5. ASSEMBLY
Connect rubber tubes from the cuff with the manometer and withthe inflation bulb with regulating valve.
6. USE INSTRUCTIONS
Insert the infusion cuff into the built-in pouch, if any doubt arises refer to the picture available in the manual.
Close the valve situated on the bulb (clockwise) and inflate the cuff till the required pressure is reached (it is possible
to control the applied pressure value on the manometer).
The value must be chosen depending on the medical substance which is to be injected and its required speed.
If necessary, release the faucet on the bulb to reach the required pressure.
In order to quickly deflate the cuff, open the faucet on the bulb entirely.
ATTENTION! Please always rely on qualified and trained personnel!
7. CLEANING
MANOMETER: Only use soft and dry cloth.
INFLATIN BULB: Clean with soft and moist cloth and dry accurately ( do NOT autoclave it)
FABRIC POUCH: Leave out the cuff from inside the pocket. Clean the cloth part with warm water and neutral soap (NO
washing machine, NO spin)
BLADDER: Clean it with warm water and neutral soap paying attention that water does not permeate. Leave it to dry
at room temperature
8. MAINTENANCE
Boscarol medical devices are accurately checked before commercialized and provided with CE mark.
Before every use, make sure the cuff is completely empty and verify that the pointer on the manometer display is at
zero position whiting tolerance of +/- mmHg.
Before utilize, in non-use conditions, always verify that pointer is acting inside tolerance area of ±3 mmHg
(recognizable by a small red line placed perpendicularly to the notches of graduate scale in proximity of zero value).
Manometers out of ±3 mmHg tolerance need to be recalibrated. Please refer to the manufacturer for technical
assistance.
9. WARRANTY
The device is indicated for professional use and has a 12 months warranty from the date of purchase. The
manufacturer, or the authorized service center, owns the right to repair or replace the parts affected by
manufacturing or material faults, or to replace the entire device to its at their discretion.
Warranty is not applicable in case of damages caused by improper use, abuse, alteration or interventions performed
by non-qualified unauthorized personnel.
With respect to this, please send the device to be repaired, together with proof of purchase, to the manufacturer or to
the service assistance center stating the encounter failure/s.
The life of the device is 3 years if used and stored as indicated by the manufacturer.