
Biotage®Flash 400 User Manual | © Biotage 2018
Safety
Safety Requirements
You must observe all safety requirements when installing and
operating the system. Failure to install or use the system in
a manner specified by Biotage may result in personal injury
and/or equipment damage.
If the system has been damaged or does not function properly,
turn off the gas supplies, depressurize the system (all pressure
gauges should read zero (0) bar/psig), open the vacuum break
valves (see MV-04 and MV-05 in Figure 14 on page 8), and
contact Biotage 1-Point Support (www.biotage.com).
Installation
»
The system must be unpacked and installed by an
authorized Biotage service engineer. Prepare the
installation site as described on page 14.
»
Follow regional safety practices when handling
and moving shipping boxes and containers,
and when moving the system.
»
The total weight of the package including the system is up to
1043 kg (2300 lbs) depending on the system configuration.
Use suitable lifting equipment when moving the package.
»
The system can weigh up to 650 kg (1433 lbs). It must
be installed on a level floor that can support the
system weight concentrated on the four wheels.
»
The system must be placed in a well-ventilated walk-in
fume hood or an equivalent enclosure that is capable
of exhausting 0.6 m3/min (21.2 ft3/min). Follow
local and national safety regulations for installing a
system inside a fume hood and the safety regulations
supplied by the fume hood manufacturer.
»
Never install or use the system near
a potential ignition source.
»
The system, when operated with non-polar solvents
such as hexane, methylene chloride, etc., can build up a
high static electricity charge, which in certain conditions
can be dangerous. To eliminate any risk of a static
discharge, the system and ancillary containers must
be grounded before use as described on page 14.
Failure to follow these grounding instructions may result
in equipment damage, personal injury, or death.
»
A trained person must verify that the system and
ancillary containers are grounded before each run.
Nominal resistance must be below 5 Ohm between
ground and each metal point in the system.
»
Only nitrogen must be used to move radial
compression fluid through its flow path due to the
risk of exothermal reaction and/or explosion.
»
Ensure that the inert gas tubing connected to the
system cannot come in contact with chemicals.
Corrosives and solvents can dissolve the tubing.
»
The outlets of the three safety relief valves (located on
the radial compression module, the air pressure regulator
to the solvent pump, and the radial compression fluid
reservoir) must be directed away from the operator.
»
To avoid injury to yourself or damage to the system,
never apply gas pressures greater than 100 psig
(6.9 bar) to the system inlet gas ports.
»
Do not over-tighten the fittings or the
tubing may become damaged.
»
Never modify the system components in any way
and always use spare parts supplied by Biotage.
Failure to follow this warning may result in serious
personal injury or death. Please contact your Biotage
representative for questions regarding this.
Operation
»
Use the system only for its intended purpose, as described
in the user documentation delivered with the system and
user documentation available at www.biotage.com. If the
system is used in a manner not specified by Biotage, the
safety features of the system may be compromised.
»
The system must not be operated unattended.
»
Never operate the system if damaged.
»
Only nitrogen must be used to move radial
compression fluid through its flow path due to the
risk of exothermal reaction and/or explosion.
»
Inspect the head clamps before and after each use.
If the clamp, threaded studs, nuts, or washers show
any sign of wear or damage, replace it/them. Only use
clamps, studs, nuts, and washers supplied by Biotage.
»
Confirm that the head clamps are tight before compressing
the radial compression module. Hand-tighten the clamps,
to a torque of 34 Nm (25 ft-lb). Ensure that the space
between the clamp halves is distributed evenly on both
sides. The gap should be approximately 10 mm (0.4")
on both sides. If the cartridge is not sealed properly,
replace the O-rings (see Figure 3 on page 3)
and retry, or contact Biotage 1-Point Support.
»
Ensure that all clamps are properly closed before
introducing pressure into the system.
»
Before removing the head assembly, ensure that
the system is depressurized (all pressure gauges
should read zero (0) bar/psig) and completely
drained, and the vacuum break valves (see MV-04
and MV-05 in Figure 14 on page 8) are open.
»
Ensure that the pump pressure gauge on the control
panel (see PG-01 in Figure 14 on page 8) reads
zero (0) bar/psig before disconnecting the stainless
steel overbraided solvent inlet tube from the three-way
injection valve, or the three-way injection valve.
»
Never remove a tube or system component
when it is under pressure.