Boscarol OB 500 Series User manual

Rev. 10-2018 USMAN500 - UK
MEDICAL SUCTION UNIT BOSCAROL
OB500
FA / FM / LINER
USER MANUAL
MANUFACTURED BY:

Rev. 04-2010 USMAN2012-IT
2 - 28
Rev. 10-2018 USMAN500 - UK

Rev. 04-2010 USMAN2012-IT
3 - 28
Rev. 10-2018 USMAN500 - UK
INDEX
INDEX.......................................................................................................................................................................................................3
SYMBOLS .................................................................................................................................................................................................4
WARNINGS, PRECAUTIONS AND IMPORTANT INFORMATION................................................................................................................6
IMPORTANT INFORMATION....................................................................................................................................................................7
OB500 SUCTION UNIT..............................................................................................................................................................................8
DESCRIPTION AND INTENDED USE STATEMENT ................................................................................................................................8
CONTRA INDICATION FOR USE...........................................................................................................................................................8
CONTROLS, INDICATIONS AND CHECK PANEL....................................................................................................................................8
CONTROL MODULE ............................................................................................................................................................................9
SUCTION UNIT....................................................................................................................................................................................9
ELECTRICAL CONNECTIONS..............................................................................................................................................................10
TEST OF THE UNIT ............................................................................................................................................................................11
COLLECTION JAR...............................................................................................................................................................................11
OB-J FA REUSABLE, AUTOCLAVABLE JAR..........................................................................................................................................11
PROTECTION FILTER .........................................................................................................................................................................11
REUSABLE JAR WITH DISPOSABLE LINER SERRES® ...........................................................................................................................12
OB-J COLLECTION JAR.......................................................................................................................................................................12
JANKAUER SUCTION TUBE AND FINGER-TYP END-PIECE .................................................................................................................12
WALL BRACKET FOR THE JAR............................................................................................................................................................12
POWER SUPPLY ................................................................................................................................................................................13
MAINTENANCE AND REUSE...................................................................................................................................................................13
AFTER EACH REUSE ..........................................................................................................................................................................13
CLEANING INSTRUCTION..................................................................................................................................................................14
REUSABLE COLLECTION JAR OB-J FA ................................................................................................................................................14
DECONTAMINATION OF THE COLLECTION JAR ................................................................................................................................14
REPLACING THE PROTECTION FILTER...............................................................................................................................................15
REUSABLE COLLECTION JAR OB-J .....................................................................................................................................................15
REUSABLE COLLECTION JAR SERRES®...............................................................................................................................................16
REASSEMBLY OF THE JAR .................................................................................................................................................................16
DISPOSAL OF CONTAMINATED PARTS..............................................................................................................................................17
CLEANING THE SUCTION UNIT .........................................................................................................................................................17
SAFETY..............................................................................................................................................................................................17
DISPOSING OF THE SUCTION UNIT...................................................................................................................................................17
ACCESSORIES AND SPARE PARTS...........................................................................................................................................................18
SERVICE..................................................................................................................................................................................................19
FAULT FINDING......................................................................................................................................................................................19
TECHNICAL DATA AND CONFORMITY TO INTERNATIONAL LAW...........................................................................................................19
THE RISKS OF RECIPROCAL INTERFERENCE WITH OTHER DEVICES........................................................................................................21
RISK OF ELECTROMAGNETIC INTERFERENCE AND POSSIBLE REMEDIES ...............................................................................................21
GUARANTEE...........................................................................................................................................................................................24
DECLARATION OF CONFORMITY............................................................................................................................................................25

Rev. 04-2010 USMAN2012-IT
4 - 28
Rev. 10-2018 USMAN500 - UK
SYMBOLS
S.1 Symbols used on the device and recalled in the user manual
Double insulation. The device is doubly electrically isolated from the user and the work
environment (there is no ground conductor or connection equipotential point).
Applied part type BF
Only use the device in the indicated temperature range. The use of the OB500 outside these
limits might hinder its operation and make internal security intervene.
Read the user manual
The accessories and/or consumables of the device, which show this symbol, are disposable.
After use, they must be disposed of and replaced with another product.
Specific warnings regarding the device, which must always be taken into consideration.
CE mark perform to the MDD 93/42/EEC for Medical devices with class above I
Device homologated under the ECE-R10 International Regulation and performed to the
95/54/EEC European Directive.
Manufacturer
Production date
The device contains electrical and/or electronic equipment that must be recycled per EC
Directive 2012/19/EC –Waste Electrical and Electronic Equipment (WEEE)
Electric and magnetic fields generated by radiographic or tomographic equipment, portable
radio equipment, RF radio and devices marked by the symbol on the side might have an impact
on the proper operation of the device. In these cases, the device should not be used or a proper
distance should be kept from this equipment.
The materials composing the device can be recycled following specific procedures provided by
national laws and local regulations.
The device complies to the European Directive 2011/65/UE (RoHS)

Rev. 04-2010 USMAN2012-IT
5 - 28
Rev. 10-2018 USMAN500 - UK
Authorized Representative in the European Community, if the manufacturer is not resident
therein.
Expiration date
Do not use if the packaging is not intact
Catalogue ID code
Sterile device. Sterilization method: ethylene oxide
Sterile device. Sterilization method: ionizing radiation
Production batch number
Serial number
S.2 Symbols used in the user manual to recall the reader’s attention
Important safety information to avoid the risk of injury to the user or patient and / or damage
to the device
Information to which you should pay special attention
Note or information for correct use of device

Rev. 04-2010 USMAN2012-IT
6 - 28
Rev. 10-2018 USMAN500 - UK
WARNINGS, PRECAUTIONS AND IMPORTANT INFORMATION
Read carefully
This User Manual has been prepared using simple, easy-to-understand language. If you have difficulty
interpreting the above, contact the manufacturer for further clarification
•Read all instructions carefully before using the unit. Careful, proper use will ensure smooth operation and protect
both patients and operators alike.
•The unit is designed exclusively to remove non-flammable organic fluids (secretions) during medical procedures.
For this reason, it should only be used by duly trained personnel.
•Never use the suction unit in the presence of flammable and/or explosive liquids, gases and mixtures as this could
lead to explosion and/or fire.
•Using the suction unit under environmental conditions other than those indicated herein can seriously compromise
function and modify its technical parameters (e.g. the maximum suction value or battery life).
•If suction is performed without the collection jar and/or filter in place, or if you suspect that substances may have
entered the suction circuit (i.e. inside the suction unit), immediately contact the nearest service centre or the
manufacturer to have the unit serviced.
•Before cleaning the unit or proceeding with any maintenance, unplug the unit from the external power supply. Do
not submerge in liquids as this could damage the suction unit and cause the safety devices to cut in.
•The unit does not require any maintenance on the operator’s part. The only operations authorized are those listed
herein. For technical support, periodic overhaul and any repairs that may be needed, contact your authorized
service centre.
•The manufacturer provides authorized personnel —who have taken a specific technical assistance training course
—with the documentation necessary to carry out the work (service manual).
•To ensure patient safety, precision of the displayed values and proper unit function, use only original spare parts.
By failing to comply with this warning, the operator assumes responsibility for any patient injury or property
damage.
•The OB500 suction unit does not perform any clinical diagnostics on the patient.
•The device cannot be used to assist patients during investigations through NMR (nuclear magnetic resonance).
WARNING ON REUSE OF DISPOSABLE PARTS
Reuse of disposable materials or parts may compromise the functionality of the device and represent direct and indirect
sources of contamination for users and patients.
Sterilization and / or cleaning of disposable parts (antibacterial filters, disposable bags, suction tubes, Jankauer, etc.)
can cause structural degradation of the material and consequent dangers arising from loss of mechanical integrity.
Tel.: +39 0471 932893
raq@boscarol.it

Rev. 04-2010 USMAN2012-IT
7 - 28
Rev. 10-2018 USMAN500 - UK
IMPORTANT INFORMATION
Note:
The suction unit was designed and tested according to the latest regulatory standards. If the suction unit
is hooked up to a non-compliant electrical system and/or if the work is not performed by professional
installer, both the suction unit and the electrical system could be damaged. Always consult a qualified
technician with knowledge of the latest requirements!
Preventive maintenance and safety inspection:
The device should be checked at least once every 12 months from the authorized service center. Every 24
months an inspection of safety and technical maintenance is required instead. Refer to authorized service
centers for planning of inspection. The periodic safety inspection of the device does not fall under
warranty.
Contamination:
Sending a contaminated suction unit to the manufacturer, installer or service centre is strictly forbidden.
Any device received in such condition will be rejected and health authorities notified of possible
contamination. Here the term contaminated means a suction unit that has not been cleaned of the
secretions aspirated from the patient. If the substances aspirated have entered the suction unit, it must
be discarded. For Oscar Boscarol srl, the safety of its employees and authorized service centre staff is
important. The suction units will not be demolished according to the WEEE Directive (Waste Electrical and
Electronic Equipment) if the suction unit is contaminated and there is the risk of infection (application of
It. Legislative Decree 81).
Operator responsibilities
•The OB500 suction unit is designed for emergency health services and must therefore be ready for use at any
time and in any situation.
•Immediately replace any components/parts that are damaged, altered or missing, and/or for which a unit
malfunction is suspected. Always replace such parts with original spares.
•Dispose of packaging in accordance with current regulations and make certain that it is out of the reach of
children.
•Read these instructions carefully before using the suction unit. Careful, proper use will ensure smooth
operation and protect both patients and operators alike.
•Operate the suction unit only in compliance with the technical specifications laid out by the manufacturer in
this manual.
Intended Use
•The suction unit can be used on all types of patients following the correct medical technique.
•The suction unit is designed to clear the upper airways. Clearing the lower respiratory tract is to be performed
by medical and/or health care professionals trained and authorized to perform this function.
•In some countries, this information must be verified according to the protocols implemented by the local
emergency medical services.

Rev. 04-2010 USMAN2012-IT
8 - 28
Rev. 10-2018 USMAN500 - UK
OB500 SUCTION UNIT
INSPECT THE SUCTION UNIT AND ALL PARTS BEFORE USE.
DO NOT ATTEMPT TO USE IF ANY PARTS ARE DAMAGED OR MISSING.
Description and intended use statement
The OB 500 is a stationary, 12 Vdc operated medical suction unit used to clear the upper airway and restore breathing
either spontaneously or with assistance. The OB500 Medical Suction Unit is specially designed for the installation on
emergency vehicles, for camp hospital and mobile aid centres. Thanks to its flexibility a perfect installation is allowed,
coming as a kit. The device can also be installed in hardly accessible areas (e.g. under a seat or cabinet) while the
regulation part (along with the jar support) may be installed nearby patient’s stretcher. It is provided with an electric
safety that is connected to the 12 Vdc power supply of the vehicle (ambulances, helicopters etc.). The suction unit is
made in compliance with ISO10079-1:2009 and it is not to be considered as a portable one.
Contra indication for use
Do not use the OB500 for draining the thorax.
Controls, indications and check panel
All the controls of the suction unit (except for the main switch) are on the front side of the control module.
The OB500 is powered by the 12 Vdc source of the emergency vehicles.
Model BSU402:
1. Suction unit
2. Control module
3. OB-J FA jar
4. Wall bracket for the jar
Model BSU412:
1. Suction unit
2. Control module
3. SERRES® jar
4. Disposable bag
Model BSU414:
1. Suction unit
2. Control module
3. OB-J jar
4. Disposable bag

Rev. 04-2010 USMAN2012-IT
9 - 28
Rev. 10-2018 USMAN500 - UK
of the wall bracket
Available spaces for the fixing
Square bracket
Control module
The regulation device is supplied already assembled and it is made in plastic material (UL white ABS).
The chassis comes with holes for screw fixing onto the wall. Vacuum regulation may occur through
the proper knob.
The vacuum adjustment is possible by turning the knob. Turning the control knob clockwise will
increase the vacuum. The monitoring of required vacuum is possible by the analog vacuum meter and
is expressed in millibars (mbar) and kilo-pascal (kPa) or millimeters of mercury (mmHg).
To prevent any possible damages on other tubes, electric cables on the back side of the wall, check the
right length of the screws before the installation of the control module. After installation proceed with
the connections of all tubes.
Suction unit
The suction unit is built in shockproof flame-retardant ABS plastic (UL). A separable power cable
and a plastic connector to the pneumatic hose onto the control module are also provided. The
fixing of the suction unit can be done with square metal bracket direct to the case of the unit
and the wall or the floor of the ambulance. Pay attention before the boring, to prevent any
damage to the suction unit. It is better to open the cover of the suction unit, verify the free
space and then bore. Do not use any type of glue, adhesives or rubber solutions. Do not cover
the air outlet on the suction unit. After the installation always check the good closure of the
cover and the fixing of the screws. Do not install the unit near high temperature sources.
Fig. 2
Fig.1

Rev. 04-2010 USMAN2012-IT
10 - 28
Rev. 10-2018 USMAN500 - UK
OB500
to the suction unit
SWITCH
Power central unit of the ambulance
Suction unit
To the control panel
BLUE= MINUS (-)
BROWN=PLUS (+)
WARNING
Open the cover and check the free spaces before the fixing of the square bracket.
Don not use screws more than 20mm long!
Electrical connections
Refer to the electric international norms (for SELV DC voltage), for the installation of the suction unit on the emergency
vehicles. See the figure 3 for the right connections of the power cable from the suction unit. The section of the main
supply power cables should not be less than 1mm ² (each wire).
The power cable of the device is flame retardant and features two wires with a section of 1 mm ². The insulated wires
are of different colors: brown must be connected to the positive 12 Vdc. The switch must
reckon on one nominal power flow rated 5 Adc at least (must be a switch for automotive
12 Vdc). Before the verification of the right functioning of the unit, control the absence of
short circuit on the connections. Never connect the unit direct to the 12 Vdc battery of the
ambulances (protect the power line with right value of fuse and with appropriate switch).
On the figure below the connections of the switch (normally available on the central power
unit of the ambulances), and the wiring on the vehicle.
Fig. 3
Fig. 4

Rev. 04-2010 USMAN2012-IT
11 - 28
Rev. 10-2018 USMAN500 - UK
Test of the unit
The operations described in this section allow the user to verify the correct functioning of the suction unit. The test
should be performed at least once a day and before use in the field:
Daily checklist:
•turn the regulator knob clockwise to maximum;
•cover the patient tube (on the collection jar) with a finger or a lint free cloth;
•check the maximum value of the vacuum (-800mbar +/-10%) on the gauge;
•turn the regulator knob anti-clockwise and check the reading on the gauge.
When finished, compare the results of this test with the value on the table below:
Test –phase
Result
Remedy
Check the maximum vacuum.
Value range between 700 and 800 mbar
(70kPa ÷ 80kPa; 525 ÷ 600mmHg).
Check that the lid on the collection jar is
tight and that all connections are secure.
Change the disposable liner.
Check the vacuum regulation.
Gauge reading ranges from maximum to
minimum.
Check the vacuum connections and/or the
regulator (anti-clockwise for minimum
vacuum).
WARNING
In the event of continued problems after taking the action outlined above, send the unit to an
authorised service centre for service or repair.
Collection jar
The device is sold with three different types of jar with a capacity of 1000ml:
- reusable, autoclavable OB-J FA jar –OB500 FA
- reusable, autoclavable OB-J jar –OB500 FM
- reusable SERRES® jar with disposable 1000ml bag –OB500 LINER
OB-J FA reusable, autoclavable jar
OB-J FA jar (Fig. 5) is made of transparent plastic material (polyprophilen). The jar includes
pot (1), cover (2), the shut-off valve (3) and the 90° plastic pipe (4). The lid of the jar
provides the direct inclusion of the protection filter.
The jar can be sterilized in an autoclave (max. pressure 2 bar (g) and maximum
temperature 121°C) for a maximum of 15 minutes. It is strongly recommended that the
collection jar is replaced after 30 autoclave cycles. The jar must always be used in the
upright position, to avoid the intervention of shut-off valve. Should this occur, switch the
device off and disconnect the patient tube (‘VACUUM’ marked on the lid).
Protection filter
To prevent fluid overflow, a special protection filter is used between the jar and the unit. The filter is
produced with PTFE hydrophobic material which prevents fluids entering the pneumatic circuit. Working
together with the shut-off valve on the jar, the filter isolates the pneumatic suction pump from gas and
fluid substances. Replace the filter when you look at contamination, discoloration or increased
resistance of aspiration.
Warnings and precautions on the use of protection filter
If the suction unit is used either in an emergency or on a patient where the risk of contamination is not
known, the filter must be changed after each use.
If instead the patient's pathology is known and / or where there is no danger of cross contamination, we
recommend replacing the filter after each shift. In case of complete inactivity of the unit is recommended
Fig. 5

Rev. 04-2010 USMAN2012-IT
12 - 28
Rev. 10-2018 USMAN500 - UK
to replace the filter once a month. The material it is made should be damaged even under specific
environmental conditions (humidity, heat, cold)
Do not use the suction unit without the protection filter or jar!
Reusable jar with disposable Liner SERRES®
The OB500 LINER is sold with a 1000 ml SERRES® secretions jar equipped with a disposable
bag (Fig.6).
The jar is made of a transparent plastic material (polycarbonate) and can hold a disposable bag
equipped antibacterial and anti-contamination filter. The jar is equipped with an angular junction
placed laterally which allows it to be connected to the suction unit. The filter integrated into the
disposable bag performs the function of preventing the backflow of aspirated liquids towards the
suction unit when it completely fills up, and to avoid cross-contamination. On the top of the
disposable bag there is a white plastic disposable corner joint, which allows to connect standard
suction catheters or cannulas Jankauer.
OB-J collection jar
OB-J includes the jar (1), adaptor for disposable bag SERRES (2) and the fitting "L" (3). The
disposable bag is provided separately. The integrated filter in the bag performs the function of
preventing backflow of aspirated liquid into the suction unit, when it is completely filled.
The OB-J jar can be sterilized by conventional steam autoclave at a maximum temperature of 121
°C and a pressure of 2 bar (200kPa). Every 30 sterilization cycles, the jar should be replaced.
Jankauer suction tube and Finger-typ end-piece
The suction unit is sold complete with a sterile, Jankauer type suction probe which is connected
to a plastic suction tube. The probe and the tube are disposable and must be changed after
each use. To facilitate correct operation, the probe is angled to allow ease of entrance to the
mouth and the respiratory airway.
A different type of suction set, which includes a silicone tube (length 130 cm –51.2 in)
and one Finger-tip sterile joint, is available at customer request. Finger-typ joint allows
the user to control the vacuum by opening or closing a venturi hole with a finger. When
the venturi hole is completely open, the vacuum will operate at the minimum value, with
the venturi hole completely closed the vacuum will operate at the value set on the unit.
Silicone tube is sterilisable, Finger-typ is disposable.
Wall bracket for the jar
The device BSU402 and BSU414 includes one metal bracket, which is necessary for wall mounting
of the bottle jar. The bracket should be installed so as to keep the bottle upright.
The wall bracket is provided with two holes allowing vertical and horizontal adjustment. The best
position, both for its usage and for cleaning and sterilization activities, lies under the control
module. The silicone cable provided with the device may be properly shortened depending on wall
brackets distance from the wall itself.
Fig.6
Fig.7

Rev. 04-2010 USMAN2012-IT
13 - 28
Rev. 10-2018 USMAN500 - UK
OB500 LINER includes one plastic bracket. The fixing procedure is the same as for OB500 FA.
Power supply
The external power source of the OB500 should be between 12÷15Vdc. Insert in series to the plus wire an appropriate
protection fuse (min. T15A). Check the correctly connections and the insulation of the cable every 2 months. Do not use
the unit if not fixed and not connected to the right power source (12Vdc).
WARNING
Check that the external power source is correctly rated. The external power source has to be protected
by a security fuse with a minimum rating of T15A (delayed type). The mains voltage can be deadly to
humans!
MAINTENANCE AND REUSE
After each reuse
After each use, switch off the suction unit; disconnect the single patient use - disposable parts (e.g. Finger-typ, catheters,
liner etc.) and dispose of them according to local regulations. Check that the suction unit operates correctly - check the
connection tubes and the controls on the front side for damage. Further, inspect the chassis for possible signs of
damage. Clean and disinfect the unit following the instructions in paragraph “Cleaning the unit”. Replace all single use
- disposable parts with new components and recharge the battery. Make a daily function test as described under "Daily
checklist".
Occupational safety and PPE (Current legislation concerning workplace safety and health)
The decontamination process is always a delicate process, which implies a specific training, especially in the field of
emergency, where the patient’s medical condition and contamination level is mostly unknown. For this reason, the user
must always use personal protective equipment (PPE) to protect himself and other people.
In case a proper PPE is not available, please contact the person who is in charge for the workers’ safety in your company
or organisation.

Rev. 04-2010 USMAN2012-IT
14 - 28
Rev. 10-2018 USMAN500 - UK
The organic secretions collected by the device’s jar might cause serious
infections to the users. For this reason, suitable PPE and disinfection
products must always be used, applying what has been established by the
competent authorities.
CLEANING INSTRUCTION
Reusable collection jar OB-J FA
1. Wear gloves and protective clothing.
2. Remove yellow angular connector with patient tube. If the tube is equipped with a Yankauer probe, it must be
disposed of together with the curved probe (disposable devices). The angular connector can be sterilized.
3. Disconnect the protection filter with silicone tube and conical connector.
4. Remove the lid of the bottle (attention to possible contamination with the liquids contents!). Empty the
contents of the bottle in accordance with local regulations or hospital practice.
5 –8 Disassemble the shut-off valve.
Dispose of disposable parts and disassembled the jar, soak the parts in cold water and rinse well. Then dip the same
parts in hot water at a temperature not exceeding 60 ° C containing a mild, non-alcoholic detergent. After washing, rinse
all parts with clean hot water (30-40°C max.) and then dry with a lint free soft cloth. Before reassembling, check that all
parts are clean, dry and undamaged.
Decontamination of the collection jar
The collection jar can be disinfected with any mild, non-abrasive detergent. Do not use any alcohol or solvent-based
cleaning agents as these may damage the collection jar. Do not use coloured disinfectant solutions as these may stain
the collection jar. Never use neat disinfectant dilute according to the manufacturer’s instructions. The jar is
manufactured from plastic and, whilst it offers good protection against water and humidity, long periods of immersion
in water or detergents may affect its integrity. For sterilisation, only use steam autoclaves at a maximum temperature
of 121°C at a maximum pressure of 2 bar (g) for 15 minutes. The jar should be placed in the autoclave upside-down to
assist draining. At the end of the autoclave cycle, place all parts on a flat, sterile (or clean) surface and leave to cool to
ambient temperature before reassembling. Check that the collection jar and all parts are undamaged.
WARNING
•Do not put weight on the parts during the sterilisation cycle.
•Observe the maximum limits for temperature, pressure and duration during the autoclave cycle.

Rev. 04-2010 USMAN2012-IT
15 - 28
Rev. 10-2018 USMAN500 - UK
•Never exceed the value of 60°C for washing or cleaning operations (with the exception of
sterilization in a steam autoclave).
•Cleaning, disinfection and sterilisation should only be carried out by trained personnel.
•The collection jar can be used for up to 30 sterilisation cycles, after which, the jar, tubes and plastic
connectors must be replaced.
•After reassembling the jar, check that the lid is properly fitted to achieve a seal.
Place all components on a flat, clean surface. During the assembly phases, check all parts for damage. Ensure that the
shut-off valve moves freely inside its cage. To remove the filter, pull gently and twist holding the outer rim of the filter
whilst supporting the collection jar by the lid. Take care not to damage the filter.
Replacing the protection filter
Disconnect the tube from the top side of the contaminated filter. Carefully remove
the filter from the collection jar lid and dispose of it in accordance
with local regulations or hospital practices. Install a new filter
ensuring that the written “IN” is on the underside, above the
‘VACUUM’ port on the lid. Incorrect installation of the filter will lead
to immediate failure when used.
NOTE
Ensure that the protection filter is correctly fitted to the collection jar (with written “IN” to the jar). The
filter may break if fitted incorrectly and there is a risk of unit contamination.
Reusable collection jar OB-J
The jar OB-J LINER is equipped with a specific disposable bag, homologated for this type of application. Unlike the OB-J
FA version, the disposable filter is inside the bag.
1. Wear gloves and protective clothing.
2. Remove white angular connector with patient tube. If the tube is equipped with a Yankauer probe, it must be
disposed of together with the curved probe (disposable devices).
3. Close the connection <PATIENT> with the plug provided on the lid.
4. Remove the sealed liner from the jar.
5. Disconnect silicone tube from the jar.
6. Remove adapter for disposable bag.
7. Disassemble the red angular connector.

Rev. 04-2010 USMAN2012-IT
16 - 28
Rev. 10-2018 USMAN500 - UK
Dispose of disposable parts and disassembled the jar, soak the parts in cold water and rinse well. Then dip the same
parts in hot water at a temperature not exceeding 60 ° C containing a mild, non-alcoholic detergent. After washing,
rinse all parts with clean hot water (30-40°C max.) and then dry with a lint free soft cloth. Before reassembling, check
that all parts are clean, dry and undamaged.
If it is necessary to sterilize the jar, proceed as described on the previous page “Decontamination of jar”.
Reusable collection jar SERRES®
The suction cannister and its grey angle connector can be washed (85°C) and autoclaved (121°C). Before washing,
disconnect the angle connector.
Wear the protection gloves.
1. Disconnect the patient tube from the suction unit.
2. Disconnect the grey connector from lid.
3. Disconnect the white connector and close the connection with the plug provided on the lid.
4. Remove the sealed liner from the jar.
Clean the canister and grey connector.
Reassembly of the jar
Extract a new disposable bag from the packing, stretch it (picture 1) and insert it into the jar (picture 2). Insert the
complete jar into the bag and connect the tube to the suction unit. Activate the suction unit. Close with a finger the
connector <PATIENT> and, at the same time, press lightly the bag from the center of the lid (picture 3). Make sure the
bag is completely swollen. Connect the patient tube (Jankauer) to the connector <PATIENT>.
NOTE
The collecting bag must be inserted into a rigid container of the same size.
It is strictly forbidden to use defective products!
Fig. 1
Fig. 3
Fig. 2

Rev. 04-2010 USMAN2012-IT
17 - 28
Rev. 10-2018 USMAN500 - UK
Disposal of contaminated parts
Always follow local regulations or hospital practices when dealing with contaminated materials. Never store
contaminated parts with new or sterile parts.
Cleaning the suction unit
To clean the chassis of the suction unit, use a damp cloth with mild detergent. Rinse the chassis with a damp cloth and
dry it with a soft towel.
WARNING
•Before cleaning the suction unit, ensure that all external power supplies have been disconnected.
•Never immerse the suction unit in water.
•Do not use abrasive substances, alcohol or solvents to clean the chassis.
To correctly disinfect and decontaminate the suction unit, we recommend using specific, approved products. These
disinfectants must be free of alcoholic and/or abrasive substances. Oscar Boscarol srl can provide specific materials for
disinfection of medical equipment, including our suction units. These disinfectants, available in different formats (wipes,
spray, liquids), have been laboratory tested and guaranteed to deactivate viruses, bacteria and microorganisms. When
used periodically, they destroy and prevent the formation of dangerous biofilms (superficial layers that easily host
bacteria, moulds, viruses and microorganisms). Our disinfectants do not contain alcohol, chlorine, phenols, aldehydes
and halogens.
Safety
The suction unit OB500 is equipped with some electrical safety devices to protect it from any failure or by bad
connection. All protections are placed inside and there is no replacement by the user (security automatic reset). The
intervention of the same, and thus the failure of the device must provide for the immediate deployment of the device
to the service.
Disposing of the suction unit
The unit contains electrical and/or electronic equipment that must be recycled according to EC Directive
2012/19/EU –Waste Electrical and Electronic Equipment (WEEE) enacted in Italy with Leg. Decree
49/2014. The suction unit is also compliant with Directive 2011/65/EC which restricts and prohibits the
use of certain hazardous substances in electrical and electronic equipment. Harmful
substances that violate the above Directive are not used in the production and
assembly of electronic boards or in the wiring and connection of electric cables.

Rev. 04-2010 USMAN2012-IT
18 - 28
Rev. 10-2018 USMAN500 - UK
ACCESSORIES AND SPARE PARTS
Index code
Description
User parts
BSU730
Protection filter for OB-J FA jar –5 pcs
BSU732
Protection filter for OB-J FA jar –15 pcs
BSU734
Protection filter for OB-J FA jar –40 pcs
BSU705
Disposable 1000 ml Liner –6 pcs
BSU706
Disposable 1000 ml Liner –12 pcs
BSU707
Disposable 1000 ml Liner –36 pcs
BSU500
Autoclavable OB-J FA jar, without protection filter
BSU506
Autoclavable OB-J jar, without disposable bag
57308
SERRES® jar, without disposable bag
126140107191
Jankauer suction tube
BSU750
End-piece sterile disposable Finger-typ –5 pcs
BSU752
End-piece sterile disposable Finger-typ –15 pcs
BSU754
End-piece sterile disposable Finger-typ –50 pcs
11214101003
Sterile suction catheter Ch.10 black
11214101004
Sterile suction catheter Ch.12 white
11214101005
Sterile suction catheter Ch.14 green
11214101006
Sterile suction catheter Ch.16 orange
11214101007
Sterile suction catheter Ch.18 red
11214101008
Sterile suction catheter Ch.20 yellow
Spare parts
ASP75300
Wall holder for jar
57820
Wall holder for jar SERRES®
BSU902
Silicone patient tube - length 130cm / 51,2inch (int.diam.6mm/ext.12mm)
SPS4003
Suction unit 2P
SPS4000
Control module
SPS4015
Power cable 2P
SPS4050
Silicone tube set (1x length 120 cm, 1x length 150 cm)
SPS6000
Bottle OB-J FA without cover
SPS6002
Shut-off valve –3 pcs
SPS6004
90° plastic joint for OB-J FA jar –3 pcs
SPS6006
Cover for OB-J FA jar complete with shut-off valve and 90° connector
SPS6014
Conical connector –3 pcs
SPS5092
Red angular joint for OB-j jar –3 pcs
-
User manual –available on the website www.boscarol.it
NOTE
The parts listed in the list may be changed without prior notice from the manufacturer and / or technical
improvements. Contact the manufacturer for more information.

Rev. 04-2010 USMAN2012-IT
19 - 28
Rev. 10-2018 USMAN500 - UK
SERVICE
There are no serviceable components inside the suction unit. Do not open the suction unit and do not modify any
electrical or mechanical parts. Always contact your service center or the manufacturer. Any even minimal intervention
on the device voids the warranty. Unauthorized intervention may compromise its conformity with the applicable laws
and reduce the security to the detriment of users and patients.
FAULT FINDING
Malfunction
Possible cause(s)
Corrective action
High noise, low suction, high vibration.
•Internal pump damaged.
•Refer to authorised service
personnel.
The unit turns on but no suction is
performed.
•Internal pump damaged.
•Refer to authorised service
personnel.
The vacuum power on the patient side is
either very low or absent.
•Vacuum regulator set to minimum.
•Protection filter blocked or damaged.
•Connection tubes blocked, kinked or
disconnected.
•Disposable liner not inserted
properly.
•Shut-off valve blocked or damaged.
•Disposable liner full of fluid
substances, protection filter blocked.
•Pump motor damaged.
•Turn the vacuum regulator clockwise
and check the value of the vacuum
on the gauge.
•Replace the filter.
•Replace or reconnect the tubes,
check the jar connections.
•Re-insert the liner in the jar correctly.
•Replace the reusable jar.
•Empty the jar, or disconnect the tube
from the jar and unblock the shut-off
valve. The jar will only work in the
upright position.
•Replace the disposable liner.
•Refer to authorised service
personnel.
Vacuum is always at maximum even if the
jar is removed.
•Fault on the internal pneumatic
circuit.
•Refer to authorised service
personnel.
When turning on the device, electric
safety arrangements always interfere.
•Internal pump damaged or short
circuits.
•Refer to authorised service
personnel.
NOTE
In presence of anomalies or faults other than those listed in the table above, contact the authorized
service center or the manufacturer of the device.
TECHNICAL DATA AND CONFORMITY TO INTERNATIONAL LAW
Classification according to the MDD93/42/EEC
Medical suction unit suitable for fixed use, in accordance with ISO10079-1:2009.
MDD Classification:
IIb
Vacuum degree:
HIGH VACUUM-HIGH FLOW
Mode of operation:
TEMPORARY (maximum continuous use 60 minutes)
Electrical requirements:
SELV (12÷15 Vdc)
Use of the device in the health emergency:
complying to IEC60601-1-12

Rev. 04-2010 USMAN2012-IT
20 - 28
Rev. 10-2018 USMAN500 - UK
Degree of protection against electric shock (IEC60601-1):
TYPE BF
With respect to protection from electric shock:
CLASS II
Degree of protection against ingress of liquids (IEC529):
IP32d
Accordance with general IEC60601-1:
Complying with the 3° Edition
Dimensions
Suction unit’s max dimensions:
188 mm (l) x 188 mm (h) x 90 mm (p)
Control modules max dimensions:
90 mm (l) x 130 mm (h) x 85 mm (p)
Weight:
2Kg max. complete with all accessories
Tolerance on all values:
±5%
Technical data
Max vacuum power:
800 mbar (80 kPa, 600 mmHg) ±10%
Vacuum Regulation:
linear
Vacuum range regulation:
30÷800 mbar (3÷80 kPa; 225÷600 mmHg)
Max flow rate:
30 litres per minute with free air ±10%
Approximate maximum noise energy:
70 dBA
Accuracy of the analogue gauge:
±5%
Power supply
Running:
12÷15 Vdc (Direct Current)
Max current load:
70 W (max. current 6 A)
Electrical safety devices:
Internal not accessible to the user
NOTE
The external power source must supply at least 6A to allow correct running.
Environmental and transport conditions
Operating temperature range:
0÷50 °C (32÷122 °F)
Storage and transport temperature range (with original
packaging):
-25÷55 °C (-77÷131 °F)
Storage and transport temperature range (without packaging):
0÷50 °C
Storage and transport relative humidity:
15÷95%, not condensed
Storage and transport atmospheric pressure:
70÷106 kPa (700÷1060 mbar; 525÷795 mmHg)
NOTE
Be aware that if the unit is operating at an altitude over 2,500 meters, the pressure value can slightly
decrease. This is caused by the decrease in the atmospheric pressure.
This manual suits for next models
3
Table of contents
Other Boscarol Medical Equipment manuals

Boscarol
Boscarol OB User manual

Boscarol
Boscarol OB 1000 User manual

Boscarol
Boscarol OB500 FA User manual

Boscarol
Boscarol OB 2012 FA User manual

Boscarol
Boscarol OB 2012 FA User manual

Boscarol
Boscarol PAL32000 User manual

Boscarol
Boscarol OB1000 Series User manual

Boscarol
Boscarol OB OXIKIT PLUS User manual

Boscarol
Boscarol OB WB User manual

Boscarol
Boscarol OB1000 FA User manual