(70kPa ÷ 80kPa; 525 ÷ 600mmHg).
tight and that all connections are secure.
Change the disposable liner.
Check the vacuum regulation.
Gauge reading ranges from maximum to
minimum.
Check the vacuum connections and/or the
regulator (anti-clockwise for minimum
vacuum).
WARNING
In the event of continued problems after taking the action outlined above, send the unit to an
authorised service centre for service or repair.
Collection jar
The device is sold with two different types of jar with a capacity of 1000ml:
- reusable, autoclavable jar (OB1000FA)
- reusable jar with disposable 1000ml Liner SERRES (OB1000 LINER).
OB-J FA reusable, autoclavable jar
OB-J FA jar (Fig. 3) is made of transparent plastic material (polyprophilen). The jar
includes pot (1), cover (2), the shut-off valve (3) and the 90° plastic pipe (4). The lid of
the jar provides the direct inclusion of the protection filter.
The jar can be sterilized in an autoclave (max. pressure 2 bar(g) and maximum
temperature 121°C) for a maximum of 15 minutes. It is strongly recommended that
the collection jar is replaced after 30 autoclave cycles. The jar must always be
used in the upright position, to avoid the intervention of shut-off valve. Should this
occur, switch the device off and disconnect the patient tube (‘VACUUM’ marked on
the lid).
Protection filter
To prevent fluid overflow, a special protection filter is used between the jar and the unit. The filter is
produced with PTFE hydrophobic material which prevents fluids entering the pneumatic circuit.
Working together with the shut-off valve on the jar, the filter isolates the pneumatic suction pump from
gas and fluid substances. Replace the filter when you look at contamination, discoloration or
increased resistance of aspiration.
Warnings and precautions on the use of protection filter
If the suction unit is used either in an emergency or on a patient where the risk of contamination is
not known, the filter must be changed after each use.
If instead the patient's pathology is known and / or where there is no danger of cross contamination, we
recommend replacing the filter after each shift. In case of complete inactivity of the unit is recommended to
replace the filter once a month. The material it is made should be damaged even under specific
environmental conditions (humidity, heat, cold)
Do not use the suction unit without the protection filter or jar!!
Reusable jar with disposable Liner SERRES®
The suction unit OB1000 LINER is sold with a 1000 ml SERRES® secretions
jar equipped with a disposable bag.
The jar is made of a transparent plastic material (polycarbonate) and can hold a
disposable bag equipped with antibacterial and anti-contamination filter. The jar is
equipped with an angular junction placed laterally which allows it to be connected to
the suction unit. The filter integrated into the disposable bag performs the function of
preventing the backflow of aspirated liquids towards the suction unit when it
completely fills up, and to avoid cross-contamination. On the top of the disposable
bag there is a white plastic disposable corner joint, which allows to connect standard
suction catheters or cannulas Jankauer.