Boscarol OB1000 Series User manual

Rev. 04-2018 USMAN1000 2P-EN
MEDICAL SUCTION UNIT
OB1000 FA - OB1000 FM
USER MANUAL
MANUFACTURED BY:

Rev. 04-2010 USMAN2012-IT
2 - 28
Rev. 04-2018 USMAN1000 2P-EN

Rev. 04-2010 USMAN2012-IT
3 - 28
Rev. 04-2018 USMAN1000 2P-EN
INDEX
SYMBOLS............................................................................................................................................................................. 4
WARNINGS, PRECAUTIONS AND IMPORTANT INFORMATION........................................................................................... 6
IMPORTANT INFORMATION ............................................................................................................................................... 7
OB1000 SUCTION UNIT....................................................................................................................................................... 8
□DESCRIPTION AND INTENDED USE ........................................................................................................................ 8
□CONTRAINDICATIONS FOR USE ............................................................................................................................. 8
□CONTROLS, OPERATIONS AND CONTROL PANEL .................................................................................................. 8
□INDICATOR LIGHTS................................................................................................................................................. 9
□UNIT TESTING ........................................................................................................................................................ 9
□COLLECTION JARS ................................................................................................................................................ 10
□OB-J FA REUSABLE, AUTOCLAVABLE COLLECTION JAR ................................................................................... 10
□PROTECTION FILTER ........................................................................................................................................ 10
□OB J COLLECTION JAR WITH SERRES DISPOSABLE LINER ................................................................................ 11
□ YANKAUER SUCTION CATHETER AND END-PIECE WITH FINGER-TIP CONTROL .................................................. 11
□POWER SUPPLY AND BATTERY CHARGING.......................................................................................................... 11
□BATTERY FUNCTION TEST .................................................................................................................................... 12
MAINTENANCE AND REUSE .............................................................................................................................................. 12
□AFTER EACH USE.................................................................................................................................................. 12
□CLEANING OF OB-J FA AUTOCLAVABLE JAR ........................................................................................................ 13
DECONTAMINATION/STERILISATION OF THE COLLECTION JAR AND SILICONE TUBING ........................................ 13
REPLACING HE FILTER.............................................................................................................................................. 14
□CLEANING OF OB-J JAR ........................................................................................................................................ 14
REASSEMBLY OF THE JAR ........................................................................................................................................ 15
□DISPOSAL OF CONTAMINATED PARTS................................................................................................................. 15
□CLEANING THE SUCTION UNIT............................................................................................................................. 15
□SUCTION UNIT SAFETY......................................................................................................................................... 16
□DISPOSING OF THE SUCTION UNIT ...................................................................................................................... 16
ACCESSORIES, CONSUMABLES AND SPARE PARTS ........................................................................................................... 17
TECHNICAL SERVICE.......................................................................................................................................................... 18
TROUBLESHOOTING ......................................................................................................................................................... 18
TECHNICAL DATA AND REFERENCES TO LEGAL REQUIREMENTS ..................................................................................... 19
RISKS OF MUTUAL INTERFERENCE WITH OTHER DEVICES........................................................................................... 21
RISK OF ELECTROMAGNETIC INTERFERENCE AND POSSIBLE SOLUTIONS ................................................................... 21
WARRANTY ....................................................................................................................................................................... 24
DECLARATION OF CONFORMITY....................................................................................................................................... 25

Rev. 04-2010 USMAN2012-IT
4 - 28
Rev. 04-2018 USMAN1000 2P-EN
SYMBOLS
S.1 Symbols used on the suction unit and accessories, and indicated in the user manual
Double insulation symbol. The suction unit has dual electrical insulation to protect user and
work environment, providing two means of protection (there is no ground conductor or
equipotential connection point).
Type BF applied part
Use suction unit only within indicated temperature range. Using the OB1000 outside these
limits may compromise its operation, reduce battery life and cause internal safety devices to
trip.
Read the User Manual
Suction unit accessories and/or consumables bearing this symbol are single use, disposable
items. After use, they must be discarded and replaced with new ones.
Specific warnings regarding the suction unit, which must always be taken into consideration.
CE marking compliant with It. Leg. Decree 46/97 for medical devices rated class I or higher
(European Directive MDD 93/42/EEC)
Device homologated under the ECE-R10 International Regulation and performed to the
95/54/EEC European Directive.
Manufacturer
Date of manufacture
The OB1000 suction unit contains electrical and/or electronic equipment that must be recycled
per European Directive 2012/19/UE –Waste Electrical and Electronic Equipment (WEEE).
Electric and magnetic fields generated by radiographic or tomographic equipment, portable
radio equipment, RF radio and devices bearing this symbol could affect suction unit operation.
In these cases, the suction unit should not be used or a proper distance should be kept from
such equipment.
The materials composing the suction unit can be recycled following specific procedures
outlined in national laws and local regulations.
The suction unit complies with European Directive 2011/65/EU (RoHS)

Rev. 04-2010 USMAN2012-IT
5 - 28
Rev. 04-2018 USMAN1000 2P-EN
Authorized Representative in the European Community, if the manufacturer is not resident
therein.
Expiration date
Do not use if the packaging is not intact
Catalogue ID code
Sterile device. Sterilization method: ethylene oxide
Sterile device. Sterilization method: ionizing radiation
Production batch number
Serial number
S.2 Symbols used in the User Manual to draw the reader’s attention
Important safety information on proper use of the unit to prevent injuring operator or patient
and/or damaging the suction unit
Warning: information requiring special attention
Notes or information on correct use of the unit

Rev. 04-2010 USMAN2012-IT
6 - 28
Rev. 04-2018 USMAN1000 2P-EN
WARNINGS, PRECAUTIONS AND IMPORTANT INFORMATION
Read carefully
This User Manual has been prepared using simple, easy-to-understand language. If you have difficulty
interpreting the above, contact the manufacturer for further clarification
•Read all instructions carefully before using the unit. Careful, proper use will ensure smooth operation and protect
both patients and operators alike.
•The unit is designed exclusively to remove non-flammable organic fluids (secretions) during medical procedures.
For this reason, it should only be used by duly trained personnel.
•Never use the suction unit in the presence of flammable and/or explosive liquids, gases and mixtures as this could
lead to explosion and/or fire.
•Using the suction unit under environmental conditions other than those indicated herein can seriously compromise
function and modify its technical parameters (e.g. the maximum suction value or battery life).
•If suction is performed without the collection jar and/or filter in place, or if you suspect that substances may have
entered the suction circuit (i.e. inside the suction unit), immediately contact the nearest service centre or the
manufacturer to have the unit serviced.
•Before cleaning the unit or proceeding with any maintenance, unplug the unit from the external power supply. Do
not submerge in liquids as this could damage the suction unit and cause the safety devices to cut in.
•The unit does not require any maintenance on the operator’s part. The only operations authorized are those listed
herein. For technical support, periodic overhaul and any repairs that may be needed, contact your authorized
service centre.
•The manufacturer provides authorized personnel —who have taken a specific technical assistance training course
—with the documentation necessary to carry out the work (service manual).
•To ensure patient safety, precision of the displayed values and proper unit function, use only original spare parts.
By failing to comply with this warning, the operator assumes responsibility for any patient injury or property
damage.
•Do not use any batteries except those approved by the manufacturer.
•Do not make any mechanical and/or electrical modifications to the wall/charging bracket. Replacing any parts
thereof, and/or modifying the bracket itself, can seriously affect unit anchoring.
•The OB1000 suction unit does not perform any clinical diagnostics on the patient.
•Il dispositivo non può essere utilizzato per assistere pazienti durante indagini attraverso RMN (risonanza magnetica
nucleare).
□BATTERY
•Before using the suction unit for the first time (and/or upon receiving it), charge the internal battery for at least 24
consecutive hours.
•Keep the device under load even when not in using. Remaining plugged into the vehicle power supply (12 ÷ 15 VDC)
does not damage the suction unit, but allows maximum autonomy of the battery.
•Failure to regularly recharge the battery will lead to a ‘deep discharge’ at which time the battery will need to be
replaced.
□WARNING ON REUSE OF DISPOSABLE PARTS
•Reuse of disposable parts may compromise the suction unit function and be direct or indirect source of operator
and patient contamination.
•Sterilization and/or cleaning of disposable parts (antibacterial filters, suction tubes, Yankauer suction catheters, etc.)
can cause structural damage leading to the risks of lost mechanical integrity.
Tel.: +39 0471 932893
raq@boscarol.it

Rev. 04-2010 USMAN2012-IT
7 - 28
Rev. 04-2018 USMAN1000 2P-EN
IMPORTANT INFORMATION
Note:
The suction unit was designed and tested according to the latest regulatory standards. If the suction unit
is hooked up to a non-compliant electrical system and/or if the work is not performed by professional
installer, both the suction unit and the electrical system could be damaged. Always consult a qualified
technician with knowledge of the latest requirements!
Preventive maintenance and safety inspection:
The device should be checked at least once every 12 months from the authorized service center. Every 24
months an inspection of safety and technical maintenance is required instead. Refer to authorized service
centers for planning of inspection. The periodic safety inspection of the device does not fall under
warranty.
Contamination:
Sending a contaminated suction unit to the manufacturer, installer or service centre is strictly forbidden.
Any device received in such condition will be rejected and health authorities notified of possible
contamination. Here the term contaminated means a suction unit that has not been cleaned of the
secretions aspirated from the patient. If the substances aspirated have entered the suction unit, it must
be discarded. For Oscar Boscarol srl, the safety of its employees and authorized service centre staff is
important. The suction units will not be demolished according to the WEEE Directive (Waste Electrical and
Electronic Equipment) if the suction unit is contaminated and there is the risk of infection (application of
It. Legislative Decree 81).
□Operator responsibilities
•The OB1000 suction unit is designed for emergency health services and must therefore be ready for use at any
time and in any situation. The suction unit is portable and cannot be wall mounted or secured with specific
restraint systems. Given that it is light weight, it can be stored in bags, backpacks and specific first aid kits.
•Always make certain that the internal battery is sufficiently charged (press the test button).
•Immediately replace any components/parts that are damaged, altered or missing, and/or for which a unit
malfunction is suspected. Always replace such parts with original spares. The suction unit should be stored in
a place inaccessible to children.
•Dispose of packaging in accordance with current regulations and make certain that it is out of the reach of
children.
•Read these instructions carefully before using the suction unit. Careful, proper use will ensure smooth
operation and protect both patients and operators alike.
•Operate the suction unit only in compliance with the technical specifications laid out by the manufacturer in
this manual.
□Intended Use
•The suction unit can be used on all types of patients following the correct medical technique.
•The suction unit is designed to clear the upper airways. Clearing the lower respiratory tract is to be performed
by medical and/or health care professionals trained and authorized to perform this function.
•In some countries, this information must be verified according to the protocols implemented by the local
emergency medical services.

Rev. 04-2010 USMAN2012-IT
8 - 28
Rev. 04-2018 USMAN1000 2P-EN
OB1000 SUCTION UNIT
INSPECT THE SUCTION UNIT AND ALL PARTS BEFORE USE.
DO NOT ATTEMPT TO USE IF ANY PARTS ARE DAMAGED OR MISSING.
For accessories and options available, see the catalogue at www.boscarol.it or send an email to
□Description and intended use
The OB1000 is a portable electrical medical suction device designed to remove fluids and substances obstructing the
upper airways and restore spontaneous and/or assisted respiration. High vacuum is normally used for oropharyngeal
tract suction while low vacuum values are used for tracheal suction and/or applications in children and infants. The
device can be used in emergency health services, first aid, home care and in hospitals and/or mobile medical units. The
device comes with a carrying/storage bag. The unit is designed to meet the classification for “HIGH VACUUM – HIGH
FLOW” medical suction equipment (see ISO10079-1).
□Contraindications for use
Do not use the OB1000 for thorax drainage.
□Controls, operations and control panel
All controls are on the front of the suction unit. The unit can be controlled when fitted on the wall-bracket or in the
carrying bag.
To activate the device, press the switch (6), which is protected against infiltration of moisture, splashing of water and
other cleansers. Vacuum can be adjusted by turning the knob (2). Turning the control knob clockwise increases the
vacuum. The vacuum produced by the internal pump can be read on the analogue vacuum gauge (1) and is expressed
in millibars (mbar) and kilo-pascals (kPa) or millimetres of mercury (mmHg). The vacuum gauge is fluorescent and can
be seen in the dark. On the back are two contacts (9) that allow the charging of the device if fitted on the wall bracket.
Alternatively, you can use the charging cable plugged into the external 2-pin connector (8).
Model BSU210:
1. Suction unit
2. Autoclavable jar
3. Protection filter
4. Angular connector
5. Silicone tube with joint
Model BSU212:
1. Suction unit
2. OB-J jar
3. Disposable bag SERRES®
4. Silicone tube with joint

Rev. 04-2010 USMAN2012-IT
9 - 28
Rev. 04-2018 USMAN1000 2P-EN
□Indicator lights
All lights are placed on the front and display the operation of the device (see Fig1): the autonomy of the battery (3-4-5)
and the recharge state (7).
The table below indicates the condition of the LEDs and the relative power of the battery:
LEDS STATUS
BATTERY POWER LEVEL
LED green on (3)
>70% maximum power
LED yellow on (4)
40-50% medium power
LED orange on (5)
10-20% battery low –the suction unit twill shut down soon
The indicator for charging ON/CHG (7), placed next to the indicators of autonomy, has two different colors: Yellow
indicates that charging is taking place; green indicates that charging is complete. The indicator lights up whenever the
device is connected for recharging. If the LED does not light up, there could be a malfunction of internal recharge circuit,
lack of power (12 Vdc) or lack of connection of external cable to a power source for 12 Vdc.
WARNING
Always check that the plug is inserted correctly into the cigarette lighter: vehicle vibration could cause it
to come out. To ensure this, check the <ON/CHG> LED: it should be on, both during charging and once
charging has been completed!
A low battery compromises suction unit function, and thus its use. It’s always necessary to recharge the
battery after each use. The suction unit can be left steadily plugged into the charge. The battery has a 2-
year life and is automatically replaced during the safety inspection. The battery cannot be replaced by the
user (contact your authorized service centre).
□Unit testing
The test should be performed daily, to ensure that the suction unit is in good working order, there are no anomalies
and/or casing breakage and that the unit is functioning properly.
Daily test
This test lets you quickly check whether the suction unit is fit for use in the field; it involves functional tests that take no
more than 5 minutes.
•Disconnect the unit from the wall bracket or from the external charging cable;
•Set the unit on a stable surface in the upright position so the front is facing you. Do not withdraw the unit from its
carrying/storage bag;
•Turn on the unit with the switch on the front panel. The suction unit should run smoothly and you should not note
any fluctuation in the external pump rpm. You should not hear any unusual noise and/or sharp vibrations;
•Check the indicator lights for battery power (when the battery is fully charged, all of the lights will be on);
•Completely close the vacuum regulator (turning it clockwise) and, with your finger, plug the transparent silicone
tubing running from the filter or disposable bag to the container. The sound of the pump should change and the
Fig.1
9

Rev. 04-2010 USMAN2012-IT
10 - 28
Rev. 04-2018 USMAN1000 2P-EN
reading on the vacuum gauge should reach maximum value (about -800 mbar, -80 kPa, -600 mmHg) in a few
seconds.;
•While keeping your finger over the silicone tubing, turn the vacuum regulator counter-clockwise and check the
reading on the instrument to ensure that suction drops to nearly 0 (40-50 mbar);
•Turn off the suction unit and turn it 180° to check the dataplate on the back and the condition of the
carrying/storage bag;
•Plug the external power supply cable into the cigarette lighter or optional adapter and check that the charging
process starts (orange LED is on).
When finished, compare the results of this test with the value on the table below:
Test –phase
Result
Remedy
Switch the unit on.
The indicator lights for the power and the
pump motor switch are on (noise from the
motor).
Pump failure and/or battery completely
discharged. Recharge the batteryand check
the indicator light goes green.
Check the battery power.
When the battery is fully charged the green
light will be on.
Recharge the unit immediately.
Check the maximum vacuum.
Value range between 700 and 800 mbar
(70kPa ÷ 80kPa; 525 ÷ 600mmHg).
Check that the lid on the collection jar is tight
and that all connections are secure. Change
the disposable liner.
Check the vacuum regulation.
Gauge reading ranges from maximum to
minimum.
Check the vacuum connections and/or the
regulator (anti-clockwise for minimum
vacuum).
If any of the tests are not passed, even after taking the steps outlined above, send the suction unit to
an authorised service centre or consult the manufacturer.
□Collection jars
The suction unit is marketed with three different types of collection jar:
- OB-J FA autoclavabile collection jar with a capacity of 1000 ml (OB1000 FA)
- OB-J collection jar with a capacity of 1000 ml, with disposable liner SERRES®(OB1000 FM)
□OB-J FA reusable, autoclavable collection jar
The jar is made of specific transparent plastic. It includes canister (1), snap-on lid (2),
overflow valve (3) and 90° plastic connector (4). The lid allows direct insertion of the
filter.
The jar can be sterilized in a conventional steam autoclave at a maximum
temperature of 121 °C and pressure of 2 bar (200kPa). The jar must be replaced if it
shows any deformation, breaks or cracks.
The jar must always be used in the upright position to prevent activating the
overflow valve. Should this occur, switch the suction unit off and disconnect the
patient catheter.
When used in a home environment, the jar can be cleaned using a special cleanser
able to guarantee medical device disinfection. The aspirated secretions must be disposed of in compliance with medical
doctrine, i.e. as prescribed by the physician according to the patient's medical condition.
□Protection filter
The filter protects the suction circuit against contaminants aspirated during use. It is made of a
hydrophobic material and prevents the passage of any atomized fluids and aerosols, thus preventing
their uptake (complete absence of patient side suction).
In case of possible contamination, discolouration and increased resistance to suction replace the filter.
Fig.3

Rev. 04-2010 USMAN2012-IT
11 - 28
Rev. 04-2018 USMAN1000 2P-EN
If the suction unit is used on a patient whose medical condition is not known, always replace the filter
after use. This will prevent contamination, even serious contamination, of the operating environment,
operator and patient. If, instead, the patient's medical condition is known and/or there is no danger of
cross contamination, we recommend replacing the filter after each shift or when the filter turns dark.
If the suction unit remains unused, it is advisable to replace the filter once a month. The material used in
its construction could even be damaged by particular environmental conditions (humidity, heat, cold).
Do not use the suction unit without the protection filter or jar!
□OB J collection jar with SERRES disposable liner
The OB-J jar for SERRES disposable liners (see Fig. 4) is made of a specific transparent plastic. It
includes a canister (1), adaptor for SERRES disposable liners (2), "L" connector (3) and disposable
liner (4). The filter integrated into the liner prevents aspirated fluids from flowing back into the
suction unit when it is completely full.
The jar can be sterilized in a conventional steam autoclave at a maximum temperature of 121 °C and
pressure of 2 bar (200kPa). The disposable liner must be replaced after use on a given patient.
When used in a home environment, the jar can be cleaned using a special cleanser able to guarantee
medical device disinfection. The disposable liner can never be emptied and reused. The liner containing the aspirated
secretions must be disposed of in compliance with medical doctrine, i.e. as prescribed by the physician according to the
patient's medical condition.
□Yankauer suction catheter and end-piece with finger-tip control
The suction unit is sold complete with a sterile, Yankauer-type suction catheter and tubing for
connection to the jar. The suction tip and catheter are disposable and must be changed after
each use. To facilitate correct operation, the rigid suction tip is angled so it can reach all parts
of the mouth and upper airways.
Upon request, the suction unit can be outfit with a silicone patient tubing (length: 130 cm) and one sterile Finger-tip
joint so that sterile suction tips of varying size can be connected. The finger-tip joint
allows the user to control vacuum directly with a finger, without requiring any
commands. The silicone tubing can be sterilised while the finger-tip joint is, instead,
disposable.
The disposable devices supplied with the medical suction unit bear labels providing all
information needed for proper use.
Warning! Do not use sterile medical devices beyond their expiration date or if the package is damaged.
□Power supply and battery charging
The suction unit has a rechargeable internal battery (which cannot be replaced by the operator The maximum battery
charging time (depending on residual charge) is about 15 consecutive hours. A fully charged battery will provide
approximately 45÷60 minutes of continuous operation (at open flow). This time may also vary, even considerably, if the
suction unit is used outside of the parameters recommended by the manufacturer (e.g. when used in the presence of
very high or very low temperatures).
When properly charged, average battery life is 24 months. After this period, we recommend replacing the battery.
If the unit is not used for a long time, run a complete check and fully charge the battery every 15 days.
The suction unit can be charged using the cable (supplied), the wall bracket or optional power supply (100÷230 Vac).
The charging cable must be connected to a 12÷15 Vdc power supply.
To be used while being charged, the suction unit must be connected to an external power supply (12÷15 Vdc) that can
provide at least 6A.

Rev. 04-2010 USMAN2012-IT
12 - 28
Rev. 04-2018 USMAN1000 2P-EN
CAUTION
Check that the external 11÷25 Vdc power supply is protected by a fuse rated at least 15A (time-delay).
Request such protection from the manufacturer if necessary.
The power supply is an exclusive accessory, available only from the manufacturer. It is approved for such function and
cannot be replaced with other brands. It can only be used indoors and on a power supply compliant with the law.
WARNING
Never tamper with and/or open the power supply. Danger of death. The power supply contains internal
electronics subject to line voltages that can be fatal.
□Battery function test
The steps required to check that the battery is fully functional and efficient are given below:
•Recharge the suction unit for at least 15 consecutive hours;
•Disconnect the suction unit from charging and set the vacuum regulator to maximum suction (turning the knob
clockwise);
•Turn the suction unit on and let it run (without closing the patient tubing). Note the operating time which must be
at least 20 consecutive minutes;
•If this time is not reached and the suction unit shuts down before 20 minutes has elapsed, the battery is damaged
and must be replaced.
NOTE
When battery life decreases significantly, it should be replaced. Contact the manufacturer or an
authorized service centre.
MAINTENANCE AND REUSE
□After each use
After each use, unplug the suction unit, disconnect the disposable parts and discard them. Check that the suction unit
is intact, check the connection tubing and check for any structural anomalies. Clean and disinfect the suction unit as
described below. Replace all single-use, disposable parts with new components and recharge the battery.
Run the daily function test as described under "Daily test” in page.9.
Occupational safety and health and PPE (It. Legislative Decree no. 81)
The decontamination process is always a delicate process, which implies specific training, especially in the emergency
field where the patient’s medical condition and degree of contamination are mostly unknown. For this reason, the
operator must always wear personal protective equipment (PPE) to protect himself and others.
Charging cable
BSU855
OB10WB wall bracket
BSU806
Power supply
BSU897EU –BSU897UK –BSU897JP

Rev. 04-2010 USMAN2012-IT
13 - 28
Rev. 04-2018 USMAN1000 2P-EN
If proper PPE is not available, please contact your safety representative
The organic secretions collected by the suction unit can cause severe
operator infection. For this reason, always use suitable PPE and
disinfection products as established by the competent authorities.
CLEANING INSTRUCTIONS
□Cleaning of OB-J FA autoclavable jar
1. Wear gloves and protective clothing.
2. Remove the yellow angular connector with patient tube, withdrawing it from the jar. If the tube is equipped
with a Yankauer suction tip, it must be disposed of together with the curved tip (disposable device). The yellow
angular connector can be autoclave sterilized.
3. Disconnect the conical connection from the suction unit connector.
4. Loose canister fixing belt.
5. Pull the jar vertically out of the unit.
6. Disconnect the filter from the lid by turning it slightly in its housing.
7. Remove the lid (paying attention to possible contamination with the contents of the jar!). Empty the contents
of the jar.
8 –9 –10 –11 Separate all parts of lid.
After having disposed of the disposable filter and Yankauer suction catheter, complete with tubing, set the reusable
parts in cold running water and rinse thoroughly. Then dip the same parts in hot water (temperature not higher than
60°C) containing a mild, non-alcoholic detergent. Rinse thoroughly and, if necessary, use a non-abrasive brush to remove
any deposits. After washing, rinse all parts with hot running water (30-40°C max.) and then dry with a soft, non-abrasive
cloth. Before reassembling, check that all parts are clean, dry and intact. If the suction unit is equipped with silicone
tubing and “finger-tip” connection, dispose of the connector and clean the silicone tubing. The tubing can be autoclave
sterilized.
Decontamination/sterilisation of the collection jar and silicone tubing
The collection jar and silicone tubing can be disinfected with any mild, non-abrasive chemical cleanser. Alcohol or
solvent-based detergents cannot be used. Do not use any coloured disinfectants as these may damage the plastic of the
jar and stain the silicone tubing, reducing its transparency (e.g. Betadine). Never use disinfectants undiluted. Sterilize
with a steam autoclave at a maximum temperature of 121°C for max. 15 minutes. Do not use pressures above 2 bar
(200 kPa). The jar should be placed in the autoclave upside-down (bottom facing upward). At the end of the autoclave
cycle, leave to cool to ambient temperature, check that it is intact and then reassemble the jar following the operations
used to dismantle it in inverse order.

Rev. 04-2010 USMAN2012-IT
14 - 28
Rev. 04-2018 USMAN1000 2P-EN
WARNING
•Do not put weight on the parts during the sterilisation cycle.
•Observe the maximum limits for temperature, pressure and duration during the autoclave cycle.
•Never exceed the value of 60°C for washing or disinfection operations (with the exception of
sterilization in a steam autoclave).
•Cleaning and/or and sterilisation operations should only be performed by trained personnel.
•Replace the collection jar if it presents fissures, cracks or even partial breakage.
•After reassembling the jar, always check that the lid is properly fitted so as to prevent loss of vacuum
and carryover of fluids.
Place all components of the jar on a flat, secure surface. During assembly and disassembly, always check all parts for
damage. The overflow valve has a float that slides on a plastic guide. Make certain that it slides easily and unhindered
and that the silicone seal is intact.
Replacing he filter
Carefully disconnect the silicone tubing from the contaminated filter and dispose
of it in accordance with current laws and regulations. Remove the filter from the
lid by screwing or unscrewing it from its housing. This operation facilitates
withdrawal and prevents it from breaking inside the lid! Install a new filter
ensuring that the part marked “IN” is connected to the jar inlet marked VACUUM.
Failure to heed this detail can cause filter failure and contamination of the suction
unit intake circuit.
NOTE
The filter must be inserted with the side marked “IN” facing toward the jar. Using the suction unit with filter
inserted incorrectly can lead to contamination of the suction circuit.
□Cleaning of OB-J jar
1. Wear gloves and protective clothing.
2. Disconnect the patient tubing together with the white angular connector for the disposable liner and discard it.
3. Close the “PATIENT” connector with the cap provided on the liner cover.
4. Disconnect the red connection from the suction unit connector.
5. Disconnect silicone tube from the red angular connector on the jar.
6. Loose canister fixing belt and pull the jar vertically out of the unit.
7. Remove the liner from the jar and dispose of it in compliance with current laws and regulations.
8. Disconnect for disposable liner adapter.

Rev. 04-2010 USMAN2012-IT
15 - 28
Rev. 04-2018 USMAN1000 2P-EN
9. Unscrew the plastic elbow connector while keeping the screw pressed inside the jar. Be careful with the seal
ring.
Dispose of the single-use parts and, after disassembling the collection jar, set the reusable parts under cold running
water and rinse thoroughly. Then dip the same parts in hot water (temperature not higher than 60°C) containing a mild,
non-alcoholic detergent. Rinse thoroughly and, if necessary, use a non-abrasive brush to remove any deposits. After
washing, rinse all parts with hot running water (30-40°C max.) and then dry with a soft, non-abrasive cloth. Before
reassembling, check that all parts are clean, dry and intact.
If the jar needs to be sterilised, proceed as described in page 13 “Decontamination/sterilisation of the collection jar and
silicone tubing”.
Reassembly of the jar
Extract a new disposable bag from the packaging, stretch it (picture 1) and insert it into the jar (picture 2). Insert the
complete jar into the bag and connect the tube to the suction unit. Activate the suction unit. Close with a finger the
connector <PATIENT> and, at the same time, press lightly the bag from the centre of the lid (picture 3). Make sure the
bag is completely swollen. Connect the patient tube (Yankauer) to the connector <PATIENT>.
NOTE
The collecting bag must be inserted into a rigid container of the same size.
Il is strictly forbidden to use defective products!
□Disposal of contaminated parts
Always follow local regulations or hospital practices when dealing with contaminated materials. Never store
contaminated parts with new or sterile parts.
□Cleaning the suction unit
Disconnect the suction unit from any external power supply. To clean the chassis of the suction unit, use a damp cloth
with mild detergent (type used for dishes and/or delicate clothing). When finished, dry the surface with a dry cloth or
paper towel.
WARNING
•Never submerge the suction unit in water or other liquids.
•Do not clean the unit with abrasive substances, alcohol or solvents that could deteriorate plastics
or remove printing/labels.
To correctly disinfect and decontaminate the suction unit, we recommend using specific, approved products. These
disinfectants must be free of alcoholic and/or abrasive substances. Oscar Boscarol srl can provide specific materials for
disinfection of medical equipment, including our suction units. These disinfectants, available in different formats (wipes,
spray, liquids), have been laboratory tested and guaranteed to deactivate viruses, bacteria and microorganisms. When
used periodically, they destroy and prevent the formation of dangerous biofilms (superficial layers that easily host
bacteria, moulds, viruses and microorganisms). Our disinfectants do not contain alcohol, chlorine, phenols, aldehydes
and halogens.
Fig. 1
Fig. 3
Fig. 2

Rev. 04-2010 USMAN2012-IT
16 - 28
Rev. 04-2018 USMAN1000 2P-EN
NOTE
For more detailed information, contact us at in[email protected] visit our website www.boscarol.it .
□Suction unit safety
All electrical suction unit guards are set on the inside and cannot be accessed by the operator. If the above safety devices
trip, if they do not reset automatically, contact your authorized service centre or Oscar Boscarol srl.
□Disposing of the suction unit
The unit contains electrical and/or electronic equipment that must be recycled according to EC
Directive 2012/19/EU –Waste Electrical and Electronic Equipment (WEEE) enacted in Italy with
Leg. Decree 49/2014. The suction unit is also compliant with Directive 2011/65/EC which
restricts and prohibits the use of certain hazardous substances in electrical and electronic
equipment. Harmful substances that violate the above Directive are not used in the production
and assembly of electronic boards or in the wiring and connection of electric cables.

Rev. 04-2010 USMAN2012-IT
17 - 28
Rev. 04-2018 USMAN1000 2P-EN
ACCESSORIES, CONSUMABLES AND SPARE PARTS
Manufacturer code
Description
Accessories
BSU806
Wall bracket OB10WB
BSU897EU
Adapter 100/240 Vac - 2 poles and Euro-plug
BSU897UK
Adapter 100/240 Vac - 2 poles and UK-plug
BSU897JP
Adapter 100/240 Vac - 2 poles and Japan/USA-plug
Parti di consumo
BSU730
Filter for autoclavable jar –5 pcs
BSU732
Filter for autoclavable jar –15 pcs
BSU734
Filter for autoclavable jar –40 pcs
BSU705
Disposable bag - 6 pcs
BSU706
Disposable bag - 12 pcs
BSU707
Disposable bag - 36 pcs
BSU500
OB-J FA autoclavable jar, without protection filter
BSU506
OB-J jar, without disposable liner
126140107191
Yankauer suction catheter
BSU750
End-piece with fingertip control –5 pcs
BSU752
End-piece with fingertip control –15 pcs
BSU754
End-piece with fingertip control –50 pcs
11214101003
Suction tip Finger Ch 10 black
11214101104
Suction tip Finger Ch 12 white
11214101005
Suction tip Finger Ch 14 green
11214101006
Suction tip Finger Ch 16 orange
11214101007
Suction tip Finger Ch 18 red
11214101008
Suction tip Finger Ch 20 yellow
Parti di ricambio
BSU855
External charging cable with cigarette lighter fitting and 2-pole plug
BSU902
Silicone patient tubing - length 130cm
SPS6000
OB-J FA jar without lid
SPS6002
Over-fill valve –3 pcs
SPS6004
Elbow connector for OB-J FA –3 pcs
SPS6006
Lid for OB-J fa jar, without protection filter
SPS6011
Red angular connector –3 pcs
SPS6014
Conical connector –5 pcs
SPS6023
Silicone tubing 15 cm with conical connector
BSU825
Protection bag for OB1000
--
User manual - can be downloaded from www.boscarol.it
NOTE
To make technical improvements, the parts listed may be changed by the manufacturer without prior
notice. Contact the manufacturer for additional information (info@boscarol.it).

Rev. 04-2010 USMAN2012-IT
18 - 28
Rev. 04-2018 USMAN1000 2P-EN
TECHNICAL SERVICE
No electrical and/or mechanical part of the suction unit OB1000 is designed to be repaired by the dealer, customer
and/or operator. Do not open the suction unit and do not tamper with the electrical and/or mechanical parts. Always
contact your authorized service centre or the manufacturer. Performing even the most minor operation on the suction
unit voids the warranty. Unauthorized access to the suction unit can jeopardize its conformity with the applicable laws
and reduce safety for both operators and patients. Contact Oscar Boscarol srl for a list of authorized service centres by
TROUBLESHOOTING
Malfunction
Possible cause(s)
Solution
The suction unit does not work
•Battery low
•Battery failure
•Internal electronic circuit failure
•Charge the suction unit with the
charging cable or power supply
adapter
•Contact an authorized service centre
to have the battery replaced
•Contact an authorized service centre
The suction unit does not work when
connected to the wall bracket.
•Cable damaged.
•Damaged wall bracket and/or
contacts on the suction unit.
•External power source failure (12÷15
Vdc –min. 6A).
•Replace the cable
•Replace the wall bracket.
•Check the external power source.
The suction unit only works if it is
mounted on the wall-bracket or fitted
with the external cable.
•Faulty battery.
•Contact an authorized service centre.
The power supply doesn't work properly.
•Faulty power supply.
•Check the Charging LED. If it is on,
but the battery is not charged, please
refer to authorised service personnel.
•Replace the power supply.
The suction unit works, but the battery
power indicator lights are off.
•Faulty internal circuit.
•Very low battery power.
•Check that the indicator lights work if
connected to the wall-bracket or to
the external charger cable. If they
work, immediately charge the
battery for at least 24 hours.
•Charge the battery for at least 24
hours.
Suction unit battery life is significantly
reduced
•Battery is dead
•Internal charge circuit failure
•Replace battery
•Contact an authorized service centre
Vacuum power on patient side is low or
absent
•Vacuum regulator completely open.
•Filter plugged
•Tubing connecting filter with device
is plugged, bent and/or disconnected
•Overflow valve for 500 ml
•Pump damaged
•Close vacuum regulator completely
and check vacuum on both
instrument and patient sides (by
turning the knob clockwise)
•Replace filter
•Connect tubing to filter and/or jar;
replace if plugged and eliminate any
bends
•Disconnect the tubing running to the
suction unit, empty the jar and check
that the valve moves properly (the
silicon gasket must be facing
upwards). The jar can only be used in
the upright position
•Contact an authorised service centre
Vacuum is always at maximum even if the
jar is removed.
•Fault on the internal pneumatic
circuit.
•Contact an authorized service centre.
High noise, low suction, high vibration.
•Internal pump is damaged
•Contact an authorized service centre

Rev. 04-2010 USMAN2012-IT
19 - 28
Rev. 04-2018 USMAN1000 2P-EN
NOTE
In the case of anomalies or malfunctions other than those indicated above, always contact only an
authorized service centre and/or the manufacturer.
Technical data and references to legal requirements
Classification according to European Directive 93/42/EEC
The secretion suction unit is a medical device suitable for fixed and portable use. Compliant with ISO 10079-1:2009.
Medical Device classification:
IIb
Vacuum degree:
HIGH VACUUM-HIGH FLOW
Mode of operation (duration):
TEMPORARY (maximum continuous use 60
minutes)
Electrical requirements:
SELV (12÷15 VDC)
Use of the device in the home environment:
IEC 60601-1-11
Use of the device in the health emergency:
IEC 60601-1-12
Applied part according to IEC 60601-1:
TYPE BF
Insulation rating:
CLASS II
Protection against ingress of liquids and solids (IEC 529):
IP34d
Compliant with general rules in IEC 60601-1:
Compliant with 3rd Edition
Dimensions
Maximum dimensions:
320 mm (l) x 250 mm (h)x100 mm (p)
Weight:
3,5 Kg max. max. complete with all accessories
Tolerance on all values:
±5%
Technical data
Max suction power:
850mbar (85kPa, 638mmHg) +/-10%
Vacuum Regulation:
linear
Vacuum range regulation:
30÷850 mbar (3÷85 kPa)
Max flow rate:
24 LPM (litres per minute) at open flow ±10%
Max running time with the maximum current-load:
> 45 minutes ±10% run time
Approximate maximum noise:
65dB
Vacuum gauge precision:
±5%
Battery power indicator precision:
±5%
Power supply
Running/charging:
12÷15 Vdc (direct current)
Max current load:
70W (max. current 6A)
Battery:
rechargeable hermetically sealed acid type,
Capacity 3,2 Ah
Max charge time:
15 hours consecutive

Rev. 04-2010 USMAN2012-IT
20 - 28
Rev. 04-2018 USMAN1000 2P-EN
Electrical safety devices:
Internal, not accessible to operator
NOTE
The external direct current power supply must provide at least 6A to enable correct unit operation or
charging. If the suction unit is plugged into the mains or an external DC power supply, the internal battery
is not used.
Special storage and operating conditions
Operating temperature range:
0÷50 °C (32÷122 °F)
Charging temperature range:
15÷30 °C (59÷86 °F)
Storage and transport temperature range (packaged unit):
-25÷55 °C (-77÷131 °F)
Storage and transport temperature range (without packaging):
0÷50 °C
Relative humidity for storage, operation and transport (unit without battery):
15–95%, not condensed
Atmospheric pressure for storage and transport:
70–106 kPa (700–1060 mbar)
Operating in the rain
The OB1000 suction unit is protected against ingress of liquids and solids. However, it is always best to
protect the unit from heavy rains. During operation and storage, the unit must be kept in its
carrying/storage bag and kept dry. If the suction unit is completely wet, move it to a dry area, dry the
outside and wait at least 30 minutes before attempting to restart.
High altitude operation
The operator must take into account the altitude when using the unit. Under such conditions, the vacuum produced by
the internal pump may drop, even considerably, as a result of the reduced atmospheric pressure.
Technical filter specifications
The antibacterial/antiviral filter consists of a PTFE filter support and an air-tight polypropylene container.
Max pressure applicable: 1bar (100kPa)
Retention capacity: for aqueous solutions - up to 0.9 bar (90 kPa); for airborne particles - 0.1 µm 99.99%
Storage of SERRES products
SERRES products are factory-sterile and should be stored in warm indoor locations. Protect the package from humidity,
dirt and dust. Disposable products can be used for 5 years after the date on the label.
Battery charger technical data
Input: 100÷240 Vac 50/60Hz, 1,5A
Output: 13,8 Vcc 4,34A, 60W
Terms and symbols
Vac Voltage (alternating current)
Vdc Voltage (direct current)
°C Unit of measure for temperature (°C = degrees
Celsius)
bar unit of measure for pressure
kPa unit of measure for pressure (kilopascal)
mmHg unit of measure for pressure (millimetres of
mercury)
Conversion formula: 1bar = 100kPa = 750mmHg
This manual suits for next models
4
Table of contents
Other Boscarol Medical Equipment manuals

Boscarol
Boscarol IMM121640 User manual

Boscarol
Boscarol OB 2012 FA User manual

Boscarol
Boscarol PAL32000 User manual

Boscarol
Boscarol OB WB User manual

Boscarol
Boscarol OB 500 Series User manual

Boscarol
Boscarol OB 1000 User manual

Boscarol
Boscarol OB500 FA User manual

Boscarol
Boscarol OB OXIKIT PLUS User manual

Boscarol
Boscarol OB 2012 FA User manual

Boscarol
Boscarol OB User manual