Boyd S2914 Dimensional drawing

1
SCHM-003
Rev 2
Product Informaon Manual
S2914 Surgery Chair
12900 44th Street North
Clearwater, Florida 33762
800-255-2693 * 727-561-9292
FAX: 727-561-9393 SALES
FAX: 727-573-1682 SERVICE
www.boydindustries.com
S2914

2
SCHM-003
Rev 2
Table of Contents
Title Page
Table of Contents
Introducon
Applicaon of Product
Warranty Informaon
Safety Warning Symbols
Safety Tips / Precauons
About Your Chair
Descripon of Product Line
Standard Safety Features
General Product Informaon
Product Specicaons
Transport & Disposal
Unpacking and Power Requirements
Installaon Instrucons
Chair Installaon
Inializaon Procedure
Operaon Instrucons
Programming Instrucons
General Operang Instrucons
General Maintenance and Care
Cleaning Instrucons
Service Informaon
Upholstery Installaon and Removal
Troubleshoong Guide
Authorized Representave (Europe Only)
1
2
3
4
5
6-9
10
11
12-15
16
17
18
19

3
SCHM-003
Rev 2
Intended Use and Application of Product
This product is designed to comfortably posion paents during dental surgical procedures. The chair back and base are adjusta-
ble to allow the most ergonomic working posion for the doctor and their assistant(s).
Warranty Statement
All Boyd manufactured products are warranted against defects in material or workmanship only. No other warranes are
expressed or implied. This warranty shall extend for three (3) years for parts and one (1) year for labor. The Boyd warranty only
covers products manufactured by Boyd. Products distributed, not manufactured, by Boyd carry the stated manufacturer’s
warranty.
The warranty period commences upon invoice date of the equipment. Wrien noce of breach must be submied to Boyd
Industries, Inc. within this period. Boyd reserves the right to repair or replace, at its sole discreon, the item in queson.
Warranty is voided if items are misused or product is improperly maintained or installed. The warranty does not cover any
alteraon of the product from its original condion or if product is used in any way other than the product’s design intent.
An authorized service technician must perform all service and maintenance. Any warranty labor must be rst approved by Boyd.
Failure to coordinate warranty service may result in denial of warranty repair. Wrien log and receipts for service must be
maintained to validate the warranty on Boyd manufactured products.
No claim for consequenal damages will be allowed. Freight damage or mishandling by any third party will not be considered a
warranty item. Improper installaon by a third party causing damage to the equipment is not a warranty item.
Introducon
Congratulaons
on your purchase of the Boyd Surgery Chair
Boyd Industries, Inc. has spent years designing and developing the best dental products available in the world today. The
following features make your Boyd chair the most advanced and funconal surgery chair available in the market:
All steel frame construcon
Motorized back, toe and canlever li base
Motorized lt control for Trendelenburg posioning
14” vercal stroke
Three (3) operator designated pre-set posions
Auto return to HOME switch
Choice of extendable headrests
Snap in / snap out upholstery
Smooth, quiet, low voltage D/C motors
Swing out Armrest
Hand Control (Oponal)

4
SCHM-003
Rev 2
Safety Warning Symbols
Cauon
Informaonal: Read Technical
Manual Before Servicing
Main Power On
Main Power O
Protecve Earth Ground
Type B Equipment
General Warning Symbol
Light

5
SCHM-003
Rev 2
1. Medical Electrical equipment requires special precauons regarding EMC and needs to be installed and put into service according to
the EMC standards.
2. Portable and mobile RF communicaons equipment can aect medical electrical equipment.
3. The use of accessories, transducers, and cables other than those specied, may result in increased emissions or decreased immunity
of the S2914 Chair.
4. Read instrucons:
All safety and operang instrucons should be read before this product is operated.
5. Follow instrucons:
All use and operang instrucons should be followed.
6. Owner’s manual:
Retain safety and operang instrucons for future reference.
7. Heed warnings:
Adhere to all warnings labeled on equipment and in owner’s manual.
8. Accessories:
Use only accessories from Boyd Industries, Inc. catalog on this chair. Third party accessories will void the warranty and may be
unsafe.
9. Cleaning:
Use mild detergent in water on all painted surfaces and plasc parts. Use only cleaners specically designed for cleaning
disinfecng vinyl upholstery. Refer to vinyl manufacturer’s documentation for specific cleaning information.
10. Line cord:
This product is equipped with a polarized, grounded, alternang current line plug. This plug will t into the power outlet only
one way. This is a safety feature. If you are unable to insert the plug into your outlet, contact a local electrician to correct the
problem with the A/C outlet.
11. Power cord protecon:
Power supply cords should be routed so that they are not walked on or pinched by any other equipment placed upon or against
them and will allow access to disconnect the device. Accessory cords aached to outlets on the chair should be clear from
pinching during operaon as well as paent’s entrance and exit of the equipment.
12. Liquid and object entry:
Do not intentionally pour liquids of any kind on this equipment or push objects of any type into openings as they could touch
moving parts or dangerous voltage points inside equipment which could cause fire hazard, electric shock and/or bodily injury.
13. Servicing:
Do not attempt to service this product yourself. This action will void any warranties. Opening covers could expose you to
dangerous current and mechanical pinch points. Service should only be performed by qualified service personnel. Contact Boyd
Technical Services Department for authorized technicians in your area.
14. Damage or failures requiring service:
Unplug equipment from power source and refer to a qualified service technicians if any of the following occur:
Power supply cord is damaged.
Liquid has been spilled into equipment.
Objects have fallen into equipment.
Equipment has been exposed to rain or water.
Equipment has been dropped or damaged.
Equipment shows significant change in operation.
Equipment does not operate normally.
15. Replacement parts:
Contact Boyd Technical Services Department to order approved replacement parts. Unauthorized substitutions could result in
electrical shock, fire hazard and/or bodily injury.
16. Safety checks:
Upon the completion of any service to equipment, service personnel must perform safety checks to insure that all equipment
is functioning properly and safely.
17. Heat / Open flame:
Equipment should be installed away from heat sources such as radiators, heat registers, stoves, or other products that produce
heat or open flames. There should be no smoking in a room where this equipment will be operated or stored.
Safety Tips and Precauons
This product has been designed and tested with your safety in mind. Improper use can result in electric shock or other physical
injuries. Do not defeat the safeguards that have been designed into this product. Please read and observe the following safety points
when installing and operang this product. Save this manual and all supplemental literature for future reference. Observing
precauons listed in this booklet will help you get many years of service and safe operaon from your equipment.

6
SCHM-003
Rev 2
Descripon of Product Line
Model: S2914
Features: Power Back
Power Toe
Power Tilt
Canlever Style Li Base
Hand Controls (Oponal)
Foot Control with Programming
Standard Safety Features
About Your Chair
1. Power Supply Chair is circuit breaker protected.
2. Hand Controls Hand Controls are oponal and connected using a coiled cable. Pressing any motor control funcon switch
will cancel any automac movement funcon.
3. Foot Controls Foot Control is integrated into the base assembly. Pressing any motor control funcon switch will cancel
any automac movement funcon. Lockout funcon disables all funcons safely hold chair posion.
4. Covers and Guards Covers and guards are designed to prevent possible injury from moving parts.
5. Base Safety Boom cover on the base li arm assembly is a safety stop device. Any obstrucon under the li arm
assembly will trip the safety system which will stop all chair movements. Chair will not operate unl
the obstrucon is removed.

7
SCHM-003
Rev 2
General Product Informaon
Frame Welded steel with powder coated nish
Upholstery Standard upholstery is vinyl
Upgrades are available
Movement CB Base: Canlevered Li - 14.00” (355mm)
Back: Electric Actuator
Tilt: Electric Actuator
Toe: Electric Actuator
Net Weight S2614: 275 lbs (125 kg)
S2615: 260 lbs (118 kg)
Electrical 24V D/C low voltage motors
Microprocessor controlled
Chair operates at 24V D/C
Power Supply 115VAC 60Hz 10A
230VAC 50-60 Hz 7A
(Refer to Serial Number Scker for required voltage)
Protecon Class: I
Electrical Shock Type: B
Liquid Protecon Ordinary equipment (IPXO)
Flammable Gases Not suitable for use in the presence of ammable gases or where such gases may accumulate in
concentraon. (Closed Space)
Operaon Mode Intermient duty cycle:
10% - 2 minutes On / 18 minutes O
Indoor use only
Operang Temperature: 5° C to 40° C (41° F to 104° F)
Temperature Relave Humidity: 20% to 90% @ 30° C - Not condensing
Atmospheric Pressure: 97kPa
Storage / Transportaon Temperature: -10° C TO 50° C (14° F TO 122° F)
Temperature Relave Humidity: 20% to 90% @ 30° C - Not condensing
Atmospheric Pressure: 45 to 115kPa
Weight Capacity 500 lbs (228 kg) Subject to normal weight distribuon
Note: Do not sit on back, foot end of seat or adjustable toe secon. These secons are not designed
to support the full weight of paent. Do not li or slide chair using back secon or seat upholstery.
Warning: Do not modify this equipment without authorizaon from Boyd Industries, Inc.
About Your Chair (connued)

8
SCHM-003
Rev 2
Product Specicaons
S2914
About Your Chair (connued)
19.90 to
top of seat
frame
33.90 to
top of seat
frame
29.60
63.90
23.25
SIDE VIEW
Lowered
Raised
Note: Headrest height is adjustable
approximately 6” plus arculang movement
TOP VIEW
64.00
29.50
34.00
20.00

9
SCHM-003
Rev 2
Transporng of Chair
When transporng the chair:
Chair base must be in full down posion, and the chair back must be in full up posion (Home Posion).
Chair frame must be secured to the chair base.
Remove the chair power cord.
Secure or remove dental light and any accessories.
Move equipment by gently lng it on the side edge of the oor plate by liing on under side of seat only enough to be able
to slide onto proper moving equipment. Do not use back secon or seat upholstery to li or move chair. Never aempt to
move chair alone and always use proper liing techniques to avoid physical injury.
Chair should be secured to transporng equipment and/or transporng vehicle.
Transporng and Storage Temperatures for chairs to be kept between -10° C TO 50° C (14° F TO 122° F).
Disposal of Equipment
Disposal of Boyd equipment:
Boyd chairs are designed and built to last and provide ecient and safe service for many years. The expected service life of this
chair is 15 years. When the me nally comes to rere the chair, please dispose of the equipment according to statutory and
regulatory requirements. Circuit boards and electrical cabling should be recycled as electrical salvage. Plasc parts should be re-
cycled accordingly. Any material unsuitable for recycling should be disposed of appropriately.
About Your Chair (connued)

10
SCHM-003
Rev 2
Unpacking
Preassembly Check: Open shipping cartons carefully and check to see that all parts are received and have not been
damaged during shipping.
Unpacking Chair: Carefully remove two (2) fasteners securing chair to pallet using nut driver supplied with order.
Slide chair from pallet to proper moving equipment and secure to move chair to nal installaon
locaon. Take care using any sharp objects or tools to avoid damage to upholstery or painted
parts.
Cauon: Chair should be lied from below main seat frame. Do not at anyme move chair by liing on
back or seat upholstery. Use proper liing techniques to avoid injury.
Contents:
Complete chair: Chair and Base assembled with upholstery
Headrest
Four (4) anchors (Used to secure to oor)
A/C power cable
Accessories Opons (I.E. Arms, Light post, etc.)
Owner’s / Operator’s manual
(Note: Accessories may be packaged separately)
Power Requirements
Check chair idencaon label located on electrical panel for required voltage specicaons. Connect chair to
hospital grade receptacle only.
Warning: To avoid risk of electric shock, this equipment must only be connected to a supply main
with protecve earth.
115VAC 60 Hz 10A
230VAC 50-60 Hz 7A
Unpacking and Power Requirements
Cauon: Do not plug external devices or equipment to the Mul-Socket Outlet located in the seat. Our Mul-
Socket GFI Outlet is a low amperage circuit not designed to handle heavy loads which can result in breaker shut
down and reduced safety when higher amperage circuits are applied. This dual outlet is designed for use with
Boyd Light Accessory only.

11
SCHM-003
Rev 2
Locaon: Full size templates are available upon request for laying out location of chair and utilities (if required) prior to actual
installation.
Cauon: Base must be anchored to floor. Failure to do so may result in serious bodily injury to operator and/or patient and
could cause unwarranted product damage. Chairs not properly anchored do not meet warranty requirements.
1. Posion chair as desired. Drill holes in oor through base plate
mounng holes. Anchor base to oor with four (4) anchors
provided.
Cauon: To prevent possible injury or damage, do not drill through
carpeting. Cut away carpeting at anchor hole locations.
2. Make ulity connecons in oor in front of base (if required).
3. Ensure safety covers are in place and removable upholstery is
secure. Shiing may have occurred during handling or shipping.
4. Plug chair into approved A/C outlet. Refer to Serial Number Scker
for required voltage.
5. Turn main power switch ON (C1) located on the electrical panel at
the back of base.
6. Perform initialization procedure. (see below)
7. Perform safety and operaonal check.
8. If chair does not operate correctly, see the “Troubleshoong”
secon of this manual.
9. When chair is operang correctly, turn main power OFF (C1) unl
chair is to be used.
10. Refer to “Upholstery removal and installation” for upholstery use.
Inializaon Procedure
1. Run the back motor to its full down position.
2. Run the base motor to its full up position.
3. Run the tilt motor to its full tilt position.
4. Run the toe motor to its full up position.
5. Press and release HOME button and allow both motors to fully return to the patient exit position (back full up, base full
down, tilt up and toe down).
Chair is now initialized and ready for use.
Installaon Instrucons
Chair Installation
Model: S2914
Back View
of Chair
Main Power
Switch (C1)
Serial Number
Scker
Light Power
Switch (C2)
Anchors
4 places
Foot Control
Connecon
Front View
of Chair
Hand Control
Connecon
A/C Input
to Chair
Home
Base Up
Hand
Control
Back Down
Tilt Back
Toe Up
(S-2614)
Foot Control
S2914
Toe Up
Tilt Up Back Down
Base Up
Home

12
SCHM-003
Rev 2
Model: S2914
Programming Instructions:
1. Read these instrucons completely before you begin programming.
2. Using the motor controls, posion the chair as desired.
3. Wait at least three (3) seconds and push and hold the save (S) buon for
four (4) seconds. The system will issue a confirmation beep. Within two
(2) addional seconds, push the buon of the desired posion (i.e.
“1”,”2”,”3”). The system will issue a second confirmation beep. This will
automacally replace any pre-programmed posion.
4. Reposion chair and recall that posion to insure that it has been
programmed correctly.
5. Repeat steps 2, 3, and 4 unl all desired posions are programmed. You
can save up to three (3) posions.
6. HOME is a preset posion that cannot be changed.
The programmed posions can be aained at any me and from any posion.
General Operating Instructions:
1. Before operang chair, watch for any obstrucons under chair.
Do not allow small children to play on, around, or underneath
chair or base.
Do not allow service animals on or underneath chair or base.
Any obstrucon that may hinder normal movement of any part of
the chair should be cleared away before inializing any
posioning of the chair.
Check to be certain that all people are clear of the chair before
operang chair funcons.
2. Chair should be turned OFF at main power switch (C1) any me it is not
being used.
Operang Instrucons
Outlet for use with Boyd Light Accessory
Only and may not be repositioned.
Improper use could lead to injury or
equipment damage.
Foot Control
Save
Posion 2
Posion 3
Position 1
S2914
Switched
Outlet
Active
Outlet
230V
Outlet
115V
Outlet

13
SCHM-003
Rev 2
Model: S2914
Note: Chair movement can be stopped by depressing any direconal control (i.e. back up / back down,
base up / base down, or any programmed posion).
Adjustments to chair posioning can be made anyme using the “Up / Down” buons or pedals.
Hand & Foot Control
Operang Instrucons (connued)
BASE UP/DOWN:
Press and hold to move chair
upward or downward
BACK UP/DOWN:
Press and hold to move back
upward or downward
POSITION #1:
Press to move to position #1
See separate instructions
POSITION #2:
Press to move to position #2
See separate instructions
POSITION #3:
Press to move to position #3
See separate instructions
HOME:
Returns chair and base to
Original entrance / exit
position
(back up / base down)
Press once to activate.
Hand
Control
TOE UP/DOWN:
Press to hold to move toe
upward or downward
TILT UP/DOWN:
Press to hold to tilt
upward or downward
SAVE:
Used for programming
See separate instructions
Foot Control
S2914
Posion 2
Posion 3
Position 1
LOCK OUT:
Used for locking chair
position
Light On/O
Toe Posion
Tilt Posion Back Posion
Base Posion
Power Light
Home

14
SCHM-003
Rev 2
Model: S2914
Headrest Release
(All Models)
Headrest height adjustment:
To lower or raise the headrest; support the headrest with one hand and push the release lever down with the other.
Lower or raise the headrest to desired height.
Release the spring loaded lever to lock in position.
Cauon:
It is recommended that the headrest height is adjusted before lowering the chair back.
Headrest
Release Lever
Be sure to support the patient’s head and neck at all times when adjusting the headrest position with patient seated.
Operang Instrucons (connued)

15
SCHM-003
Rev 2
Model: S2914
Armrest Adjustment and Removal
Operang Instrucons (connued)
Armrest Positioning:
The armrest for the S2914 has two (2) set positions. One
position is for patient use, holding arms parallel to seat
and the second position is to help with ease of entering
and exiting the chair. A locking pin on the armrest is
placed into the pocket in the armrest holder for the
desired position. Simply lift the armrest up and swivel
into Use or Enter/Exit position and lower into pocket to
lock.
Armrest Removal:
Pull the spring loaded Release Knob and lift the entire
Armrest Assembly out of the Armrest Holder to remove
the armrest.
Release Knob Armrest Holder
Armrest Assembly
‘Use’ Posion
‘Enter/Exit’ Posion
Pockets
Locking Pin
Swivel Release and Lock (Optional)
Swivel Positioning:
Unlock swivel by turning the Release Lever to the left.
Position chair as desired and lock in position by turning
the Release Lever to the right. Swivel allows chair
movement approximately 42.5° both right and left of
center.
Release Lever

16
SCHM-003
Rev 2
Infecon Control
All established infection control policies should be followed. All work surfaces and patient contact surfaces should be disinfected
between patients and at the end of the day using a product that is EPA registered as ‘hospital disinfectant’ and labeled for
‘tuberculocidal’ activity.
Two (2) such products are:
Tri-cide (800)342-3096
Medocadcio (800)777-7072
Technical Service
For parts, service, or technical information contact:
It is recommended that the chair and equipment be thoroughly cleaned and disinfected in accordance with the appropriate FDA
guidelines. Use only cleaners and disinfectants approved for cleaning vinyl. Do not use strong or harsh cleaners as they may cause
the vinyl color to fade. Apply the cleaner to a towel, not directly to the vinyl. Refer to vinyl manufacturer’s data on Boyd’s website
for cleaning products. (www.boydindustries.com/support)
General Maintenance and Care
Cleaning
Worldwide: Boyd Industries, Inc.
12900 44th Street North
Clearwater, Florida USA 33762
tel: 1-727-471-5071 (direct service line)
1-800-255-2693
fax: 1-727-561-9393

17
SCHM-003
Rev 2
Installation: S2914 Upholstery
Back Upholstery – Place the Hangers located near the top of the Back Upholstery into the slots on the Back Frame and slide
upholstery downward until it stops on Hangers. Then, press the Bullets located near the bottom of the upholstery into the
Receiving Clips on the Back Frame. The Back Upholstery Flap attaches to the rear Velcro Strip before installing the Seat
Upholstery.
Toe Upholstery – Place the two upper ‘Z’ brackets under the upper cross member of the Toe Frame. Then, press the Bullets
located near bottom of the upholstery into the Receiving Clips on the bottom of the Toe Frame. The Toe Upholstery Flap attaches
the front Velcro Strip before installing the Seat Upholstery.
Seat Upholstery – Place the two Angle Brackets against the rear cross member of Seat Frame. Lower the upholstery so that the
Locking Angle engages tightly into slot at front of Seat Frame. The angle must be in the slot and not outside of the frame.
Note: Seat Upholstery Brackets can be adjusted as required to reduce any Seat Upholstery movement.
Caution: Move the Seat Upholstery back & forth and side to side to verify that it is securely fastened to the Seat Frame before
operating.
Upholstery Installation and Removal
Velcro for securing
Back Flap
Back Frame
Slots
Receiving
Clips
Hangers
Bullets
Back Upholstery
Aachment Same as Above
Seat Upholstery
Toe Upholstery
‘Z’ Brackets
Bullets
Seat Locking Angle
Seat Locang Angles
Toe Flap
Velcro for securing
Toe Flap
Receiving Clips

18
SCHM-003
Rev 2
Trouble Shooting Guide
Appropriate Action:
Check the main ON/OFF switch (C1) to ensure that it is ON, it will be illuminated if it is on.
Check the A/C outlet that the chair is plugged into to insure that there is power going to the chair. Also
check the building’s circuit breakers.
1. Check to make sure that the cable plug on the motor is pushed all the way into the control box and
into the extension cable under the seat.
2. Check the cables connected to and from the DJB box (refer to Base Parts List) that they are
completely engaged.
3. Press and hold both position ‘1’ & position ‘2’ switches down simultaneously for five to ten seconds
or until beeping stops and changes to a steady tone. Depress the HOME switch to reset actuators.
Damaged gears - call for service.
Jammed lock mechanism. Pull headrest outward with both hands. If this problem persists - call for
service.
Symptoms:
Chair will not operate
Chair will not operate
Main On/Off switch (C1) is “ON”
Outlet on chair works
Motors will not work
Noisy motor(s)
Headrest will not release
Cauon: Always disconnect main power before performing any electrical related troubleshoong and stand clear from chair
upon compleon before restoring main power.

19
SCHM-003
Rev 2
Boyd Industries, Inc
12900 44th Street N
Clearwater, FL 33762 U.S.A
Table of contents
Other Boyd Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Anest Iwata
Anest Iwata Cliq DV-350 manual

IceCure
IceCure IceSense3 Instructions for use

Siemens
Siemens POLYMOBIL III Service instructions

Mercado Medic
Mercado Medic REAL 6100 PLUS Battery Replacement

marquette hellige
marquette hellige MicroSmart Servicing instructions

Moretti
Moretti SKEMA FORM MOC330X instruction manual

CSZ
CSZ Blanketrol III 233 Operation and technical manual

Kartsana
Kartsana BRAVA TG-1000 instruction manual

Stryker
Stryker Speed-Sheet 6010 Operation manual

Orliman
Orliman 2010-S INSTRUCTIONS FOR USE AND PRESERVATION

Bioness
Bioness NESS L300 Plus user guide

Orliman
Orliman LTG-305 Use and maintenance instructions