
13
OPERATING INFORMATION
{TIMING} The Timing parameter menu has three selections: Single/Double Pulse. Delay,
Width, a Single/Double Pulse.
Delay determines the time duration between the mainframe TRIG OUT pulse and
the PULSE OUT and LIGHT OUT pulses (see Figure 1-1).
In Single Pulse operation, Delay may be specified over the entire timing range of
the mainframe, from zero delay to a delay of 640 seconds. The time interval
between the leading edge of the output pulse (mainframe PULSE OUT and
module PULSE OUT or LIGHT OUT) and the TRIG OUT pulse will be equal to
Delay plus the insertion delay.
In Double Pulse operation, Delay may range from 500 us to 640 seconds. Two
identical pulses are provided from each output (mainframe PULSE OUT, module
PULSE OUT and LIGHT OUT). The leading edge of the initial pulse, which occurs
at the zero delay position, and the leading edge of the second pulse are separated
by the selected Delay.
Width allows the user to select one of four fixed pulse widths for the output: either
10 ns, 15 ns, 20 ns or 25 ns (or from an alternate set of widths ordered from the
factory). The desired width is selected by pressing 1, 2, 3, or 4 (corresponding to
10 ns, 15 ns, 20 ns or 25 ns respectively) followed by the ENTER or EXEC key (if
alternate widths have been supplied, 1, 2, 3, and 4 correspond in order to
increasing pulse widths). Pressing 0, followed by ENTER or EXEC, will result in
zero output from the LIGHT OUT jack, the module PULSE OUT jack, and the
mainframe PULSE OUT jack; the mainframe, however, will continue to produce
TRIG OUT pulses.
{LEVEL} The Level parameter menu for the 106H has only the Peak Level available.
Peak Level selects the output power level for the optical pulses. This level may be
adjusted between 5 mW and 50 mW in 0.2 mW steps. No light output is present
between pulses.
{UNITS} This key is not used with the106H module (and has no effect).
FUNCTION KEYS These keys are not used with the 106H module (and have no effect).
{A}, {B}, {C}
Remote Programming
The remote programming commands that are specific to this module correspond to the module
dependent front panel commands, as described in the previous section. Consult the 6040 manual and
the above section for details on controlling the instrument with remote programming. The Module Status
command, the only command that is specific to the module, is not used with the 106H and will return an
undefined value.