
10 2022/04 | IFU-0229.AEN
3. The me display will ash and the default me can now be changed by using the up or down arrows. The
me will change in increments of 10 seconds. The me cannot be set lower than 20 seconds, nor higher
than 270 seconds.
4. To save the set default me, the display will stop ashing aer 5 seconds and the new default will be set.
Operate Without Temperature Reading
If the cryoICE BOX does not display a temperature and the cryoICE probe is properly plugged in (red and black connec-
tors) the cryoICE probe should not be used. If the Acvaon Buon is pressed with this condion, the cryoICE BOX will
ash and beep for 5-seconds. If the Acvaon Buon is pressed again within 5-seconds, the cryoICE BOX will sequence
into FREEZE mode and the counter will start the countdown immediately. This should only be done at the discreon of
a physician as there will not be temperature feedback.
6. SYSTEM DISASSEMBLY AFTER USE
Check to see that the service icon is not illuminated. If so, nofy AtriCure service to correct the problem.
cryoICE Probe
1. The cryoICE probe can only be removed in the READY mode.
2. Remove the cryoICE probes pneumac connecons by pushing in the sliding ring on the receptacle while
pulling out the cryoICE probe side of the connector.
3. Remove the black and red connecons for the thermocouples.
N2O Cylinder Removal
1. Turn-O the N2O cylinder by turning the knob clockwise.
2. Purge the N2O from the system by pressing and holding the N2O Exhaust Switch in the back of the unit.
Watch the pressure gauge on the cylinder to see that all the pressure has been released. If the cryoICE
BOX is powered o, pull and hold the N2O Manual Exhaust Knob unl the pressure is relieved.
3. Disconnect the gas cylinder inlet ng on the back of the cryoICE BOX by sliding the collar back.
4. Disconnect the hose from the N2O cylinder by unscrewing the black knob.
5. Turn-O power and unplug the cryoICE BOX.
7. PREVENTIVE MAINTENANCE AND CLEANING OF THE AtriCure cryoICE BOX
Note: Do not spray or pour liquids directly on the unit.
Note: The unit and/or accessories cannot be sterilized.
WARNING
Ensure Isopropyl Alcohol (IPA) is completely dry before operang the unit.
CAUTION: Avoid causc or abrasive cleaners to avoid damage to ACM chassis.
Guidelines
The following guidelines are recommended for cleaning the unit. It is the user’s responsibility to qualify any
deviaons from these processing methods.
1. Disconnect the unit or cart from the outlet before cleaning.
2. If the unit and/or accessories are contaminated with blood or other body uids, they shall be cleaned
before the contaminaon can dry (within two hours of contaminaon).
3. The outer surfaces of the unit and/or accessories shall be cleaned with 70% -90% Isopropyl alcohol (IPA)
wipes for a minimum of two minutes. Do not allow uids to enter the chassis.
4. Pay aenon to all areas where uids or soil may gather, such as under/ around the handles or any ght
crevices/ grooves.
5. Dry the unit and/or accessories with a dry, white lint-free cloth.
6. Conduct a nal conrmaon of the cleaning process by visually inspecng the white cloth for remaining soil.
7. If soil remains on the white cloth, repeat steps 3 through 6.
8. Aer cleaning is complete, turn the unit on to perform Power On Self-Test (POST). If any errors are
received, contact AtriCure to begin return process.