If InBody230 is dismantled, the warranty is void, and service costs will be
charged.
DANGER
NOTECAUTION
Please keep the packaging for future events of repacking InBody230 and sending
it for servicing or otherwise transporting InBody230 elsewhere.
Head
Stand
Base
Hand Electrode
Input Buttons
Foot Electrode
A
B
C
❶
Basic
❶ InBody230
❷ Install Manual 1EA
❸ User’s Manual CD 1EA
❹ Warranty
❺ Adapter 1EA(DC 12V, 3.4A)
❻ Power Cable 1EA
Optional
❼ Thermal printer
❽ SD400 Set
❾ Result Sheet
❿ Carrying Bag
⓫ InBody Tissue
Contents of the Box
A
❷ ❸ ❹
B
❺
C
❼
❻ ❽ ❾
❿ ⓫
2. Step on the foot plate and measure weight.
3. Enter your personal information and press “ENTER” button.
DELETE
ABC
QZ DEF
GHI JK
PRS TUV
DANGER
NOTECAUTION
Failure to connect the device to a properly grounded outlet may result in damage to
the device, inaccurate measurement and/or inappropriate operation due to possible
electrical surges.
DANGER
NOTECAUTION
Do not plug the device to a shared power source with any other electrical devices such
as a treadmill, refrigerator, air conditioning unit, microwave or RF medical diathermies
that can be affected by electromagnetic interference. It may result in inaccurate
measurements.
1. Turn on the InBody230 and keep the power on until the home screen
initializes. It will take 15 seconds. During the warming, do not touch
the base.
Measurement
Installation
3. If the air bubble is not in the center of the level indicator, tighten or loosen
the leveling screws(4EA) on the foot plate to adjust the level of InBody230.
InBody230 must be leveled properly for accurate weight measurement.
<Leveled state>
Level Indicator
Leveling Screws
1. After opening the product box, please remove the accessories box and
the center Air-cell cushion that is covering the InBody.
2. Carefully remove the InBody from the box and place
the InBody230 horizontal to the ground. Raise the
stand, lift the head, and slightly spread the arms.
4. Plug the power adapter cable into the InBody230.
<Rear view of InBody230>
Power switch
* Please use the device after
properly connecting the power
and printer cables to the
appropriate matching port.
케이블고정장치
<Rear view of InBody230>
<Printer Cable / Option>
❺
* Make sure the power cord is rmly plugged into the power adapter.
*
Please plug the power cord and the printer power cord to an
electrical outlet. A surge protector is highly recommended.
❺Earth Ground Pole
Three Pole Socket
❻
Cautions before Measurement
Do not use this equipment in combination with the following electronic medical devices.
- Medical devices, such as a pacemaker.
- Electronic life support systems, such as an articial heart/lung.
- Portable electronic medical devices, such as an electrocardiograph.
Callouses on the palm or sole or particularly dry hands and feet may lead to difculties in measuring.
If this is the case, please wipe the palm and sole with an InBody Tissue before testing.
Proper Posture
Proper posture is essential to achieve reliable results and high reproducibility.
Do not talk or move during the InBody test.
Gently grip the hand
electrodes so that all the
four ngers of each hand
are touching the surface
of the either hand
electrode. Touch the
circular electrode with
the thumb.
If the examinee’s
thighs are in contact
with each other, have
the examinee wear a
pair of knee-length
shorts so that the
thighs are not in contact
each other.
Step on the foot plate with bare feet,
ensuring that both heels are touching
the smaller foot electrode on either side
as illustrated.
Do not wear socks
during measurement.
Armpits are not closed up.
Straight both arms.
ⓒ
2014 BIOSPACE Co., Ltd. All rights reserved. BM-ENG-40-G-140818
DANGER
NOTECAUTION
If you are using a printer, a result sheet will be printed automatically.
< Option >
4. Grab the hand electrode and hold your arms slightly away from
the torso.
5. Remain still during measurement and the test takes about 30 seconds.
For any problems with InBody230 or for any clinical information, contact:
BIOSPACE Co., Ltd. [HEAD OFFICE]
InBody Bldg., 54, Nonhyeon-ro 2-gil, Gangnam-gu, Seoul 135-960 KOREA
TEL: +82-2-501-3939 FAX: +82-2-578-2716
Website: http://www.inbody.com E-mail: info@inbody.com
For more information on the functions of InBody230 and descriptions on the thermal results sheet items,
see the user’s manual CD included.
For purpose of improvement, specications and design are subject to change without notice.
FAQs
Q1. I do not know how to congure the settings.
A. Use the left right buttons (◀,▶) or number to select a category among
Region, Printer, Result, Other and Interface. Then the selected item’s
sub-categories are displayed on the screen.
Q2. Can I use my printer?
A. There is a list of compatible printers.
Please contact BIOSPACE or the local distributor.
Q3. I do not know why error messages are shown on the LCD.
A. The InBody230 displays an error message to alert the operator of
problems and to recommend the correct action.
Turn the InBody230 off and on again
after removing the objects from the
foot plate.
If entered subject data are beyond
acceptable ranges, this error message
appears.
A subject should wipe his/her soles and
palms with wet tissue and try again.
Q4. Where can I buy printer paper and other optional devices?
A. Please contact BIOSPACE or the local distributor.
Q5. How do I know the measurement taken is correct?
A. After taking the measurement, check the impedance at the bottom of the
results sheet printed. The impedance value must decrease in the
arrow direction.
Q6. How do I clean hand/foot electrode?
A. For cleaning of hand/foot electrode, please use InBody tissue, which
also has antibacterial functionality.

Other manuals for InBody230

2

Other Biospace Measuring Instrument manuals

Biospace InBody170 User manual

Biospace

Biospace InBody170 User manual

Biospace InBody170 User manual

Biospace

Biospace InBody170 User manual

Biospace InBody230 Manual

Biospace

Biospace InBody230 Manual

Biospace inbody s10 User manual

Biospace

Biospace inbody s10 User manual

Popular Measuring Instrument manuals by other brands

Rockwell Automation Allen-Bradley 1440-TBS-J installation instructions

Rockwell Automation

Rockwell Automation Allen-Bradley 1440-TBS-J installation instructions

HTA Instrumentation Equinox-COND-101 user guide

HTA Instrumentation

HTA Instrumentation Equinox-COND-101 user guide

Water Specialties Propeller meter ML03 Operation and maintenance manual

Water Specialties

Water Specialties Propeller meter ML03 Operation and maintenance manual

Badger Meter HR-LCD Pulse Programming manual

Badger Meter

Badger Meter HR-LCD Pulse Programming manual

Leviton 4000 Series installation guide

Leviton

Leviton 4000 Series installation guide

Sanwa DCM301 user manual

Sanwa

Sanwa DCM301 user manual

YOKOGAWA AQ6360 user manual

YOKOGAWA

YOKOGAWA AQ6360 user manual

CS Instruments DS 350-P Operation manuals

CS Instruments

CS Instruments DS 350-P Operation manuals

Hospitex LICG892 Service manual

Hospitex

Hospitex LICG892 Service manual

Master Meter Register Install & Remove

Master Meter

Master Meter Register Install & Remove

Satec PRO EM235 Quick start manual

Satec

Satec PRO EM235 Quick start manual

D'Amore engine'ring SMDOM-1 user manual

D'Amore engine'ring

D'Amore engine'ring SMDOM-1 user manual

ring RBAG700 instructions

ring

ring RBAG700 instructions

Opsytec Dr. Grobel RM-32 operating instructions

Opsytec Dr. Grobel

Opsytec Dr. Grobel RM-32 operating instructions

Omron Z300 V3 - SETUP Setup manual

Omron

Omron Z300 V3 - SETUP Setup manual

Keysight N2795A user guide

Keysight

Keysight N2795A user guide

HEIDENHAIN ND 221 user manual

HEIDENHAIN

HEIDENHAIN ND 221 user manual

PCB Piezotronics M355B04 Installation and operating manual

PCB Piezotronics

PCB Piezotronics M355B04 Installation and operating manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.